Endres+Hauser Proline Promag 53 Modbus RS485 Operating instructions

Category
Measuring & layout tools
Type
Operating instructions
BA117D/06/en/10.05
71005396
Valid as of version
V 3.01.XX (Device software)
Operating Instructions
Proline Promag 53
MODBUS RS485
Electromagnetic Flow Measuring System
Brief operating instructions Proline Promag 53 MODBUS RS485
2Endress+Hauser
Brief operating instructions
These brief operating instructions show you how to configure the measuring device quickly and
easily:
!Note!
Always start troubleshooting with the checklist on Page 105, if faults occur after commissioning or
during operation. The routine takes you directly to the cause of the problem and the appropriate
remedial measures.
Safety instructions Page 7
First of all, familiarize yourself with the safety instructions: here you can find information about e.g. the designated use of
the measuring device, the operational safety and the safety icons and symbols used in the document.
Installation Page 13
In the “Installation” section you can find all of the necessary information for incoming acceptance, the installation
conditions that have to be observed (orientation, installation site, vibrations, etc.), through to the actual installation of the
measuring device, including information about seals, grounding and the tightening torques that have to be maintained.
Wiring Page 47
Electrical connection of the measuring device, connection of the remote version connecting cable and connection of the
fieldbus cable are described in the “Wiring” section. Other subjects addressed in this section are, for example,
specifications of the cables used, terminal assignment, potential matching and degree of protection.
Display and operating elements Page 62
The available display and operating elements of the local display and working with the function matrix are presented
here.
Configuration programs Page 81
The measuring device can be configured and operated not only via the local display, but also via software programs.
Hardware settings Page 82
Here you can find information about setting the HW write protection, the device address, etc.
Basic configuration (device parameters, automation functions) Page 88
Device-specific parameters and functions can be configured quickly and easily via the “Commissioning” Quick Setup, e.g.
display language, measurands, units, signal type, etc.
Application-specific commissioning Page 99
Empty-pipe/full-pipe adjustment for acquisition of partially filled or empty pipes (Empty Pipe Detection).
Data storage Page 100
The configuration of the transmitter can be stored on the integrated T-DAT data memory.
!Note!
For time-saving commissioning, the settings stored in the T-DAT can be transmitted to equivalent measuring points (same
parameterization) and if a device or printed circuit board is changed.
More detailed configuration Page 85
The outputs can be changed by means of configuration of the current output and relay contacts.
The F-CHIP module gives the user the added option of using software packages for diagnosis, concentration measurement
and viscosity. Page 100
Proline Promag 53 MODBUS RS485 Brief operating instructions
Endress+Hauser 3
QUICK SETUP “Commissioning”
!Note!
More detailed information on running Quick Setup menus, especially for devices without a local
display, can be found in the “Commissioning” section.
a0004280-en
Fig. 1: Quick Setup for fast commissioning
+
++
EE
Esc
E
+
-
XXX.XXX.XX
➅➆
Pulsating Flow
Batching
Carrying out the selected Quick Setup
0402
3001
Totalizer
Unit
Unit
Volume Flow
Commission
Language
Defaults
Quick Setup
HOME-POSITION
QS
Volume Mass Quit
Configure another unit? NOYES
Unit
Density
Value
Density
Unit
Unit
Mass flow
Totalizer
Selection
System units
1002
2000
0420
0700
0400
3001
B
Frequency Pulse
Current Output 1 Freq.-/ Pulse Output 1 Quit
Assign
Current
Current
Span
Value
0_4 mA
Value
20 mA
Measuring
Mode
Time
Constant
Failsafe
Mode
Failsafe
Mode
Failsafe
Mode
4000
4001
4002
4003
4004
4005
4006
Operation
Mode
4200
Selection
Output type
Assign
Frequency
End
Value Freq.
Value
F low
Value
F high
Measuring
Mode
Output
Signal
Time
Constant
4201
4203
4204
4205
4206
4207
4208
4209
Pulse
Value
Pulse
Width
Measuring
Mode
Output
Signal
4221
4222
4223
4225
4226
4227
Assign
Pulse
Configure another Output?
Autom. Configuration of Display? NO
NO
YES
Automatic parameterization
of the display
Another
Quick Setup?
NO
YES
Communication
Brief operating instructions Proline Promag 53 MODBUS RS485
4Endress+Hauser
!Note!
The display returns to the cell SETUP COMMISSIONING (1002) if you press the ESC key combination during
parameter interrogation. The stored parameters remain valid.
The “Commissioning” Quick Setup must be carried out before one of the other Quick Setups described in these
Operating Instructions is run.
mOnly units not yet configured in the current Setup are offered for selection in each cycle. The unit for mass and
volume is derived from the corresponding flow unit.
nThe “YES” option remains visible until all the units have been configured.
“NO” is the only option displayed when no further units are available.
oThis prompt only appears if a current output and/or pulse/frequency output is available. Only the outputs not yet
configured in the current Setup are offered for selection in each cycle.
pThe “YES” option remains visible until all the outputs have been parameterized.
“NO” is the only option displayed when no further outputs are available.
qThe “automatic parameterization of the display” option contains the following basic settings/factory settings
YES Main line = volume flow
Additional line = totalizer 1
Information line = operating/system condition
NO The existing (selected) settings remain.
The BATCHING QUICK SETUP is only available when the optional software package BATCHING is installed.
The PULSATING FLOW QUICK SETUP is only available if the device has a current output or a pulse/frequency
output.
Proline Promag 53 MODBUS RS485 Table of contents
Endress+Hauser 5
Table of contents
1 Safety instructions . . . . . . . . . . . . . . . . 7
1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Installation, commissioning and operation . . . . . . . . 7
1.3 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
1.5 Notes on safety conventions and icons . . . . . . . . . . . 8
2 Identification . . . . . . . . . . . . . . . . . . . . 9
2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1.1 Nameplate of the transmitter . . . . . . . . . . . . 9
2.1.2 Nameplate of the sensor . . . . . . . . . . . . . . 10
2.1.3 Nameplate for connections . . . . . . . . . . . . 11
2.2 Certificates and approvals . . . . . . . . . . . . . . . . . . . 12
2.3 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . 12
3 Installation . . . . . . . . . . . . . . . . . . . . . 13
3.1 Incoming acceptance, transport and storage . . . . . . 13
3.1.1 Incoming acceptance . . . . . . . . . . . . . . . . . 13
3.1.2 Transport . . . . . . . . . . . . . . . . . . . . . . . . . 13
3.1.3 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
3.2 Installation conditions . . . . . . . . . . . . . . . . . . . . . . 15
3.2.1 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2.2 Location . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3.2.3 Orientation . . . . . . . . . . . . . . . . . . . . . . . . 17
3.2.4 Inlet and outlet runs . . . . . . . . . . . . . . . . . 18
3.2.5 Vibrations . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2.6 Foundations, supports . . . . . . . . . . . . . . . . 19
3.2.7 Adapters . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.2.8 Nominal diameter and flow rate . . . . . . . . 20
3.2.9 Length of connecting cable . . . . . . . . . . . . 25
3.3 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.3.1 Installing the Promag W sensor . . . . . . . . . 26
3.3.2 Installing the Promag P sensor . . . . . . . . . . 33
3.3.3 Installing the Promag H sensor . . . . . . . . . . 39
3.3.4 Turning the transmitter housing . . . . . . . . 42
3.3.5 Turning the local display . . . . . . . . . . . . . . 43
3.3.6 Installing the wall-mount
transmitter housing . . . . . . . . . . . . . . . . . . 44
3.4 Post-installation check . . . . . . . . . . . . . . . . . . . . . . 46
4 Wiring . . . . . . . . . . . . . . . . . . . . . . . . 47
4.1 Cable specification MODBUS RS485 . . . . . . . . . . . 47
4.1.1 Cable type . . . . . . . . . . . . . . . . . . . . . . . . . 47
4.1.2 Shielding and grounding . . . . . . . . . . . . . . 48
4.2 Connecting the remote version . . . . . . . . . . . . . . . 49
4.2.1 Connecting Promag W / P / H . . . . . . . . . 49
4.2.2 Cable specifications . . . . . . . . . . . . . . . . . . 53
4.3 Connecting the measuring unit . . . . . . . . . . . . . . . 54
4.3.1 Terminal assignment . . . . . . . . . . . . . . . . . 54
4.3.2 Transmitter connection . . . . . . . . . . . . . . . 54
4.4 Potential matching . . . . . . . . . . . . . . . . . . . . . . . . 56
4.4.1 Standard case . . . . . . . . . . . . . . . . . . . . . . 56
4.4.2 Special cases . . . . . . . . . . . . . . . . . . . . . . . 57
4.5 Degree of protection . . . . . . . . . . . . . . . . . . . . . . . 59
4.6 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 60
5 Operation . . . . . . . . . . . . . . . . . . . . . . 61
5.1 Quick operation guide . . . . . . . . . . . . . . . . . . . . . . 61
5.2 Local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
5.2.1 Display and operating elements . . . . . . . . . 62
5.2.2 Readings displayed (operation mode) . . . . . 63
5.2.3 Additional display functions . . . . . . . . . . . . 63
5.2.4 Icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
5.2.5 Controlling the batching processes
using the local display . . . . . . . . . . . . . . . . 65
5.3 Brief operating instructions to the
function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
5.3.1 General notes . . . . . . . . . . . . . . . . . . . . . . 67
5.3.2 Enabling the programming mode . . . . . . . . 67
5.3.3 Disabling the programming mode . . . . . . . . 67
5.4 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
5.4.1 Type of error . . . . . . . . . . . . . . . . . . . . . . . 68
5.4.2 Error message type . . . . . . . . . . . . . . . . . . . 68
5.5 MODBUS RS485 communication . . . . . . . . . . . . . 69
5.5.1 MODBUS RS485 technology . . . . . . . . . . . 69
5.5.2 MODBUS telegram . . . . . . . . . . . . . . . . . . 71
5.5.3 MODBUS function codes . . . . . . . . . . . . . . 72
5.5.4 Maximum number of writes . . . . . . . . . . . 73
5.5.5 MODBUS register addresses . . . . . . . . . . . . 73
5.5.6 MODBUS error messages . . . . . . . . . . . . . 76
5.5.7 MODBUS auto-scan buffer . . . . . . . . . . . . . 76
5.6 Operating options . . . . . . . . . . . . . . . . . . . . . . . . . 81
5.6.1 Operating program
“ToF Tool - Fieldtool Package” . . . . . . . . . . 81
5.6.2 FieldCare . . . . . . . . . . . . . . . . . . . . . . . . . . 81
5.6.3 Device drivers for operating programs . . . . 81
5.7 Hardware settings . . . . . . . . . . . . . . . . . . . . . . . . . 82
5.7.1 Hardware write protection,
switching on and off . . . . . . . . . . . . . . . . . 82
5.7.2 Configuring the device address . . . . . . . . . . 83
5.7.3 Configuring the terminating resistors . . . . . 84
5.7.4 Current output configuration . . . . . . . . . . . 85
5.7.5 Relay output configuration . . . . . . . . . . . . . 86
6 Commissioning . . . . . . . . . . . . . . . . . . 87
6.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
6.2 Switching on the measuring device . . . . . . . . . . . . 87
6.3 Quick Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 88
6.3.1 Quick Setup “Commissioning” . . . . . . . . . . 88
6.3.2 Quick Setup “Pulsating Flow” . . . . . . . . . . 90
6.3.3 Quick Setup “Batching” . . . . . . . . . . . . . . . 93
6.3.4 Quick Setup “Communication” . . . . . . . . . 96
6.3.5 Data back-up with
“T-DAT SAVE/LOAD” . . . . . . . . . . . . . . . 98
6.4 Adjustment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
6.4.1 Empty-pipe/Full-pipe adjustment . . . . . . . . 99
Proline Promag 53 MODBUS RS485 Table of contents
6Endress+Hauser
6.5 Data memory (HistoROM), F-CHIP . . . . . . . . . . . 100
6.5.1 HistoROM/S-DAT (sensor-DAT) . . . . . . . 100
6.5.2 HistoROM/T-DAT (transmitter-DAT) . . . 100
6.5.3 F-CHIP (Function-Chip) . . . . . . . . . . . . . 100
7 Maintenance . . . . . . . . . . . . . . . . . . . 101
7.1 Exterior cleaning . . . . . . . . . . . . . . . . . . . . . . . . . 101
7.2 Seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
8 Accessories . . . . . . . . . . . . . . . . . . . . 102
8.1 Device-specific accessories . . . . . . . . . . . . . . . . . . 102
8.2 Measuring principle-specific accessories . . . . . . . . 103
8.3 Service-specific accessories . . . . . . . . . . . . . . . . . 104
9 Troubleshooting . . . . . . . . . . . . . . . . 105
9.1 Troubleshooting instructions . . . . . . . . . . . . . . . . 105
9.2 System error messages . . . . . . . . . . . . . . . . . . . . . 106
9.3 Process error messages . . . . . . . . . . . . . . . . . . . . . 111
9.4 Process errors without messages . . . . . . . . . . . . . 112
9.5 Response of outputs to errors . . . . . . . . . . . . . . . . 113
9.6 Spare parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 115
9.6.1 Removing and installing
printed circuit boards . . . . . . . . . . . . . . . . 116
9.6.2 Replacing the device fuse . . . . . . . . . . . . . 120
9.6.3 Replacement of exchangeable
measuring electrodes . . . . . . . . . . . . . . . . 121
9.7 Return . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
9.8 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 123
9.9 Software history . . . . . . . . . . . . . . . . . . . . . . . . . 123
10 Technical data . . . . . . . . . . . . . . . . . . 124
10.1 Technical data at a glance . . . . . . . . . . . . . . . . . . 124
10.1.1 Applications . . . . . . . . . . . . . . . . . . . . . . . 124
10.1.2 Function and system design . . . . . . . . . . . 124
10.1.3 Input variables . . . . . . . . . . . . . . . . . . . . . 124
10.1.4 Output variables . . . . . . . . . . . . . . . . . . . 125
10.1.5 Power supply . . . . . . . . . . . . . . . . . . . . . . 126
10.1.6 Measuring accuracy . . . . . . . . . . . . . . . . . 127
10.1.7 Operating conditions: Installation . . . . . . . 127
10.1.8 Operating conditions: Environment . . . . . 128
10.1.9 Operating conditions: Process . . . . . . . . . 128
10.1.10 Mechanical construction . . . . . . . . . . . . . 131
10.1.11 Human interface . . . . . . . . . . . . . . . . . . . 136
10.1.12 Certificates and approvals . . . . . . . . . . . . 136
10.1.13 Ordering information . . . . . . . . . . . . . . . 137
10.1.14 Accessories . . . . . . . . . . . . . . . . . . . . . . . 137
10.1.15 Documentation . . . . . . . . . . . . . . . . . . . 138
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 139
Proline Promag 53 MODBUS RS485 Safety instructions
Endress+Hauser 7
1 Safety instructions
1.1 Designated use
The measuring device described in these Operating Instructions is to be used only for measuring the
flow rate of conductive liquids in closed pipes.
A minimum conductivity of 20 µS/cm is required for measuring demineralized water. Most liquids
can be metered, provided they have a minimum conductivity of 5 µS/cm, for example:
• acids, alkalis, pastes, mashes, pulps,
• drinking water, wastewater, sewage sludge,
• milk, beer, wine, mineral water, yogurt, molasses, etc.
Resulting from incorrect use or from use other than that designated, the operational safety of the
measuring devices can be suspended. The manufacturer accepts no liability for damages being
produced from this.
1.2 Installation, commissioning and operation
Note the following points:
• Installation, connection to the electricity supply, commissioning and maintenance of the device
must be carried out by trained, qualified specialists authorized to perform such work by the
facility's owner-operator. The specialist must have read and understood these Operating
Instructions and must follow the instructions it contains.
The device must be operated by persons authorized and trained by the facility's owner-operator.
Strict compliance with the instructions in these Operating Instructions is mandatory.
• Endress+Hauser will be happy to assist in clarifying the chemical resistance properties of parts
wetted by special fluids, including fluids used for cleaning. However, the user is responsible for
the choice of fluid wetted materials as regards to their in-process resistance to corrosion. The
manufacturer refuses to accept liability.
If carrying out welding work on the piping, the welding unit must not be grounded by means of
the measuring device.
• The installer must ensure that the measuring system is correctly wired in accordance with the
wiring diagrams. The transmitter must be grounded, unless the power supply is galvanically
insulated.
• Invariably, local regulations governing the opening and repair of electrical devices apply.
1.3 Operational safety
Note the following points:
• Measuring systems for use in hazardous environments are accompanied by separate “Ex
documentation”, which is an integral part of these Operating Instructions. Strict compliance with
the installation instructions and ratings as stated in this supplementary documentation is
mandatory. The symbol on the front of this supplementary Ex documentation indicates the
approval and the certification body ( 0Europe, 1USA, 1Canada).
• The measuring device complies with the general safety requirements in accordance with
EN 61010, the EMC requirements of EN 61326/A1 (IEC 1326), and NAMUR recommendations
NE 21, NE 43 and NE 53.
Depending on the application, the seals of the process connections of the Promag H sensor require
periodic replacement.
• The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser distributor will supply you with current information and updates to these
Operating Instructions.
Safety instructions Proline Promag 53 MODBUS RS485
8Endress+Hauser
1.4 Return
The following procedures must be carried out before a flowmeter requiring repair or calibration, for
example, is returned to Endress+Hauser:
• Always enclose a duly completed “Declaration of contamination” form. Only then can
Endress+Hauser transport, examine and repair a returned device.
• Enclose special handling instructions if necessary, for example a safety data sheet as per EN 91/
155/EEC.
• Remove all residues. Pay special attention to the grooves for seals and crevices which could
contain residues. This is particularly important if the substance is hazardous to health, e.g.
flammable, toxic, caustic, carcinogenic, etc.
!Note!
You will find a preprinted “Declaration of contamination” form at the back of these Operating
Instructions.
#Warning!
• Do not return a measuring device if you are not absolutely certain that all traces of hazardous
substances have been removed, e.g. substances which have penetrated crevices or diffused
through plastic.
• Costs incurred for waste disposal and injury (burns, etc.) due to inadequate cleaning will be
charged to the owner-operator.
1.5 Notes on safety conventions and icons
The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the
factory in a condition in which they are safe to operate. The devices comply with the applicable
standards and regulations in accordance with EN 61010 “Protection Measures for Electrical
Equipment for Measurement, Control, Regulation and Laboratory Procedures”. The devices can,
however, be a source of danger if used incorrectly or for other than the designated use.
Consequently, always pay particular attention to the safety instructions indicated in these Operating
Instructions by the following icons:
#Warning!
“Warning” indicates an action or procedure which, if not performed correctly, can result in injury
or a safety hazard. Comply strictly with the instructions and proceed with care.
"Caution!
“Caution” indicates an action or procedure which, if not performed correctly, can result in incorrect
operation or destruction of the device. Comply strictly with the instructions.
!Note!
“Note” indicates an action or procedure which, if not performed correctly, can have an indirect
effect on operation or trigger an unexpected response on the part of the device.
Proline Promag 53 MODBUS RS485 Identification
Endress+Hauser 9
2 Identification
2.1 Device designation
The flow measuring system consists of the following components:
• Promag 53 transmitter
• Promag W, Promag P or Promag H sensor
Two versions are available:
• Compact version: transmitter and sensor form a single mechanical unit.
• Remote version: transmitter and sensor are installed separately.
2.1.1 Nameplate of the transmitter
a0004363
Fig. 2: Nameplate specifications for the “Promag 53” transmitter (example)
1 Order code/serial number: See the specifications on the order confirmation for the meanings of the individual
letters and digits.
2 Power supply/frequency: 16 to 62 V DC/20 to 55 V AC/50 to 60 Hz
Power consumption: 15 VA/W
3 Additional functions and software
4 Available inputs and outputs
5 Reserved for information on special products
6 Ambient temperature range
7 Degree of protection
PROMAG 53
-20°C (-4°F) <Tamb<+60°C (+140°F)
Pat. US 5,323,156 5,479,007
Pat. US 4,382,387 4,704,908 5,351,554
IP67 / NEMA/Type 4X
Order Code:
Ser.No.:
TAG No.:
53P1H-XXXXXXXXXXXX
12345678901
ABCDEFGHJKLMNPQRST
16-62VDC/20-55VAC
50-60Hz
MODBUS RS485, STATUS-IN
15VA/W
i
Pat. UK 2 084 740 EP 219 725 EP 521 169
Pat. UK EP 541 878 EP 618 680
f-OUT, I-OUT
N12895
ECC
EPD/MSU
1
2
3
4
5
67
Identification Proline Promag 53 MODBUS RS485
10 Endress+Hauser
2.1.2 Nameplate of the sensor
a0004374
Fig. 3: Nameplate specifications for the “Promag” sensor (example)
1 Order code/serial number: See the specifications on the order confirmation for the meanings of the individual
letters and digits.
2 Calibration factor with zero point
3 Nominal diameter/nominal pressure
4 Medium temperature range
5 Materials: lining/measuring electrode
6 Reserved for information on special products
7 Ambient temperature range
8 Calibration tolerance
9 Additional information
EPD: with empty pipe detection electrode
R/B: with reference electrode (only for Promag P)
10 Degree of protection
11 Reserved for additional information on device version (approvals, certificates)
12 Flow direction
-20°C (-4°F)<Tamb<+60°C (+140°F) NEMA/Type4X
Pat. US 4,382,387 4,704,908 5,540,103
XXXXX-XXXXXXXXXXXX
0.5328/5
PFA
-10°C...150°C/+14°F...300°F
12345678901 RY
ABCDEFGHJKLMNPQRST
DN100 DIN/EN PN40
EPD/MSÜ, R/B
Materials:
Order Code:
TM:
K-factor:
Ser.No.:
TAG No.:
PROMAG P
1.4435/316L
Electrode:
0.5CAL
N12895
1
2
3
4
59
6
7
10
11
12
8
Proline Promag 53 MODBUS RS485 Identification
Endress+Hauser 11
2.1.3 Nameplate for connections
a0004362
Fig. 4: Nameplate specifications for Proline transmitter connections (example)
1 Serial number
2 Possible configuration of current output
3 Possible configuration of relay contacts
4 Terminal assignment, cable for power supply: 85 to 260 V AC, 20 to 55 V AC, 16 to 62 V DC
Terminal No. 1: L1 for AC, L+ for DC
Terminal No. 2: N for AC, L for DC
5 Pending signals at the inputs and outputs, possible configurations and terminal assignment Page 54 ff.
6 Version of device software currently installed
7 Installed communication type
8 Information on communication driver
9 Date of installation
10 Language group
11 Current updates to data specified in 6 to 9
Communication:
Drivers:
Device SW:
–––
3.01.00
MODBUS RS485
Date: 01. OCT 05
SW–version
ex–works
26(+) / 27(-)
NC:
Versorgung /
Tension d'alimentation
Observer manuel d'instruction
See operating manual
Betriebsanleitung beachten
max. 30 VAC / 0.5 A
26 = B (RxD/TxD-P)
27 = A (RxD/TxD-N)
max. 60 VDC / 0.1 A
max. 30 VAC / 0.5 A
3...30 VDC, Ri = 3 kOhm
max. 60 VDC / 0.1 A
STATUS–IN
MODBUS RS485
XXXXXXXXXXX
Ser.No.:
Supply /
24(+) / 25(-)
22(+) / 23(-)
20(+) / 21(-)
N/L-
PE
A:
NO:
P:
L1/L+
12
FEK1512–2
X
X
active
passive
normally open contact
normally closed contact
NO
RELAY
RELAY NC
Update 1 Update 2
2
3
1
4
5
6
7
8
9
11
(WEA)
10
Identification Proline Promag 53 MODBUS RS485
12 Endress+Hauser
2.2 Certificates and approvals
The devices are designed in accordance with good engineering practice to meet state-of-the-art
safety requirements, have been tested, and left the factory in a condition in which they are safe to
operate. The devices comply with the applicable standards and regulations in accordance with EN
61010 “Protection Measures for Electrical Equipment for Measurement, Control, Regulation and
Laboratory Procedures” and with the EMC requirements of EN 61326/A1.
The measuring system described in these Operating Instructions is therefore in conformity with the
statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the
device by affixing to it the CE mark.
The measuring system complies with the EMC requirements of the “Australian Communications
Authority (ACA)”.
The measuring device meets all the requirements of the MODBUS/TCP conformity test and holds
the “MODBUS/TCP Conformance Test Policy, Version 2.0”. The measuring device has successfully
passed all the test procedures carried out and is certified by the “MODBUS/TCP Conformance Test
Laboratory” of the University of Michigan.
2.3 Registered trademarks
KALREZ® and VITON®
Registered trademarks of E.I. Du Pont de Nemours & Co., Wilmington, USA
TRI-CLAMP®
Registered trademark of Ladish & Co., Inc., Kenosha, USA
HistoROM™, S-DAT®, T-DAT™, F-CHIP®, ToF Tool - Fieldtool® Package, Fieldcheck®,
Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
MODBUS®
Registered trademark of the MODBUS Organization
Proline Promag 53 MODBUS RS485 Installation
Endress+Hauser 13
3 Installation
3.1 Incoming acceptance, transport and storage
3.1.1 Incoming acceptance
On receipt of the goods, check the following points:
• Check the packaging and the contents for damage.
• Check the shipment, make sure nothing is missing and that the scope of supply matches your
order.
3.1.2 Transport
The following instructions apply to unpacking and to transporting the device to its final location:
• Transport the devices in the containers in which they are delivered.
Do not remove the protective plates or caps on the process connections until the device is ready
to install. This is particularly important in the case of sensors with PTFE linings.
Special notes on flanged devices
"Caution!
The wooden covers mounted on the flanges before the device leaves the factory protect the linings
on the flanges during storage and transportation. Do not remove these covers until immediately
before the device is installed in the pipe.
• Do not lift flanged devices by the transmitter housing, or the connection housing in the case of
the remote version.
Transporting flanged devices (DN 300)
Use webbing slings slung round the two process connections. Do not use chains, as they could
damage the housing.
#Warning!
Risk of injury if the measuring device slips. The center of gravity of the assembled measuring device
might be higher than the points around which the slings are slung.
At all times, therefore, make sure that the device does not unexpectedly turn around its axis or slip.
a0004294
Fig. 5: Transporting sensors with DN 300
Installation Proline Promag 53 MODBUS RS485
14 Endress+Hauser
Transporting flanged devices (DN > 300)
Use only the metal eyes on the flanges for transporting the device, lifting it and positioning the
sensor in the piping.
"Caution!
Do not attempt to lift the sensor with the tines of a fork-lift truck beneath the metal casing. This
would buckle the casing and damage the internal magnetic coils.
a0004295
Fig. 6: Transporting sensors with DN > 300
3.1.3 Storage
Note the following points:
• Pack the measuring device in such a way as to protect it reliably against impact for storage (and
transportation). The original packaging provides optimum protection.
• The storage temperature corresponds to the ambient temperature range of the measuring
transmitter and the appropriate measuring sensors. Page 128
• The measuring device must be protected against direct sunlight during storage in order to avoid
unacceptably high surface temperatures.
Choose a storage location where moisture does not collect in the measuring device. This will help
prevent fungus and bacteria infestation which can damage the liner.
Do not remove the protective plates or caps on the process connections until the device is ready
to install. This is particularly important in the case of sensors with PTFE linings.
Proline Promag 53 MODBUS RS485 Installation
Endress+Hauser 15
3.2 Installation conditions
3.2.1 Dimensions
All the dimensions and lengths of the sensor and transmitter are provided in the separate
documentation “Technical Information”.
3.2.2 Location
Entrained gas bubbles in the measuring tube can result in an increase in measuring errors.
Avoid the following locations:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipeline.
a0003202
Fig. 7: Location
Installation of pumps
Do not install the sensor on the intake side of a pump. This precaution is to avoid low pressure and
the consequent risk of damage to the lining of the measuring tube. Information on pressure tightness
of the measuring tube lining Page 130
It might be necessary to install pulse dampers in systems incorporating reciprocating, diaphragm or
peristaltic pumps. Information on vibration and shock resistance of the measuring system
Page 128
a0003203
Fig. 8: Installation of pumps
h 2xDN³
Installation Proline Promag 53 MODBUS RS485
16 Endress+Hauser
Partially filled pipes
Partially filled pipes with gradients necessitate a drain-type configuration. The Empty Pipe Detection
function offers additional certainty in detecting empty or partially filled pipes. Page 99
"Caution!
Risk of solids accumulating. Do not install the sensor at the lowest point in the drain. It is advisable
to install a cleaning valve.
a0003204
Fig. 9: Installation in partially filled pipe
Down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes longer than 5 meters. This
precaution is to avoid low pressure and the consequent risk of damage to the lining of the measuring
tube. This measure also prevents the system losing prime, which could cause air inclusions.
Information on pressure tightness of the measuring tube lining Page 130
a0003205
Fig. 10: Installation measures for down pipes
1Vent valve
2 Siphon
³5xDN
³2xDN
>5m
2
1
Proline Promag 53 MODBUS RS485 Installation
Endress+Hauser 17
3.2.3 Orientation
An optimum orientation position helps avoid gas and air accumulations and deposits in the
measuring tube. Promag, nevertheless, supplies a range of functions and accessories for correct
measuring of problematic fluids:
• Electrode Cleaning Circuit (ECC) for applications with accretive fluids, e.g. electrically
conductive deposits “Description of Device Functions” manual.
• Empty Pipe Detection (EPD) ensures the detection of partially filled measuring tubes, e.g. in the
case of degassing fluids or varying process pressures Page 99
• Exchangeable measuring electrodes for abrasive fluids (only Promag W) Page 121
Vertical orientation
This is the ideal orientation for self-emptying piping systems and for use in conjunction with Empty
Pipe Detection.
a0003206
Fig. 11: Vertical orientation
Horizontal orientation
The measuring electrode plane should be horizontal. This prevents brief insulation of the two
electrodes by entrained air bubbles.
"Caution!
Empty Pipe Detection functions correctly with the measuring device installed horizontally only
when the transmitter housing is facing upward (see Figure). Otherwise there is no guarantee that
Empty Pipe Detection will respond if the measuring tube is only partially filled or empty.
a0003207
Fig. 12: Horizontal orientation
1 EPD electrode for empty pipe detection (not for Promag H, DN 2 to 4)
2 Measuring electrodes for the signal acquisition
3 Reference electrode for the potential matching (not with Promag H)
A1
22
A
3
Installation Proline Promag 53 MODBUS RS485
18 Endress+Hauser
3.2.4 Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following requirements for the inlet and outlet runs is necessary in order to
ensure measuring accuracy.
• Inlet run: 5 x DN
•Outlet run: 2 x DN
a0003210
Fig. 13: Inlet and outlet runs
3.2.5 Vibrations
Secure the piping and the sensor if vibration is severe.
"Caution!
It is advisable to install sensor and transmitter separately if vibration is excessively severe.
Information about the permitted shock and vibration resistance Page 128
a0003208
Fig. 14: Measures to prevent vibration of the measuring device
5xDN2xDN
>10m
Proline Promag 53 MODBUS RS485 Installation
Endress+Hauser 19
3.2.6 Foundations, supports
If the nominal diameter is DN 350, mount the transmitter on a foundation of adequate load-
bearing strength.
"Caution!
Risk of damage.
Do not support the weight of the sensor on the metal casing: the casing would buckle and damage
the internal magnetic coils.
a0003209
Fig. 15: Correct support for large nominal diameters (DN 350)
Installation Proline Promag 53 MODBUS RS485
20 Endress+Hauser
3.2.7 Adapters
Suitable adapters to DIN EN 545 (double-flange reducers) can be used to install the sensor in larger-
diameter pipes. The resultant increase in the rate of flow improves measuring accuracy with very
slow-moving fluids.
The nomogram shown here can be used to calculate the pressure loss caused by cross-section
reduction.
!Note!
The nomogram applies to liquids of viscosity similar to water.
1. Calculate the ratio of the diameters d/D.
2. From the nomogram, read off the pressure loss as a function of flow velocity (downstream from
the reduction) and the d/D ratio.
a0003213
Fig. 16: Pressure loss due to adapters
3.2.8 Nominal diameter and flow rate
The diameter of the pipe and the flow rate determine the nominal diameter of the sensor. The
optimum velocity of flow is between 2 and 3 m/s. The velocity of flow (v), moreover, has to be
matched to the physical properties of the fluid:
• v < 2 m/s: for abrasive fluids
• v > 2 m/s: for accretive fluids
!Note!
Flow velocity can be increased, if necessary, by reducing the nominal diameter of the sensor using
adapters. Page 20
100
10
0.5
d/D
[mbar]
0.6 0.7 0.8 0.9
1 m/s
2 m/s
3 m/s
4 m/s
5 m/s
6 m/s
7 m/s
8 m/s
1
D
d
max. 8°
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144

Endres+Hauser Proline Promag 53 Modbus RS485 Operating instructions

Category
Measuring & layout tools
Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI