Struers LaboPol 30 User manual

Type
User manual
LaboPol-30
Instruction Manual
Original Instructions
Doc. no.: 16337025-01_B_en
Date of release: 2022.10.13
Copyright
The contents of this manual are the property of Struers ApS. Reproduction of any part of this manual without
the written permission of Struers ApS is not allowed.
All rights reserved. © Struers ApS 2022.10.13.
2 LaboPol-30
Table of Contents
1 About this manual 8
2 Safety 8
2.1 Intended use 8
2.2 LaboPol-30safety precautions 9
2.2.1 Read carefully before use 9
2.3 Safety messages 10
2.4 Safety messages in this manual 11
3 Getting started 13
3.1 Device description 13
3.2 LaboPol-30 - Front view 14
3.2.1 Control panel/Specimen movers 15
3.3 LaboPol-30 - Rear view 16
3.4 Struers knowledge 16
3.5 Accessories and consumables 16
4 Transport and storage 17
4.1 Transport 17
4.2 Long-term storage or shipping 17
5 Installation 18
5.1 Unpacking 18
5.2 Checking the packing list 18
5.3 Lifting 19
5.4 Location 19
5.5 Power supply 20
5.5.1 Single-phase supply 20
5.5.2 2-phase supply 21
5.5.3 Connection to the machine 21
5.6 Water supply and water outlet 21
5.6.1 Connecting to the water supply 21
5.6.2 Connecting to the waste water outlet 22
5.7 Recirculation unit 22
5.7.1 Connecting the recirculation cooling unit to the water inlet 22
5.7.2 Connecting the recirculation cooling unit to the water outlet 23
5.7.3 Connecting the communication cable 23
5.8 Installing the preparation disc 23
LaboPol-30 3
5.8.1 Preparation disc types 23
5.9 Noise 23
5.10 Vibration 24
6 LaboUI 24
6.1 Installation 25
6.1.1 Unpacking 25
6.1.2 Checking the packing list 25
6.1.3 Installation - LaboUI 25
6.2 Operating the device 26
6.2.1 Control panel functions 26
6.2.2 Water tap 27
6.2.3 The Spin function 27
6.2.4 The splash guard 28
6.2.5 Manual preparation 28
6.2.6 Starting and stopping the machine 28
7 LaboForce-50 30
7.1 Installation 30
7.1.1 Unpacking 30
7.1.2 Checking the packing list 31
7.1.3 Installation - LaboForce-50 31
7.1.4 Adjusting the specimen mover 32
7.2 Operating the device 34
7.2.1 Control panel functions 34
7.2.2 Water tap 35
7.2.3 The Spin function 35
7.2.4 The splash guard 36
7.2.5 Inserting a specimen 36
7.2.6 Adjusting the force 36
7.2.7 Manual preparation 37
7.2.8 Starting and stopping the machine 38
7.2.9 Removing the specimens 39
7.2.10 Changing the specimen mover plate 39
8 LaboForce-100 40
8.1 Installation 40
8.1.1 Unpacking 40
8.1.2 Checking the packing list 41
8.1.3 Installation - LaboForce-100 41
8.1.4 Electrical connection to the machine 41
8.1.5 Compressed-air connections 42
4 LaboPol-30
8.1.6 The specimen mover 42
8.1.7 The flexible specimen holder 46
8.1.8 LaboDoser-100 with LaboForce-100 49
8.1.9 LaboDoser-10 with LaboForce-100 49
8.2 Operating the device 49
8.2.1 Control panel functions 49
8.2.2 Water tap 51
8.2.3 The Spin function 51
8.2.4 The splash guard 52
8.2.5 The display 52
8.2.6 Main menu 53
8.2.7 Navigating in the display 53
8.2.8 Changing settings and text 54
8.2.9 Software settings 54
8.2.10 Configuration 55
8.2.11 Maintenance menu 58
8.3 The preparation process 59
8.3.1 Preparation modes 59
8.3.2 Starting and stopping the preparation process 63
9 LaboForce-Mi 64
9.1 Installation 65
9.1.1 Unpacking 65
9.1.2 Checking the packing list 65
9.1.3 Installation - LaboForce-Mi 65
9.1.4 The specimen mover 66
9.2 Operating the device 68
9.2.1 Control panel functions 68
9.2.2 Water tap 69
9.2.3 The Spin function 69
9.2.4 The splash guard 70
9.2.5 Inserting a specimen 70
9.2.6 Adjusting the force 70
9.2.7 Forced rotation of specimens 71
9.2.8 Manual preparation 71
9.2.9 Starting and stopping the machine 72
9.2.10 Removing the specimens 73
9.2.11 Changing the specimen mover plate 73
10 LaboDoser-10 74
10.1 Unpacking 75
LaboPol-30 5
10.2 Checking the packing list 75
10.3 Installation 75
10.4 Operating LaboDoser-10 76
10.5 Changing lubricant/suspension 76
11 LaboDoser-100 77
11.1 Installation 77
11.1.1 Unpacking 77
11.1.2 Checking the packing list 78
11.1.3 Installing LaboDoser-100 78
11.2 Operating LaboDoser-100 79
11.2.1 Changing lubricant/suspension 79
11.2.2 Cleaning the tubes 80
11.2.3 Changing the pump tubes 80
12 Maintenance and service - LaboPol-30 82
12.1 General cleaning 82
12.2 Daily 83
12.3 Weekly 83
12.3.1 LaboForce-100 - the specimen mover head 83
12.4 Monthly 84
12.4.1 LaboForce-50 - pressure feet 84
12.4.2 LaboForce-100 - Emptying the water/oil filter 84
12.5 Annually 85
12.5.1 Testing safety devices 85
12.5.2 Emergency stop 85
12.6 Spare parts 85
12.7 Service and repair 86
12.7.1 Service check - LaboForce-100 86
12.8 Disposal 87
13 Troubleshooting - LaboPol-30 87
13.1 Troubleshooting - LaboPol-30 87
13.2 LaboForce-50 88
13.3 LaboForce-100 89
13.3.1 Messages and errors - LaboForce-100 89
13.4 LaboForce-Mi 92
14 Technical data 93
14.1 Technical data 93
14.2 Safety Circuit Categories/Performance Level 94
14.3 Noise and vibration levels 94
14.4 Technical data - equipment units 94
6 LaboPol-30
14.5 Safety Related Parts of the Control System (SRP/CS) 94
14.6 Diagrams 95
14.6.1 Diagrams - LaboPol-30 95
14.6.2 Diagrams - equipment units 100
14.7 Legal and regulatory information 100
15 Pre-installation checklist 100
15.1 Installation requirements 100
15.2 Packaging specifications 101
15.3 Location 103
15.4 Dimensions 104
15.5 Recommended space 106
15.6 Transport and storage 107
15.6.1 Transport 107
15.6.2 Long-term storage or shipping 107
15.7 Unpacking 108
15.8 Lifting 108
15.9 Power supply 109
15.10 Connection to the machine 110
15.11 Safety specifications 110
15.12 Water supply 110
15.13 Compressed air 111
15.14 Recirculation cooling unit 111
16 Manufacturer 111
Declaration of Conformity 113
LaboPol-30 7
1 About this manual
Instruction Manuals
Struers equipment must only be used in connection with and as described in the Instruction Manual
supplied with the equipment.
Note
Read the instruction manual carefully before use.
Note
If you wish to view specific information in detail, see the online version of this manual.
2 Safety
2.1 Intended use
The machine is for use in a professional working environment (e.g. a materialographic laboratory).
The equipment is designed to be used only with Struers consumables specifically designed for this
purpose and this type of machine.
The machine is for professional manual or semi-automatic materialographic preparation (grinding or
polishing) of materials for further materialographic inspection.
The machine must be operated only by skilled/trained personnel.
Manual preparation LaboPol-30 is intended to be used in combination with:
• LaboUI
Semi-automatic preparation LaboPol-30 is intended to be used in combination with:
• LaboForce-50
LaboForce-100 with or without LaboDoser-100
• LaboForce-Mi
8 LaboPol-30
1 About this manual
Do not use the machine for
the following
Preparation (grinding or polishing) of materials other than solid
materials suitable for materialographic studies.
The machine must not be used for any type of explosive and/or
flammable material, or materials which are not stable during
machining, heating or pressure.
Model LaboPol-30
2.2 LaboPol-30safety precautions
2.2.1 Read carefully before use
In combination with: LaboUI, LaboForce-50, LaboForce-100, LaboForce-Mi, LaboDoser-100.
Ignoring this information and mishandling of the equipment can lead to severe bodily injuries and
material damage.
Specific safety precautions - residual risks
1. The operator must read the Instruction Manual and, where applicable, the Safety Data
Sheets for the applied consumables.
2. The machine must be placed on a safe and stable table with an adequate working height.
The table must be able to carry at least the weight of the machine and the accessories.
3. Connect the machine to a cold water tap. Make sure that the water connections are leak-
proof and that the water outlet is working.
4. Keep clear of rotating parts during operation. When you perform manual grinding or
polishing, be careful not to touch the disc. Do not try to collect a specimen from the tray while
the disc is rotating.
5. To prevent specimens from detaching from the specimen holder, make sure that the
specimen or specimens are clamped securely in the specimen holder.
6. Wear suitable gloves to protect fingers from abrasives and warm/sharp specimens.
7. Protective footwear is recommended when you handle heavy specimen holders.
8. When working at machines with rotating parts, take care to prevent clothes and/or hair from
being caught by the rotating parts. Appropriate safety clothing must be used.
9. Prolonged exposure to loud noises may cause permanent damage to a person’s hearing.
Use hearing protection if the exposure to noise exceeds the levels set by local regulations.
10. Risk of hand to arm vibration during manual preparation.Prolonged exposure to vibration
may cause discomfort, joint damage or even neurological damage.
11. The machine must be disconnected from the electrical power supply prior to any service.
Wait 5 minutes until residual potential on the capacitors is discharged.
12. The machine must not be used for any type of explosive and/or flammable material, or
materials which are not stable during machining, heating or pressure.
2 Safety
LaboPol-30 9
General safety precautions
1. The machine must be installed in compliance with local safety regulations. All functions on the
machine and any connected equipment must be in working order.
2. The operator must read the safety precautions and Instruction Manual, as well as relevant
sections of the manuals for any connected equipment and accessories.
3. This machine must be operated and maintained only by skilled/trained personnel.
4. The machine must always be used with the splash guard in place.
5. Make sure that the actual electrical power supply voltage corresponds to the voltage stated on
the type plate of the machine. The machine must be earthed (grounded). Always follow local
regulations. Always switch off the electrical power supply and remove the plug or power cable
before dismantling the machine or installing additional components.
6. Struers recommends that the main water supply is shut off or disconnected if the machine is to
be left unattended.
7. Consumables: only use consumables specifically developed for use with this type of
materialographic machine. Alcohol-based consumables: follow the current safety rules for
handling, mixing, filling, emptying and disposing of alcohol-based liquids.
8. Keep your hands clear of the specimen holder (if applicable) or specimen mover plate when
lowering the specimen mover unit, if installed.
9. If you observe malfunctions or hear unusual noises, switch off the machine and call technical
service.
10. Do not switch the machine on and off more than once every five minutes. Damage to the
electrical components could occur.
11. In case of fire, alert bystanders and the fire brigade. Disconnect the electrical power supply. Use
a powder fire extinguisher.Do not use water.
12. Struers equipment must only be used in connection with and as described in the Instruction
Manual supplied with the equipment.
13. The equipment is designed to be used only with Struers consumables specifically designed for
this purpose and this type of machine.
14. If the equipment is subjected to misuse, incorrect installation, alteration, neglect, accident or
incorrect repair, Struers will accept no responsibility for damage to the user or the equipment.
15. Dismantling of any part of the equipment, during service or repair, should always be performed
by a qualified technician (electromechanical, electronic, mechanical, pneumatic, etc.).
2.3 Safety messages
Struers uses the following signs to indicate potential hazards.
ELECTRICAL HAZARD
This sign indicates an electrical hazard which, if not avoided, will result in death or
serious injury.
10 LaboPol-30
2 Safety
DANGER
This sign indicates a hazard with a high level of risk which, if not avoided, will
result in death or serious injury.
WARNING
This sign indicates a hazard with a medium level of risk which, if not avoided,
could result in death or serious injury.
CAUTION
This sign indicates a hazard with a low level of risk which, if not avoided, could
result in minor or moderate injury.
CRUSHING HAZARD
This sign indicates a crushing hazard which, if not avoided, could result in minor,
moderate or serious injury.
HEAT HAZARD
This sign indicates a heat hazard which, if not avoided, can result in minor,
moderate or serious injury.
General messages
Note
This sign indicates that there is a risk of damage to property, or a need to proceed
with special care.
Hint
This sign indicates that additional information and hints are available.
2.4 Safety messages in this manual
WARNING
Struers equipment must only be used in connection with and as described in the
Instruction Manual supplied with the equipment.
CRUSHING HAZARD
Take care of your fingers when handling the machine.
Wear safety shoes when handling heavy machinery.
WARNING
Switch off the machine, disconnect the electrical power cable and wait 5 minutes
before you dismantle the machine or install additional components.
ELECTRICAL HAZARD
Switch off the electrical power supply before installing electrical equipment.
The machine must be earthed (grounded).
Make sure that the actual electrical power supply voltage corresponds to the
Incorrect voltage can damage the electrical circuit.
2 Safety
LaboPol-30 11
ELECTRICAL HAZARD
The pump of the recirculation cooling unit must be earthed (grounded).
Make sure that the electrical power supply voltage corresponds to the voltage stated
on the type plate of the pump.
Incorrect voltage can damage the electrical circuit.
CAUTION
Prolonged exposure to loud noises may cause permanent damage to a person’s
hearing.
Use hearing protection if the exposure to noise exceeds the levels set by local
regulations.
CAUTION
Risk of hand to arm vibration during manual preparation.
Prolonged exposure to vibration may cause discomfort, joint damage or even
neurological damage.
CAUTION
Keep clear of rotating parts during operation.
CAUTION
When working at machines with rotating parts, take care to prevent clothes and/or
hair from being caught by the rotating parts.
CAUTION
For manual preparation use the switch on the side of the specimen mover head to
disable rotation of LaboForce-50
CAUTION
For manual preparation use the switch on the side of the specimen mover head to
disable rotation of LaboForce-Mi.
CAUTION
Wear suitable gloves to protect fingers from abrasives and warm/sharp specimens.
CAUTION
When you perform manual grinding or polishing, be careful not to touch the disc.
CAUTION
Do not try to collect a specimen from the tray while the disc is rotating.
12 LaboPol-30
2 Safety
CAUTION
When the disc is rotating, make sure your hands are kept completely clear of its
periphery and out of the splash bowl.
WARNING
Do not use the machine with defective safety devices.
Contact Struers Service.
WARNING
Before you release the emergency stop, investigate the reason for activating the
emergency stop and take any necessary corrective action.
CAUTION
To prevent specimens from detaching from the specimen holder, make sure that
the specimen or specimens are clamped securely in the specimen holder.
CAUTION
Protective footwear is recommended when you handle heavy specimen holders.
CRUSHING HAZARD
Keep your hands clear of the specimen holder or specimen mover plate when
lowering the specimen mover.
WARNING
Safety critical components must be replaced after a maximum lifetime of 20 years.
Contact Struers Service.
CRUSHING HAZARD
Take care of your fingers when handling the machine.
Wear safety shoes when handling heavy machinery.
3 Getting started
3.1 Device description
LaboPol-30 is for materialographic preparation (grinding/polishing) using preparation discs of
230, 250 or 300 mm in diameter.
LaboPol-30 is for manual preparation when used with LaboUI.
LaboPol-30 is for semi-automatic preparation when used with LaboForce-50, LaboForce-100 or
LaboForce-Mi. The specimen mover is always installed on the left side of the machine.
3 Getting started
LaboPol-30 13
The operator selects the grinding/polishing surface and the cooling fluid/abrasive suspension to
be applied.
Cooling water is applied when the operator opens the water tap. Other fluids are applied
manually or by a separate dosing unit.
With LaboUI the operator holds the specimens during the preparation.
With LaboForce-50, LaboForce-100 and LaboForce-Mi the operator places the specimens in a
device, the specimen mover plate or the specimen holder.
With LaboUI, LaboForce-50 and LaboForce-Mi the operator sets the rotation speed of the
preparation disc before starting the process.
With LaboForce-100 the operator sets the process parameters before starting the process.
The machine must always be used with the splash guard in place.
The splash guard for manual preparation is supplied with the machine. (For the 300 mm
diameter disc)
The splash guard for other types of preparation must be ordered separately.
The operator starts the machine by pressing the Start button on the control panel.
With LaboUI and LaboForce-50 and LaboForce-Mi the operator stops the machine by pressing
stop on the control panel.
With LaboForce-100 the machine stops automatically when the process is completed.
The operator cleans the specimens before the next preparation step or inspection.
We recommend that an exhaust system is applied when alcohol-based suspension or lubricants
are used.
If the emergency stop is activated, the power to all moving parts is cut.
3.2 LaboPol-30 - Front view
AConnection for control
panel/specimen mover
BSplash guard for manual
preparation (both discs)
CPreparation disc location
(both discs)
DEmergency stop
EWater tap
FStorage pad
Emergency stop
If LaboForce-50, LaboForce-100, LaboDoser-100, or LaboForce-Mi is mounted on
the machine, activating the emergency stop on the machine will also stop LaboForce-
50, LaboForce-100, LaboDoser-100, or LaboForce-Mi.
14 LaboPol-30
3 Getting started
Note
Do not use the emergency stop for operational stop of the machine during normal
operation.
Before you release the emergency stop, investigate the reason for activating the
emergency stop and take any necessary corrective action.
To activate the emergency stop, press the red emergency stop button.
To release the emergency stop, turn the red emergency stop button clockwise.
3.2.1 Control panel/Specimen movers
The machine can be mounted with one of the following units. See the specific sections for
installation instructions.
LaboUI
See Control panel functions ►26.
LaboForce-50
See Control panel functions ►34.
LaboForce-100
See Control panel functions ►49.
LaboForce-Mi
See Control panel functions ►68.
3 Getting started
LaboPol-30 15
3.3 LaboPol-30 - Rear view
AMain switch
BFuses
CElectrical power socket
DWater inlet from
recirculation cooling unit
EWater inlet from main
water supply
FWaste water outlet
GLaboForce-100 socket
HRecirculation unit socket
IType plate
3.4 Struers knowledge
Mechanical preparation is the most common method of preparing materialographic specimens for
microscopic examination.
The specific requirement of the prepared surface is determined by the particular type of analysis or
examination.
Specimens can be prepared to the perfect finish, the true structure, or the preparation can be
stopped when the surface is acceptable for a specific examination.
Hint
For further information, see the section on Grinding and Polishing on the Struers
website.
3.5 Accessories and consumables
Accessories
For information about the available range, see:
The LaboSystem brochure (https://www.struers.com/Products/Grinding-and-
Polishing/Grinding-and-polishing-equipment/LaboSystem)
Consumables
The use of Struers consumables is recommended.
Other products may contain aggressive solvents, which dissolve e.g. rubber seals. The warranty may
not cover damaged machine parts (e.g. seals and tubes), where the damage can be directly related
to the use of consumables not supplied by Struers.
For information about the available range, see:
The Struers Consumables Catalogue (http://www.struers.com/Library)
16 LaboPol-30
3 Getting started
4 Transport and storage
If, at any time after the installation, you have to move the unit or place it in storage, there is a
number of guidelines we recommend that you follow.
Package the unit securely before transportation.
Insufficient packaging could cause damage to the unit and will void the warranty. Contact
Struers Service.
Struers recommends that all original packaging and fittings are kept for future use.
4.1 Transport
Disconnect the unit from the electrical power supply.
Disconnect the water inlet and the water outlet.
Disconnect the cooling system, if installed. See the instructions for the specific unit.
Remove the splash guard, the preparation disc and the bowl liner.
Lift the machine by gripping the base of the
machine from beneath, from both the left
side and the right side.
Alternatively, use a crane and two
lifting straps to lift the machine.
Position the straps under the
machine, so that they are on the outer
side of the feet.
Lift the machine onto a stable surface.
4.2 Long-term storage or shipping
Note
Struers recommends that all original packaging and fittings are kept for future use.
Clean the machine and all accessories thoroughly.
Disconnect the unit from the electrical power supply.
Disconnect the water inlet and the water outlet.
Disconnect the cooling system, if installed. See the instructions for the specific unit.
Remove the splash guard, the preparation disc and the bowl liner.
Remove the control panel or specimen mover.
4 Transport and storage
LaboPol-30 17
Remove any accessories.
Lift the machine by gripping the base of the machine from beneath, from both the left side and
the right side.
Place the machine and accessories in their original packaging.
Secure the boxes on a pallet with straps.
At the new location
At the new location, make sure that the facilities required are in place.
5 Installation
5.1 Unpacking
Note
Struers recommends that all original packaging and fittings are kept for future use.
1. Cut the packing tape on the top of the box.
2. Remove the loose parts.
3. Remove the unit from the box.
5.2 Checking the packing list
Optional accessories may be included in the packing box.
The packing box contains the following items:
Pcs. Description
1 LaboPol-30
2 Electrical power supply cables
1 Disposable bowl liner, clear plastic
1 Splash guard for manual preparation (for 300 mm disc)
1 Water inlet hose. Diameter: 19 mm/¾". Length: 2 m/6.6'
1 Filter gasket
1 Reduction ring with gasket, ¾" to ½"
1 Water outlet hose. Diameter: 40 mm/1.6". Length: 1.5 m/4.9'
1 Elbow pipe for water outlet
1 Hose clamp
18 LaboPol-30
5 Installation
Pcs. Description
1 Allen key with cross handle, 6x150 mm/0.23x6"
2 Caps for use after mounting LaboUI, LaboForce-50, LaboForce-100, or LaboForce-Mi
1 Yellow cap for use with a Recirculation Cooling Unit
1 Instruction Manual set
5.3 Lifting
CRUSHING HAZARD
Take care of your fingers when handling the machine.
Wear safety shoes when handling heavy machinery.
Note
Do not lift the machine by the light gray top part or by the water tap.
Always lift the machine from beneath.
Weight
LaboPol-30 33 kg (73 lbs)
1. Lift the machine by gripping the base of the
machine from beneath, from both the left
side and the right side.
Alternatively, use a crane and two
lifting straps to lift the machine.
Position the straps under the
machine, so that they are on the outer
side of the feet.
2. Lift the machine onto the table.
3. The machine must rest securely with all 4
feet on the table.
5.4 Location
CRUSHING HAZARD
Take care of your fingers when handling the machine.
Wear safety shoes when handling heavy machinery.
5 Installation
LaboPol-30 19
The machine must be placed close to the electrical power supply, main water supply and waste
water drain.
To facilitate easy access for service technicians, allow sufficient space around the machine.
Place the machine on a rigid, stable workbench with a horizontal surface and an adequate
height.
To move the machine, lift the front of the machine and use the rollers to carefully move it into
place.
The machine must rest securely with all 4 feet on the table.
To level the machine, turn the adjustable rubber feet.
5.5 Power supply
WARNING
Switch off the machine, disconnect the electrical power cable and wait 5 minutes
before you dismantle the machine or install additional components.
ELECTRICAL HAZARD
Switch off the electrical power supply before installing electrical equipment.
The machine must be earthed (grounded).
Make sure that the actual electrical power supply voltage corresponds to the voltage
stated on the type plate of the machine.
Incorrect voltage can damage the electrical circuit.
Note
An autotransformer is required in countries with a 110 V electrical power supply.
Power socket
The electrical power supply socket must be easy to access. The electrical power supply socket must
be located at a height ranging from 0.6 m to 1.9 m (2½” to 6’) above floor level. Not higher than 1.7 m
(5’ 6”) is recommended.
Note
The equipment is shipped with 2 types of electrical power cables. If the plug supplied
on these cables is not approved in your country, the plug must be replaced with an
approved plug.
5.5.1 Single-phase supply
Single-phase supply
The 2-pin plug (European Schuko) is for use on single-phase
electrical power connections.
The leads must be connected as follows:
20 LaboPol-30
5 Installation
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114

Struers LaboPol 30 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI