Micro Motion LNG-Series Meters Owner's manual

Type
Owner's manual
Configuration and Use Manual
MMI-20032809, Rev AA
February 2017
Micro Motion
®
LNG Series Meters
Safety messages
Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully
before proceeding to the next step.
Emerson Flow customer service
Email:
Worldwide: [email protected]
Asia-Pacific: [email protected]
Telephone:
North and South America Europe and Middle East Asia Pacific
United States 800-522-6277 U.K. 0870 240 1978 Australia 800 158 727
Canada +1 303-527-5200 The Netherlands +31 (0) 704 136 666 New Zealand 099 128 804
Mexico +41 (0) 41 7686 111 France 0800 917 901 India 800 440 1468
Argentina +54 11 4837 7000 Germany 0800 182 5347 Pakistan 888 550 2682
Brazil +55 15 3413 8000 Italy 8008 77334 China +86 21 2892 9000
Venezuela +58 26 1731 3446 Central & Eastern +41 (0) 41 7686 111 Japan +81 3 5769 6803
Russia/CIS +7 495 981 9811 South Korea +82 2 3438 4600
Egypt 0800 000 0015 Singapore +65 6 777 8211
Oman 800 70101 Thailand 001 800 441 6426
Qatar 431 0044 Malaysia 800 814 008
Kuwait 663 299 01
South Africa 800 991 390
Saudi Arabia 800 844 9564
UAE 800 0444 0684
Contents
Part I Getting Started
Chapter 1 Before you begin ............................................................................................................3
1.1 About this manual ....................................................................................................................... 3
1.2 Meter model codes ......................................................................................................................3
1.3 Communications tools and protocols .......................................................................................... 4
1.4 Additional documentation .......................................................................................................... 4
Chapter 2 Quick start .....................................................................................................................7
2.1 Power up the meter .....................................................................................................................7
2.2 Check meter status ......................................................................................................................7
2.3 Make a startup connection to the transmitter ..............................................................................8
2.4 Characterize the flowmeter (if required) ......................................................................................8
2.4.1 Sensor tags ................................................................................................................... 9
2.4.2 Density calibration parameters (D1, D2, K1, K2, FD, TC) .................................................. 9
2.5 Verify mass flow measurement ....................................................................................................9
2.6 Verify the zero ............................................................................................................................. 9
2.6.1 Verify the zero using ProLink III ...................................................................................10
2.6.2 Terminology used with zero verification and zero calibration ......................................11
Part II Configuration and commissioning
Chapter 3 Introduction to configuration and commissioning ....................................................... 15
3.1 Default values and ranges ..........................................................................................................15
3.2 Disable write-protection on the transmitter configuration ........................................................ 15
3.3 Restore the factory configuration .............................................................................................. 15
Chapter 4 Configure process measurement ..................................................................................17
4.1 Configure mass flow measurement ........................................................................................... 17
4.1.1 Configure Mass Flow Measurement Unit ...........................................................................17
4.1.2 Configure Flow Damping ............................................................................................... 19
4.1.3 Configure Mass Flow Cutoff ........................................................................................... 20
4.2 Configure volume flow measurement for liquid applications ..................................................... 21
4.2.1 Configure Volume Flow Type for liquid applications ........................................................21
4.2.2 Configure Volume Flow Measurement Unit for liquid applications ......................................21
4.2.3 Configure Volume Flow Cutoff ........................................................................................ 24
4.3 Configure gas standard volume (GSV) flow measurement .........................................................24
4.3.1 Configure Volume Flow Type for gas applications ........................................................... 25
4.3.2 Configure Standard Gas Density .....................................................................................25
4.3.3 Configure Gas Standard Volume Flow Measurement Unit .................................................... 26
4.3.4 Configure Gas Standard Volume Flow Cutoff .....................................................................28
4.4 Configure flow direction ........................................................................................................... 28
4.4.1 Options for Flow Direction ............................................................................................. 29
4.5 Configure density measurement .............................................................................................. 30
4.5.1 Configure Density Measurement Unit ...............................................................................30
Contents
Configuration and Use Manual i
4.5.2 Configure two-phase flow parameters ........................................................................ 31
4.5.3 Configure density damping ........................................................................................33
4.5.4 Configure Density Cutoff ............................................................................................... 34
4.6 Configure temperature measurement .......................................................................................34
4.6.1 Configure Temperature Measurement Unit ........................................................................34
4.6.2 Configure Temperature Damping .................................................................................... 35
Chapter 5 Configure device options and preferences ....................................................................37
5.1 Configure response time parameters ........................................................................................ 37
5.1.1 Configure Calculation Speed (Response Time) .................................................................. 37
5.2 Configure alert handling ............................................................................................................38
5.2.1 Configure Fault Timeout ................................................................................................38
5.2.2 Configure Status Alert Severity ....................................................................................... 38
5.3 Configure informational parameters ......................................................................................... 40
5.3.1 Configure the sensor index ......................................................................................... 40
5.3.2 Configure the RTD cable length .................................................................................. 41
5.3.3 Configure Sensor Serial Number .....................................................................................41
5.3.4 Configure Sensor Material ............................................................................................. 41
5.3.5 Configure Sensor Liner Material ......................................................................................42
5.3.6 Configure Sensor Flange Type ....................................................................................... 42
5.3.7 Configure Descriptor .................................................................................................... 43
5.3.8 Configure Message ...................................................................................................... 43
5.3.9 Configure Date ............................................................................................................43
Chapter 6 Integrate the meter with the control system ................................................................45
6.1 Configure events ....................................................................................................................... 45
6.1.1 Configure an enhanced event ..................................................................................... 45
6.2 Configure digital communications ............................................................................................ 46
6.2.1 Configure Modbus/RS-485 communications .............................................................. 47
6.2.2 Configure RS-485 terminals ........................................................................................ 48
6.2.3 Configure Digital Communications Fault Action .................................................................. 49
Chapter 7 Complete the configuration ......................................................................................... 51
7.1 Back up transmitter configuration ............................................................................................. 51
7.2 Enable write-protection on the transmitter configuration ......................................................... 51
Part III Operations, maintenance, and troubleshooting
Chapter 8 Transmitter operation ................................................................................................. 55
8.1 Custody transfer ........................................................................................................................55
8.2 Record the process variables ..................................................................................................... 55
8.3 View process variables ...............................................................................................................56
8.3.1 View process variables and other data using ProLink III ...............................................56
8.4 View transmitter status using the status LED ............................................................................. 56
8.5 View and acknowledge status alerts .......................................................................................... 57
8.5.1 View and acknowledge alerts using ProLink III ............................................................ 57
8.5.2 Alert data in transmitter memory ............................................................................... 58
8.6 Read totalizer and inventory values ........................................................................................... 59
8.7 Start and stop totalizers and inventories ....................................................................................59
8.8 Reset totalizers ..........................................................................................................................60
8.9 Reset inventories .......................................................................................................................60
Contents
ii Micro Motion
®
LNG Series Meters
Chapter 9 Measurement support ................................................................................................. 61
9.1 Options for measurement support ............................................................................................ 61
9.2 Zero the meter .......................................................................................................................... 61
9.2.1 Zero the meter using ProLink III ..................................................................................61
9.2.2 Record the zeros ......................................................................................................... 63
9.3 Validate the meter .....................................................................................................................63
9.3.1 Alternate method for calculating the meter factor for volume flow .............................65
9.4 Perform a (standard) D1 and D2 density calibration ...................................................................65
9.4.1 Perform a D1 and D2 density calibration using ProLink III ........................................... 66
Chapter 10 Troubleshooting .......................................................................................................... 69
10.1 Status LED states ....................................................................................................................... 69
10.2 Status alerts, causes, and recommendations ............................................................................. 70
10.3 Flow measurement problems ................................................................................................... 75
10.4 Density measurement problems ............................................................................................... 77
10.5 Temperature measurement problems .......................................................................................78
10.6 Check power supply wiring ........................................................................................................78
10.7 Check grounding .......................................................................................................................79
10.8 Check for radio frequency interference (RFI) ..............................................................................79
10.9 Check for two-phase flow (slug flow) ......................................................................................... 80
10.10 Check the drive gain .................................................................................................................. 80
10.10.1 Collect drive gain data ................................................................................................ 81
10.11 Check the pickoff voltage .......................................................................................................... 81
10.11.1 Collect pickoff voltage data ........................................................................................ 82
10.12 Check for internal electrical problems ....................................................................................... 82
10.12.1 Check the sensor coils .................................................................................................83
Appendices and reference
Appendix A Using ProLink III with the transmitter ...........................................................................85
A.1 Connect with ProLink III ............................................................................................................ 85
A.1.1 Connection types supported by ProLink III ..................................................................85
A.1.2 Connect with ProLink III to the service port ................................................................. 85
A.1.3 Connect with ProLink III to the RS-485 port .................................................................86
Appendix B Default values and ranges ............................................................................................ 89
B.1 Default values and ranges ..........................................................................................................89
Contents
Configuration and Use Manual iii
Contents
iv Micro Motion
®
LNG Series Meters
Part I
Getting Started
Chapters covered in this part:
Before you begin
Quick start
Getting Started
Configuration and Use Manual 1
Getting Started
2 Micro Motion
®
LNG Series Meters
1 Before you begin
Topics covered in this chapter:
About this manual
Meter model codes
Communications tools and protocols
Additional documentation
1.1 About this manual
This manual provides information to help you configure, commission, use, maintain, and
troubleshoot the Micro Motion LNG Series meter.
Important
This manual assumes that the following conditions apply:
The meter has been installed correctly and completely according to the instructions in the
installation manual
The installation complies with all applicable safety requirements
The user is trained in applicable safety standards
1.2
Meter model codes
Your device can be identified by the model code on the device tag.
The transmitter model number has the following form:
LNGM10S***N(P/Z)(D/I/N)***********
Model codes and device typesTable 1-1:
Model code Description I/O and communications
N Standard case One dedicated service port/
Modbus
One dedicated RS-485 port/
Modbus
P LNGS06S and LNGM10S
are paired
D With a dual core processor
Z Standalone sensor
I With enhanced core pro-
cessor for direct host con-
nection with IS barrier
Before you begin
Configuration and Use Manual 3
Model codes and device types (continued)Table 1-1:
Model code Description I/O and communications
N Spare sensor without elec-
tronics
1.3 Communications tools and protocols
You can use these communications tools and protocols to interface with the transmitter.
Communications tools, protocols, and related informationTable 1-2:
Communica-
tions tool Supported protocols Scope In this manual For more information
ProLink III Modbus/RS-485
Service port
Complete configuration
and commissioning
Basic user information.
See Appendix A.
User manual
Installed with soft-
ware
Micro Motion user
documentation CD
www.micromo‐
tion.com
Modbus host Service port
Modbus/RS-485
Complete configuration
and commissioning
None.
Modbus Interface Tool
(MIT) — available at
www.micromotion.com
1.4 Additional documentation
Micro Motion provides additional documentation to support the installation and operation
of the meter.
Additional documentation and resourcesTable 1-3:
Topic Document
Transmitter installation
Micro Motion Liquified Natural Gas Dispenser Installation Manual
Product Data Sheet
Micro Motion LNG Series Meters Product Data Sheet (PDS)
Hazardous area installa-
tion
See the approval documentation shipped with the transmitter, or
download the appropriate documentation from the Micro Motion web
site at www.micromotion.com.
Before you begin
4 Micro Motion
®
LNG Series Meters
All documentation resources are available on the Micro Motion web site at
www.micromotion.com.cn or on the Micro Motion user documentation DVD.
Before you begin
Configuration and Use Manual 5
Before you begin
6 Micro Motion
®
LNG Series Meters
2 Quick start
Topics covered in this chapter:
Power up the meter
Check meter status
Make a startup connection to the transmitter
Characterize the flowmeter (if required)
Verify mass flow measurement
Verify the zero
2.1 Power up the meter
The meter must be powered up for all configuration and commissioning tasks, or for
process measurement.
1. Ensure that all device covers and seals are closed.
WARNING!
To prevent ignition of flammable or combustible atmospheres, ensure that all covers
and seals are tightly closed. For hazardous area installations, applying power while
housing covers are removed or loose can cause an explosion.
2. Turn on the electrical power at the power supply.
Postrequisites
Although the sensor is ready to measure shortly after power-up, the electronics can take
up to 10 minutes to reach thermal equilibrium. Therefore, if this is the initial startup, or if
power has been off long enough to allow components to reach ambient temperature,
allow the electronics equalize for approximately 10 minutes.
2.2
Check meter status
Check the meter for any error conditions that require user action or that affect
measurement accuracy.
1. Wait approximately 10 seconds for the power-up sequence to complete.
Immediately after power-up, the transmitter runs through diagnostic routines and
checks for error conditions.
2. Connect to the transmitter and check for active alerts.
Quick start
Configuration and Use Manual 7
Related information
View and acknowledge status alerts
Status alerts, causes, and recommendations
2.3 Make a startup connection to the transmitter
Identify the connection type to use, and follow the instructions for that connection type in
the appropriate appendix.
Communications tool Connection type to use Instructions
ProLink III Service Port Section A.1.2
RS-485 port Section A.1.3
2.4 Characterize the flowmeter (if required)
ProLink III Device Tools > Calibration Data
Overview
Characterizing the flowmeter adjusts your transmitter to match the unique traits of the
sensor it is paired with. The characterization parameters (also called calibration
parameters) describe the sensor’s sensitivity to flow, density, and temperature.
Depending on your sensor type, different parameters are required. Values for your sensor
are provided by Micro Motion on the sensor tag or the calibration certificate.
Tip
If your meter was ordered as a unit, it has already been characterized at the factory. However, you
should still verify the characterization parameters.
Procedure
1. Specify Sensor Type.
LNG-Series (LNGS06S, LNGM10S)
2. Set the flow characterization parameters. Be sure to include all decimal points.
Set Flow Cal (Flow Calibration Factor).
3. Set the density characterization parameters.
Set D1, D2, TC, K1, K2, and FD.
Quick start
8 Micro Motion
®
LNG Series Meters
2.4.1 Sensor tags
Tag on sensorsFigure 2-1:
2.4.2 Density calibration parameters (D1, D2, K1, K2, FD, TC)
Density calibration parameters are typically on the sensor tag and the calibration
certificate.
2.5
Verify mass flow measurement
Check to see that the mass flow rate reported by the transmitter is accurate. You can use
any available method.
Connect to the transmitter with ProLink III and read the value for Mass Flow Rate in the
Process Variables panel.
Postrequisites
If the reported mass flow rate is not accurate:
Check the characterization parameters.
Review the troubleshooting suggestions for flow measurement issues.
2.6
Verify the zero
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
Quick start
Configuration and Use Manual 9
The zero verification procedure analyzes the Live Zero value under conditions of zero flow,
and compares it to the Zero Stability range for the sensor. If the average Live Zero value is
within a reasonable range, the zero value stored in the transmitter is valid. Performing a
field calibration will not improve measurement accuracy.
2.6.1 Verify the zero using ProLink III
Verifying the zero helps you determine if the stored zero value is appropriate to your
installation, or if a field zero can improve measurement accuracy.
Important
In most cases, the factory zero is more accurate than the field zero. Do not zero the meter unless one
of the following is true:
The zero is required by site procedures.
The stored zero value fails the zero verification procedure.
Prerequisites
ProLink III v3.5, or a later release
Important
Do not verify the zero or zero the meter if a high-severity alert is active. Correct the problem, then
verify the zero or zero the meter. You may verify the zero or zero the meter if a low-severity alert is
active.
Procedure
1. Prepare the meter:
a. Allow the meter to warm up for at least 20 minutes after applying power.
b. Run the process fluid through the sensor until the sensor temperature reaches
the normal process operating temperature.
c. Stop flow through the sensor by shutting the downstream valve, and then the
upstream valve if available.
d. Verify that the sensor is blocked in, that flow has stopped, and that the sensor is
completely full of process fluid.
2. Choose Device Tools > Device Calibration > Zero Verification and Calibration > Verify Zero and
wait until the procedure completes.
3. If the zero verification procedure fails:
a. Confirm that the sensor is completely blocked in, that flow has stopped, and that
the sensor is completely full of process fluid.
b. Verify that the process fluid is not flashing or condensing, and that it does not
contain particles that can settle out.
c. Repeat the zero verification procedure.
d. If it fails again, zero the meter.
Quick start
10 Micro Motion
®
LNG Series Meters
Postrequisites
Restore normal flow through the sensor by opening the valves.
2.6.2 Terminology used with zero verification and zero
calibration
Terminology used with zero verification and zero calibrationTable 2-1:
Term Definition
Zero In general, the offset required to synchronize the left pickoff and the right pickoff under
conditions of zero flow. Unit = microseconds.
Factory Zero The zero value obtained at the factory, under laboratory conditions.
Field Zero The zero value obtained by performing a zero calibration outside the factory.
Prior Zero The zero value stored in the transmitter at the time a field zero calibration is begun. May
be the factory zero or a previous field zero.
Manual Zero The zero value stored in the transmitter, typically obtained from a zero calibration proce-
dure. It may also be configured manually. Also called “mechanical zero” or “stored zero.”
Live Zero The real-time bidirectional mass flow rate with no flow damping or mass flow cutoff ap-
plied. An adaptive damping value is applied only when the mass flow rate changes dra-
matically over a very short interval. Unit = configured mass flow measurement unit.
Zero Stability A laboratory-derived value used to calculate the expected accuracy for a sensor. Under
laboratory conditions at zero flow, the average flow rate is expected to fall within the
range defined by the Zero Stability value (0 ± Zero Stability). Each sensor size and model
has a unique Zero Stability value. Statistically, 95% of all data points should fall within the
range defined by the Zero Stability value.
Zero Calibration The procedure used to determine the zero value.
Zero Time The time period over which the Zero Calibration procedure is performed. Unit = seconds.
Field Verification Zero A 3-minute running average of the Live Zero value, calculated by the transmitter. Unit =
configured mass flow measurement unit.
Zero Verification A procedure used to evaluate the stored zero and determine whether or not a field zero
can improve measurement accuracy.
Quick start
Configuration and Use Manual 11
Quick start
12 Micro Motion
®
LNG Series Meters
Part II
Configuration and commissioning
Chapters covered in this part:
Introduction to configuration and commissioning
Configure process measurement
Configure device options and preferences
Integrate the meter with the control system
Complete the configuration
Configuration and commissioning
Configuration and Use Manual 13
Configuration and commissioning
14 Micro Motion
®
LNG Series Meters
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92