Tektronix DPO4104B Technical Reference

Type
Technical Reference
xx
MSO4000B and DPO4000B Series
Digital Phosphor Oscilloscopes
Specications and Performance Verication
ZZZ
Technical Reference
*P077050903*
077-0509-03
MSO4000B and DPO4000B Series
Digital Phosphor Oscilloscopes
Specications and Performance Verication
ZZZ
Technical Reference
xx
Revision B
This document supports rmware version 1.00 and above for
both MSO4000B Series instruments and DPO4000B Series
instruments.
Warning
The servicing instructions are for use by qualied personnel
only. To avoid personal injury, do not perform any servicing
unless you are qualied to do so. Refer to all safety summaries
prior to performing service.
www.tek.com
077-0509-03
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX a nd TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverto
n, OR 97077
USA
For product information, sales, service, and technical support:
In North America, c all 1-800-833-9200.
Worldwide, visit www.tek.com to nd contacts in your area.
Table of Contents
General safety summary .......................................................................................... iii
Specications ....................................................................................................... 1
Analog Channel Input and Vertical Specications ......................................................... 2
Horizontal
and Acquisition System Specications......................................................... 9
Trigger Specications ........................................................................................ 12
Digital Acquisition Specications, MSO4000B Series................................................... 18
P6616 Digital Probe Specications......................................................................... 19
Display Specications........................................................................................ 20
Input/Output Port Specications ............................................................................ 21
Data Sto
rage Specications.................................................................................. 23
Power Source Specications ................................................................................ 23
Environmental Specications ............................................................................... 24
Mechanical Specications ................................................................................... 25
Performance Verication ......................................................................................... 27
Test Record .................................................................................................... 28
Perf
ormance Verication Procedures....................................................................... 54
Self Test................................................................................................... 55
Check Input Impedance (Resistance) .................................................................. 56
Check DC Balance....................................................................................... 57
Check DC Gain Accuracy............................................................................... 59
Check Offset Accuracy.................................................................................. 63
Ch
eck Analog Bandwidth............................................................................... 66
Check Random Noise, Sample Acquisition Mode ........... ........................................ 69
Check Sample Rate and Delay Time Accuracy....................................................... 71
Check Delta Time Measurement Accuracy ........................................................... 73
Check Digital Threshold Accuracy (MSO4000B Only)............................................. 75
Check Trigger Out ....................................................................................... 79
MSO4000B and DPO4000B Series Specications and Performance Verication i
Table of Contents
List of Tables
Table 1: Analog channel input and vertical specications..................................................... 2
Table 2: Hor
izontal and acquisition system specications..................................................... 9
Table 3: Delta-Time measurement accuracy formula ......................................................... 11
Table 4: Trigger specications ................................................................................... 12
Table 5: Digital acquisition specications, MSO4000B Series .............................................. 18
Table 6: P6616 digital p robe specications..................................................................... 19
Table 7: Display specications................................................................................... 20
Table 8: I
nput/Output port specications ....................................................................... 21
Table9:Datastoragespecications............................................................................. 23
Table 10: Power source specications .......................................................................... 23
Table 11: Environmental specications......................................................................... 24
Table 12: Mechanical specications............................................................................. 25
Table 13: Required equipment ................................................................................... 27
Table
14: Gain expected worksheet ............................................................................. 61
Table 15: Maximum bandwidth frequency worksheet ........................................................ 67
ii MSO4000B and DPO4000B Series Specications and Performance Verication
General safety summary
General safet
y summary
Review the fo
llowing safety precautions to avoid injury and prevent damag e to
this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
To avoid re or personal
injury
Use proper power cord. Use only the power cord specied for this product and
certied for the country of use.
Connect and disconnect properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Connect and disconnect properly. De-energize the circuit under test before
connecting or disconnecting the current probe.
Ground the product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input o r output
terminals o f the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
information before making connections to the product.
Connect the probe reference lead to earth ground only.
Do not apply a potential to any terminal, including the common terminal, that
exceeds the maximum rating of that terminal.
Power disconnect. The power cord disconnects the product from the power source.
Do not block the power cord; it must remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels
removed.
Do not operate with suspected failures. If you suspect that there is damage to this
product, have it inspected by qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide proper ventilation. Refer to the manual's installation instructions for details
on installing the product so it has p roper ventilation.
MSO4000B and DPO4000B Series Specications and Performance Verication iii
General safety summary
Termsinthismanual
These terms may
appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Symbols and terms on the
product
These terms may appear on the product:
DANGER in
dicates an injury hazard immediately accessible as you read
the marking.
WARNING
indicates a n injury hazard not immediatel y accessible as you
read the marking.
CAUTIO
N indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
iv MSO4000B and DPO4000B Series Specications and Performance Verication
Specications
This chapter contains specications for the MSO4000B and the DPO4000B series
oscilloscopes. All specications are guaranteed unless noted as "typical."
Typical spec
ications are provided for your convenience but are not guaranteed.
Specications that are marked with the
symbol are checked in Performance
Verication.
All specications apply to all MSO4000B and DPO4000B models unless noted
otherwise. To meet specications, two conditions must rst be met:
The oscilloscope must have been operating continuously for twenty minutes
within the specied operating temperature range. (See Table 11 on page 24.)
You must perform the Signal Path Compensation (SPC) operation described
instep2oftheSelf Test before evaluating specications. (See page 55, Self
Test.) If the operating temperature changes by more than 1 0 °C (18 °F), you
must perform the SPC operation again.
Model Bandwi
dth
Analog
Channe
ls
Sample
Rate
(1 ch)
Sample
Rate
(2 ch)
Sample
Rate
(4 ch)
Record
Length
(1 ch)
Record
Length
(2 ch)
Record
Length
(4 ch)
MSO40
00B models:
MSO4
104B
1GHz
4
5GS/
s
5GS/
s
5GS/
s
20M 20M 20M
MSO4
104B-L
1GHz
4
5GS/
s
5GS/
s
2.5 G
S/s
5M 5M 5M
MSO
4102B
1GH
z
2
5GS
/s
5GS
/s
20M 20M
MSO
4102B-L
1GH
z
2
5GS
/s
2.5
GS/s
5M 5M
DP
O4000B models:
DP
O4104B
1G
Hz
4
5G
S/s
5G
S/s
5G
S/s
20M 20M 20M
D
PO4104B-L
1
GHz
4
5
GS/s
5
GS/s
2
.5 GS/s
5M 5M 5M
DPO4102B 1 GHz
2
5GS/s 5GS/s
20M 20M
DPO4102B-L 1 GHz
2
5 GS/s 2.5 GS/s
5M 5M
DPO4054B
500 MHz 4
2.5 GS/s 2.5 GS/s 2.5 GS/s
20M 20M 20M
DPO4034B
350 MHz 4
2.5 GS/s 2.5 GS/s 2.5 GS/s
20M 20M 20M
DPO4014B
100 MHz 4
2.5 GS/s 2.5 GS/s 2.5 GS/s
20M 20M 20M
MSO4000B and DPO4000B Series Specications and Performance Verication 1
Specications
Analog Channe
l Input and Vertical Specications
The following table shows the analog channel input and vertical specications for
the MSO4000B Series and the DPO4000B Series oscilloscopes.
Table 1: Ana
log channel input and vertical specications
Characteristic Description
Number of input
channels
MSO/DPO4XX4B, MSO/DPO4XX4B-L: Four analog channels, digitized simultaneously
MSO/DPO4102B, MSO/DPO4102B-L Two analog channels, digitized simultaneously
Input coupling
DC or AC
Input resistance
selection
1M or 50
1M 1M ±1%
50 ±1%
MSO4104B,
DPO410
4B,
MSO4104B-L
DPO4104B-L
MSO41
02B-L
DPO4102B-L
MSO4102B
DPO41
02B
VSWR 1.5:1 from DC to 1 G Hz, typical
DPO4
054B
VSWR
1.5:1 from DC to 500 MHz, typical
DPO4
034B
VSWR
1.5:1 from DC to 350 MHz, typical
Input impedance,
DC coupled
50
DPO
4014B
VSW
R 1.5:1 from DC to 100 MHz, typical
Inp
ut Capacitance,
1M DC coupled
13 pF ± 2 pF
1M 300 V
RMS
at the BNC with peaks ±425 V
Installation Category II
F
or <100 mV/div, derate at 20 dB/decade above 100 kHz to 30 V
RMS
at 1 MHz,
10 dB/decade above 1 MHz
For 100 mV/div, derate at 20 dB/decade above 3 MHz to 30 V
RMS
at 30 MHz,
1
0 dB/decade above 30 MHz
Maximum peak input voltage at the BNC, ±425 V
Ma
ximum input
voltage
50 5V
RMS
with peaks ±20 V (Duty Factor 6.25%)
Overvoltage trip is intended to protect against overloads that might damage
termination resistors. A sufciently large impulse might cause damage
regardless of the overvoltage protection circuitry because of the nite time
required to detect and respond.
DC Balance
0.1 div with the input DC 50 coupled and 50 terminated
0.2 div at 1 mV/div with the input DC 50 coupled and 50 terminated
0.2 div with the input DC 1 M coupled and 50 terminated
2 MSO4000B and DPO4000B Series Specications and Performance Verication
Specications
Table 1: Analog channel input and vertical specications (cont.)
Characteristic Description
Number of digitized
bits
8bits
Displayed vertically with 25 dig itiz ation lev els (DL) per division, 10.24 divisions dynamic range.
"DL" is the abbreviation for "digitization level." A DL is the smallest voltage level change that can be
resolved by an 8-bit A-D Converter. This value is also known as the least signicant bit (LS B).
1M 1 mV/div to 1 0 V/div in a 1-2-5 sequenceSensitivity r ange
(coarse)
50 1 mV/div to 1 V/div in a 1-2-5 sequence
1 mV/div to 5 V/div <–50% to >+50% of selected setting
10 V/div <–50% to 0%
1M
Allows continuous adjustment from 1 mV/div to 10 V/div
1 mV /div to 500 mV/div <–50% to >+50% of selected setting
1 V/div <–50% to 0%
Sensitivity range (ne)
50
Allows continuous adjustment from 1 mV/div to 1 V/div
Sensitivity resolution
(ne), typical
1% of current setting
DC ga
in ac curacy
For 5 0 ,1M, TPP0500, and TPP1000 path:
±1.5%, derated at 0.100%/°C above 30 °C
±2.0%, derated at 0.100% /°C above 30 °C, 1 mV /Div setting
±3.0% variable gain, derated at 0.100%/°C above 30 °C
Offset range
Volts/div setting
1M input 50 input
1 mV /div to 50 mV/div
±1 V ±1 V
50.5 mV/div to 99.5 mV/div
±0.5 V ±0.5 V
100 mV/div to 500 mV/div
±10 V ±10 V
505 mV/div to 995 mV/div
±5 V ±5 V
1 V/div to 5 V/div
±100 V ±5 V
5.05 V/div to 10 V/div
±50 V Not applicable
Offset ranges,
minimum
For 50 path, 1 V/div is the m aximum vertical setting.
The i nput signal cannot exceed Max Input Voltage for the 50 input path. Refer to the Max Input Voltage
specication for more information.
Position range ±5 divisions
Offset accuracy
±[0.005 × | offset position | + DC Balance]
Both the position and constant offset term must be converted to volts by multiplying by the appropriate
volts/div term.
MSO4000B and DPO4000B Series Specications and Performance Verication 3
Specications
Table 1: Analog channel input and vertical specications (cont.)
Characteristic Description
Number of Waveforms
for Average
Acquisition Mode
2to512waveforms
Default of 16 waveforms
Measurement type
DC Ac curacy (in Volts)
Average of 16 waveforms ±[DC Gain Accuracy × | reading (offset -
position) | + offset accuracy + 0.1 division ]
Refer to D C Gain Accuracy for temperature
derating information.
Delta Volts between any two averages of 16
waveforms acquired with the same oscilloscope
setup and ambient conditions
±[DC gain accuracy × | reading | + 0.05 div]
Refer to D C Gain Accuracy for temperature
derating information.
DC voltage
measurement
accuracy
Average acquisition
mode
Offset, position, and the constant offset term must be converted to volts by multiplying by the appropriate
volts/div term.
The basic accuracy specication applies directly to any sample and to the following measurements: High,
Low, Max, Min, M ean, Cycle Mean, RMS, and Cycle RMS. The delta volt accuracy specication applies to
subtractive calculations involving two of these m easurements.
The delta volts (difference voltage) accuracy specication applies directly to the following measurements:
Positive Overshoot, Negative O ve rshoot, Pk-Pk, and Amplitude.
Measurement type
DC Accuracy (in volts)
Any sample
±[DC gain accuracy × | reading (offset
position) | + Offset Accuracy + 0.15 div + 0.6 mV]
Refer to D C Gain Accuracy for temperature
derating information.
Delta volts between any two samples acquired
with the same oscilloscope setup and ambient
conditions
±[DC gain accuracy × | reading | + 0.15 div +
1.2 mV]
Refer to D C Gain Accuracy for temperature
derating information.
DC voltage
measurement
accuracy
Sample acquisition
mode, typical
Offset, position, and the constant offset term must be converted to volts by multiplying by the appropriate
volts/div term.
Analog bandwidth
selections
DPO4104X, DPO4054B, DPO4034B: 20 MHz, 250 MHz, and Full
DPO4014B: 20 MHz and Full
4 MSO4000B and DPO4000B Series Specications and Performance Verication
Specications
Table 1: Analog channel input and vertical specications (cont.)
Characteristic Description
These limits are for ambient temperature of 30°C and the bandwidth selection set to FULL. Reduce the
upper bandwidth frequency by 1% for each °C above 30°C
Instrument
Volts/Div setting
Bandwidth
5 mV/div 1 V/div DC to 1.00 GHz
2 m V/div 4.98 mV/div DC to 350 MHz
MSO4104B,
DPO4104B,
MSO4104B-L,
DPO4104B-L,
MSO4102B-L,
DPO4102B-L,
MSO4102B,
DPO4102B
1 m V/div 1.99 mV/div DC to 175 MHz
5 mV/div 1 V/div DC to 500 MH z
2 m V/div 4.98 mV/div DC to 350 MHz
DPO4054B
1 m V/div 1.99 mV/div DC to 175 MHz
2 mV/div 1 V/div DC to 350 MH zDPO4034B
1 m V/div 1.99 mV/div DC to 175 MHz
Analog bandw
idth,
50 , DC coupled
DPO4010B 1 mV/div 1 V/div DC to 100 MHz
These limits are for ambient temperature of 30°C and the bandwidth selection set to FULL. Reduce the
upper bandwidth frequency by 1% for each °C above 30°C
Instrument
Volts/Div setting
Bandwidth
5 m V/div 10 V/div DC to 500 MHz
2 m V/div 4.98 mV/div DC to 350 MHz
MSO4104B,
DPO4104B,
MSO4104B-L,
DPO4104B-L,
MSO4102B-L,
DPO4102B-L,
MSO4102B,
DPO4102B
1 m V/div 1.99 mV/div DC to 175 MHz
5 m V/div 10 V/div DC to 500 MHz
2 m V/div 4.98 mV/div DC to 350 MHz
MSO4054B,
DPO4054B
1 m V/div 1.99 mV/div DC to 175 MHz
DPO4034B 5 m V/div 10 V/div DC to 350 MHz
2 m V/div 4.98 mV/div DC to 175 MHz
1 m V/div 1.99 mV/div DC to 175 MHz
DPO4014B 1 m V/div 1.99 mV/div DC to 100 MHz
Analog bandwidth,
1M, DC coupled,
typical
The D PO 4104B, MSO4104B-L, DPO4104B-L, MSO4102B-L, DPO4102B-L, MSO4102B, and DP O4102B
models have separate 50 and 1 M analog paths, so the 1 M path is not checked when the 50 path
is checked in the performance test procedure. All other models have only one analog path, so they are
fully tested with the 50 termination.
MSO4000B and DPO4000B Series Specications and Performance Verication 5
Specications
Table 1: Analog channel input and vertical specications (cont.)
Characteristic Description
These l imits are for ambient temperature of 30°C and the bandwidth selection set to FULL. Reduce the
upper bandwidth frequency by 1% for each °C above 30°C
Instrument
Volts/Div setting
Bandwidth
50 mV/div 100 V/div DC to 1 GHz (TPP1000
probe)
DC to 500 MHz
(TPP0500 probe)
20 mV/div 49.8 mV/div DC to 350 MHz
MSO4104B,
DPO4104B,
MSO4104B-L,
DPO4104B-L,
MSO4102B-L,
DPO4102B-L,
MSO4102B,
DPO4102B
10 mV/div 19.9 mV/div DC to 175 MHz
50 mV/div 100 V/div DC to 500 MHz
20 mV/div 49.8 mV/div DC to 350 MHz
DPO4054B
10 mV/div 19.9 mV/div DC to 175 MHz
50 mV/div 100 V/div DC to 350 MHz
20 mV/div 49.8 mV/div DC to 350 MHz
DPO4034B
10 mV/div 19.9 mV/div DC to 175 MHz
Analog bandwidth with
TPP0500 or TP P1000
probe, typical
DPO4014B 10 mV/div 100 V/div DC to 100 MHz
Lower frequency limit,
AC coupled, typical
< 10 Hz when AC, 1 M coupled
The AC coupled lower frequency limits are reduced by a factor of 10 when 10X passive probes are used.
Upper frequency limit,
250 MHz bandwidth
limited, typical
250 MHz, ±20%
Not available on the DPO4014B
Upper frequency limit,
20 MHz bandwidth
limited, typical
20 MHz, ±20%
6 MSO4000B and DPO4000B Series Specications and Performance Verication
Specications
Table 1: Analog channel input and vertical specications (cont.)
Characteristic Description
The formula is calculated by measuring -3 dB bandwidth of the oscilloscope. The formula accounts for the
rise time contribution of the oscilloscope independent of the rise time of the signal source.
Instrument
50
1 mV/div to
1.99 mV/div
50
2 mV/div to
4.99 mV/div
50
5 mV/div to 1 V /div
MSO4104B,
DPO4104B,
MSO4104B-L,
DPO4104B-L,
MSO4102B-L,
DPO4102B-L,
MSO4102B,
DPO4102B
2 ns 1 ns 350 ps
DPO4054B
2 ns 1 ns 700 ps
DPO4034B
2ns 1ns 1ns
DPO4014B
3.5 ns 3.5 ns 3.5 ns
Instrument TPP1000 probe
10 mV/div to
19.9 mV/div
TPP1000 probe
20 mV/div to
49.8 mV/div
TPP1000 probe
50 mV/div to 10 V/div
MSO4104B,
DPO4104B,
MSO4104B-L,
DPO4104B-L,
MSO4102B-L,
DPO4102B-L,
MSO4102B,
DPO4102B
2 ns 1 ns 350 ps
DPO4054B
2 ns 1 ns 700 ps
DPO4034B
2ns 1ns 1ns
Calculated rise time at
0.350/BW = t
r
,typical
DPO4014B
3.5 ns 3.5 ns 3.5 ns
Instrument TPP0500 probe
10 mV/div to
19.9 mV/div
TPP0500 probe
20 mV/div to
49.8 mV/div
TPP0500 probe
50 mV/div to 10 V/div
MSO4104B,
DPO4104B,
MSO4104B-L,
DPO4104B-L,
MSO4102B-L,
DPO4102B-L,
MSO4102B,
DPO4102B
2 ns 1 ns 700 ps
DPO4054B
2 ns 1 ns 700 ps
DPO4034B
2ns 1ns 1ns
Calculated rise time at
0.350/BW = t
r
,typical
DPO4014B
3.5 ns 3.5 ns 3.5 ns
MSO4000B and DPO4000B Series Specications and Performance Verication 7
Specications
Table 1: Analog channel input and vertical specications (cont.)
Characteristic Description
Peak Detect or
Envelope mode pulse
response, typical
Instrument (Sample rate maximum)
Minimum pulse width
MSO4104B, DPO4104B, , MSO4102B, DPO4102B
MSO4104B-L DPO4104B-L, ( 2 channels
enabled)
MSO4102B-L, DPO4102B-L, (1 channel enabled)
>800 ps (5 GS/s)
DPO4054B, DPO4034B, DPO4014B
DPO4104B-L , ( 3 channels enabled)
MSO/DPO4102B-L, (2 channels enabled)
>1.6 ns (2.5 GS /s)
RMS noise (mV)
Instrument Bandwidth limit
1M 50
Full bandwidth
(300 V + 8.0% of
Volts/div setting)
(75 V + 6.0% of
Volts/div setting)
250 MHz bandwidth
(100 V + 5.0% of
Volts/div setting)
(50 V + 4.0% of
Volts/div setting)
MSO4104B,
DPO4104B,
MSO4104B-L,
DPO4104B-L,
MSO4102B-L,
DPO4102B-L,
MSO4102B,
DPO4102B
20 MHz bandwidth
(100 V + 5.0% of
Volts/div setting)
(50 V + 4.0% of
Volts/div setting)
Full bandwidth
(130 V + 8.0% of
Volts/div setting)
(130 V + 8.0% of
Volts/div setting)
250 MHz bandwidth
(100 V + 6.0% of
Volts/div setting)
(100 V + 6.0% of
Volts/div setting)
DPO4054B
20 MHz bandwidth
(100 V + 4.0% of
Volts/div setting)
(100 V + 4.0% of
Volts/div setting)
Full bandwidth
(130 V + 8.0% of
Volts/div setting)
(130 V + 8.0% of
Volts/div setting)
250 MHz bandwidth
(100 V + 6.0% of
Volts/div setting)
(100 V + 6.0% of
Volts/div setting)
DPO4034B
20 MHz bandwidth
(100 V + 4.0% of
Volts/div setting)
(100 V + 4.0% of
Volts/div setting)
DPO4014B
Full bandwidth
(130 V + 8.0% of
Volts/div setting)
(130 V + 8.0% of
Volts/div setting)
Random N
oise,
Sample Acquisition
Mode
20 MHz bandwidth
(100 V + 4.0% of
Volts/div setting)
(100 V + 4.0% of
Volts/div setting)
Delay between
channels, full
bandwidth, typical
100 ps between any two channels with input impedance set to 50 , DC coupling, with equal volts/division
setting or above 10 m V/div
All settings in the instrument can be m anually time aligned using the Probe Deskew function from –125 ns
to +125 ns with a resolution of 20 ps
Deskew range
–125 ns to +125 ns with a resolution of 20 ps
8 MSO4000B and DPO4000B Series Specications and Performance Verication
Specications
Table 1: Analog channel input and vertical specications (cont.)
Characteristic Description
Crosstalk (channel
isolation), typical
100:1 at 100 MHz and 30:1 at >100 MHz u p to the rated bandwidth for any two channels having
equal Volts/Div settings
TekVPI Interface The probe interface allows installing, powering, compensating, and controlling a wide range of probes
offering a variety of features.
The interface is available on all front panel inputs including Aux In. Aux In only provides 1 M input
impedance and does not offer 50 as the other input channels do.
Total probe power
Five Tek VPI-compliant probe interfaces (one per channel)
50 W maximum internal probe power
Voltage Max Amperage Voltage Tolerance
5V
50 mA (250 mW) ±5%
Probe power per
channel
12 V
2 A (24 W) ±10%
Horizontal and Acquisition System Specications
The following table shows the horizontal and acquisitions system specications
for the MSO4000B Series and the DPO4000B Series oscilloscopes.
Table 2: Horizontal and acquisition system specications
Characteristic Description
MSO4104B, DPO4104B,
MSO4104B-L, DPO4104B-L,
MSO4102B-L, DPO4102B-L,
MSO4102B, DPO4102B
5 GS/s 2.5 S/sSample-rate range
DPO4054B,
DPO4034B
DPO4014B
2.5 GS/s —2.5 S/s
Record Length Range Non-L models: 20 M, 10 M, 5 M, 1 M, 100 k, 10 k, 1 k
L m odels: 5 M, 1 M, 100 k, 10 k, 1 k
Instrument 1 k 10 k 100 k 20 M
MSO4104B, DPO4104B,
MSO4102B-L, DPO4102B-L,
MSO4102B, DPO4102B
MSO4104B-L, DPO4104B-L (2
channels enabled)
400 ps 40 s 400 ps
400 s
400 p s 1,000 s
Seconds/Division range
DPO4054B,
DPO4034B
DPO4014B
MSO4104B-L, DPO4104B-L (4
channels enabled)
1 ns 40 s 1 ns 400 s 1 ns 1,000 s
Maximum triggered acquisition rate
> 50,000 wfm/s
MSO4000B and DPO4000B Series Specications and Performance Verication 9
Specications
Table 2: Horizontal and acquisition system specications (cont.)
Characteristic Description
Aperture Uncertainty
(3 ps + 0.1 ppm × record duration)
RMS
, for records having 1 minute duration
Long-term s
ample rate and
delay time accuracy
±5 ppm over any 1mstimeinterval
10 MSO4000B and DPO4000B Series Specications and Performance Verication
Specications
Table 2: Horizontal and acquisition system specications (cont.)
Characteristic Description
The formula to calculate the delta-time measurement accuracy (DTA) for a given instrument
setting and input signal is given in the following table. (See Table 3.) The formula assumes
insignicant signal content above Nyquist and insignicant error due to aliasing. The
abbreviations used in the formula are as follows:
SR
1
= slew rate around 1st point in measurement (1
st
edge)
SR
2
= slew rate around 2nd point in measurement (2
nd
edge)
N =input-referred noise (V
RMS
)(RefertoRandom Noise and Sample Acquisition Mode
specications.)
TBA = time base accuracy (5 ppm) (Refer to Long-term Sample Rate and Delay Time
Accuracy specications.)
t
p
= delta-time measurement duration (sec)
RD = (record length)/(sample rate)
t
sr
= 1/(sample rate)
assume edge shape that results from Gaussian lter response
Delta-time m
easurement
accuracy
The term under the square-root sign is the stability and is due to TIE (Time Interval Error). The
errors due to this term occur throughout a single-shot measurement. The second term is due
to both the absolute center-frequency accuracy and the center-frequency stability of the time
base and varies between multiple single-shot measurements over the observation interval (the
amount of time from the rst single-shot measurement to the nal single-shot measurement).
Table
3: Delta-Time measurement accuracy formula
The terms used in these formulas are dened under Delta-time measurement accuracy, in the preceding table. (See Table 2.)
MSO4000B and DPO4000B Series Specications and Performance Verication 11
Specications
Trigger Speci
cations
The following table shows the trigger specications for the MSO4000B Series
and the DPO4000B Series oscilloscopes.
Table 4: Tri
gger specications
Characteristic Description
Aux In (External) trigger input
impedance, typical
1M ±1%inparallelwith13p2pF
Aux In (External) trigger maximum
input vol
tage
The maximum input voltage at the BNC 300 V
RMS
Installa
tion Category II
Derate at 20 dB/decade above 3 MHz to 30 V
RMS
at 30 MHz, 10 dB/decade above 30 M Hz.
Aux In (External) trigger bandwidth,
typical
250 MHz ±20%
MSO4104B, DPO4104B,
MSO4104
B-L,
DPO4104B-L,
MSO4102B-L,
DPO410
2B-L, MSO4102B,
DPO4102B
1GHz
DPO4054B
500 MH
z
DPO4034B
350 MH
z
Trigger bandwidth, Edge, typical
DPO4014B
100 M
Hz
MSO4104B, DPO4104B,
MSO4104B-L,
DPO
4104B-L,
MSO4102B-L,
DPO4102B-L, MSO4102B,
DPO
4102B
1GHz
DP
O4054B
500 MHz
D
PO4034B
350 MHz
Trig
ger bandwidth, Pulse and Logic,
typical
D
PO4014B
100 MHz
12 MSO4000B and DPO4000B Series Specications and Performance Verication
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87

Tektronix DPO4104B Technical Reference

Type
Technical Reference

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI