Lakeshore CRX-VF Cryogenic Probe Station, CPX Cryogenic Probe Station, CPX-VF Cryogenic Probe Station, CRX-4K Cryogenic Probe Station, CRX-6.5K Cryogenic Probe Station, CRX-EM-HF Cryogenic Probe Station, EMPX-H2 Cryogenic Probe Station, EMPX-HF Cryogenic Probe Station, FWPX Cryogenic Probe Station, PS-100 Tabletop Probe Station, TTPX Cryogenic Probe Station Installation guide

  • I've reviewed the Installation Instructions for the Probe Blade Spare Tip. This document outlines the process for replacing the tip, including cleaning, soldering, and bending. I'm ready to answer your questions about this procedure.
  • What should be used to clean the blade?
    What type of solder should be used?
    What should the tip bend degree be?
Lake Shore Cryotronics, Inc. 575 McCorkle Boulevard, Westerville, Ohio 43082
Phone: 614-891-2244; Fax: 614-818-1600; Email: info@lakeshore.com; Website: www.lakeshore.com
Form F009-10-00 Revision C; 2 May 2014
Installation Instructions: Probe Blade Spare Tip
1. Remove the old tip and completely remove the remaining solder from the original tip installation.
2. Thoroughly clean the blade, preferably in an ultrasonic cleaner using de-ionized water, then rinse with
isopropyl alcohol.
3. Use Kester® no clean solder (containing Sn96.5Ag3.0Cu0.5 alloy) or equivalent and Kester®
flux pen 2331-ZX or equivalent.
4. Hold the tip to the bottom of the blade, allowing the excess length of the tip to extend off the back of the
blade.
5. Using the flux pen, apply the flux to the bottom of the blade and to the portion of the tip contacting the blade;
do not contact the sharpened tip. Flux should pool around the base of the tip on the blade.
6. Solder the tip to the blade using a steady swipe of solder from the iron. Do not rest the iron on the blade; the
solder should flow instantly.
7. Wet a swab with isopropyl alcohol and lightly clean the flux off the blade around the soldered area. Do not
use the swab on the tip portion extending past the blade.
8. Trim excess length off the back of the blade.
9. Place bend to the end of the tip for probing at approximately 0.75 mm from sharp end of tip at a 60° angle.
/