Installation Manual for SRS 300 Gas Assisted Regeneration Kit Page 6 of 7
Document No. 031665 Rev. 06 ©2006 Dionex Corporation October 2006
8.1 ICS-300, ICS-2500, ICS-2000, ICS-1500, ICS-1000, and DX 500, DX 600 Systems
The following items are recommended for operation with DX 500 and DX 600 systems to automatically turn ON and OFF
the flow of gas.
AC2 Power Control Accessory P/N 046057
CAM Controlled Air Module, 115V P/N 037125
The AC2 module controls the CAM module, turning the gas ON and OFF.
8.2 DX 120 Systems
The following items are recommended for operation with DX 120 systems to automatically turn ON and OFF the flow of
gas.
CAM Controlled Air Module, 115 V P/N 037125
Accessory Power Adapter Cable P/N 960749
The cable connects the CAM directly to the AC output on the back of the DX 120, which is activated through a solid
state relay.
8.3 Gas Requirements
It is important to turn off the suppressor current when the analytical pump is off; otherwise, the presence of gas flow will
cause a high voltage alarm and in some instances will damage the suppressor device. In order to reduce gas waste, turn
off the gas flow when the system is not operational.
WARNING: Nitrogen, helium or argon gases are recommended for the above application. Compressed air is not
recommended for the above application due to potential contamination. When using in-house
generated gases, suitable traps to remove moisture, hydrocarbons, carbon dioxide and other
particulate matter are recommended.
SECTION 8 - GAS FLOW CONTROL