04 Series for CO(-H2) Chapter 3: Measuring Mode • 15
3. If CAL.RMDR is set to ON (factory setting) and a calibration is due, the screen that
appears next depends on how CAL.EXPD is set in User Mode (see pg.53). The three
possible screens are described below. If a calibration is not due, the instrument displays
the number of days left until a calibration is due.
CAL.EXPD set to
CONFIRM (factory setting)
CAL.EXPD set to
CANT.USE
CAL.EXPD set to
NONE
LCD
Sound Buzzer sounds double pulsing
tone
Buzzer sounds double pulsing
tone
None
Action • Option A, Perform calibration
:
Press and release POWER
MODE to enter User Mode
and perform a calibration. The
instrument takes you straight
to the calibration start screen
in User Mode’s GAS CAL\A-
CAL(E-CAL) item (if
Password Protection is set to
On using the 04 Series Setup
Program, you must enter a
password first). See pg.42 for
calibration instructions.
If the calibration is success-
ful, the screen above will not
appear again until the unit is
due for calibration. If the cali-
bration is not successful, the
screen above will again appear
in the startup sequence.
• Option B, Bypass message
: To
continue without performing a
calibration, press and release
AIR.
The CO-04C cannot be used until
a successful calibration is per-
formed. Press and release
POWER MODE to enter User
Mode and perform a calibration.
The instrument takes you straight
to the calibration start screen in
User Mode’s GAS CAL/A-CAL
(E-CAL) item (if Password Pro-
tection is set to On using the 04
Series Setup Program, you must
enter a password first). If you
don’t press POWER MODE, the
instrument automatically goes to
the calibration start screen after 6
seconds (if Password Protec-
tion is set to On using the 04
Series Setup Program, you must
enter a password first). See pg.42
for calibration instructions.
If the calibration is successful,
the screen above will not appear
again until the unit is due for cali-
bration. If the calibration is not
successful, the screen above will
again appear in the startup
sequence.
• Option A, Perform calibration
:
If you want to enter User
Mode and perform a
calibration, press and release
POWER MODE. The
instrument takes you straight
to the calibration start screen
in User Mode’s GAS CAL/A-
CAL (E-CAL) item (if
Password Protection is set to
On using the 04 Series Setup
Program, you must enter a
password first).
• Option B, Bypass message
:
To continue without
performing a calibration, wait
a few seconds for the
instrument to continue with its
startup sequence.