1000DF00483 Issue 2 15/46
Options
DoseRateCalculator(DRC)
This feature allows the clinician to calculate a rate or a dose rate for continuous drug infusions, and is based on parameters such
as drug dosage, patient weight, concentration, etc. Once calculated, the pump will display (Dose Calculator).
When the DRC VTBI has counted down to 0.0 ml, the channel will switch to the pre-set KVO rate or remain at the current rate,
whichever is less.
Qualied service personnel can turn the Dose Rate Calculator feature on or o.
A
• Thepatientweight,drugconcentration,anddiluentvolumecannotbechangedwhileinfusing.
Changestoanyoftheseitemswhileonholdwillrecalculatevolumetricratetomaintaindoserate.
• Whenadrugamountisgreaterthan10,000units(Un),aKisusedtoindicateavaluemultipliedby
1,000.(e.g.,1,000,000=1,000K).
• DRCcannotbeusedinconjunctionwithsecondaryorotheroperatingmodes.
1. Select desired channel, as necessary. Channel must be infusing in primary mode or on hold in primary mode, secondary
mode or a loading dose program.
2. Press OPTIONS key.
3. Press DoseRateCalculator soft key. Dose Rate Menu will appear.
4. Press EnterNewProgram soft key.
5. If current dose unit is appropriate press ENTER key or to enter a new dose unit, use the
soft key and ENTER key to
select and conrm each segment of the dose unit.
6. If current concentration unit is appropriate press ENTER key or to enter a concentration unit, use the
soft key and
ENTER key to select and conrm.
7. If current weight or height unit is appropriate press ENTER key or to enter a weight or height unit, use the
soft key
and ENTER key to select and conrm.
8. Verify that all parameters are correct then press ok soft key.
9. To enter dose rate, use numeric keypad and press ENTER key or press Rate soft key if Dose Rate is required to be
calculated. To enter rate, use numeric keypad and press ENTER key.
10. To enter concentration, use numeric keypad. Press ENTER key.
11. To enter diluent volume, use numeric keypad. Press ENTER key.
12. To enter height and/or weight, use numeric keypad. Press ENTER key.
A
áááááorâââââwillappearifacalculatedvalueisoutsidethedisplay’srange:
• Usethesoftkeytohighlightthevalueyouwanttochange.
• Usethenumerickeypadtoenterthevalue.
• Presstoacceptthechange.
13. Verify that all parameters are correct then press ok soft key.
14. To enter VTBI value, use numeric keypad. Press ENTER key.
15. To clear VI, if required, press CLEAR or 0 (zero) key. Press ENTER key.
16. Verify that all parameters are correct then press ok soft key.
17. Press run soft key or RUN/HOLD key to start infusion.
18. To briey view current setup information press
4
soft key.
MakingChangesDuringDRCProgram
1. Select desired channel, as necessary.
2. Press soft key next to parameter to be edited. (Press twice to get VI) Current value is highlighted.
3. Make changes: To enter a new value, use numeric keypad or to reset Volume Infused to 0.0 ml, press CLEAR or 0 (zero) key.
4. To accept new value, press ENTER key.
To change the Concentration, Weight or Height the pump must be on hold.
1. Select desired channel, as necessary. To place channel on hold, press RUN/HOLD key.
2. Press soft key next to parameter to be edited. Current value is highlighted.
3. Make changes: To enter a new value, use numeric keypad.
4. To accept new value, press ENTER key.
5. Verify that all parameters are correct then press ok soft key.
6. Press run soft key or RUN/HOLD key to resume the infusion.
QuittingtheDRCProgram
The channel must be on hold.
1. Press menu soft key.
2. To return to primary set-up page, press QuitProgram soft key.