  • Hello, I'm your chat assistant. I have reviewed the programming manual for the UNI-T UTE9802 power meter. This document details the commands used to interact with the device for reading measurements like voltage, current, power, and frequency as well as setting different alarm limits. I'm ready to answer any questions you might have about its capabilities based on this manual.
  • What is the command to read the identification string of the instrument?
    How do I read all measured values at one time?
    What command is used to set the measurement mode?
    How do I set the upper limit of the current alarm?
UTE9802 Programming manual
UTE9802 Programming manual
REV.1
2022. 4
UNI-T TECHNOLOGY(China) Co.Lt
UTE9802 Programming manual
*IDN?
This query command reads the identification string of the instrument. Return value example
UTE9802SN202111051001V1.12V0.02
MEAS:VOLT
Read the measured voltage value. Return value example220.0VOLOut of
range, the same below
MEAS:CURR
Read the measured current value. Return value example5.12A or 500.1mA
MEAS:POW
Read the measured power value. Return value example350.02W
MEAS:PF
Read the measured power factor value. Return value example0.501
MEAS:FREQ
Read the measured frequency value. Return value example50.01Hz
MEAS:ALL
Read back all measured values at one time: measurement mode, voltage,
current, power, power factor, frequency, separated by commas.
ExampleAC,220.1V5.01A1103.2W0.90149.99Hz
ExampleDC,220.1V5.01A1103.2W
ExampleRMS,220.1V5.01A1103.2W0.99849.99Hz
MEAS:MODE
Read the measurement mode. Return value exampleACDCRMSAC+DC
SET:MODE XXX
Set the measurement mode. Return value exampleSUCCESSERR
RANG:VOLT
Read the voltage range. Return value exampleAUTO 75V150V300V600V
RANG:VOLT XXXX
Set the voltage range. Return value exampleSUCCERR
RANG:CURR
Read the current range. Return value exampleAUTO 500mA2A8A20A
RANG:CURR XXXX
UTE9802 Programming manual
Set the current range. Return value exampleSUCCERR
CURR:HIG XXXX
Set the upper limit of current alarm.
ExampleCURR:HIG 5A. Return value exampleSUCCERR
CURR:HIG?
Read current alarm upper limit value.
CURR:LOW XXXX
Set the lower limit of current alarm.
ExampleCURR:LOW 5A. Return value exampleSUCCERR
CURR:LOW?
Read the lower limit value of current alarm.
POW:HIG XXXX
Set the upper limit of power alarm.
ExamplePOW:HIG 5W. Return value exampleSUCCERR
POW:HIG?
Read the upper limit value of power alarm.
POW:LOW XXXX
Set the lower limit of power alarm.
ExamplePOW:LOW 5W. Return value exampleSUCCERR
POW:LOW?
Read the lower limit value of power alarm.
ALAR:DELA?
Read the alarm delay value. Return value example0.2S
ALAR:DELA XXX
Set the alarm delay value. Return value exampleSUCCERR
STA?
Read alarm signs, return current and power alarm signs. Separated by commas.
Example
Return NORNORIndicates normal.
Return LOWHIGIndicates current lower limit alarm and power upper limit
alarm.
/