  • Hello! I've reviewed the product details for the Potter PNA Portable Nitrogen Analyzer. This document outlines its features, such as one-button calibration and its durable design. It also includes information about operating temperatures, storage conditions, and how to calibrate the device. I'm ready to answer any questions you have.
  • How do I calibrate the N2 analyzer?
    When is a new calibration required?
    What is the operating temperature range?
    What type of battery does the device use?
Portable Nitrogen Analyzer
5401581 - REV A 09/17 PAGE 1 OF 1
Potter Electric Signal Company, LLC St. Louis, MO Phone: 800-325-3936 www.pottersignal.com
Features
• Durable N2 analyzer
• One-button calibration
• Auto-off
• 3-digit display
• Maintenance free
Description
The Portable Nitrogen Analyzer is an easy way to determine the purity
of nitrogen that the Potter Nitrogen Generator is producing. Since
the analyzer is portable, it gives you the ability to check the end of
the line purity of the re sprinkler system, ensuring the system is
protected. Store the N2 analyzer in a temperature similar to its ambient
environment of daily use.
Operation
Prior to turning on the unit, wait 5 minutes to allow the sensor to reach
equilibrium. Using the “ON/OFF” key, make sure the unit is in the
power “ON” mode. Press and hold the Calibration Key for 3 seconds
until the display reads “CAL”. This will calibrate the N2 Analyzer
to room air. A new calibration is required when the measured N2
percentage is above 80.1% N2 or below 78.1% N2. The N2 Analyzer
is ready to use.
Technical Specications
Measurement
Range 0-99.9%
Resolution 0.1% Nitrogen
Operating
Temperature 15°C to 40°C (59°F to 104°F)
Storage
Temperature -15°C to 50°C (5°F to 122°F)
Sensor Type Galvanic fuel cell
Battery Type 3V Lithium button cell (non-replaceable)
Model Description Stock No.
PNA Portable Nitrogen Analyzer 1119504
Ordering Information