vse VSI 0.02 Operating instructions

Type
Operating instructions

This manual is also suitable for

1
 



 
 
 




2
Table of Contents


Importantbasicinformation3
Generalfunctiondescriptionofowmeter4
Generaldescription4
Flowmeterselection4
DeclarationofConformity4
Generalconditionsforinitialstart-up  4
Maximumoperatingpressure 5
StatementtoEU-Directive97/23/EG,Pressurizeddevices5
Flowmeterrange 6
Assemblyoftheowmeter  6
Cleaningandushingofpipelinesbeforeinitialstart-up 7
Filteringofliquid7
Flowmeterswithhighdenitionofvolumemeasurement8
Technicalspecicationsofpreamplier  13
Plugassignmentofpreamplier14
Maintenance14
Sendingbackofrepairsandsampledevices15
TechnicalspecicationsVSI002/IPF–VSI4/IPF  15
FlowresponsecurvesVSI002–VSI4 16
DimensionsVSI002–VSI417
Dimensions,subplatesAP02-4 18
TechnicalspecicationsVSI10/IPF  19
FlowresponsecurvesVSI10 19
DimensionsVSI10 20
Dimensions,subplateAPG10  20
Typekey 21
Plugassignment  22
Preamplier-blockwiringdiagram  22
Connectiondiagram  23
EC-DeclarationofConformity23
3
Important basic information
Dear customer, dear user,
These installation and operating instructions should provide you with the information you need to properly install and commission
the ow meter. The installation, commissioning and testing are to be performed by trained and qualied personnel only. These ope-
rating instructions must be read and applied carefully to ensure proper, trouble-free and safe operation of the ow meter. VSE is not
liable for any damage incurred resulting from not complying with the instructions in this operating instruction. It is not permitted in
any case to open the device.
These operating instructions for the ow meters of the series VSI High Denition Flow Meterfrom VSE must be stored, so that
they can be read by the group of authorized personnel at any time. Chapters may not be taken of these instructions at any time. A
missing operating instructions manual or missing pages must be replaced immediately. VSE can supply you with new instructions or
you can download the operating instructions from the internet (www.vse-ow.com). The operating instructions must be given to each
subsequent user of this product.
Legal information
This document is not managed by an updating service of VSE Volumentechnik GmbH.
Changes to this document may be made without notice.
VSE Volumentechnik GmbH does not provide any implicit guarantees of commercial qualities and suitability for a specic purpose.
If the device has been opened, modied or incorrectly connected to the electrical circuits, the guarantee of VSE Volumentechnik
GmbH for safe operation is void. VSE Volumentechnik GmbH is not liable in any way for personal injuries or damage to goods
resulting from improper installation or improper operating of the ow meter.
Operating manual–no.: E060024 (E) Date 31
st
May 2006
4

 Istheowmeter?
 Istheuid?
 Istheuidoristherein?
 Whichdoesthesolidmatterpossessandcanit?
 Doestheuidhaveorother?
 Isitnecessarytoinstallapre-switched?
 Aretheandfreeofassemblyresiduessuchasswarf,weldchips?
 Istheandisitensuredthatcangetintothepipelinesystemfromthetank?
 Istheuidoftenchangedandis?
 Arethepipelinesandtheentiresystemcompletely?
 Whatisbeingused?
 Aretheuidandthecleaningagentcompatiblewiththe?
 Arethefortheuidundergoingmeasurement?
Flowmetersmadebymeasurethevolume
owofliquidsaccordingtothetoothedwheelprincipleApairofverypre-
ciselyadjustedtoothedwheelsinthehousingconstitutesthemeterAsignal
pick-upsystemregistersmeterrotationfreeofcontactandtoothbytooth
Inowmetersofhighresolution(VSI),eachtoothisoutputasamultipleof
digitalpulses,dependingoninterpolationsetting
Thegapswithintheteethofthemeterwheels,formmeterchambersinthe
areas,inwhichtheyarecompletelyenclosedbythehousingwalls;these
chambersdigitaliseliquidowdependingontheirchambervolume
Theliquidowwithinonemeterrotationofatoothdivisionisdividedbythe
setinterpolationfactorThisgivesthevolumemeasurementperpulse(V
m
)
andisdenedincm
3
/pulseItidentiestheconstructionalsizeofaow
meter(egVSI1/16)
• General description
Pleasefollowallinstructionsinthisoperatingmanual;onlythisguarantees
trouble-freeoperationoftheowmetersVSEisnotliableforanydamage
ensuingfromnotfollowingoftheseinstructions
Openingthedevicesduringthetermofguaranteeisonlyauthorisedafter
consultationandapprovalofVSE
• Flow meter selection
         
          





• Declaration of Conformity
FlowmetersoftheVSI”productlinearetestedfortheirelectromagnetic
compatibilityandinterferencetransmissionintermsofthelawonelectro-
magneticcompatibilityandcorrespondtothelegalprescriptionsenforced
byEMCdirectivesTheymaynotbeoperatedindependentlyandaretobe
connectedviacabletoapowersourceandsupplydigitalelectricsignals
for electronicevaluationA declarationofconformityissubmittedforall
owmeters,whichyoucanrequestifyourequire


          



Allowmetersaretestedaccordingtothevalid,legallyprescribedelectro-
magneticcompatibilitydirectivesEN55011andEN61000andpossess
theCE-certicationTheEC-declarationofconformityistheCE-labelatta-
chedtoallowmeters
• General conditions for initial start-up
Beforeassemblyandbeforeinitialstart-up,youhavetonotethefollowing
propertiesandaspectsofthecorrespondingcharacteristicsofyoursystem,
sothatatrouble-freeandsafeoperationispossible
• General function description of ow meter
5

 Isthelowerthanthemaxpermittedoperatingpressureoftheowmeter?
 Isthe(onowmeter)belowthemaxpermittedfallofpressure?
 Doesanexcessivelyoccurontheowmeteratmaxow(egwithhigherviscosity)?
 Doestheowrangeoftheowmeter(dependingonviscosity)correspondtothe?
 Notethatowrangedecreasesthe!
 Doesthetemperaturerangeoftheowmetercorrespondtotheofthemedium?
 Istheofthepipelinelargeenoughandarethefallsinpressureinthesystemnotexcessive?
 Isthe(supplyandreverseow)correctlyconnectedandleak-proof?
 Hasthesufcientpowertooperatethesystem?
 AblockingowmetercanstopthewholeowIsaprovidedinthesystem?
• Maximum operating pressure
Beforeassembling the ow meter,you haveto test thatthe maxopera-
tingpressureofthesystemdoesnotexceedthemaxpermittedoperating
pressureoftheowmeterMeanwhile,observethetoppressuresthatcan
occur,whenoperatingthesystem

 Haveyouselectedtheoptimalowmeterandisthisequippedwiththe?
 Doestheoftheowmetercorrespondtotheprovidedvoltage?
 Isthepowersupplyvoltagesuppliedbythemainsorevaluationdevicesufciently?
 Doestheofthepowersupplyvoltagecorrespondtotherequiredpoweroutput?
 Hastheelectricconnectionbeeninstalledbasedontheenclosed?
 IsthecorrectlyconnectedonbothsidesontheearthconductorPE?
 IsthereabetweentheearthconductorPEontheowmeterandtheearthconductorPEontheevaluationdevice?
 Doesacorrectingleadhavetobelaidtoeliminatethebetweentheowmeterandtheevaluationdevice?
 Istheowmeterconnectedrmlytothe(egviathepipelines)?
 IsthemeterconstructedtobetotheearthconductorPE(egconnectionviahoses)?Ifthisisthecase,themeterhastobeconnected
withtheearthconductorPE!
 Isthereaofthecableprotectiveshielding(earthconductorPE)viathehousing,ofthe4or5-pinroundplugtothe
housingoftheowmeter?
 Isthecablelaidfault-freeandtheinstallationsecuredfrominputof?
 Istheoftheconnectioncablermlyscrewedtogetherwiththeplugoftheowmeter?
 Arethewiresonthecorrectlyandproperlyconnected?
 Doestheentiresystemcorrespondtothedirectivesoftheelectromagneticcompatibilitylaws()?
 Havealllocalvalidregulations,,guidelinesandbackgroundconditionsofthe
beenmaintainedandobserved?
 SystemsthatcanleadtopersonalinjurythroughmalfunctionorfailurearetobeequippedwiththeThefunctioning
ofthesesafetydevicesistobecheckedatregularintervals
Thefollowingoperatingpressuresarepermitteddependingonowmeter
version:
 Flowmeteringreycastironversion p
max
=315bar/4500psi
 Flowmeterinstainlesssteelversion p
max
=450bar/6500psi
Important:
Please consult VSE for all operating pressures > 450 bar / 6500 psi and for special versions.
• Statement to EU-Directive 97/23/EG, Pressurized devices
VSEowmeterarepressurizeddevicesaccordingtoarticle1,paragraph
214ofabovementioneddirectiveThereforetheyaresubjecttotheregu-
lationstothisdirective
Accordingtoarticle3,paragraph14,VSEowmetershavetoconform
withthetechnicalrequirementsoftheguidelineTheuidstobemeasured
arebelonginginmostofallcasestotheclass2,denedinarticle9,para-
graph22VSEowmeterdonotreachthelimitvaluesasdenedinarticle
3,paragraph11
ThetechnicalrequirementsforVSEowmetersthereforearelimitedtothe
parts indicated in article 3,paragraph 3 It means the devices have to
be designed and manufactured in conformity with acknowledged engi-
neering, suchas practiced inone of the member states This is herewith
conrmed
BesidethistheparagraphdeclaresthatthesedevicesmustnothaveaCE-
markingaccordingtoDirective97/23/EGThereforewedonotissuede-
clarationsofCEandourproductsarenotlabelledaccto97/23/EG
6
• Assembly of the ow meter
Theowmetershouldbemountedonaneasilyaccessiblelocation,sothat
dismantlingforcleaningthemeterpresentsnoproblemSinceowmeters
canworkinanyinstallationpositionandowdirection,youcanmountiton
anylocationofyoursystemTakecare,wheninstallingtheowmeterthat
liquidalwaysremainsintheowmeter,evenatsystemstandstillandthatit
canneverrunemptyTheoutowoftheowmetershouldthereforealways
showacertainbackpressure,sincethisclampstheowmeterrmlyinthe
liquidcolumn(themetersupportsitselfthroughthisontheliquidcolumn)
andthepipelinecannotrunemptyIncriticalcasesorwhenthepipeline
isatstandstillorstandbyandcanrunempty,werecommendinstallingan
extranon-returnvalveintheoutowline
Flowmetersofthe“VSI”productlinecanbemounteddirectlyontoablock
orintothepipelineusingfourscrewsAlwaysselectlargecrosssections
forthehydraulicsupplyandreturnowrespectivelyfortheentirepipeline
system(ifpossible)Thislowersthefallinpressureandtheowrateinthe
totalsystem
Important:
Make sure that the specied maximum permitted operating pressure of the ow meter
cannot be exceeded, whatever the operating mode of the system. Note the ow meter
range that is dependent on the viscosity of the uid to be measured.
Important:
Make sure that the ow meter is always completely lled both in inow and outow
and that the outow has a little back pressure. This prevents the meter being damaged by
a sudden and steep increase of ow and at the same time improves measurement accuracy.
Fig. 1: Flow meter installation with non-return valve
 
VSI002;VSI004;VSI01;VSI02 15Nm
VSI04;VSI1;VSI2 35Nm
VSI4 120Nm
VSI10 250Nm
VSEsuppliessubplatesforallowmetersoftheVSI”productline;they
havevariouspipethreadsandsideorrearsideconnection(seesubplates
datasheet)Dependingontheprovidedconditions,theinstalledpipeline,
thepipecrosssectionorpipethread,theoperatorcanchoosethesuitable
subplateandincorporatethisintothesystemormachinewithoutadditional
reductions
TheowmeterisscrewedontotheblockorsubplatewithfourDIN912
cheeseheadscrewsThescrewsaretobeevenlypre-tensedcrosswisewith
thefollowingtorques
When changing the fastening screws you must take great care that the
screwsareofpropertyclass109and129

The ow meter is directly mounted onto a subplate or manifold, extra
componentsarenotneededTheblockcontainsthehydraulicsupplyand
outow of the ow meter and the xing bore holes (see ow meter
dimensionsheet)

Please note the special instructions for mounting sizes VSI 4 and VSI 10
(see appendix)
• Flow meter range
Theowmeterrangespeciedintheowmeterdatasheet(Q
min
-Q
max
)
referstothetestinguid“hydraulicoil“withaviscosityof21mm
2
/sata
temperatureof20°CForthisowmeterrange,VSEspeciesmeasurement
accuracyofupto03%ofthemeasurementvalueandarepetitionaccu-
racyof005%
Foruidsof lowerviscosity(<21mm
2
/s)measurement accuracydeteri-
orates, while for uids of higher viscosity (> 21 mm
2
/s) it can improve
Alsonote,however,thattheowmeterrangeisrestrictedincaseofhigher
viscosity(seeowmeterdatasheet)
Non-returnvalve
Flowmeter
Tank
7
• Filtering of liquid
Stronglycontaminateduidsorextraneousmaterialintheuidcanblock,
damageorevendestroytheowmeterAlwaysinstallasufcientlylarge
lterforthesecasesinfrontoftheowmetertopreventdamagetotheow
meterThenecessarylteringdependsonsize,bearingsystemandmodel
ofowmeter
Important:
Please ush out the pipe lines and the tank thoroughly, to prevent contamination
within the ow meter.
Important:
When mounting the ow meter, you must take great care that the seals are not damaged and
correctly placed in the hydraulic connections of the ow meter. Wrongly installed or dama-
ged seals lead to leakage and to a leaky system, which may have dire consequences.
Please make sure that ow meters with EPDM seals do not come into contact with oil and
greases on a mineral oil basis. These uids can decompose the seals.
The yellow plastic plugs in the hydraulic connections of the ow meter protect the meter
against dirt and contamination during storage and shipping. Before mounting the ow meter
you have to remove these plugs so that in- and outow is free and open.
• Cleaning and ushing of pipe lines before initial start-up
Beforeinitialstart-upoftheowmeter,youmustushandcleanthewhole
systemContaminateduidcaneffectthecorrectfunctionoftheowmeter
orseriouslydamagethemeter
Afterpreparingandconnectingupthesystempipes,youmustrstcarefully
ushandcleanthewholepipelinesystemandthe tankTodothis,you
havetomountadiversionplateontotheblockorsubplateinsteadofthe
owmeter,sothattheuidcanowthroughthediversionplateandallex-
traneousmaterial(egswarf,metalchips,etc)canbeushedoutwithout
obstructionUseauidascleansingagent,whichiscompatiblewiththe
uidbeingusedlaterandwhichdoesnotcauseundesirablereactionsYou
canconsultthesuppliersandmanufacturersoftheuidorcontactVSEfor
thecorrespondinginformationVSEsuppliesbypassplatesforallVSI-ow
metersizes
Flowmetersaremeasurementpick-upsystemsmadewithhigh-levelprecisi-
onTheyhaveamechanicalmeterconsistingoftwotoothedwheels,which
isadaptedtothehousingwithnarrowslotsEventhetiniestdamagetothe
toothedwheelsandbearingscancauseameasurementerrorSoalways
makesurethatnoextraneousmaterialgetsintothemeterandthattheuid
owingthroughisalwaysfreefromdirtandcontamination
Afterthesystemhasbeencarefullyushedoutandnoextraneousmaterial
isinthepipeline,youcanmounttheowmeterandcommencetheinitial
start-up
 
VSI002/004/01 10µm
VSI02/04 20µm
VSI1/2/4/10 50µm

For information on lter size for ow meters with plain bearings, in spe-
cial version, or with specially adjusted meter tolerances, please consult
VSE GmbH.
Important:
A blocking ow meter can stop the whole ow. You have to provide a control
valve / bypass for the system.
8
45°
90°
135°
180°
225°
270°
315°
360°
pulse signal from channel 1
pulse signal from channel 2
zero signal pulse (zero pulse)
one measurement value
interpolation factor IPF = 8
V
m
*
= V
m
/ 8
1/2 V
m *
1/4 V
m*
pulse signal from channel 1 without interpolation
pulse signal from channel 2 without interpolation
45°
180°
360°
90°
270°
1 V
m
*
1
1
Division of a single pulse into 360°.
All other signal pulses can be regarded in this way.
Evaluation electronics recognise flow direction from the channel offset of 90°.
Each individual pulse flank is offset 90° and has a value of 1/4 V
m*
.
22.5°
11.25°
33.75°
1.pulse
1 2 3 45 67 8
ThepreampliersofthestandardversionforowmetersoftheVS”pro-
ductlineoutputapulsepertoothgapvolumeV
z
,whichcorrespondstothe
volumemeasurementV
m
(V
m
=V
z
/pulse)Thistakesplaceintwochannels,
sothatyoucanattainamaximumresolutionof1/4V
z
fortheevaluationof
allanksAhigherresolutionisnotpossiblewiththesepreampliers
Butsinceashigharesolutionaspossibleisnecessaryforpreciseandexact
owandvolumemeasurements,thevolumemeasurementV
m
mustberesol-
vedevenmorethanisthecasewithconventionalpreampliersVSEhas
thereforedevelopedthepreamplierwithinterpolation,withwhichaselec-
tableresolutionofupto64anks(16pulses)perperiodcanbeattained
(seetable 3) Thismeans thatyoucan resolve the volume measurement
V
m
withthispreampliertoamaximumof1/64V
m
Fortheevaluation,this
meansthatapartvolumeof1/64V
m
frompulseanktopulseank(for
quadrupleevaluationorankcount)ismeasured,orafullsignalpulseis
countedaspartvolumeof1/16V
m
(pulsecount)(seeg3,interpolation
V
m
/16)
This individually programmed high resolution enables you to set the vo-
lume measurement V
m
optimally for each provided case of application
Furthermore,newapplicationscanbeavailedofwiththehigherresolution
potential
Fig2showstheresolutionofthevolumemeasurementV
m
withaninterpola-
tionfactorof8Thisresolveseachvolumemeasurementintoeightindividu-
alpartvolumesApulseonthesignaloutputofchannel1orchannel2thus
hasavalueofV
m*
=V
m
/8=1/8V
m
perpulseIndoubleevaluation(ank
evaluationofonechannel)thisresultsinavalueof1/2V
m*
=V
m
/16=
1/16V
m
andforquadrupleevaluation(ankevaluationofbothchannels)
theresultisavalueof1/4V
m*
=V
m
/32=1/32V
m
perankEvaluation
electronicscanrecogniseowdirectionfromsignalsoffset90°
• Flow meters with high denition of volume measurement
Measuring,controllingandregulatinginthelowerowrange
Measuring,controllingandregulatinginzeroow
Measuring,controllingandregulatinginbothowdirections
Measuring,controlling,dosingandllingofsmallvolumes
Flowmeterswithinterpolationelectronics(VSI)outputtwodigitalsignals
withprogrammablehighresolutionthatarephase-offset90°(seeg3)
Inadditiontothesignalemission,azerosignalemissionisprovided,which
emitsazerosignalateachfullyregisteredvolumemeasurementV
m
(see
g2)
Fig. 2: Signal emission of the preamplier with interpolation
(interpolation factor 8)
9
Thepreamplierofthe“VSI”productlinehasaprogrammedinterpolation
factor(IPF)withwhichyoucanprogramnew,differentresolutionsHence
you can program a resolution of4 to 64 angular steps (see g 4) per
volumemeasurementV
m
Thefrequencymultiplication“f*”liesbetween1
and16(seetable3)
Column1:ProgrammableinterpolationfactorIPF(programmedinthefactory)
Column2:PulsespervolumemeasurementV
m
Column3:MaximumresolutionofthesignalanksThesignalanksofchannels1and2areevaluated
Column4:VolumemeasurementV
m*
resultingfromthemaximumresolutionofthesignalanks
Column5:Maximumresolutioninangledegreesatresolutionofsignalanks
Column6:Maximumfrequencyf
max*
atmaximumowQ
max
andprogrammedinterpolationfactorIPF
Inpractice,themaximumowQ
max
oftheowmeterisseldomrun,sothatalowerfrequencycanbecalculatedThemaximumfrequencyisthen
calculatedaccordingtothefollowingformula:
f
max
°
Maximumfrequencyoftheowmetersignals
Q
max
°
Maximumowattainedinthecaseofapplicationdescribedhere
IPF Programmedinterpolationfactor
V
m
Volumemeasurementoftheowmeter
Example:Flowmeter1/10…maxowthesystemcanberunonatamaximum
Q
max
°
=40l/min=666667ml/sec;IPF=10;V
m
=1ml/pulse;f
max
°
=666667Hz=666667kHz
Atmaxow
max
°
=40l/min,theowmeter1/10…outputsafrequencyoff
max
°
=666667Hz







*


*




*
1 1 4(quadrupling) V
m
/4 9 f
max
x1
2 2 8 V
m
/8 4 f
max
x2
3 3 12 V
m
/12 30° f
max
x3
4 4 16 V
m
/16 22 f
max
x4
5 5 20 V
m
/20 18° f
max
x5
8 8 32 V
m
/32 11 2 5° f
max
x8
10 10 40 V
m
/40 f
max
x10
12 12 48 V
m
/48 7 5° f
max
x12
16 16 64 V
m
/64 562 f
max
x16

Only the marked lines are illustrated in the diagram of g. 3
°

°


10
Fig. 3: Interpolation of the volume measurement V
m
Important:
Test the connected evaluation electronic system as to whether it can process the maximum
frequency f
max
of the ow meter. Check the data from the following table for the relevant
ow meter, or calculate the maximum frequency data f
max°
with formula 1.
{
Vm /1
{
Vm /2
{
Vm /4
{
Vm /8
{
Vm /16
channel
channel
channel
channel
channel
channel
channel
channel
channel
channel
°0/°063°0 180°
45° 90° 135° 225° 270° 315° 45° 90° 135° 180°
22.5°
22.5°
5.625°
11.25°
16.875°
If the resolution is sixteen-fold, this gives
a recognition of 22.5° (360°/16)
This results in a resolution of 5.625° (22.5°/4)
when evaluating the flanks of channel 1 and 2
The highest resolution permits recognition of
a
measurement volume of 1/64 V .
one measurement volume Vm
360° angle degrees
1. pulse
eslup .2eslup .1
1. pulse 2. pulse 3. pulse 4. pulse
1. 2. 3.
4.
5. 6. 7.
8.
1. 2.
3. 4. 5. 6. 7. 8.
9. 10 11 12
13
14 15
16
1.
2.
m
At initial start-upof the system, you have to program the 

(seetable4,column4)inyourevaluationelectronicsas
parametervalue(egmultiplier)Theevaluationelectronicsthenmultiply
everypulsetheowmeteroutputsbythevolumemeasurementV
m*
and
thuscalculatestheowandthevolumeForowmeterswithhighresolu-
tion, the parameter value  
is dependent on
volumemeasurementV
m
(seetable4,column2)andontheprogrammed
interpolationfactorIPF(seetable4,column3)
PleaseconsultthisrstofallforthevolumemeasurementV
m*
andprogram
thisvalueasparameterintoyourevaluationelectronics
Ontable4youcanviewthecorrespondingvolumemeasurementV
m*
(see
table 4, column 4) with the real, programmed interpolation factor IPF*
(seetable4,column3)TheK-factorincolumn5isameasurementforthe
resolutionperlitre
Youthengetthemaximumfrequencyf
max*
(seetable4,column8)ofthe
signalpulsesforthemaximumowQ
max
(seetable4,column6)withpro-
grammedinterpolationfactorIPF*Thisfrequencyisdependentonthepro-
grammedinterpolationfactorIPFandincreaseswithrisingresolution
V
m
1
st
1
st
1
st
2
nd
2
nd
2
nd
1
st
2
nd
1
st
2
nd





11

 











  1 002 50000 


 16667
2 001 100000 33333
3 000666667 150000 50000
4 0005 200000 66667
5 0004 250000 83333
8 00025 400000 133333
10 0002 500000 166667
12 000166667 600000 200000
16 000125 800000 266667
  1 004 25000 


 16667
2 002 50000 33333
3 001333333 75000 50000
4 001 100000 66667
5 0008 125000 83333
8 0005 200000 133333
10 0004 250000 166667
12 000333333 300000 200000
16 00025 400000 266667
  1 01 10000 


 16667
2 005 20000 33333
3 003333333 30000 50000
4 0025 40000 66667
5 002 50000 83333
8 00125 80000 133333
10 001 100000 166667
12 000833333 120000 200000
16 000625 160000 266667
  1 02 5000 


 15000
2 01 10000 30000
3 006666667 15000 45000
4 005 20000 60000
5 004 25000 75000
8 0025 40000 120000
10 002 50000 150000
12 001666667 60000 180000
16 00125 80000 240000
  1 04 2500 


 16667
2 02 5000 33333
3 013333333 7500 50000
4 01 10000 66667
5 008 12500 83333
8 005 20000 133333
10 004 25000 166667
12 003333333 30000 200000
16 0025 40000 266667
12
 











  1 10 1000 


 13333
2 05 2000 26667
3 033333333 3000 40000
4 025 4000 53333
5 02 5000 66667
8 0125 8000 106667
10 01 10000 133333
12 008333333 12000 160000
16 00625 16000 213333
  1 20 500 


 10000
2 10 1000 20000
3 066666667 1500 30000
4 05 2000 40000
5 04 2500 50000
8 025 4000 80000
10 02 5000 100000
12 016666667 6000 120000
16 0125 8000 160000
  1 40 250 


 10417
2 20 500 20833
3 133333333 750 31250
4 10 1000 41667
5 08 1250 52083
8 05 2000 83333
10 04 2500 104167
12 033333333 3000 125000
16 025 4000 166667
  1 333333333 300 


 26250
2 166666667 600 52500
3 111111111 900 78750
4 083333333 1200 105000
5 066666666 1500 131250
8 041666666 2400 210000
10 033333333 3000 262500
12 027777777 3600 315000
16 020833333 4800 420000
V
m
 = physicalvolumemeasurement(size)oftheowmeter(volumepertoothandpertoothgap)
Q
max
= maxow(testconditions)
f
max
 = maxfrequencyatQ
max
 f
max
=Q
max
/V
m
IPF* = programmableinterpolationfactor
* = allcharactersmarkedwith*refertoIPF*
V
m*
= interpolatedvolumemeasurement V
m*
=V
m
/IPF*;K-factor*=1/V
m*
f
max*
= maxinterpolatedfrequencyatQ
max
 f
max*
=Q
max
/ V
m*
13

1Column: Flowmeter,versionandsize VSI01
2Column: physicalvolumemeasurement
V
m
= 01ml/pulse
(correspondstothevolumemeasurementV
m
atinterpolationfactorIPF*=1)
3Column: Interpolationfactor=hardwareprogrammed IPF*= 10
4Column: Volumemeasurement
V
m*
= 001ml/pulse
5Column: ;reciprocalvalueofvolumemeasurementV
m
*
K-factor*=100000pulse/l
6Column: maximumow

oftheowmeter Q
max
= 10l/min
7Column: maximumfrequency

atinterpolationfactorIPF=1(seecolumn2) f
max
= 16667Hz
(correspondstothevolumemeasurementV
m
atinterpolationfactorIPF=1)
8Column: maximumfrequency
*
atprogrammedinterpolationfactor(seecolumn3) f
max*
= 166667Hz
Thehigh-levelstagesofthepreamplierarefastlinedriversforthreechan-
nelswithacharacteristicwaveimpedanceadjustmentfor75ΩlinesThey
emitthesignalsofchannelA,channelBandthezerochannelZThepush-
pullhigh-levelstagesaredesignedforhighdriverpowerofapprox300mA
on24V;theyarecurrentlimitingandshort-circuit-proofbyswitch-offatex-
cessivetemperatureCapdiodesagainstV
b
andagainstGNDprotectthe
emittanceagainst“echoes”througherroneouslyadjustedlinesandagainst
destructionthroughESD
Aswitchingcircuitinthehigh-levelstagemonitorsthevoltagesupplyU
b
as
wellasthetemperatureandincaseoferrorswitchesallhigh-levelstages
to high impedance The error output is designed as open collector and
likewiseshort-circuit-proof;itsendsanerrormessagetotheexternalelec-
tronicsifthereisamalfunction
Datatransferwith24Vsignalsisdoneasarulewithoutcircuitterminati-
on with awave impedance An erroneously adjusted circuit termination
causesreections,whichcanrunrepeatedlyhitherandthitherifnoadjust-
mentisprovidedonthesendersideIfthereisafastsequenceofpulses,
transmissionisinterferedwithbythesereections
Iftherearelongconnectionlinesbetweenthepreamplierandtheevalu-
ationelectronicsandalsoahighresolution,lineswithawaveimpedance
ofapprox40Ωto150Ωshouldbeusedandacorrespondingtermination
impedanceconnectedtotheevaluationelectronicsAnoptimaladjustment
ofthetransmissionlineandtheterminationwaveimpedancescanpermit
linelengthsofuptoapprox150…200m
Highinterferenceprotectionisachievedbyalargeoutputamplitudeand
bytheintegratedwaveadjustmentWhenthesignalsonthereceiverside
areguidedthroughoptocouplers,youfurthermoregetagalvanicsepara-
tionbetweensenderandreceiversideandpotentialdifferencescanthus
beavoided
Thereectionofreturnsignalsispreventedinthehigh-levelstageofthe
preamplierbyanintegratedwaveimpedanceadjustmentThisadjustment
enhancesinterference-proofperformance
Scansensor: GMRsensorwithdiplexer;orotherscansensors,whichsupplyasineandcosinesignal;
Numberofsensors: Twoscansensorsforgeneratingthesineandcosinesignal;
Adjustment: Offsetadjustmentbymeansoftwopotentiometers;
Resolution: Programmableinarangeof1–64ankspervolumemeasurementV
m
;
Frequency: Frequencymultiplication:programmableinarangeof1–16timesthefrequencyofthescansensors
Emissionsignals: ChannelA,channelB,zerochannelZ
ChannelAandB: Twosignaloutputsforemittingthedigitalowsensorsignals;betweenchannelAandchannelBthereisachanneloffsetof90°;
Flowdirection: RecognitionofowdirectionfromchanneloffsetofthesignalsfromchannelAtochannelB
ZerosignalZ: Zerosignal,markstheowofonevolumemeasurementV
m
;
Outputs: 3currentlimitingandshort-circuit-proofhigh-levelstages(channelA,channelB,zerosignalZ);integratedadjustmenttoa75Ω
waveimpedance;drivercurrentapprox300mAat24Vpowersupply;smallsaturationvoltageupto30mAloadcurrent;short
switchingtimes;integratedfree-rundiodesagainstV
b
andGND;temperatureprotectionswitchingwithhysteresis;outputsareof
highimpedanceincaseoferror;
The24Vlinedriversaredesignedforcontroltechnologywithcableadjustment;
Errormessages: Short-circuit-proofoutputforissuingerrormessagesatexcessivetemperatureorlowvoltage,lowactive
Operatingvoltage: V
b
=8…28VDC
Currentconsumption:I
noload
=approx40mA;totalcurrentconsumptiondependingonloadingofoutputs;
• Technical specications of preamplier
14
Fig4showstheplugassignmentofthepreamplierAsyoucansee,this
plughasvepins,theouterfourassignedexactlyastheonesinstandard
preampliersInadditiontothepowersupplyandsignaloutputofchannel
1and2,thereisafthpin,whichemitsthezerosignal
Asarule,onlythefourouterpinsareneededtoevaluatetheowmeter
signals;pinassignmentisthereforethesameasinpreviouslydescribedpre-
ampliersForthisreasonyoucanuseanormal4-wiredconnectioncable
fortheconnectionoftheowmeterButpleasemakesurethattheshielding
ofthecableislaidontheplugsideontothemetalhousingoftheplug
TheshieldingoftheconnectioncableisplacedonbothsidesTheearth
conductorPEisconnectedviatheshieldingfromtheevaluationelectronics
tothepreamplierhousingandthemeterThecableshieldingshouldal-
waysbelaidcontinuouslyasfarastheowmeterandnotinterruptedin
crossconnectorsorbranchsocketsLaytheconnectioncableasdirectly
aspossiblefromtheevaluatingdevicetotheowmeter,sinceinterruptions
arealwaysapotentialsourceoferror
TheowmetermustbeconnectedelectricallywiththeearthconductorPE
Thisisnormallysecuredbytheearthedpipelines



• Plug assignment of preamplier
Important:
Only use well-shielded cables for the connection cable, with a wire cross section of
≥ 4 or 5 x 0.25 mm
2
. Please make sure that the housing of the round plug is metallic, that it
has a connection for the shielding and that the potential of the earth conductor PE is
connected with the cable shielding and the housing of the preamplier.
Important:
Please make sure that no extra inductive elements are connected in the power supply of the
ow meter, such as contactors, relays, valves etc.
These components are potential sources of interference (especially if the inductive elements
are not provided with an adequate protective circuit), generate high interference pulses,
when switched and can interfere with the functioning of the ow meter, although this
complies with the electromagnetic compatibility directives.
Working life is dependent on operating conditions and thus the specic properties of the
devices, limited through wear, corrosion, deposits or age. The operator is responsible for
regular control, maintenance and recalibration. Any indication of a malfunction or damage
prohibits any further use. On request, we can supply you with a borrowed device for the
duration of repair or overhauling. We advice to a yearly control and recalibration.
• Maintenance
pin 1
power supply
U = 10 – 28 V DC
pin 4
digital signal
2. channel
pin 3
power supply
GND (-U = 0 V)
pin 2
digital signal
1. channel
Top view of plug
metal housing connected
with the shielding and the
earth conductor PE
pin 5
digital signal
zero signal
Fig. 4: Flange plug installed in the preamplier housing of the ow meter
b
b
15
 








 0002…2 1667*IPF…166667*IPF 002/IPF 50000*IPF
 0004…4 1667*IPF…166667*IPF 004/IPF 25000*IPF
 001…10 1667*IPF…166667*IPF 01/IPF 10000*IPF
 002…18 1667*IPF…150000*IPF 02/IPF 5000*IPF
 003…40 1250*IPF…166667*IPF 04/IPF 2500*IPF
 005…80 0833*IPF…133333*IPF 1/IPF 1000*IPF
 01…120 0833*IPF…100000*IPF 2/IPF 500*IPF
 10…250 4167*IPF…104167*IPF 4/IPF 250*IPF
 :upto03%ofmeasurementvalue(withviscosity>20mm
2
/s)
 :±005%underthesameoperatingconditions
 :CastironEN-GJS-400-15(EN1563)or
Stainlesssteel14305
 :Ballbearingsorsteelplainbearings(medium-dependent)
 :FPM(standard),NBR,PTFEorEPDM
 :CastironEN-GJS-400-15(EN1563)315bar/4500psi
Stainlesssteel14305 450bar/6500psi
 :–40…+120°C(–40°F…248°F)
 :–20…+50°C(–4°F…122°F)
 :1…100000mm
2
/s
 :any
 :any
 :max72db(A)
 :10to28volts/DC
 :3currentlimitingandshort-circuit-proofhigh-levelstages
lowsignal:0=GND;highsignal:1=U
b
-1
 :90°±30°max
 :1/1±15°max
 :Aluminium
 :IP65
Itisimperativethatyouencloseanexactdescriptionofthecomplaint,ob-
jection or fault, when returning the device so as to ensure a rapid and
economic repairof theow meters andother componentsFurthermore,
youmustincludeasecuritysheet,whichinformsunambiguously,whichuid
wasrunwiththeowmeterandhowdangerousthisuidis
Themaintenanceoflegalregulationsasregardsworksafety,suchaswork-
placeregulations,accidentpreventionregulations,andstipulationsonen-
vironmental protection, waste disposal and the water management law,
obligesindustrial corporations to protect theiremployeesand otherper-
sonsandenvironmentagainstharmfuleffects,whenhandlinghazardous
materialsIf furthersafetyprecautionsare stillnecessary despitecareful
emptyingandcleaningoftheowmeter,informationonthisisimperative
andmustbeincludedwiththereturneddespatch
When returning ow meters toVSE Volumentechnik GmbH, please note
thatinspectionandrepairwillonlybeperformedifthe
oftheutiliseduidisenclosedandtheowmeterscompletely
cleanedandushedThisprotectsouremployeesandsimpliesourwork


• Sending back of repairs and sample devices
• Technical specications VSI 0.02 / IPF – VSI 4 / IPF
Adjustable interpolation factors IPF: 1; 2; 3; 4; 5; 8; 10; 12; 16
16
• Flow response curves VSI 0.02 – VSI 4
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
s
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
FlowQ
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
FlowQ
Flowresistance∆p
17
• Dimensions VSI 0.02 – VSI 4


 




 100 80 91 M6 120 9 114 58 70 40 20 11x2 28 34
 100 80 92 M6 11 5 9 115 59 70 40 20 11x2 28 34
 100 80 94 M6 9 9 117 61 70 40 20 11x2 28 34
 100 80 94 M6 95 9 117 61 70 40 20 11x2 30 37
 115 90 965 M8 11 5 16 120 635 80 38 34 1796x262 40 50
130 100 101 M8 12 5 16 124 68 84 72 34 1796x262 53 68
130 100 118 M8 15 16 141 85 84 72 34 1796x262 67 84
180 140 145 M12 20 30 168 110 46 95 45 3617x262 147 184

Dimensions in mm


Stainlesssteelversion
Connectiondiagram
Housingwithoutmillededge
Castironversion
Connectiondiagram
O-ring
Castironversion
Plugpositionfor
VSI002toVSI04andVSI4
Earth
Centringborehole
Centringborehole
PlugpositionforVSI1/VSI2
M12x1
M12x1
18
• Dimensions, subplates AP. 02 - 4






 






G1/4“
G3/8“
G1/2“
35
20
80 90 40 70
26
100 M6/12 18
23 30
28
38

G1/2“ 35 28
90 100 38 80
46
115 M8/15 27
G3/4“ 40 33 52
G1/2“ 35 28
100 110 72 84
46
130 M8/15 36G3/4“ 40 33 52
G1 55 41 55
G11/4“ 70 51
120
130 100
110 60
M8/15
7 4
*G11/2 70
56
120
72
G11/2“ 80 140 110 180 120
only for AP . 4 U...
Flowmeter
Subplate
Flowmeter
Subplate
19
 








VSI10 15…525 750*IPF…262567*IPF 3333/IPF 300*IPF
 :upto05%ofmeasurementvalue(withviscosity>20mm
2
/s)
 :±005%underthesameoperatingconditions
 :CastironEN-GJS-400-15(EN1563)
 :Ballbearingsorsteelplainbearings(medium-dependent)
 :70kgwithoutsubplate
 :FPM(standard),NBR,PTFEorEPDM
 :420bar/6000psi
 :–40…+120°C(–40°F…248°F)
 :–20…+50°C(–4°F…122°F)
 :5…100000mm
2
/s
 :any
 :any
 :db(A)
 :10to28volts/DC
 :0…2625Hz
 :3currentlimitingandshort-circuit-proofhigh-levelstages
lowsignal:0=GND;highsignal:1=U
b
-1
 :90°±30°max
 :1/1±15°max
 :Aluminium
 :IP65
• Technical specications VSI 10 / IPF
Adjustable interpolation factors IPF: 1; 2; 3; 4; 5; 8; 10; 12; 16
• Flow response curves VSI 10
FlowQ
Flowresistance∆p
20
• Dimensions VSI 10

• Dimensions, subplate APG 10.

Dimensions in mm
Dimensions in mm
Weight70kg
ConnectionsurfaceVSI10
View X
Weight21kg
Weight30kg
ConnectionsurfaceVSI10
View X
View X
Liftingeyebolt
Barhandle
Liftingeyebolt
O-ring
Connectiondiagram
Earth
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57

vse VSI 0.02 Operating instructions

Type
Operating instructions
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI