GW Instek GPM-8213 User manual

Category
Measuring, testing & control
Type
User manual
Digital Power Meter
GPM-8213
USER MANUAL
ISO-9001 CERTIFIED MANUFACTURER
This manual contains proprietary information, which is protected by
copyright. All rights are reserved. No part of this manual may be
photocopied, reproduced or translated to another language without
prior written consent of Good Will company.
The information in this manual was correct at the time of printing.
However, Good Will continues to improve products and reserves the
rights to change specification, equipment, and maintenance
procedures at any time without notice.
Good Will Instrument Co., Ltd.
No. 7-1, Jhongsing Rd., Tucheng Dist., New Taipei City 236, Taiwan.
TABLE OF CONTENTS
3
Table of Contents
SAFETY INSTRUCTIONS .................................................. 5
GETTING STARTED ........................................................ 11
Characteristics .......................................... 12
Appearance ............................................... 15
Set Up ....................................................... 26
BASIC SETTING ............................................................. 29
Setting up measurement range ................. 30
Setting up measurement status ................ 34
Setting up System status .......................... 42
MEASUREMENT AND OTHER FUNCTIONS .................... 47
Measurement function .............................. 48
Other functions ........................................ 51
Integration measurement function ............ 53
REMOTE CONTROL ....................................................... 60
Configure Remote Control Interface .......... 61
Return to Local Control ............................. 62
COMMAND OVERVIEW ................................................. 65
Command Syntax ...................................... 65
Command List .......................................... 69
APPENDIX .................................................................... 102
Specifications ......................................... 103
Dimensions ............................................ 108
Declaration of Conformity ....................... 109
Power measurement ............................... 110
Introduction to IEC-62301 ....................... 112
GPM-8213 User Manual
4
EUP Directive Lot6 specifications ........... 112
Connection Guide ................................... 113
SAFETY INSTRUCTIONS
5
SAFETY INSTRUCTIONS
This chapter contains important safety
instructions that you must follow during
operation and storage. Read the following before
any operation to ensure your safety and to keep
the instrument in the best possible condition.
Safety Symbols
These safety symbols may appear in this manual or on the
instrument.
WARNING
Warning: Identifies conditions or practices that
could result in injury or loss of life.
CAUTION
Caution: Identifies conditions or practices that
could result in damage to the GPM-8213
or to other properties.
DANGER High Voltage
Attention Refer to the Manual
Protective Conductor Terminal
Earth (ground) Terminal
GPM-8213 User Manual
6
Do not dispose electronic equipment as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased.
SAFETY INSTRUCTIONS
7
Safety Guidelines
General Guideline
CAUTION
Make sure that the voltage input level does not
exceed DC848V/AC600V.
Make sure the current input level does not
exceed 20A.
Do not place any heavy object on the
instrument.
Avoid severe impact or rough handling that can
lead to damaging the instrument.
Do not discharge static electricity to the
instrument.
Use only mating connectors, not bare wires, for
the terminals.
Do not block or obstruct the cooling fan vent
opening.
Do not perform measurement at the source of a
low-voltage installation or at building
installations (Note below).
Do not disassemble the instrument unless you
are qualified as service personnel.
Make sure that the COM terminal to earth is
limited to 300Vpk.
Remove all test leads before disconnecting the
mains power cord from the socket.
If the equipment is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
The device should be placed in a place where
the plug connected to it can be removed easily.
GPM-8213 User Manual
8
(Note) EN 61010-1:2010 specifies the measurement categories and
their requirements as follows. The GPM-8213 falls under category II
300V.
Measurement category IV is for measurement performed at the
source of low-voltage installation.
Measurement category III is for measurement performed in the
building installation.
Measurement category II is for measurement performed on the
circuits directly connected to the low voltage installation.
AC Input voltage: 100-240 VAC 50/60Hz
The power supply voltage should not fluctuate
more than 10%.
Connect the protective grounding conductor of
the AC power cord to an earth ground, to avoid
electrical shock.
Disconnect the power cord before cleaning.
Use a soft cloth dampened in a solution of mild
detergent and water. Do not spray any liquid.
Do not use chemicals containing harsh material
such as benzene, toluene, xylene, and acetone.
Location: Indoor, no direct sunlight, dust free,
almost non-conductive pollution (Note below)
Temperature: 0°C to 40°C
Humidity: < 30°C: < 80%RH(non-condensing);
30°C~40°C:<70%RH(non-condensing);
>40°C: <50%RH (non-condensing)
Altitude: <2000m
SAFETY INSTRUCTIONS
9
(Note) EN 61010-1:2010 specifies the pollution degrees and their
requirements as follows. The GPM-8213 falls under degree 2.
Pollution refers to “addition of foreign matter, solid, liquid, or
gaseous (ionized gases), that may produce a reduction of
dielectric strength or surface resistivity”.
Pollution degree 1: No pollution or only dry, non-conductive
pollution occurs. The pollution has no influence.
Pollution degree 2: Normally only non-conductive pollution
occurs. Occasionally, however, a temporary conductivity caused
by condensation must be expected.
Pollution degree 3: Conductive pollution occurs, or dry, non-
conductive pollution occurs which becomes conductive due to
condensation which is expected. In such conditions, equipment
is normally protected against exposure to direct sunlight,
precipitation, and full wind pressure, but neither temperature
nor humidity is controlled.
Storage
environment
Location: Indoor
Temperature: -40°C to 70°C
Humidity: <90%RH(non-condensing)
Disposal
Do not dispose this instrument as unsorted
municipal waste. Please use a separate collection
facility or contact the supplier from which this
instrument was purchased. Please make sure
discarded electrical waste is properly recycled to
reduce environmental impact.
GPM-8213 User Manual
10
Power cord for the United Kingdom
When using the unit in the United Kingdom, make sure the power
cord meets the following safety instructions.
NOTE: This lead/appliance must only be wired by competent persons
WARNING: THIS APPLIANCE MUST BE EARTHED
IMPORTANT: The wires in this lead are coloured in accordance with the
following code:
Green/ Yellow:
Earth
Blue:
Neutral
Brown:
Live (Phase)
As the colours of the wires in main leads may not correspond with
the coloured marking identified in your plug/appliance, proceed
as follows:
The wire which is coloured Green & Yellow must be connected to
the Earth terminal marked with either the letter E, the earth symbol
or coloured Green/Green & Yellow.
The wire which is coloured Blue must be connected to the terminal
which is marked with the letter N or coloured Blue or Black.
The wire which is coloured Brown must be connected to the
terminal marked with the letter L or P or coloured Brown or Red.
If in doubt, consult the instructions provided with the equipment
or contact the supplier.
This cable/appliance should be protected by a suitably rated and
approved HBC mains fuse: refer to the rating information on the
equipment and/or user instructions for details. As a guide, a cable
of 0.75mm2 should be protected by a 3A or 5A fuse. Larger
conductors would normally require 13A types, depending on the
connection method used.
Any exposed wiring from a cable, plug or connection that is
engaged in a live socket is extremely hazardous. If a cable or plug is
deemed hazardous, turn off the mains power and remove the cable,
any fuses and fuse assemblies. All hazardous wiring must be
immediately destroyed and replaced in accordance to the above
standard.
GETTING STARTED
11
GETTING STARTED
This chapter describes the GPM-8213 in a nutshell,
including accessories, package contents, its main
features and front / rear panel introduction.
Characteristics .......................................................... 12
Accessories ................................................................. 15
Package Contents ....................................................... 16
Appearance ............................................................... 17
Front Panel ................................................................. 17
Display Overview ........................................................ 21
Rear Panel................................................................... 24
Set Up ...................................................................... 26
Tilting the Stand ......................................................... 26
Power Up .................................................................... 27
Connect the wires to the GPM-8213 ......................... 28
GPM-8213 User Manual
12
Characteristics
The GPM-8213 is a high-precision, programmable power meter for
using in standby measuring the device with low power such as
switching power supplies, transformers, power supplies, adapter
and other devices. It equips with a color TFT-LCD screen which is
very convenient for reading the measurement results. The GPM-
8213 has become a reliable power measurement instruments
because of its simple operation, excellent performance and
automatic measurement interface.
Press the buttons on the front panel to easily
turn on the GPM-8213 measurement function.
All settings and measurements results are
displayed on the TFT-LCD screen panel for easy
use of each function.
Standard display mode: 2 main measurement
results and 6 secondary measurement results are
displayed in this screen.
Simple display mode: 4 major measurement
results are displayed in this screen.
GETTING STARTED
13
Performance
6 selectable voltage ranges available from 15V to
300V with 0.1% of reading + 0.1% of range.
12 selectable current ranges available from 5mA
to 20A with 0.1% of reading + 0.1% of range.
It can even measure the voltage of abnormal
wave of CF 3. The half-range CF is up to 6.
It can even measure the current of abnormal
wave of CF 3. The half-range CF is up to 6.
Test terminals in the front panel.
Total harmonic distortion measurement.
Features
Full five-digit measurement.
Voltage measurement range: 15V ~ 600V or
automatic switching
Current measurement range: 5mA ~ 20A or
automatic switching
Maximum accuracy of 0.1% of reading + 0.1% of
range
2 main measurement readings and 6 minor
measurement readings are displayed in the
screen of standard display mode.
4 main measurement readings are displayed in
the screen of simple display mode.
Added stand-alone display of total harmonic
distortion measurement function (13 steps)
Test bandwidth of voltage and current: DC ~
6kHz.
Added W-h power time integrator function
Selectable boot settings (Previous / Default)
Interface
Standard interface: USB / RS232 / LAN
Optional interface: GPIB
GPM-8213 User Manual
14
It can be applied to production test such as
power supplies, transformers, motors, electrical
equipment and other equipment with low
standby power.
It can be applied to power measurement
conforms to IEC 62301
It can be applied to assess the power
consumption of product design.
GETTING STARTED
15
Accessories
Standard Accessories
Part number
Description
82PM-82130Ex1
User Manual CD
82DM-83421Mx1
Safety Instruction Sheet
Region dependent
Power Cord
GTL-209
Test leads: 2x red, 2x black
Optional Accessories
Part number
Description
GPM-001
Test Fixture
GTL-232
RS232C cable
GTL-246
USB cable
GTL-248
GPIB cable
GRA-422
Rack Adapter Panel (19, 2U)
Option
Name
Description
Opt.01
GPIB (Factory installed)
GPM-8213 User Manual
16
Package Contents
Check the contents before using the instrument.
Main unit
Test leads (red x2,
black x2)
Power cord x1 (region
dependent)
User manual CD
Safety instruction
sheet
GETTING STARTED
1
17
Appearance
Front Panel



 



POWER
INTEGRATOR
  
CURRENT VOLTAGE
I


I
V
V


CAT II 300V

GPM-8213

Power Switch
POWER
Turns On
or Off the main
power. For the power up sequence,
see page
27.
Current
,
Voltage
Terminals
CURRENT VOLTAGE
I


I
V
V


CAT II 300V
Current input:
I+ and I
terminals
; Voltage
input: V+
and V-
terminals.
GPM-8213 User Manual
1
18
Note
If the measurement power supply has
positive and negative
electrode,
please
connect
+ to the positive
electrode of
power supply
and - to the negative
electrode of power supply.
Warning
The m
aximum measurable
current and
voltage
are 600 V and 10A for
voltage
and current
terminals of the front
panel
of the GPM-8213. Do not
input
exceeded
voltage and current,
otherwise it will burn the device.
Function keys
 

V
-Range key,
up/down arrow keys
and Enter key
can be used together
to select
a voltage range or auto
range
measurement mode.
See page
30
.
 

I
-Range key,
up/down arrow keys
and Enter key
can be used together
to select
a current range or auto
range
measurement mode.
See page
30
.
GETTING STARTED
1
19

Press this button to
display the
maximum
measurement reading.
See page 51.

Press this key to select measure
mode
(DC/AC/AC+DC)
. See page
52.

Press this key to e
nter the
measurement settings menu
. See
page 34.

Press this key to
switch
window and
stop
refreshing. See page 52.

 
INTEGRATOR
ġ
Use the left and right arrow
keys to select Integrator
mode
, and press Enter
button
to enter the time
integrator
function.
See page
53
. ġ


Press this key to toggle to key lock.
In
Remote
control mode, press this
button to
switch to local mode.
See
page
52.
Confirm button

This button is used to e
nter the
menu, confirm the settings
and
switch
between
the standard display
mode and
simple
display mode (no
function table and
display icon)
. See
page 52.
GPM-8213 User Manual
2
20
Cancel (Exit)
button

Press this button to c
ancel the
current setting
. T
he cursor returns to
the default position or return to the
previous menu according to the
situation. See page 52.
Arrow Keys
ġ
This for arrow keys are used
to e
dit the parameters,
browse the menu system
and
select the parameter
range.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114

GW Instek GPM-8213 User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI