  • Hello! I'm a chat assistant, and I have analyzed the operating instructions for the FRITSCH Gas Pressure and Temperature Measuring System models 50.2510.00 and 50.2540.00. This system is designed for use with Planetary Mills to measure gas pressure and temperature during grinding. I am ready to answer any questions you may have about the device, its installation, usage or features.
  • What is the primary purpose of the GTM System?
    What does the measured gas pressure indicate?
    How is the data transmitted from the sensors in the grinding bowl?
    What is the maximum measuring rate of the system?
Gas Pressure and Temperature Measuring System
Planetary Mills
Operang instrucons
Gas Pressure and Temperature Measuring System for controlling the
grinding process for mechanical alloying
Valid for the following arcle numbers: 50.2510.00 / 50.2540.00
Read the instrucons prior to performing any task!
Translaon of the original operang instrucons
Version 03/2020 Index 006
Fritsch GmbH
Milling and Sizing
Industriestrasse 8
55743 Idar-Oberstein
Telephone: +49 (0)6784/ 70-0
Fax: +49 (0)6784/ 70-11
Internet: www.fritsch.de
Cercaons and CE conformity
Fritsch GmbH has been cered by the SGS-TÜV Saar GmbH.
An audit cered that Fritsch GmbH conforms to the requirements of the
DIN EN ISO 9001:2015.
The enclosed Conformity Declaraon lists the guidelines the FRITSCH
instrument conforms to, to be able to bear the CE mark.
Cercaon
CE Conformity
Cercaons and CE conformity
- 3 -
Table of contents
1 Brief descripon of the GTM System............................................ 6
1.1 Scope of applicaon............................................................... 6
1.2 Operang principle................................................................ 6
1.3 The measuring principle......................................................... 6
1.4 Design.................................................................................... 7
2 Technical data................................................................................ 8
3 Safety instrucons for using the GTM System.............................. 9
3.1 Requirements for the user..................................................... 9
3.2 Electrical safety.................................................................... 10
4Installaon................................................................................... 11
4.1 Unpacking............................................................................ 11
4.2 Seng up............................................................................. 11
4.3 Electrical connecon............................................................ 11
4.4 Charging the baery............................................................ 11
4.4.1 Undervoltage display........................................................ 11
4.4.2 Charging rechargeable baeries....................................... 12
4.4.3 Rechargeable baery lifespan........................................... 12
4.5 Inial start-up....................................................................... 12
5 Installing the soware................................................................. 14
5.1 Seng up the interface........................................................ 17
5.2 Conguring the transmission unit........................................ 17
5.2.1 Conguraon.................................................................... 17
5.2.2 Setup - conguraon......................................................... 18
6 Performing a measurement......................................................... 19
6.1 Filter..................................................................................... 19
6.2 Measuring............................................................................ 19
6.3 Transferring measurement data........................................... 20
6.4 Exing the measurement..................................................... 21
6.5 Saving / loading measurements........................................... 21
6.6 Evaluang the measurement............................................... 21
7 Detailed informaon................................................................... 24
7.1 GTM soware...................................................................... 24
7.2 Operang modes and conguraon.................................... 24
7.2.1 Data reducon.................................................................. 24
7.2.2 Standby through conguraon......................................... 25
7.2.3 Automac scan rate adjustment....................................... 25
8 Cleaning........................................................................................ 26
9 Maintenance................................................................................ 27
10 Checklist for troubleshoong...................................................... 28
Table of contents
- 4 -
11 Guarantee terms.......................................................................... 29
12 Index............................................................................................. 31
Table of contents
- 5 -
1 Brief descripon of the GTM System
1.1 Scope of applicaon
The GTM System records the gas pressure and temperature in the
grinding bowl during the grinding process in Planetary Mills. This pro-
vides a wide range of informaon for describing the communion,
mixing, acvaon, amorphisaon or phase transion processes that have
occurred.
nThe measured gas pressure contains informaon about temperature
changes due to heang of the grinding container. It describes the
interacon of the conned gas with the surfaces created during the
grinding process (adsorpon and desorpon of gases) while taking
into account the whole powder volume.
nBy measuring the basic temperature rise in the system using a tem-
perature sensor, the thermal eect on the gas pressure can be taken
into account to a limited extent.
One essenal applicaon is measuring slow changes of material states
during the grinding process, such as the amorphisaon of metals. Such
tests can take many hours, and rapid data acquision is not necessary.
Another important applicaon for high-energy grinding processes is to
achieve phase formaons or changes as a result of mechanical alloying.
During this processes, extremely fast reacons may occur that can be
recorded by the GTM System at a high measuring rate.
1.2 Operang principle
A microelectronic silicon membrane pressure sensor and a temperature
sensor are used as sensing components that are integrated into a modi-
ed grinding bowl lid (transmission unit). The measured values emied
by the sensors are digitally transmied to a staonary receiver located
outside of the Planetary Mill.
The received data telegrams are checked for completeness and correct-
ness in the receiver and transferred to a PC via a serial RS232 interface.
The computer programme calculates the measured values from the
received data and represents them graphically as a funcon of me. The
measured values can be saved and processed with a spreadsheet applica-
on, for example.
Power is supplied to the transmier and the sensors by rechargeable bat-
teries in the transmission unit. The data of two grinding bowls can also
be evaluated simultaneously. Two transmission units are required for
this.
1.3 The measuring principle
The measuring principle is based on the fact that gas, pressure and tem-
perature are directly proporonal in an ideal gas. (volume and mass of
the gas are constant p = const * T).
Brief descripon of the GTM System
- 6 -
The transfer of the process heat of the powder to the gas takes place due
to the large powder-gas contact surface and the close intermixing directly
during the grinding process so that temperature changes in the powder
are detected as pressure changes, almost without a me delay.
The release of heat to the grinding bowl causes a change in the relave
temperature of the grinding bowl. To measure this temperature change,
the temperature of the grinding container lid is also measured.
1.4 Design
The GTM System consists of the following parts:
1 Cylinder screws M6 x 50
2 GTM transmier
3 Cylinder screws M5 x 14
4 Filter lid
5 O-ring 78 x 2.5
6 O-ring 78 x 2.5
7 Grinding bowl
Brief descripon of the GTM System
- 7 -
2 Technical data
Analysis soware runs on WINDOWS XP / 7 / 8 / 10©
GTM pressure measuring range 0 bar - approx. 8 bar
Resoluon of pressure signal < 0.024 kPa
Resoluon of temperature signal 0.025 K
Measuring rate Up to 200 measured values / s in single and double trans-
mission mode
Temperature range of transmission unit 10 °C - 70 °C
Permissible temporary heat of reacon 15 kJ / s
Operang me with full baery Approx. 80 h
Data transmission RS232 connecon to PC (enclosed adapter RS232 to USB)
Radio transmission UHF radio transmier, 10 mW
Range of radio transmission Typically 15 m in buildings
Technical data
- 8 -
3 Safety instrucons for using the GTM System
nThe device may only be used for the purpose described.
nOnly use original accessories and original spare parts. Failure to
observe this instrucon can compromise the safety of the machine.
nThe operang personnel must be familiar with the content of the
operang manual. This includes ensuring that the operang manual
is provided with the device.
nDo not remove the informaon signs.
nSafe conduct must be strictly observed during all work.
nUnauthorised alteraon of the device will void Fritsch's declaraon
of conformity to European direcves and void the guarantee.
nThe device may only be operated indoors. The surrounding air may
not carry any electrically conducve dust.
nOnly use the measuring system in Planetary Mills from Fritsch GmbH.
nThe GTM System and the laboratory Planetary Mill are not explosion-
protected and are not suitable for grinding explosive materials.
nOperate the measuring system within the dened operang parame-
ters.
nDo not use the measuring system unaended.
nWatch out for jerking movements of the Planetary Mill and correct
the mass distribuons as necessary.
nWatch out for slipping movements of the grinding bowls and
increase the clamping force of the clamping device, if necessary.
nReplace screws that have worn-out threads.
nOnly use the measuring system in a dry environment.
nObserve the safety instrucons in the operang manual of your Plan-
etary Mill.
nBefore using the measuring system, consider the enthalpy of forma-
on of the new compound that could be created. Ensure that only so
much sample material is added to the grinding bowl so that the
quanty of heat released does not exceed 15 kJ per second!
nBefore switching on the Planetary Mill, disconnect the charging cable
from the transmission unit!
3.1 Requirements for the user
This operang manual is intended for persons assigned with operang
and monitoring the Fritsch of the GTM System. The operang manual
and especially its safety instrucons are to be observed by all persons
working on or with this device. In addion, the applicable rules and regu-
laons for accident prevenon at the installaon site are to be observed.
Always keep the operang manual at the installaon site of the of the
GTM System.
People with health problems or under the inuence of medicaon,
drugs, alcohol or exhauson must not operate this device.
Safety instrucons for using the GTM System
- 9 -
The of the GTM System may only be operated by authorised persons and
serviced or repaired by trained specialists. All commissioning, mainte-
nance and repair work may only be carried out by technically qualied
personnel. Qualied personnel are persons who, because of their educa-
on, experience and training as well as their knowledge of relevant
standards, regulaons, accident prevenon guidelines and operang
condions, are authorised by those responsible for the safety of the
machine to carry out the required work and are able to recognize and
avoid possible hazards as dened for skilled workers in IEC 364.
In order to prevent hazards to users, follow the instrucons in this
manual.
Malfuncons that impair the safety of persons, the of the GTM System or
other material property must be reced immediately. The following
informaon serves both the personal safety of operang personnel as
well as the safety of the products described and any devices connected
to them: All maintenance and repair work may only be performed by
technically qualied personnel.
This operang manual is not a complete technical descripon. Only the
details required for operaon and maintaining usability are described.
Fritsch has prepared and reviewed this operang manual with the
greatest care. However, no guarantee is made for its completeness or
accuracy.
Subject to technical modicaons.
3.2 Electrical safety
Note the input voltage of the receiver! Only connect plugs and cables to
the sockets intended for this purpose.
Safety instrucons for using the GTM System
- 10 -
4 Installaon
4.1 Unpacking
Compare the contents of the delivery with your order.
4.2 Seng up
nOperate indoors only
nOperang temperature of receiver 20 - 40 °C
nPlace the receiver next to the PC to be used. Connect the RS232
cable (9-pin) with the supplied adapter and the USB plug to your PC.
Then connect the cable to the interface of the receiver. Adjust the
interface used in the soware's default sengs.
4.3 Electrical connecon
Before connecng the power adapter, compare the voltage and current
values stated on the type plate with the values of the mains system to be
used.
If the mains voltage is not 230 V AC, use a power adapter that delivers
12 V DC of output voltage and 300 mW of power.
Plug the jack plug of the power adapter into the rear power supply
socket of the receiver and plug the power adapter into a power outlet.
Switch on the receiver (press the "Power" buon), the green display next
to the switch lights up.
4.4 Charging the baery
4.4.1 Undervoltage display
nIf the voltage falls below a dened value, the soware will display
the message ‘Charge rechargeable baeries / baeries. The trans-
mission unit can measure for several more hours aer the message
appears for the rst me. The rechargeable baery will then be 2/3
depleted.
nAer the message Cauon! Transmier in undervoltage
transmission mode’ is displayed, measuring and transmission mode
will be shut o aer approximately 1000 data transmissions.
Installaon
- 11 -
4.4.2 Charging rechargeable baeries
nRecharge the rechargeable baeries prior to long grinding processes.
nThe transmission unit must be at rest when the rechargeable bat-
teries are recharged. Plug the charging cable into the transmier and
into one of the two charging sockets on the receiver.
nThe assigned green display ashes briey three mes and lights up
during the charging process.
nThe green display goes out when the rechargeable baeries are fully
charged.
nIt takes approx. 17 hours to charge discharged baeries.
nStop charging the rechargeable baeries aer one day.
NOTICE!
Aer switching on the ON / OFF switch and subsequent
charging, the rechargeable baeries are always charged
for approx. 5 hours. If this is done several mes in short
succession, there is a risk of overcharging. The
rechargeable baeries may lose capacity as a result.
4.4.3 Rechargeable baery lifespan
nThe lifespan of the baeries will be shortened if the transmission
unit overheats.
nThe rechargeable baeries should be discharged as much as possible
prior to recharging.
nThe transmier rechargeable baeries can last up to 1000 charging/
discharging cycles if used properly.
4.5 Inial start-up
NOTICE!
To use the GTM system, the pressure pad on the Safe-
Lock locking device must be replaced. Simply remove
the original pressure pad and place the pressure pad
supplied with the GTM system on it.
Installaon
- 12 -
1. Unscrew the transmission unit from the grinding bowl.
2. Plug the switch plug of the charging cable into the socket of the
transmission unit.
3. Unscrew the cover of the transmission unit.
4. Switch on the start-up switch so that the point is visible (the
switched-o state is shown in the photo).
Installaon
- 13 -
5 Installing the soware
The soware for receiving the GTM data can only be
executed under WINDOWS 10 as administrator!
nStandard PC with MS Windows operang system (95 or higher).
nA serial interface with a transfer rate of 19,200 baud.
nUSB interface
1. Start the PC.
2. Insert USB ash drive and start the Fritsch Soware Installer.
3. Select ‘GTM System’ .
4. Next, select ‘GTM System’ again and the soware installaon will
begin.
5. Start by selecng the desired language.
Prerequisites for installaon
Installing the soware
Installing the soware
- 14 -
6. The Setup Wizard starts.
7. Accept the licence agreements.
8. Choose a storage locaon for the GTM soware.
9. Select the points at which you would like to set up links for the
GTM soware.
Installing the soware
- 15 -
10. Specify the start menu entry in which you want to install the link.
11. Press the ‘Install’ buon to nish the installaon.
12. A message will nofy you whether the installaon was successful.
13. Start the soware from the start menu or the GTM 2 link on your
desktop.
Installing the soware
- 16 -
5.1 Seng up the interface
1. In the system sengs, verify the COM port to which the receiver
unit is connected.
2. In the GTM 2 soware, open the Parameters menu item in the Set-
tings tab.
3. Here you can select the correct COM port. This seng will be kept
when the programme is restarted.
5.2 Conguring the transmission unit
The standard parameter set is loaded by default. If you require dierent
sengs, the transmission unit can be in changed in the conguraon
menu. The conguraon must be retained during a measurement. The
standard parameter set is sucient for a wide range of applicaons.
5.2.1 Conguraon
nSerial number:
Can only be changed in administrator mode.
nStand by:
Period of me unl transmier module is temporarily switched o
for energy-saving purposes. Data is only exchanged again if impor-
tant changes have been made to measured values.
nCheck transmission:
Aer the me period of (stand by) x (check transmission), data is
cyclically exchanged even if measured values are idencal.
nScan rate:
Dynamic scan rate in the case of changing measured values.
Example: Measured values that barely change are transmied at
larger me intervals, whereas values that change more dramacally
are transmied at larger intervals. In this way, limited baery charge
is preserved.
n‘p:T’ transmission:
Rao of pressure to temperature measurement.
Example: ‘p:T’ = 15, then the pressure, which can change quickly
under certain circumstances, is measured 15 mes, aer which the
relavely slow temperature change is measured once.
Transmission unit
Pressure
Temperature
Installing the soware
- 17 -
nUser-dened descripon of the seng.
nThe dened sengs apply to either channel 1 or the (oponal)
channel 2.
nSave sengs as data set or load other sengs.
5.2.2 Setup - conguraon
The values set under Conguraon must be transferred to the trans-
mission unit in order to take eect.
To do this, proceed as follows:
1. Connect the transmission unit to the receiver using the charging
cable.
2. Note the receiver channel used in the conguraon
3. Acvate the interface via Connecon in the ‘Measurement’ tab.
4. Then select the Setup broadcast-unit transfer tool under
‘Measurement’ and ‘Setup’ .
5. Choose the ‘Setup’ buon to transfer the new conguraon data.
Comment
Sengs
Save / load
Installing the soware
- 18 -
6 Performing a measurement
6.1 Filter
NOTICE!
Only use the transmission unit with a lter
inserted.
Change the lter when you grind dierent mate-
rials.
Check the lter aer each measurement and
replace it, if necessary.
nUnscrew the two sloed screws (boom of transmission unit)
nRemove the plate
nInsert the lter (a lter is already inserted at the factory)
nScrew the plate back on ghtly
6.2 Measuring
1. Check the following:
nStart-up switch of the transmission unit
nCharge level of the baery
nFilter
nScrew connecon of the transmission unit
2. Fill the grinding bowl (see the operang manual of the Planetary
Mill).
3. Check the lter of the grinding bowl. Replace it if damaged.
4. Screw the transmission unit onto the grinding bowl. (4 screws)
5. Place the grinding bowl into the Planetary Mill.
6. Follow the instrucons for ghtening the grinding bowl in the
operang manual of the Planetary Mill!
CAUTION!
The Planetary Mill must never be loaded on one
side. It must always be loaded symmetrically,
meaning the combined weight of the GTM
System, the grinding balls and the sample must
be the same as a counterweight.
7. Compensate for the weight dierence to the GTM System.
8. Switch on the receiver and the PC.
9. Start the GTM II programme.
Performing a measurement
- 19 -
6.3 Transferring measurement data
1. Press the buon in order to acvate the connecon to the
receiver.
2. Select the ‘1 channel’ menu item in the ‘Measurement’ menu. A
graph window for data recording appears.
3. The ‘CHECK IN’ message prompts you to disconnect the plug from
the transmission unit. This acvates the unit and starts the trans-
mission of data to the receiver.
4. The yellow LED on the receiver indicates that measurement data is
being received. (yellow display changes the switching status in
me intervals according to conguraon)
5. Set the rotaonal speed and grinding me on the Planetary Mill,
clamp the transmission unit into the bowl holder.
6. As soon as the CHECK IN is completed, data transmission will
begin. If you choose the ‘Start Measurement’ buon (arrow), the
recording is deleted and is started again.
7. Start the grinding process, observing the instrucons in the oper-
ang manual for the Planetary Mill.
NOTICE!
Monitor the temperature of the transmission
unit. Do not allow temperatures to exceed 70 °C
and plan a cool-down phase, if necessary.
The computer programme will display the following message if the tem-
perature is too high: Temperature exceeded on transmission unit
Performing a measurement
- 20 -
/