Azbil F4H User manual

Type
User manual

This manual is also suitable for

Thank you for purchasing your Azbil
Corporation product.
This manual contains information for
ensuring the safe and correct use of
the product.
Those designing or maintaining
equipment that uses this product
should first read and understand
this manual. This manual contains
information not only for installation,
but also for maintenance,
troubleshooting, etc. Be sure to keep
it nearby for handy reference.
(Not for use in Japan) No. CP-SP-1405E
Compact Digital Mass Flow Controller
Model F4H9050/9200/9500/0002/0005/0020
User's Manual
for Installation and Configuration
NOTICE
Please make sure that this manual is available to the user of the
product.
Unauthorized duplication of this user’s manual in part or in whole
is forbidden. The information and specifications in this manual are
subject to change without notice.
Considerable effort has been made to ensure that this manual is
complete and accurate, but if you should find an omission or error,
please contact us.
In no event is Azbil Corporation liable to anyone for any indirect,
special, or consequential damages as a result of using this product.
Please read “Terms and Conditions” from the following URL before ordering and use.
https://www.azbil.com/products/factory/order.html
© 2017–2023 Azbil Corporation. All Rights Reserved.
Modbus™ is a trademark and the property of Schneider Electric SE,
its subsidiaries and affiliated companies.
i
Conventions Used in This Manual
The safety precautions explained in the following section aim to prevent injury to the operator
and others, and to prevent property damage.
WARNING
Warnings are indicated when mishandling this
product might result in death or serious injury.
CAUTION
Cautions are indicated when mishandling this
product might result in minor injury to the user, or
only physical damage to the product.
In describing the product, this manual uses the icons and conventions listed below.
Use caution when handling the product.
The indicated action is prohibited.
Always follow the indicated instructions.
Handling Precautions
Handling Precautions indicate items that the user should pay attention to
when handling the F4H 9050/9200/9500/0002/0005/0020.
Note Notes indicate information that might benefit the user.
This indicates the item or page that the user is requested to refer to.
(1), (2), (3): Numbers within parentheses indicate steps in a sequence or parts of an
explanation.
ii
Safety Precautions
WARNING
Never allow a gas that is within explosive limits to pass through this device.
Otherwise, an explosion may result.
To control the flow of oxygen, be sure to use the oxygen model.
Do not use the oxygen model for gas that contains oil. If this should happen,
never use the flow controller again for oxygen. The oil on parts that come into
contact with the gas may catch fire.
Do not use this device in locations containing explosive gases or near flam-
mable liquids or gases.
CAUTION
Prevent foreign matter from entering the flow path of this device.
If rust, water droplets, oil mist, or dust from the pipes enters the device, a mea-
surement or control error may occur, or the device may be damaged.
If there is a possibility of foreign matter entering the device, install an upstream
filter, strainer, or mist trap capable of eliminating foreign matter 0.1 µm and
greater in diameter. Be sure to inspect and replace the filter at regular intervals.
Do not allow lint, metal shavings, water, etc., to enter the case of the device.
Otherwise, a device malfunction or failure may result.
Do not subject the product to pressure beyond the rated pressure resistance range.
Otherwise, damage to the product may result.
Be sure to use this device within the flow-rate range stipulated in the product
specifications. To prevent excessive flow, use a suitable means to control the
supply pressure or use a throttle valve or the like to control the flow rate. If the
flow rate exceeds the upper limit, both the flow rate display and the output
voltage/current may indicate lower values than the actual flow rate.
If damage could result from the abnormal functioning of this device, include ap-
propriate redundancy in the system design.
When using this device for air-fuel ratio control of a burner, prevent the occur-
rence of flashback and also take countermeasures for the instrumentation to
protect the device even if flashback occurs.
A fire or an increase in pressure in the pipes caused by flashback from the
burner will cause device failure.
Use this device within the operating differential pressure range. Failure to do so
may result in hunting.
Continuous hunting causes valve failure. It also may prevent the controlled flow
rate from reaching the preset value.
iii
CAUTION
When controlling oxygen flow, use the oxygen model and follow the precau-
tions below:
The piping work should be carried out by a specialist with expertise in han-
dling oxygen.
• Use degreased pipes and parts.
Be sure to remove foreign matter, burrs, etc., from the pipes before connecting
this device.
Do not drop this device or subject it to impact.
Otherwise, damage to this precision instrument may result.
Mount this device securely so that it will not vibrate.
Otherwise, a malfunction or failure may result.
When mounting this device, make sure that the top panel does not face down-
ward. If the top panel faces downward, device failure may result.
When connecting the device to the pipes, hold the hexagonal part of the fitting
in place and turn the pipe.
After connecting the piping, check for gas leaks.
If using Rc connections, exercise care not to use an excessive amount of sealant.
Foreign matter or burrs in the pipes may result in measurement errors.
Before connecting pipes with Swagelok or VCR fittings, check the instructions in
the manual provided by the fitting manufacturer.
Ensure that the wiring is correctly installed before turning the power on.
Incorrect wiring may result in damage or malfunction.
Do not apply excessive force to the connector or cables while the connector
cable or AC adapter is connected.
Otherwise, damage to the connector or circuit board may result.
When disposing of this device, please observe local regulations.
Do not connect equipment that causes a large throttling or pressure loss near
the piping that is downstream of this device. Otherwise, hunting may result.
When storing and transferring the device, place it in a plastic bag, etc., to pre-
vent foreign matter from entering the flow path.
If the device is installed in an environment subject to large temperature
changes, introduce a flow of sufficiently dry gas through the pipes to remove
any remaining wet gas in order to prevent water from condensing.
Condensation may result in device malfunction.
Do not blow air through the device or wipe the inside of the flow path with a
rag.
iv
CAUTION
Do not wash the device or steam-clean the inside of the flow path.
Otherwise, a sensor malfunction may result.
Do not apply a negative voltage or a voltage exceeding 5 V to the flow rate
setup input (+) terminal.
Otherwise, a device malfunction or failure may result.
Always use the specified fittings and gaskets. After the piping work has been
completed, check that there are no gas leaks before operating the device.
Failure to do so may result in gas leakage.
The valve on this device cannot completely stop a flow.
If complete shutoff is required, provide a separate shutoff valve.
In addition, if an external shutoff valve is closed, fully close the valve of this de-
vice using either one of the following methods while it is standing by:
• Set the flow rate to zero.
• Set the operation mode to “valve fully closed.
Even though the external shutoff valve is closed (flow rate is zero), leaving the
device in control mode will cause a momentary excessive flow rate when the
external shutoff valve is opened. Also, in control mode or with the valve forced
fully open, if the external shutoff valve is kept closed, the valve overheating
prevention function will be activated and the electrical current driving the
valve will be forcibly limited.
If there is a risk of a lightning surge, use a surge absorber (surge protector).
Failure to do so may result in a fire or device failure.
Do not turn off the power when settings are being written to EEPROM via
RS-485 communication, when settings are being changed using the PC loader,
or when the zero point adjustment switch is being used. Doing so might cause
a problem.
v
The Role of This Manual
A total of 4 different manuals are available for the Compact Digital Mass Flow Controller (Model
No.: F4H). Read them as necessary for your specific requirements. If a manual you require is not
available, contact the azbil Group or its dealer.
Compact Digital Mass Flow Controller Model F4H9050/9200/9500/
0002/0005/0020 User’s Manual for Installation and Configuration
Manual No. CP-SP-1405E
This manual.
This manual describes the hardware and all function of this unit.
Personnel in charge of design and/or manufacture of a system using
this unit should read this manual thoroughly.
This manual covers mounting, connections for wiring, all functions and
settings of this unit, operating procedures, troubleshooting, and de-
tailed specifications.
Smart Loader Package Model MLP300A000 for Model F4H-
Series Compact Digital Mass Flow Controller User’s Manual
Manual No. CP-SP-1415E
Running the Smart Loader Package on a personal computer en-
ables you to set the parameters of the F4H on the PC.
This manual describes operations on the personal computer.
Compact Digital Mass Flow Controller Model F4H9050/9200/9500
/0002/0005/0020 User’s Manual
Manual No. CP-UM-5905JECK
This manual is supplied with the product.
Personnel in charge of design and/or manufacture of a system
using this unit should read this manual thoroughly.
This manual covers safety precautions, mounting, wiring,
and main specifications. For more detailed information, see
CP-SP-1405E.
Compact Digital Mass Flow Controller Model F4H Series
User’s Manual for RS-485 for Communication Functions
Manual No. CP-SP-1408E
Those using the RS-485 communication functions should read
this manual.
This manual gives an overview of communications, wiring,
communication procedures, a list of F4H communication data,
instructions on correcting problems, and communications
specifications.
vi
Copyrights, licenses, and registered trademarks
- Notice regarding the use of software licensed from ARM
This product includes software licensed from ARM Limited. The following information describes the conditions of the
above-mentioned license and is not for the purpose of restricting your use of the product, etc.
Copyright © 2009–2015 ARM LIMITED All rights reserved.
Redistribution and use in source and binary forms, with or without modification, are permitted provided that the follow-
ing conditions are met:
- Redistributions of source code must retain the above copyright notice, this list of conditions and the following
disclaimer.
- Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following dis-
claimer in the documentation and/or other materials provided with the distribution.
- Neither the name of ARM nor the names of its contributors may be used to endorse or promote products derived from
this software without specific prior written permission.
THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS AS IS” AND ANY EXPRESS OR IMPLIED
WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL COPYRIGHT HOLDERS AND CONTRIBUTORS BE LIABLE FOR
ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED
TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS
INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR
TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF
ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
- Notice regarding the use of software licensed from STMicroelectronics
This product includes software licensed from STMicroelectronics. The following information describes the conditions
of the above-mentioned license and is not for the purpose of restricting your use of the product, etc.
COPYRIGHT© 2014 STMicroelectronics
Redistribution and use in source and binary forms, with or without modification, are permitted provided that the follow-
ing conditions are met:
1. Redistributions of source code must retain the above copyright notice, this list of conditions and the following
disclaimer.
2. Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following dis-
claimer in the documentation and/or other materials provided with the distribution.
3. Neither the name of STMicroelectronics nor the names of its contributors may be used to endorse or promote products
derived from this software without specific prior written permission.
THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS AS IS” AND ANY EXPRESS OR IMPLIED
WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A
PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT HOLDER OR CONTRIBUTORS BE LIABLE FOR
ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL
DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE,
DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN
CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE
OF THIS SOFTWARE, EVEN IF ADVISED OF THE POSSIBILITY OF SUCH DAMAGE.
vii
Contents
Conventions Used in This Manual
Safety Precautions
The Role of This Manual
Copyrights, licenses, and registered trademarks
Chapter 1. OVERVIEW 1-1
Overview 1-1
Features 1-1
Model selection guide 1-3
Chapter 2. NAMES AND FUNCTIONS OF EACH PART 2-1
Top panel 2-1
Main unit 2-1
Chapter 3. MOUNTING AND WIRING 3-1
3-1 Mounting 3-4
Installation location 3-4
Installation method 3-4
Piping 3-6
3-2 Wiring 3-7
Connector for external connection 3-7
Communications connector (RS-485) 3-8
Wiring 3-8
Wiring Examples 3-9
Chapter 4. BASIC SETTING 4-1
Operation Mode 4-1
Function for controlling the flow rate (SP) 4-1
Flow rate setup by analog signal 4-1
Flow rate setup by communication 4-2
About control range and setup range 4-2
Regarding operation outside the control range 4-2
Selection function for valve operation 4-3
Gas type selection function 4-3
LED indication 4-4
Communication functions 4-4
Chapter 5. DETAILED SETTING 5-1
5-1 Function Code Settings 5-1
Function code settings 5-1
viii
5-2 Parameter Settings 5-9
Parameter code settings 5-9
5-3 Other Detailed Setting 5-13
SP ramp control function 5-13
Zero point adjustment (when hydrogen/He gas is used) 5-14
Multipoint flow rate correction 5-15
Flow rate range change function 5-16
Resolutions after changing range 5-17
Functions related to maintenance 5-31
Chapter 6. TROUBLESHOOTING 6-1
Operation at alarm/event occurrence 6-1
Other problems 6-4
Chapter 7. Specifications 7-1
F4H9050/9200/9500/0002/0005/0020 7-1
Applied voltage and maximum differential pressure (air) for F4H0020 7-4
Korean KC mark 7-5
Relationship between differential pressure and flow rate when
the valve is fully open (for air) 7-6
Optional parts (sold separately) 7-7
External Dimensions 7-8
1-1
Chapter 1.
OVERVIEW
Overview
This device is a high-performance compact digital mass flow control-
ler with advanced functions for the general industrial market. It fea-
tures high speed and wide range flow rate control.
This product incorporates a Micro Flow sensor, which is an ex-
tremely high-speed response flow speed sensor developed by Azbil
Corporation, a proportional solenoid valve, and advanced actuator
control technology in order to achieve high-speed control.
Cables with dedicated
connectors (optional parts)
This device
Controller
(e.g. SDC25/26)
24 V DC power supply
RS-485 device
Device Conguration
Features
High-speed controllability
A high-speed response of 0.3 seconds* is realized.
* When control begins from the fully closed state or when the set-
ting is changed during control, the controlled flow rate reaches
the set value ± 2 % within 0.3 s.
Wide control range
Provides a wide control range of 1 to 100 % full scale (FS).
(For F4H9050, 2 to 100 % FS)
1-2
Chapter 1. OVERVIEW
Ease of use
Product runs on 24 V DC from a single general-purpose power
supply. In addition, the power supply circuit and the input/output
circuit in this device are isolated.
When multiple F4Hs are driven through analog input/output using
a programmable logic controller (PLC) or the like, even if the analog
module channels on the PLC side are not isolated, a common power
supply can be used to power this device. Even without using an in-
dividual power supply for each device, there is no need for concern
regarding effects from one circuit on adjacent ones.
In addition, for the case of simple use in a laboratory, a convenient
AC adapter (optional) is available.
1-3
Chapter 1. OVERVIEW
Model selection guide
zSUS316 Air/nitrogen model
Basic model No.
Standard flow rate range
Flow path material
Piping connection method
Gas type
Communication type
O-ring material
Default gas type setting
Option 1
Option 2
Option 3
Appended No.
Remarks
F4H Compact Digital Mass Flow
Controller
9050 1.00 to 50.00 [mL/min]*
9200 2.0 to 200.0 [mL/min]*
9500 5.0 to 500.0 [mL/min]*
0002 0.020 to 2.000 [L/min]*
0005 0.050 to 5.000 [L/min]*
0020 0.20 to 20.00 [L/min]*
6SUS316 (with degreasing of
gas-contacting parts)
UUNF (female thread is cut
into the main unit)
T Rc fitting
SSwagelok fitting or
equivalent
V VCR fitting or equivalent
N Air/nitrogen
2 RS-485 CPL model
3 RS-485 Modbus RTU model
0 Fluoroelastomer
N Factory setting: Air/nitrogen
0 No optional functions
0 No optional functions
0 No optional functions
D Inspection certificate
YInspection certificate +
traceability certification
0 Product version
* Flow rate range for air, nitrogen, and argon.
1-4
Chapter 1. OVERVIEW
zSUS316 Oxygen model
Basic model No.
Standard flow rate range
Flow path material
Piping connection method
Gas type
Communication type
O-ring material
Default gas type setting
Option 1
Option 2
Option 3
Appended No.
Remarks
F4H Compact Digital Mass Flow
Controller
9050 1.00 to 50.00 [mL/min]*1
9200 2.0 to 200.0 [mL/min]*1
9500 5.0 to 500.0 [mL/min]*1
0002 0.020 to 2.000 [L/min]*1
0005 0.050 to 5.000 [L/min]*1
0020 0.20 to 20.00 [L/min]*1
6SUS316 (with degreasing of
gas-contacting parts)
UUNF (female thread is cut
into the main unit)
T Rc fitting
SSwagelok fitting or
equivalent
V VCR fitting or equivalent
S Oxygen
2 RS-485 CPL model
3 RS-485 Modbus RTU model
0 Fluoroelastomer
S Factory setting: oxygen
0 No optional functions
0 No optional functions
0 No optional functions
D Inspection certificate
YInspection certificate +
Traceability certification
0 Product version
* Flow rate range for air, nitrogen, argon, and oxygen.
2-1
Chapter 2.
NAMES AND FUNCTIONS OF EACH PART
Top panel
NET indicator
Loader jack
Rotary switch SW1
(for communication parameters)
Rotary switch SW2
(for communication address)
Rotary switch SW3
(for communication address)
Connector for external connection:
D-sub 9-pin male
PWR indicator
Communications
connectors (RJ45)
BRATE ×10 ×1
NETPWR
F4H
Main unit
Top panel
Jack for
AC adapter
Zero point adjustment switch
(when hydrogen/helium gas is used):
when the F4H valve is closed,
the currently detected sensor signal
can be stored as zero.
Outlet pipe connection port:
gas ow outlet
Inlet pipe connection port:
gas ow inlet
To operate, press for 5 s to adjust the
zero point.
Press for 20 s to restore the
factory-set value.
Memo
3-1
Chapter 3.
MOUNTING AND WIRING
WARNING
Never allow a gas that is within explosive limits to pass through this device.
Otherwise, an explosion may result.
To control the flow of oxygen, be sure to use the oxygen model.
Do not use the oxygen model for gas that contains oil. If this should happen,
never use the flow controller again for oxygen. The oil on parts that come into
contact with the gas may catch fire.
Do not use this device in locations containing explosive gases or near flam-
mable liquids or gases.
CAUTION
Prevent foreign matter from entering the flow path of this device.
If rust, water droplets, oil mist, or dust from the pipes enters the device, a mea-
surement or control error may occur, or the device may be damaged.
If there is a possibility of foreign matter entering the device, install an up-
stream filter, strainer, or mist trap capable of eliminating foreign matter 0.1
µm and greater in diameter. Be sure to inspect and replace the filter at regular
intervals.
Do not allow lint, metal shavings, water, etc., to enter the case of the device.
Otherwise, a device malfunction or failure may result.
Do not subject the product to pressure beyond the rated pressure resistance
range.
Otherwise, damage to the product may result.
Be sure to use this device within the flow-rate range stipulated in the product
specifications. To prevent excessive flow, use a suitable means to control the
supply pressure or use a throttle valve or the like to control the flow rate. If the
flow rate exceeds the upper limit, both the flow rate display and the output
voltage/current may indicate lower values than the actual flow rate.
If damage could result from the abnormal functioning of this device, include
appropriate redundancy in the system design.
When using this device for air-fuel ratio control of a burner, prevent the occur-
rence of flashback and also take countermeasures for the instrumentation to
protect the device even if flashback occurs.
A fire or an increase in pressure in the pipes caused by flashback from the
burner will cause device failure.
Use this device within the operating differential pressure range. Failure to do
so may result in hunting.
Continuous hunting causes valve failure. It also may prevent the controlled
flow rate from reaching the preset flow rate.
3-2
Chapter 3. MOUNTING AND WIRING
CAUTION
When controlling oxygen flow, use the oxygen model and follow the precau-
tions below:
The piping work should be carried out by a specialist with expertise in han-
dling oxygen.
• Use degreased pipes and parts.
Be sure to remove foreign matter, burrs, etc., from the pipes before connect-
ing this device.
Do not drop this device or subject it to impact.
Otherwise, damage to this precision instrument may result.
Mount this device securely so that it will not vibrate.
Otherwise, a device malfunction or failure may result.
When mounting this device, make sure that the top panel does not face down-
ward. If the top panel faces downward, device failure may result.
When connecting the device to the pipes, hold the hexagonal part of the fitting
in place and turn the pipe.
After connecting the piping, check for gas leaks.
If using Rc connections, exercise care not to use an excessive amount of seal-
ant. Foreign matter or burrs in the pipes may result in measurement errors.
Before connecting pipes with Swagelok or VCR fittings, check the instructions
in the manual provided by the fitting manufacturer.
Ensure that the wiring is correctly installed before turning the power on.
Incorrect wiring may result in damage or malfunction.
Do not apply excessive force to the connector or cables while the connector
cable or AC adapter is connected.
Otherwise, damage to the connector or circuit board may result.
When disposing of this device, please observe local regulations.
Do not connect equipment that causes a large throttling or pressure loss near
the piping that is downstream of this device.
Otherwise, hunting may result.
When storing and transferring the device, place it in a plastic bag, etc., to pre-
vent foreign matter from entering the flow path.
If the device is installed in an environment subject to large temperature
changes, introduce a flow of sufficiently dry gas through the pipes to remove
any remaining wet gas in order to prevent water from condensing.
Condensation may result in device malfunction.
3-3
Chapter 3. MOUNTING AND WIRING
CAUTION
Do not blow air through the device or wipe the inside of the flow path with a
rag.
Do not wash the device or steam-clean the inside of the flow path.
Otherwise, a sensor malfunction may result.
Do not apply a negative voltage or a voltage exceeding 5 V to the flow rate
setup input (+) terminal.
Otherwise, a device malfunction or failure may result.
Always use the specified fittings and gaskets. After the piping work has been
completed, check that there are no gas leaks before operating the device.
Failure to do so may result in gas leakage.
The valve on this device cannot completely stop a flow.
If complete shutoff is required, provide a separate shutoff valve.
In addition, if an external shutoff valve is closed, fully close the valve of this de-
vice using either one of the following methods while it is standing by:
• Set the flow rate to zero.
• Set the operation mode to "valve fully closed."
Even though the external shutoff valve is closed (flow rate is zero), leaving the
device in control mode will cause a momentary excessive flow rate when the
external shutoff valve is opened. Also, in control mode or with the valve forced
fully open, if the external shutoff valve is kept closed, the valve overheating
prevention function will be activated and the electrical current driving the
valve will be forcibly limited.
If there is a risk of a lightning surge, use a surge absorber (surge protector).
Failure to do so may result in a fire or device failure.
3-4
Chapter 3. MOUNTING AND WIRING
3-1 Mounting
Installation location
Install the device in a location not subject to the following:
• High or low temperature and/or humidity
• Sudden temperature changes that result in condensation
• Corrosive and/or flammable gases
Large amounts of dust, salt, iron powder or other conductive sub-
stances in the atmosphere, or water droplets, oil mist, or organic
solvents
• Direct vibration or shock
• Direct sunlight, wind, or rain
• Splashing of fluids, such as oil or chemicals
• Sources of electrical noise
• Strong magnetic or electrical fields
Installation method
CAUTION
When mounting, be sure to secure the product sufficiently so that it does not
vibrate.
Otherwise, a device failure may result.
Using the mounting screw holes on the base of the device, mount
using two M4 screws. Unit: mm
15
20
4.5
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82

Azbil F4H User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI