12
• Electrode: Connect the pH or ORP electrode to the BNC
socket (#1 at page 9).
To benefit from the differential (symmetrical) input, con-
nect the proper electrode wire (if available) or a cable with
a potential matching pin (grounding bar) to the relevant
terminal (#2 at page 9) and enable the differential input
through the setup code I.04.
Note When it is not possible to immerse the Potential Matching Pin
together with the pH electrode in the solution (asymmetrical
input), disable the differential input through the setup code
I.04. In this case, detach the Potential Matching Pin from
its connector. Failure to do so may cause damage of the
reference electrode.
• Pt 100/Pt1000 Terminals: these contacts (#4 at page 9)
connect the Pt 100/Pt 1000 temperature sensor for au-
tomatic temperature compensation of pH measurement.
In the case of shielded wire, connect the shield to pin 4.
In the case of a 2-wire sensor
connect the Pt 100/Pt 1000 to
pins 1 and 3, and short pins 2
and 3 with a jumper wire.
If the Pt 100/Pt 1000 has more
than 2 wires, connect the two
wires of one end to pins 2 and
3 (pin 2 is an auxiliary input to
compensate for the cable resis-
tance) and one wire from the
other end to pin 1. Leave the fourth wire unconnected, if
present.
Note The instrument automatically recognizes the sensor type
(Pt100 or Pt1000).
• Analog Outputs: when using shielded cable, if the shield
is not connected at the other end of the cable, then con-
nect it to the “+” terminal, otherwise leave it floating.
All cables connected to rear panel should end with cable
lugs.
A circuit breaker (rated 10A max.) must be connected in
close proximity to the equipment, and in a position easy to
reach by the operator, for disconnection of the instrument
and of all the devices connected to the relays.