Fluidwell F018 Owner's manual

Category
Measuring, testing & control
Type
Owner's manual
User Manual
F018-P
FLOW RATE MONITOR / TOTALIZER
WITH LINEARIZATION, PULSE/ALARM AND ANALOG
OUTPUTS AND HART COMMUNICATION
Signal input: pulse, NAMUR or coil flowmeter signal
Signal output: one analog reflecting flow rate
and one flow rate alarm
or scaled/unscaled pulse
Options: intrinsically safe
backlight
HART communication
F-Series - Field mounted indicators for safe and hazardous areas
More info: www.fluidwell.com/fseries
User Manual F018-P
Page 2 FW_F018-P_M_v0501-03_EN
TABLE OF CONTENTS
1 ABOUT THIS MANUAL............................................................................................................ 4
1.1 How to use this manual .................................................................................................. 4
1.2 Use of pictograms........................................................................................................... 4
1.3 Warranty and technical support ...................................................................................... 4
1.4 Model Reference ............................................................................................................ 4
2 SAFETY.................................................................................................................................... 5
2.1 Personal safety ............................................................................................................... 5
2.2 End-user responsibilities................................................................................................. 5
2.3 Potential equipment damage .......................................................................................... 5
2.4 Disposal of electronic waste ........................................................................................... 5
3 INTRODUCTION....................................................................................................................... 6
3.1 System description ......................................................................................................... 6
3.2 Product features ............................................................................................................. 6
3.3 Installation example ........................................................................................................ 7
4 OPERATION............................................................................................................................. 8
4.1 Introduction ..................................................................................................................... 8
4.1.1 Operating modes ............................................................................................................ 8
4.2 Control panel .................................................................................................................. 8
4.2.1 Display............................................................................................................................ 8
4.2.2 Backlight (Type ZB) ........................................................................................................ 9
4.3 Displayed information ..................................................................................................... 9
4.3.1 Flow rate and total or flow rate ....................................................................................... 9
4.3.2 Clear total ....................................................................................................................... 9
4.3.3 Accumulated total ........................................................................................................... 9
4.4 Setting alarm thresholds ................................................................................................. 9
4.5 Operator alarms.............................................................................................................. 10
4.5.1 Flow rate alarm............................................................................................................... 10
4.5.2 Low battery ..................................................................................................................... 10
4.5.3 Internal error ................................................................................................................... 10
5 CONFIGURATION.................................................................................................................... 11
5.1 Introduction ..................................................................................................................... 11
5.2 Configuring using SETUP mode..................................................................................... 11
5.2.1 Entering SETUP mode ................................................................................................... 11
5.2.2 Navigating the SETUP menu.......................................................................................... 11
5.2.3 Changing configuration settings ..................................................................................... 12
5.2.4 Returning to Operator Mode........................................................................................... 12
5.3 SETUP menu overview................................................................................................... 13
5.4 SETUP menu explanations............................................................................................. 14
5.4.1 Menu 1: Total ................................................................................................................. 14
5.4.2 Menu 2: Flow rate .......................................................................................................... 15
5.4.3 Menu 3: Alarm ............................................................................................................... 16
5.4.4 Menu 4: Display ............................................................................................................. 17
5.4.5 Menu 5: Flowmeter ........................................................................................................ 18
5.4.6 Menu 6: Linearization .................................................................................................... 18
5.4.7 Menu 7: Analog output ................................................................................................... 20
5.4.8 Menu 8: Pulse output ..................................................................................................... 21
5.4.9 Menu 9: Others .............................................................................................................. 21
6 INSTALLATION........................................................................................................................ 22
6.1 Installation / environmental conditions............................................................................ 22
6.2 Handling the F018-P enclosure ...................................................................................... 23
6.2.1 Identification.................................................................................................................... 23
6.2.2 Opening, assembling and closing the F-Series .............................................................. 24
6.3 Mechanical installation.................................................................................................... 25
6.3.1 Mechanical dimensions .................................................................................................. 25
6.3.2 Mounting the F-Series .................................................................................................... 27
6.4 Electrical installation ....................................................................................................... 28
6.4.1 Electrical safety............................................................................................................... 28
6.4.2 Protective earth (PE) connections .................................................................................. 28
6.4.3 Field wiring connections ................................................................................................. 29
6.4.4 Power supply .................................................................................................................. 30
6.4.5 Sensor supply................................................................................................................. 30
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 3
6.5 Terminal connectors safe area applications - Type PX, PD ........................................... 30
6.5.1 Terminals 1-3 (6): flowmeter input.................................................................................. 31
6.5.2 Terminals 4-5: Power supply - type PX and PD ............................................................. 34
6.5.3 Terminal 6: Sensor power supply - Type PD.................................................................. 34
6.5.4 Terminals 7-8: Pulse or Alarm output - Type OT............................................................ 34
6.5.5 Terminals 9-10: Backlight power supply - Type ZB ........................................................ 34
6.5.6 Terminals 11-12: Analog output + output loop power - Type AH.................................... 35
6.5.7 Terminals 11-12: HART communication - Type CR ....................................................... 35
7 INTRINSICALLY SAFE APPLICATIONS ................................................................................ 36
7.1 Identification.................................................................................................................... 36
7.2 Electrical installation in hazardous areas........................................................................ 37
7.2.1 Installations based on FM or CSA Certificate................................................................. 37
7.2.2 Electrical data - Control drawing..................................................................................... 38
7.2.3 Installations based on ATEX or IECEx Certificate.......................................................... 39
7.2.4 Electrical data - Annex 1................................................................................................. 40
7.2.5 Power supply .................................................................................................................. 41
7.2.6 Sensor supply................................................................................................................. 41
7.3 Terminal connectors hazardous area applications ......................................................... 41
7.4 Configuration examples Intrinsically Safe applications................................................... 42
7.4.1 F018-P-AH-CR-OT-PX-XI-(ZB) - Ex ia IIC/IIIC............................................................... 42
7.4.2 F018-P-AH-CR-OT-PX-XI-ZB - Ex ia IIC/IIIC ................................................................. 43
7.4.3 F018-P-AH-CR-OT-PD-XI-ZB - Ex ia IIC/IIIC ................................................................. 43
8 MAINTENANCE........................................................................................................................ 44
8.1 General directions........................................................................................................... 44
8.2 Instructions for repair ...................................................................................................... 44
8.3 Battery replacement........................................................................................................ 45
8.3.1 Safety instructions .......................................................................................................... 45
8.3.2 Battery replacement procedure ...................................................................................... 46
8.3.3 Disposal of batteries ....................................................................................................... 46
APPENDIX A - TECHNICAL SPECIFICATION .............................................................................. 47
A.1 General ........................................................................................................................... 47
A.2 Input................................................................................................................................ 49
A.3 Output ............................................................................................................................. 49
A.4 Operational ..................................................................................................................... 50
APPENDIX B - TROUBLESHOOTING ........................................................................................... 51
APPENDIX C - HART COMMUNICATION...................................................................................... 52
APPENDIX D - LEGAL INFORMATION ......................................................................................... 55
D.1 Declarations of Conformity ............................................................................................. 55
User Manual F018-P
Page 4 FW_F018-P_M_v0501-03_EN
1 ABOUT THIS MANUAL
1.1 HOW TO USE THIS MANUAL
This operation manual is divided into two main parts:
The daily use of the F018-P is described in Section 4: Operation[»8]. These instructions are
meant for users / operators.
All subsequent sections and appendices are meant for electricians/technicians. These provide a
detailed description of all software settings and hardware installation guidance.
Always follow the instructions in this manual, especially the safety information. For additional
information, please contact your supplier.
1.2 USE OF PICTOGRAMS
A hazardous situation may occur if the unit is not used for the purpose it was designed for or is used
incorrectly. The following pictograms inform you of important information:
A warning indicates a hazardous situation that could cause injury or death.
A caution indicates a condition that could cause:
moderate/minor personal injury or equipment damage.
incorrect functioning of the unit or connected instruments.
A note informs you of important information.
1.3 WARRANTY AND TECHNICAL SUPPORT
For warranty and technical support on your Fluidwell products, visit our internet site
www.fluidwell.com or contact us at [email protected].
1.4 MODEL REFERENCE
Hardware version: 03.04.xx
Software version: 03.06.xx
Document version: FW_F018-P_M_v0501-03_EN
© Copyright 2023 Fluidwell B.V. - The Netherlands
Information in this document is subject to change without prior notice. The manufacturer is
not responsible for mistakes in this material or for incidental damage caused as a direct or
indirect result of the delivery, performance or use of this material.
© All rights reserved. No parts of this publication may be reproduced or used in any form or
by any means without written permission of your supplier.
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 5
2 SAFETY
2.1 PERSONAL SAFETY
Explosion hazard: Never open the unit when explosive atmosphere is present, and the unit is
connected to a power supply or consuming device like the internal battery supply.
Risk of electric shock: Only open the unit if all leads are free of potential electrical energy.
Immediately inform the person responsible for the installation if you disagree with the safety
precautions or if you detect errors or danger.
Intrinsically Safe applications: follow the instructions as mentioned in Section 7: Intrinsically safe
applications[»36].
Certificates and related documents are available on our website or from your distributor.
2.2 END-USER RESPONSIBILITIES
Installation, use, and servicing of the unit must be done by authorized technicians.
This unit is not designed for use in life support appliances, devices, or systems where
malfunction of the product could result in personal injury. Customers using or selling these
products for use in such applications do so at their own risk and agree to fully indemnify the
manufacturer and supplier for any damages resulting from such improper use or sale.
Check all connections, settings and technical specifications of all peripheral devices supplied
with the unit.
The unit must be installed in accordance with the EMC (Electro Magnetic Compatibility)
guidelines.
Connect Protective Earth (PE) / proper grounding to the metal enclosure as indicated in
Section 6.4.2: Protective earth (PE) connections[»28].
If the unit is used on a ship, truck or somewhere else with no Protective Earth (PE) connection
available, connect an alternative grounding suitable for the application to the metal enclosure.
The Protective Earth (PE) wire (or alternative grounding) may never be disconnected or
removed.
The local labor and safety laws, and regulations must be adhered to.
Modifying the unit without prior written consent from the manufacturer will result in the
immediate termination of product liability and warranty period.
The manufacturer accepts no responsibility whatsoever if the safety information and instructions
in this manual are not observed.
2.3 POTENTIAL EQUIPMENT DAMAGE
Before installing the unit, check the mains voltage and information on the manufacturer's label
for product configuration.
Electro Static Discharge (ESD) causes irreparable damage to electronics. When installing or
opening the unit, the technician should first discharge him- or herself by touching a well-
grounded object.
Never touch the electronic components.
Never expose the unit to conditions that may exceed the enclosure classifications as indicated
by the manufacturer's label and Section 6.1: Installation / environmental conditions[»22].
2.4 DISPOSAL OF ELECTRONIC WASTE
At the end of its life this product should be disposed of according to the
(inter)national regulations regarding waste electronic equipment. If a
battery is installed in this product, it should be disposed of separately. The
separate collection and recycling of your waste equipment will help to
conserve natural resources and ensure that it is recycled in a manner that
protects the environment.
User Manual F018-P
Page 6 FW_F018-P_M_v0501-03_EN
3 INTRODUCTION
3.1 SYSTEM DESCRIPTION
The flow rate monitor / totalizer model F018-P is a microprocessor-driven instrument designed to
linearize the flowmeter’s flow curve and to display the current flow rate, resettable total and non-
resettable accumulated total. At the output side, it is able to generate a scaled pulse according to the
accumulated total or an unscaled pulse reflecting the flowmeter pulse signal, generate high and/or
low flow rate alarms and is supplied with a HART- enabled analog output representing the current
flow rate.
Fig.1: The F018
This manual describes the daily use, configuration and installation of the F018-P with pulse input
from a flowmeter and its available options.
The following figure shows the F018-P used in a typical application.
Alarm output Pulse output
mA
Analog output
Linearization
or
Fig.2: Typical application
3.2 PRODUCT FEATURES
General characteristics
15 point linearization table
One output reflecting totalized flow or flow alarm
Possibility to transmit unscaled flowmeter signal
Ultra-low power consumption to allow for long battery life (Type PB/PC)
Various mounting possibilities
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 7
A choice of enclosure materials: GRP, aluminum or stainless steel (316L), suitable for every
industrial environment
Ability to process a broad range of sensor signals
Simple and standard configuration and operator interface.
Configuration
The F018 can be used for many types of application and has a SETUP mode to allow for configuring
it to your requirements. See Section 5: Configuration[»11] for further information.
All settings are stored in an EEPROM memory so they will not be lost if there is a power failure.
Input
One flowmeter with a passive or active pulse, NAMUR or coil signal output
Several options are available for powering the sensor.
Output
A pulse output is available from the F018-P, reflecting the passing of a programmable quantity
or the presence of an alarm.
Pulse output can also be configured to retransmit the flow meter input pulse signal.
An analog output (4 - 20 mA) reflecting the linearized flow rate is available.
Several driving options are available, depending on the ordered Type.
Options
Intrinsic safety (Type XI)
HART-enabled analog output (4 - 20 mA, Type CR)
Multidrop communication
Device description (DD) available for use with a HART handheld or HART Host
Power and sensor supply (Types PB, PC, PD, PX, in various combinations)
Panel mount enclosure (Types HB, HC)
Flame-proof enclosure (Type XF)
Display backlight (Type ZB).
3.3 INSTALLATION EXAMPLE
Following parts can be recognized in below installation example.
13
24
56
Fig.3: Installation example F018-P
1. Front cover 3. Label 5. Keys
2. Display 4. Back cover 6. Cable entry (process mount)
User Manual F018-P
Page 8 FW_F018-P_M_v0501-03_EN
4 OPERATION
This unit may only be operated by authorized and trained personnel who have read and understood
this manual, particularly Section 2: Safety[»5].
4.1 INTRODUCTION
This chapter describes the daily use and operation of the F018-P. For this, the F-Series is equipped
with a control panel that provides the operator with various functions, information and operating
modes.
4.1.1 OPERATING MODES
The following operating mode are available:
OPERATOR-mode
This is the default mode of operation and mode indicator
RUN
is shown on the display. This
mode is available for all operators and gives access to the other modes. The available functions
and rights depend on the type of unit and configured settings. The unit will always return to this
mode when there is no user interaction for some time in any of the other modes.
SETUP-mode
This mode allows to review all configuration settings using a SETUP-menu. The mode indicator
SETUP
is shown on the display and access can be restricted via a password. After selecting a
configuration item, it can be changed using the PROGRAM-mode.
PROGRAM-mode
This mode is used to change settings in the SETUP-menu, enter values in OPERATOR-mode or
enter a password. When a setting is being changed, this is indicated by the mode indicator
PROGRAM
on the display.
The following paragraphs describe the modes of operation in more detail. SETUP-mode and
PROGRAM-mode are described in Section 5: Configuration[»11].
4.2 CONTROL PANEL
The control panel is located on the front side of the F-Series.
A Liquid Cristal Display (LCD) to show process values and various other information.
Three keys (PROG/ENTER, SELECT, CLEAR) to control the unit during operation.
An optional backlight to illuminate the display (Type ZB).
Functions of the keys in Operator Mode
PROG/ENTER-key
This key is used to program and save new values or settings.
The PROG-key is also used to gain access to SETUP mode, see Section 5:
Configuration[»11].
SELECT-key
This key is used to select the displayed information.
CLEAR-key
This key is used to clear the value of total.
4.2.1 DISPLAY
In the F018-P, all inputs and outputs are continuously processed in the background. The relevant
process information is then displayed on a selection of screens when the product is in OPERATOR-
mode.
The display contains one line with larger digits and one line with smaller characters below that. The
top line displays key process information, while the bottom line usually displays additional
information or system messages. Also, measurement unit and symbols to indicate operating mode,
function, and status may be displayed.
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 9
Normally, the display is updated depending on the refresh rate selected in the configuration settings.
By pressing any key, the display switches to refreshing the information 8 times per second. After 30
seconds of key inactivity, the display returns to the configured refresh rate.
RATE
TOTAL
RUN
3
m
Fig.4: Example of display information during process
4.2.2 BACKLIGHT (TYPE ZB)
The F018-P with Type ZB is equipped with a backlight to illuminate the display for improved
readability. The backlight can be turned on or off, and its intensity can be set in the SETUP- menu.
4.3 DISPLAYED INFORMATION
4.3.1 FLOW RATE AND TOTAL OR FLOW RATE
The main display shows the totalized flow with the large digit line and the actual flow rate on the
bottom line. After selecting any other information, the F018 will always return to this display
automatically.
The configured measurement unit for Total is indicated on the top line of the display, the unit for flow
is included in the bottom line. Please note that the display can be configured to show the actual flow
rate only, using the large digits (see
SETUP
4.1: DISPLAY > FUNCTION). In both cases, accumulated total
will be shown after pressing the SELECT key.
When “-------” is shown, the flow rate is too high to be displayed.
The arrows ▲▼ indicate the process value trend: increasing or decreasing.
4.3.2 CLEAR TOTAL
The value for total can be re-initialized. To do so, press CLEAR twice. After pressing CLEAR once,
the flashing text "PUSH CLEAR" is displayed. To avoid re-initialization at this stage, press another
key than CLEAR or wait for 20 seconds.
Re-initialization of total DOES NOT influence the accumulated total.
4.3.3 ACCUMULATED TOTAL
After pressing SELECT, accumulated total is displayed.
The accumulated total cannot be re-initialized. The value will count up to 99,999,999,999. The unit
and number of decimals are displayed according to the configuration settings for total.
4.4 SETTING ALARM THRESHOLDS
Setting the threshold values might not be available in Operator Mode. This is set in
SETUP
3.1: ALARM >
ALARMSET.
When the SELECT key is pressed several times, the set alarm thresholds are displayed. The
measuring unit applied is set in
SETUP
2.1: FLOW RATE > UNIT.
To enter alarm values, follow these steps:
1. Press PROG. The text PROGRAM will flash or a pass code is requested
2. Use ►to select a digit and ▲to increase it
3. Confirm the new value by pressing ENTER.
User Manual F018-P
Page 10 FW_F018-P_M_v0501-03_EN
PROGRAM
Fig.5: Example of display information during programming minimum flow rate alarm level
When a value was changed but ENTER has not been pressed, the change can be canceled by
waiting for 20 seconds or by pressing ENTER and holding for three seconds.
4.5 OPERATOR ALARMS
4.5.1 FLOW RATE ALARM
When the actual flow rate is outside the allowed range, an alarm message will be displayed
indicating the type of alarm: "LO RATE", "HI RATE". The alarm is terminated automatically as soon
as the flow rate is within its range again.
4.5.2 LOW BATTERY
When the battery voltage drops, it must be replaced. To warn in time, an icon “LOW BATTERY” will
flash on the left side of the display. The F018 will continue to function correctly for several days up to
some weeks, but as soon as the warning stop flashing and is displayed continuously the battery
must be replaced shortly after, or it will run out.
See Section 8.3: Battery replacement[»45] for further information.
RUN
Fig.6: Example of low battery alarm
4.5.3 INTERNAL ERROR
When an internal error occurs, “ALARM” will be displayed. Press the SELECT key several times to
display the 4-digit error code. Consult Section B: Troubleshooting[»51] for more information.
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 11
5 CONFIGURATION
5.1 INTRODUCTION
This chapter describes how technicians can use configuration settings to configure the F-Series for
optimal functionality. Configuration of the F018-P can be done in SETUP mode, using the front keys.
5.2 CONFIGURING USING SETUP MODE
For an overview of Operating modes, see Section 4.1.1: Operating modes[»8].
Changing settings may influence current operation
In SETUP mode the unit remains fully operational. Make sure the product is not being used for any
application when changing settings.
5.2.1 ENTERING SETUP MODE
1. Press the PROG-key for 7 seconds. During this, the symbols ▲▼ are displayed. You may be
prompted to enter a password.
2. The
SETUP
indicator is now activated on the display and the
RUN
indicator is deactivated.
5.2.2 NAVIGATING THE SETUP MENU
The SETUP menu contains several menu groups, which each contain various menu items. Each
menu item has a unique number which is displayed at the bottom of the display. Additionally, each
menu item is expressed with a keyword.
The menu number is a combination of two figures separated by a dot:
the first figure indicates the selected menu group (e.g. 1. TOTAL).
the second figure indicates the menu item (e.g. 1.1 UNIT).
The following image shows the layout of the menu structure:
Fig.7: SETUP menu layout and navigation
Navigate the SETUP menu with the following functions:
PROG-key
When a menu item is selected (e.g. 1.1), this key is used to start the programming
sequence.
When a menu group is selected (and no function), this key is used to scroll back a group
(e.g. 3. à 2. à 1. à 3. à 2. ...)
SELECT-key
This key is used to select the next menu item in the list (e.g. 1. à 1.1 à 1.2 à 1.3 à
1. ...). When the top of the list is reached, it will wrap around and return to the menu group
selection.
CLEAR-key
This key is used to select the previous menu item in the list (e.g. 1.3 à 1.2 à 1.1 à 1. à
2. ...). When the bottom of the list is reached, it returns to the menu group selection.
When a menu group is selected (and no item), this key is used to scroll to the next group
(e.g. 1. à 2. à 3. à 1. à 2. ...)
User Manual F018-P
Page 12 FW_F018-P_M_v0501-03_EN
5.2.3 CHANGING CONFIGURATION SETTINGS
A menu item either contains a value (a number with optionally a decimal point and sign, e.g.
-123.45) or a selection list (e.g. L – m3 - USGAL). After a menu item is selected in the SETUP menu,
a new value can be programmed by performing the following steps. During the programming
sequence, the
PROGRAM
indicator is shown on the display.
Step 1: Start the programming sequence
PROG-key
When a menu item is selected in the SETUP menu, this key is used to start the
programming sequence.
Step 2a: Change a value
SELECT-key
This key is used to increment the selected digit or to select de next position of the decimal
point.
When the entered value is out of range, the increase sign ▲ or decrease sign ▼ will be
displayed; the value will be automatically brought within range when confirmed later.
CLEAR-key
This key is used to select the next digit. If a decimal point can be set, this will be included
in the sequence as well (e.g. [decimal point] à digit 1 à digit 2 à digit 3 à [decimal
point]).
SELECT-key + CLEAR-key
When a value can also be entered as a negative number, pressing the SELECT-key and
CLEAR-key simultaneously will toggle the ‘–‘ (minus) sign on and off.
Step 2b: Change the selection in a list
SELECT-key
This key is used to select the next item in the list (e.g. L àà USGAL).
At the end of the list, the selection will wrap around to the first selection.
CLEAR-key
This key is used to select the previous item in the list (e.g. USGAL àà L).
At the bottom of the list, the selection will wrap around to the last selection.
Step 3: Finish the programming sequence
PROG-key
During the programming sequence, this key is used to confirm or cancel the new value
and return to SETUP mode. To cancel the operation, either press the PROG-key for 3
seconds or wait for 20 seconds: the programming sequence is cancelled and the former
value is reinstated.
5.2.4 RETURNING TO OPERATOR MODE
When all configuration settings are set correctly, return to OPERATOR mode by pressing the
PROG-key for three seconds, or do not press any key for 2 minutes to return automatically. The
RUN
indicator is now activated on the display and the
SETUP
indicator is deactivated.
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 13
5.3 SETUP MENU OVERVIEW
1 TOTAL DEFAULT
1.1 UNIT L - m3 - kg - lb - GAL - USGAL - bbl - no unit L
1.2 DECIMALS 0 - 3 0
1.3 K-FACTOR 0.000010 - 9,999,999 1
1.4 DECIMALS K-FACTOR 0 - 6 0
2 FLOW RATE DEFAULT
2.1 UNIT L/min - L/hr - m3/sec - m3/min - m3/hr - m3/day - g/sec - g/
min - g/hr - kg/sec - kg/min - kg/hr - kg/day - ton/min - ton/
hr - ton/day - GAL/sec - GAL/min - GAL/hr - GAL/day - bbl/
sec - bbl/min - bbl/hr - bbl/day - lb/sec - lb/min - lb/hr - lb/
day - cf/sec - cf/min - cf/hr - rev/sec - rev/min - none - L/sec
L/min
2.2 DECIMALS 0 - 3 0
2.3 K-FACTOR 0.000010 - 9,999,999 1
2.4 DECIMALS K-FACTOR 0 - 6 0
2.5 CALCULATION per 1 - 255 pulses 10
2.6 CUT-OFF 0.1 - 999.9 seconds 30
2.7 DAMPING 0.1 - 999.9 seconds 0.1
3 ALARM DEFAULT
3.1 ALARMSET Operate - setup - hide Operate
3.2 FLOWZERO default - no relays - ignore default
3.3 ALARM LOW -999,999 ... 999,999 0
3.4 ALARM HIGH -999,999 ... 999,999 0
3.5 DELAY ALARM LOW 0 - 999.9 seconds 0.0
3.6 DELAY ALARM HIGH 0 - 999.9 seconds 0.0
3.7 REL OUT Lo - Hi - Lo+Hi - Pulse - Off Lo+Hi
4 DISPLAY DEFAULT
4.1 FUNCTION total - flow rate total
4.2 LCD NEW fast - 1 sec - 3 sec - 15 sec - 30 sec - off 1 sec
4.3 BACKLIGHT off - on off
4.4 BACKLIGHT ALARM off - on - flash off
4.5 BACKLIGHT
BRIGHTNESS
1 - 5 5
4.6 BATTERY MODE operational - shelf operational
5 FLOWMETER DEFAULT
5.1 SIGNAL npn - npn_lp - reed - reed_lp - pnp - pnp_lp - namur -
coil_hi - coil_lo - active
coil_lo
5.2 RETRANSMIT enable - disable disable
6 LINEARIZATION DEFAULT
6.1..F FREQ. / M-FACTOR
1..15
Frequency: 0.1 - 9,999.9 Hz
M-Factor: 0.000001 - 9.999999
0.0 Hz
1.000000
6.G LINEARIZATION Enable / disable Disable
6.H DECIMALS
FREQUENCY
0 - 3 1
User Manual F018-P
Page 14 FW_F018-P_M_v0501-03_EN
7 ANALOG OUTPUT DEFAULT
7.1 PV LRV -999,999 ... 999,999 0
7.2 PV URV -999,999 ... 999,999 1600
7.3 LOOP CURRENT
MODE
Enabled - disabled Enabled
7.4 LOOP TEST Off - set 4 mA - set 20 mA Off
7.5 TUNE LO 0000 - 9999 660
7.6 TUNE HI 0000 - 9999 3300
8 PULSE OUTPUT DEFAULT
8.1 PULSE WIDTH 0.001 - 9.999 sec 0
8.2 DECIMALS 0 - 3 0
8.3 IMPULSE PER X-quantity 1000
8.4 REL OUT Lo - Hi - Lo+Hi - Pulse - Off Lo+Hi
9 OTHERS DEFAULT
9.1 TYPE / MODEL fixed F018-P
9.2 SOFTWARE VERSION fixed 03.xx.xx
9.3 SERIAL NUMBER fixed xxxxxxx
9.4 PASS CODE 0000 - 9999 0000
9.5 POLL ADRESS 0 ... 63 0
9.6 TAG NUMBER 0000000 - 9999999 0000000
5.4 SETUP MENU EXPLANATIONS
5.4.1 MENU 1: TOTAL
Calculating the K-factor for Total
The K-factor is used to convert the flowmeter pulse frequency to a quantity. Two examples are
given.
Example 1
The flowmeter generates 2.4813 pulses per liter.
The flow will be displayed using cubic meters (m3; 1.1: TOTAL > UNIT).
A cubic meter equals 1000 liters, thus the flowmeter generates 2,481.3 pulses per m3.
Enter the K-factor as 248130 at 1.3: TOTAL > K-FACTOR and 2 at 1.4: TOTAL > DECIMALS K-FACTOR.
Example 2
The flowmeter generates 6.5231 pulses per gallon.
The flow will be displayed using gallons (1.1: TOTAL > UNIT).
Enter the K-factor as 652310 at 1.3: TOTAL > K-FACTOR and 5 at 1.4: TOTAL > DECIMALS K-FACTOR.
When a different unit is selected, the displayed values are NOT recalculated.
1 TOTAL
1.1 UNIT Determines the measurement unit for total and accumulated total. The
following units can be selected:
L - m3 - kg - lb - GAL - USGAL - bbl - no unit
Changing the measurement unit will have consequences for operator
and setup values. Please note that the K-factor should be checked also.
Recalculation is not done automatically.
1.2 DECIMALS Determines for Total and Accumulated Total the number of decimals.
The following can be selected:
0000000 – 111111.1 – 22222.22 – 3333.333
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 15
1 TOTAL
1.3 K-FACTOR This value is used to convert the flowmeter pulse signals into a Total.
The K-Factor is based on the number of pulses generated by the
flowmeter per measurement unit selected in 1.1: TOTAL > UNIT.
Accuracy of the measurement system depends on the accuracy of the
K-factor.
Examples are given just above this table.
1.4 DECIMALS K-FACTOR This setting determines the number of decimals for the K-factor
(1.3: TOTAL > K-FACTOR). The following can be selected:
0 - 1 – 2 – 3 – 4 – 5 - 6
Please note that this setting influences the accuracy of the K-factor
indirectly. This setting has NO influence on the displayed number of
digits set for Unit (1.1: TOTAL > UNIT).
5.4.2 MENU 2: FLOW RATE
Settings for total and flow rate are independent of each other. In this way, different settings like units
of measurement can be used for each.
The display update time for Flow rate is one second or more.
Calculating the K-factor for Flow rate
The K-factor is used to convert the flowmeter pulse frequency to a flow. Two examples are given.
Example 1
The flowmeter generates 2.4813 pulses per liter.
The flow will be displayed using cubic meters (m3; 2.1: FLOW RATE > UNIT).
A cubic meter equals 1000 liters, thus the flowmeter generates 2,481.3 pulses per m3.
Enter the K-factor as 248130 at 2.3: FLOW RATE > K-FACTOR and 2 at 2.4: FLOW RATE > DECIMALS K-
FACTOR.
Example 2
The flowmeter generates 6.5231 pulses per gallon.
The flow will be displayed using gallons (2.1: FLOW RATE > UNIT).
Enter the K-factor as 652310 at 2.3: FLOW RATE > K-FACTOR and 5 at 2.4: FLOW RATE > DECIMALS K-
FACTOR.
2 FLOW RATE
2.1 UNIT Determines the measurement unit for the flow rate. The following can be
selected:
L/min - L/hr - m3/sec - m3/min - m3/hr - m3/day - g/sec - g/min - g/hr -
kg/sec - kg/min - kg/hr - kg/day - ton/min - ton/hr - ton/day - GAL/sec -
GAL/min - GAL/hr - GAL/day - bbl/sec - bbl/min - bbl/hr - bbl/day - lb/sec
- lb/min - lb/hr - lb/day - cf/sec - cf/min - cf/hr - rev/sec - rev/min - none -
L/sec
Changing the measurement unit will have consequences for operator
and setup values. Please note that the K-factor should be checked also.
Recalculation is not done automatically.
2.2 DECIMALS Determines the number of decimals for the flowrate. The following can
be selected:
0000000 - 111111.1 - 22222.22 - 3333.333
2.3 K-FACTOR The K-factor is used to convert the flowmeter pulse signals to a flowrate
unit. The K-factor is based on the number of pulses generated by the
flowmeter per measurement unit selected in 2.1: FLOW RATE > UNIT.
Accuracy of the measurement system depends on the accuracy of the
K-factor.
Examples are given just above this table.
User Manual F018-P
Page 16 FW_F018-P_M_v0501-03_EN
2 FLOW RATE
2.4 DECIMALS K-FACTOR This setting determines the number of decimals for the K-factor
(2.3: FLOW RATE > K-FACTOR). The following can be selected:
0 - 1 - 2 - 3 - 4 - 5 - 6
Please note that this parameter influences the accuracy of the K-factor
indirectly. This setting has NO influence on the displayed number of
digits for “flow rate” (2.2: FLOW RATE > DECIMALS).
2.5 CALCULATION The flowrate is calculated by measuring the time passed between a set
number of pulses. The more pulses counted, the more accurate the
displayed flowrate will be, but the update period will become longer. The
maximum number of pulses to be counted for a calculation is 255.
The lower the number of pulses, the higher the power consumption of
the F018-P will be (important for battery powered applications).
For low frequency applications (below 10 Hz), do not program more
than 10 pulses, to prevent a very slow update time. For
high‑frequency applications (above 1 kHz), enter a value of 100
pulses or more.
2.6 CUT-OFF Sets a minimum threshold for the flow to be displayed. If during the set
time less than the number of pulses set at 2.5: FLOW RATE > CALCULATION
are detected, the flowrate will be displayed as zero.
The cut-off time has to be entered in seconds. The maximum time is 999
seconds (approximately 16.5 minutes).
2.7 DAMPING Damping is used in case the flowmeter output is not stable, leading to
fast changes in the displayed flow value.
The damping value entered sets the step-response time of the flow rate
measurement. This is the time it takes for the displayed value (and thus
the analog output value) to reach 63% of its new value.
For example, the flow rate suddenly increases from 30 to 100 L/s and
the damping is set to 10 seconds. Initially, the displayed flow rate value
will be 30 L/sec. Then, after 10 seconds, the displayed value will reach
30 + 0,63 x (100-30) = 74 L/sec
The value will keep increasing and eventually it will reach 100 L/sec.
5.4.3 MENU 3: ALARM
3 ALARM
3.1 ALARMSET This function determines if the alarm thresholds can be set at both
Operator level and SETUP-level or SETUP-level only, and if the alarm
values are visible at operator level. Choices are:
Operate - setup - hide
If SETUP has been selected, the alarm values are still visible for the
Operator but cannot be changed.
If Hide has been selected, the alarm values are invisible at Operator
level.
3.2 FLOWZERO When the sensor signal is zero, the monitoring can be ignored or
disabled. Select the desired action when the signal becomes zero:
Default: the alarm output is switched, and the alarm is indicated on
the display.
No relay: the alarm output is not switched, but the alarm is indicated
on the display.
Ignore: the alarm output is not switched, and no alarm is indicated
on the display.
3.3 ALARM LOW Sets the low alarm threshold. An alarm is generated as long as the Flow
rate is lower than this setting. The function is disabled by entering 0.0.
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 17
3 ALARM
3.4 ALARM HIGH Sets the high alarm threshold. An alarm is generated as long as the
Flow rate is higher than this setting. The function is disabled by entering
0.0.
Both flow rate alarms are signaled in HART.
A low flow alarm will set device status bit 14.5
A high flow alarm will set device status bit 14.6.
3.5 DELAY ALARM LOW An alarm generated by 3.3: ALARM > ALARM LOW can be ignored during X-
time period. If the Flow rate is still below the Alarm Low level after this
delay time, an alarm is generated.
3.6 DELAY ALARM HIGH An alarm generated by 3.4: ALARM > ALARM HIGH can be ignored during X-
time period. If the Flow rate is still above the Alarm High level after this
delay time, an alarm is generated.
3.7 REL OUT One transistor output is available to transmit the alarm condition or to
generate the scaled pulse output.
Assign with this function the type of signal to be transmitted:
low flow rate alarm, high flow rate alarm, both alarms or pulse output.
This setting is equal to setting 8.4: PULSE OUTPUT > REL OUT.
This setting is overruled by
SETUP
5.2: FLOWMETER > RETRANSMIT[»18].
5.4.4 MENU 4: DISPLAY
4 DISPLAY
4.1 FUNCTION Determines which information is shown with the 17mm large digits on
the main screen: flow rate, total or accumulated total.
By pressing SELECT you can scroll though the various values. After 30
seconds the display will return to the set information.
4.2 LCD NEW Sets the LCD display refresh rate. The following can be selected:
fast - 1 sec - 3 sec - 15 sec - 30 sec - off
When a key is pressed, the display refresh rate will switch to ‘fast’ for 30
seconds. After this the refresh rate will return to the set update rate.
Please note that information will be processed in the background and
will not be lost at low refresh rates.
Calculating the display information significantly affects the power
consumption. When the application does not require a fast display
refresh rate, it is strongly advised to select a slow rate.
Indication of expected battery life:
4.2: DISPLAY > LCD NEW set to fast: 2 years
4.2: DISPLAY > LCD NEW set to 1sec: 5 years.
Display backlight is only available with Type ZB. The following functions
will affect this optional backlight.
4.3 BACKLIGHT With this setting, the optional backlight can be turned on or off.
4.4 BACKLIGHT ALARM If the F018-P generates an alarm, the backlight can be set to change to
red. The following settings are available:
OFF: during an alarm the color is white.
ON: during an alarm the color is red.
FLASH: during an alarm the color flashes red.
4.5 BACKLIGHT
BRIGHTNESS
Backlight brightness can be adjusted with this setting. A value from 1 to
5 can be selected, where 1 is minimum, and 5 maximum brightness.
4.6 BATTERY MODE The F018-P has two main modes: operational and shelf mode. In shelf
mode the product can be stored for several years with extremely low
power consumption from the battery. All settings and totals are
maintained in memory.
To wake up the F018 from shelf mode, simply press the SELECT key
twice.
User Manual F018-P
Page 18 FW_F018-P_M_v0501-03_EN
5.4.5 MENU 5: FLOWMETER
5 FLOWMETER
5.1 SIGNAL Selects the type of input sensor pickup / signal. The F018-P can process
several types of input signal. The settings with LP are used to apply a
built-in low-pass filter.
FLOWMETER SELECTION CHARACTERISTICS
TYPE OF SIGNAL EXPLANATION RESISTANCE FREQ / mVpp REMARKS
NPN NPN input 100kΩ pull-up max.6kHz
threshold1.2V Open collector
NPN - LP NPN input
with low pass filter 100kΩ pull-up max.1.2kHz
threshold1.2V
Open collector
less sensitive
REED Reed-switch input 1MΩ pull-up max.600Hz
threshold1.2V
REED - LP Reed-switch input
with low pass filter 1MΩ pull-up max.120Hz
threshold1.2V Less sensitive
PNP PNP input 47kΩ pull-down max.6kHz
threshold1.2V
PNP - LP PNP input
with low pass filter 100kΩ pull-down max.1.2kHz
threshold1.2V Less sensitive
NAMUR NAMUR input 820Ω pull-down max.4kHz External power
required
COIL HI Coil input
high sensitivity -
min.20mVpp Sensitive for
interference
Type ZF min.10mVpp
Type ZG min.5mVpp
COIL LO Coil input
low sensitivity - min.90mVpp Normal sensitivity
ACTIVE
Active pulse input
detection level
1.2VDC
47 kΩ max.10kHz External power
required
5.2 RETRANSMIT The conditioned flow meter input pulse can be retransmitted via the
transistor output. This can be useful for calibration purposes or
amplification and squaring of sinewave like signals.
With this setting enabled, for every pulse that is received on the input a
pulse of 1 ms will be generated on the output. The maximum frequency
for this function is 500 Hz, pulses at higher frequencies will not be
transmitted.
With this function disabled, the function of the transistor output is set by
SETUP
3.7: ALARM > REL OUT[»17] and/or
SETUP
8.4: PULSE OUTPUT > REL
OUT[»21].
5.4.6 MENU 6: LINEARIZATION
Explanation of principle
In normal situations, the calculation of flow is based on flowmeters that follow an ‘ideal line’ and give
a fixed number of pulses over the entire frequency range to represent a certain amount. The
calculation uses an average K-factor KF0, which is entered at 2.3: FLOW RATE > K-FACTOR.
Since many flowmeters do not follow this ‘ideal line’, the linearization function can be used to reflect
the actual flow curve better. This allows for more accurate flowrate and totalization values, as well as
improved analog and pulse output values over the frequency range of the flowmeter. The
linearization function uses linearization points to calculate new K-factors based on the measured
flowmeter frequency. As shown in the chart below, each point consists of a frequency and a Meter
Factor (MFx, indicating the deviation of the K-factor KFx from the average K-factor KF0).
F018-P User Manual
FW_F018-P_M_v0501-03_EN Page 19
KF0 KF0
64 93 161 336 514 10,000
Frequency
Hz
K-Factor
KF0=51.64178
KF1=35.7 @ 64Hz.
KF2=47.5 @ 93Hz.
KF3=53.8 @ 161Hz.
KF4=49.2 @ 336Hz.
KF5=52.9 @ 514Hz.
KF1
KF2
KF3
KF4
KF5
MF1=0.691300
MF2=0.919798
MF3=1.041792
MF4=0.952717
MF5=1.024364
0
0
35.7
51.6
Fig.8: Example of K-Factors and linearization points
When a new frequency is measured, the corresponding Meter Factor is calculated by interpolating
between the linearization points. The new Meter Factor is then used to calculate the actual K-factors
for flowrate and total using this formula:
When the measured frequency is below the lowest or above the highest frequency given in the
linearization points, the Meter Factor belonging to that frequency is used.
The following formula can be used to calculate the Meter Factors for the linearization points:
In this formula, the value of KF0 represents the average K-factor of the flowmeter and KFx @ flow x
represents the actual K-factor (derived from the actual flow) at the selected frequency (note that the
K-factors must be in the same engineering units (e.g. pulses / GAL) to perform the calculation).
Entering linearization points
A table of multiple linearization points can be entered into the F018-P. It is advised to enter the
linearization points in increasing order of frequency; however, this is not mandatory. The
linearization points are usually located at frequencies where the linear behavior of the flowmeter
changes. Linearization points that are not used can be disabled by entering a frequency of 0 Hz.
6 LINEARIZATION
6.1..F FREQ. / M-FACTOR
1..15
The frequency is displayed on the bottom line of the display. The
maximum frequency is 9,999.9 Hz. When 0.0 Hz is selected, the M
Factor is disabled. Please check also 6.H: LINEARIZATION > DECIMALS
FREQUENCY before entering frequencies.
The M Factor is displayed on the top line of the display. A value between
0.000001 and 9.999999 can be entered. Most M-factors will be around
1.000000, for example 0.945354 or 1.132573.
The M-factor is always defined with six decimals. Please note that the
decimal point is not displayed!
6.G LINEARIZATION With this setting the linearization function is switched off or on.
User Manual F018-P
Page 20 FW_F018-P_M_v0501-03_EN
6 LINEARIZATION
6.H DECIMALS
FREQUENCY
Determines the number of decimals for the frequency entered at
6.1..F: LINEARIZATION > FREQ. / M-FACTOR 1..15. The following can be selected:
00000 – 1111.1 – 222.22 – 33.333
5.4.7 MENU 7: ANALOG OUTPUT
An analog 4-20 mA signal is generated according to the flowrate. The settings for flow rate (2: FLOW
RATE) influence the analog output directly and should be configured first. The relationship between
rate and analog output is set with the following functions.
Example: Calculating the actual output current. Suppose the PV-LRV is set at 0L/sec and PV-URV
at 1600 L/sec. Then the analog output at an actual flowrate of 1000L/sec will be
1000/1600 * (20mA-4mA) + 4mA = 14mA
7 ANALOG OUTPUT
7.1 PV LRV Primary variable Lower Range Value.
This setting sets the flow rate value at which the analog output should
generate 4mA.
The unit, time and decimals are according to 2.1: FLOW RATE > UNIT and
2.2: FLOW RATE > DECIMALS.
The analog output value can be programmed “upside down” if desired
by switching the PV LRV and PV URV value. Please enter the high flow
rate value here.
7.2 PV URV Primary variable Upper Range Value.
This setting sets the flow rate value at which the analog output should
generate 20 mA.
The unit, time and decimals are according to 2.1: FLOW RATE > UNIT and
2.2: FLOW RATE > DECIMALS.
The analog output value can be programmed “upside down” if desired
by switching the PV LRV and PV URV value. Please enter the low flow
rate value here.
7.3 LOOP CURRENT
MODE
Loop current mode sets the output current fixed to 4mA (for multidrop
purposes) or in normal mode.
7.4 LOOP TEST The loop test is meant to be able to calibrate the 4 and 20 mA values
with 7.5: ANALOG OUTPUT > TUNE LO and 7.6: ANALOG OUTPUT > TUNE HI.
The loop current mode has to be set to ‘enabled’ for this loop test to
function! After a reset or power loss, this setting is reset to off, which
means normal operation.
7.5 TUNE LO The initial minimum analog output value is 4mA. However, this value
might differ slightly due to external influences such as temperature for
example. The 4mA value can be tuned precisely with this setting.
After pressing PROG, the current will be about 4mA. The current can be
increased / decreased with the arrow-keys.
The new value will become active after pressing ENTER.
7.6 TUNE HI The initial maximum analog output value is 20mA. However, this value
might differ slightly due to external influences such as temperature for
example. The 20mA value can be tuned precisely with this setting.
After pressing PROG, the current will be about 20mA. The current can
be increased / decreased with the arrow-keys. The new value will
become active after pressing ENTER.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60

Fluidwell F018 Owner's manual

Category
Measuring, testing & control
Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI