Lansinoh affinity User manual

Category
Breast pumps
Type
User manual
Bisphenol A
FREE
Double Electric Breast Pump
Custom Expression™ Technology
ComfortFit™ Breast Cushions • Independently Adjustable Cycle and Suction Levels
Easy-to-Use • Easy-to-Clean
The Affinity Breast Pump provides all the features mums are looking for in a high-quality
double electric breast pump.
•Allowsforsingleordoublepumpingaccordingtoyourpreference
•Expressmoremilkinhalfthetimethroughdoublepumping
•LansinohCustomExpression™technologyallowsyoutoindependentlyadjustthe
cycleandsuctionsettingsforcomfortableandefcientmilkexpression
•Providestwophasesofexpression–theLet-DownPhaseandtheExpressionPhase–
to simulate the natural rhythm of breastfeeding
•Speciallydevelopedtopreventmilkbackupintubingandpump
•ACorbatteryoperated—twooptionstotyourneeds(batteriesnotincluded)
•Small,quietandlightweight
•Easytoclean
If you have any questions, please call our customer service department on
0113 259 1425 or email [email protected]
IMPORTANT SAFEGUARDS
TheAfnityBreastPumpisapersonalcareitem.Ifyouareamotherwhoisinfectedwith
HepatitisB,HepatitisCorHumanImmunodeciencyVirus(HIV),pumpingbreastmilkusing
theLansinohAfnityBreastPumpwillnotreduceorremovetheriskoftransmittingthevirus
to your baby through your breastmilk.
Thisequipmentgenerates,usesandcanradiateradiofrequencyenergyand,ifnotinstalled
andusedinaccordancewiththeinstructions,maycauseharmfulinterferencetoother
devicesinthevicinity.However,thereisnoguaranteethatinterferencewillnotoccurina
particularinstallation.Ifthisequipmentdoescauseharmfulinterferencetootherdevices,
whichcanbedeterminedbyturningtheequipmentoffandon,theuserisencouragedtotry
tocorrecttheinterferencebyoneormoreofthefollowingmeasures:
• Reorientorrelocatethereceivingdevice.
• Increasetheseparationbetweentheequipment.
• Connecttheequipmentintoanoutletonacircuitdifferentfromthattowhichtheother
device(s)areconnected.
• Consultthemanufactureroreldservicetechnicianforhelp.
Whenusingelectricalproducts,especiallywhenchildrenarepresent,basicsafetyprecautions
shouldalwaysbemaintained,includingthefollowing:
READ ALL INSTRUCTIONS BEFORE USING
Danger: To reduce the risk of electrocution:
• Alwaysunplugthisproductimmediatelyafterusing.
• Donotusewhilebathing.
• Donotplaceorstorethisproductwhereitcanfallorbepulledintoabathorsink.
• Donotplaceordroppumpintowaterorotherliquid.
• Donotreachforanelectricalproductthathasfallenintowater.Unplugimmediately.
Warning: To reduce the risk of burns, electrocution, fire, or injury to persons:
• Thisproductshouldneverbeleftunattendedwhenpluggedin.
• Closesupervisionisnecessarywhenthisproductisusednearchildren.
• Usethisproductonlyforitsintendeduseasdescribedinthismanual.Donotuse
attachments not recommended by the manufacturer.
• Neveroperatethisproductifithasadamagedcordorplug,ifitisnotworkingproperly,
ifithasbeendroppedordamaged,ordroppedintowater.
• Keepthecordawayfromheatedsurfaces.
• Neverusewhilesleepingordrowsy.
• Neverdroporinsertanyobjectintoanyopeningortubing.
• DonotuseoutdoorswiththeACadapteroroperatewhereaerosolsprayproductsare
beingusedorwhereoxygenisbeingadministered.
SAVE THESE INSTRUCTIONS
along with your receipt for future reference.
1 2
Important Safeguards
Contents
2 –ImportantSafeguards
3 –PumpDiagram
5 – Affinity Breast Pump Parts
6 –CleaningInstructions
7 –AssemblingtheBreastmilkExpressionSet
8 – Pumping Tips
8 –UsingTheAfnityBreastPump
10 –PowerOptions
11 –BreastmilkStorageGuidelines
12 –BreastmilkStorageBags
13 – Troubleshooting
14 – Warranty
15 –DeclarationsforElectromagnetic
Compatibility(EMC)
17 –TechnicalData
Congratulations On Your New
Lansinoh Anity Double Electric Breast Pump!
Quick-Change button
Quick-Change icon
Quick-Change button Quick-Change icon
Power, On/Off icon
Cycle icon
Suction/Vacuum icon
Quick-Change icon
Power, On/Off button
Cycle increase/decrease button
LCD screen
Suction increase/decrease button
Quick-Change button
1.
2.
3.
4.
5.
Cycle indicator
Time/level indicator
Low battery indicator
Let-down/Stimulation Phase indicator
Expression Phase indicator
Suction indicator
Tubing Adapter Input
6.
7.
8.
9.
10.
11.
Bottle holder (2)
AC adapter input
Tubing adapter input
Battery compartment for 6 AA batteries (not included)
12.
13.
14.
15.
AC Adapter AA Battery Compartment (underneath pump)
3 4
Pump Diagram Pump Diagram
Anity Breast Pump Parts:
Item: Item Number:
a. Tubing #50403
b.BreastmilkExpressionSetincludes:
b1.ComfortFit™breastcushion(2) #50408
b2.Breastcushionbody(2) #50415
b3.Diaphragm(2) #50402
b4.DiaphragmCap(2) #50410
b5.WhiteValve(2+2extra) #50401
b6.150mlBottles(2) #50414
c.BreastmilkStorageBottleRing(2)andSealingDisk(2) #50406&#50413
d.ACAdapter/PowerCord #50455
e.VelcroStrapforTubing #50417
Twoextravalvesareincluded.Largersizecushionsareseparatelyavailableforsale.
Toordertheseitemsorreplacementparts,pleasevisitwww.lansinoh.co.uk/customerservice
orcall01132591425andreferencetheitemnumbersabove.
Before the rst use and aer each use thoroughly clean and dry the pump
• AlwayswashhandsthoroughlywithsoapandwaterbeforehandlingtheBreastmilk
ExpressionSet,thepump,orotherattachments.
• TaketheBreastmilkExpressionSetaparttoclean.
• Ensurethatthetubingisfreeofanyliquid.Donotoperatethepumpifliquidisvisiblein
the tubing.
Cleaning Instructions
Unlessotherwiseinstructedbyyourhealthcareprovider,followthesesterilisingandcleaning
instructions.
Sterilisation Prior To First Use
1.Priortousingthesetforthersttime,disassembleandboilallpartsoftheBreastmilk
ExpressionSetfor10minutes.
•Note:Ensureyouuseenoughwatertokeeppartsaoatwhileboiling.
•Note: Do not boil the tubing and tubing connector.
2.Removeallpartsfromwaterimmediatelyafterboiling.
3.Completelyairortoweldrybeforeassemblyandnextuse.
•Note:Partiallydriedpartsmayaffectoperationofthepump.Pleaseensurepartsare
completelydrybeforeuse.
Cleaning
• Setasidethetubingandtubingconnector.Theseitemsdonotneedtobewashed
orsterilisedbecausethediaphragmpreventsthemfromcomingintocontactwith
the breastmilk.
• AllpartsoftheBreastmilkExpressionSetmaybecleanedintheuppercompartmentof
yourdishwasherorhandwashedwithwashingupliquidinhotwater.DONOTusesolvents
orabrasives.Rinsewithhot,cleanwater.
• TocleanyourBreastPump,wipethebasewithadampcloth,usingonlywaterormildsoap.
• Never wash the tubing or the tubing connector.
IT IS VERY IMPORTANT THAT ALL PARTS OF YOUR AFFINITY PUMP
ARE COMPLETELY DRIED BEFORE EACH USE. IF ALL PARTS ARE NOT
COMPLETELY DRY THIS WILL AFFECT THE PERFORMANCE.
NEVER WASH OR BOIL TUBING AND TUBING CONNECTOR AS THIS WILL
PREVENT YOUR PUMP FROM FUNCTIONING PROPERLY.
5 6
Anity Breast Pump Parts Cleaning Instructions
Diaphragm Cap
Diaphragm
White Valve
ComfortFit™ Breast Cushion
Breast Cushion Body
150 ml Breastmilk Storage Bottle
Assembling e Breastmilk Expression Set
YourBreastmilkExpressionSetcomespreassembledinLansinohproductpackaging.
CleanBreastmilkExpressionSetparts(excepttubingandtubingconnector)asdescribedin
the“CleaningInstructions”section(seepage6).Onceallpartsarecleananddry,assemble
theexpressionsetasfollows:
1. Placepurplediaphragmontopofbreastcushionbody,twistcleardiaphragmcaponto
breastcushionbody,enclosingthepurplediaphragm.Diaphragmcanbeusedoneither
side.Ensurediaphragmcapistwistedtightlyintoplace.
2. Pushbreastcushionintobreastcushionbody.
3. Gentlysecurewhitevalveintothebottomofbreastcushion.
Note:Usecautionwhenhandlingandcleaningthesmall
whitevalve(seepage6).
4. Screwbreastmilkstoragebottleontobreastcushionbody.
5.Inserttubingintobackofcleardiaphragmcap,ensuring
a tight fit.
6. Inserttubingconnector,atendoftubing,intotubingadapterinputonthepump.
7. Nowyouarereadytousethepump.
Pumping Tips
Themostimportantpartofsuccessfulpumpingisthelet-downreex.Thisiswhenyour
milkbeginstoowfreely.Warmwetcompressesappliedtothebreastsandlight,circular
massagingofthebreastspriortoand/orduringapumpingsessionwillhelpthisnaturalreex
occur.Thinkingoforlookingatyourbaby,oraphotoofyourbaby,mayalsobehelpful.
Somemumscanachievemorethanonelet-downduringapumpingsession.Youmaybe
abletoachieveasecondorthirdlet-downduringyourpumpingsessionbypressingthe
Quick-Changebutton.
How Long To Pump
Doublepumping,comparedtosinglepumping,willreducethetimeyouspend.Itisimportant
to remember that successful pumping is a learned skill. Breastmilk is produced on a supply
anddemandbasis:expressingyourbreastmilkregularlystimulatesandmaintainsyourmilk
supply.Ifbreastfeedingproblemsoccur,contactyourmidwife,breastfeedingcounselloror
other healthcare professional.
Using e Lansinoh Anity Breast Pump
For Double Pumping:
1. Makesurebothtubesareattachedtothetubing
connector and insert into pump base.
2. HoldtheBreastmilkExpressionSettoyourbreasts
so that your nipples are centred.
3. PresstheOn/Offbuttontobeginpumping.Lansinoh
AfnityPumpwillstartintheLet-DownPhase.
Youwillhearahigherpitchedsoundinshortcycles.
Aftertwominutes,theLet-DownPhasewillautomatically
changetotheExpressionPhase.Thisischaracterisedby
alowerpitchsoundandlongercycles.
Ifyouprefertobypassthepre-setLet-DownPhaseor
ifyourmilkletsdownsoonerthantwominutes,press
theQuick-ChangebuttontomovetotheExpression
Phase.
Note – the Expression Icon will appear in the
display to signify that you are in the Expression Phase.
Diaphragm Cap
Diaphragm
White Valve
ComfortFit™ Breast Cushion
Breast Cushion Body
150 ml Breastmilk Storage Bottle
-
+
-
+
Figure 1
Figure 2
-
+
-
+
Figure 1
Figure 2
-
+
-
+
Figure 1
Figure 2
7 8
Assembling e Breastmilk Expression Set Pumping Tips / Using e Anity Breast Pump
4. WhileintheExpressionPhase,youcanindependently
adjustboththecycleandsuctionsettings.Adjustthe
suctionandcycletoyourcomfortlevelwhilepumping
bypressingtheincrease(+)ordecrease(–)buttonnext
tothesuctionicon,orcycleicon.
Pleasemakeonesettingchangeatatimetoexperience
thedifferencebeforemakingfurtheradjustments.
5. Formaximummilkow,thecyclesetting(ontheleft
sideoftheLCDscreen)shouldbeincreasedtoL6,as
indicatedinFigure1.Thisactuallyreducesthecycle
speed,butincreasesthecyclelength.Alongercycle
length(andslowerpumpingrhythm),afterlet-down
hasoccurred,allowsthepumptomoreeffectively
drawoutmilk.
AfteryouincreasethecyclesettingtoL6,thesuction
setting(ontherightsideoftheLCDscreen)shouldbe
increasedtoL8,asindicatedinFigure2,orthe
maximumlevelthatyoundcomfortable.Thisincreases
thesuctionpowerofthebreastpumpandcombinedwith
the longer cycle length provides the greatest level for
milkexpression.
6. Whenapumpingsessionisnished,turnthepumpoffwiththeon/offbutton.Afterturning
thepumpoff,thedisplaywillremainonforafewseconds.Makesurethereisnosuction
onthenipple(s)beforeremovingthebreastcushions.Ifnecessary,breakthesuctionby
insertingangerbetweenthebreastcushionandthebreast.
Note: LCDdisplaylightwillautomaticallyswitchoffifnobuttonsarepressedfor20
secondsinordertosavepower.Ifthepumpisleftonbutunattendedformorethan30
minutes,thepumpwillautomaticallypoweroff.
For Single Pumping:
Removeonetubeandclosethetubingconnector.Make
sure that the plug is placed into the pipe outlet properly and
pushed in securely.
Please don’t forget to thoroughly clean all parts
(otherthantubingandtubingconnector)aftereachuseas
indicated on page 6.
Adjustable for Maximum Milk Flow and Comfort
YourLansinohAfnityBreastPumpallowsyoutoadjustbothcycleandsuctionsettings.
Choosethesettingsthataremostcomfortableforyouandresultinthegreatest
milkexpression.
Cycle Buttons
Thesecontrolthecyclelengthofthebreastpump.Thelowercyclesettingscreateashorter
cyclelengthandfasterpumpingrhythm,whichishelpfulforelicitingmilklet-down.
Oncelet-downhasoccurred,youmaypushthe(+)buttontoincreasethecyclelength,
whichcreatesaslowerpumpingrhythm,formorepowerfulmilkexpression.
Note: Thehighestcyclesetting(L6)actuallycreatestheslowestpumpingrhythm,whilethe
lowestcyclesetting(L1)createsthefastestpumpingrhythm.Thehighestcycle setting
combinedwiththehighestsuctionlevelprovidesthegreatestlevelformilkexpression.
Suction Buttons
Thesecontrolthesuctionstrength.Pushthe(+)buttontoincreasethesuction.Continueto
pushthe(+)buttontoincreasethesuctiontothehighestlevelcomfortableforyou.Pushthe
(-)buttontodecreasethesuction.Yourpumpcanbeadjustedfromsuctionlevels1to8.
Note:Ifthesuctiononyourbreastreachesuncomfortablelevels,reducethesuctionlevel
bypushingthe(-)Suctionbutton,ordisconnectthepumpconnectorfromitssocketonthe
electricpump.
Power Options
Thepumpmaybeoperatedunderanyofthefollowingpowersupplymodes:
-
+
-
+
Figure 1
Figure 2
-
+
-
+
Figure 1
Figure 2
-
+
-
+
Figure 1
Figure 2
9 10
AC Adapter(included)
ConnecttheACadaptertothepower
socket located on the side of the pump.
ThenconnecttheACadapterintothe
mainpowersupply.Electricalsafetycan
onlybeachievedwhenthebreastpump
isconnectedtotheACPowerAdapter
suppliedwiththeproduct(See“Technical
Data”section,page17)
6 AA Alkaline Or Rechargeable Batteries
(notincluded)Openthebatterycoverunderthepump
andinsert6AAbatteries.Makesurethepolarity(+)
and(-)iscorrectasindicated.Closethecompartment
by gently pressing the cover until it snaps into place.
Removebatteriesifpumpisnotinuseforalongperiod
oftime.Disposeofbatteriesaccordingtoregulations
forenvironmentalprotectionandrecycling.Replace
batteriesasasimultaneousset.Donotmixoldandnew
batteries.Donotmixalkaline,standard(carbon-zinc),or
rechargeable(nickel-cadmium)batteries.
Using e Anity Breast Pump Using e Anity Breast Pump / Power Options
Quick-Change button
Quick-Change icon
Quick-Change button Quick-Change icon
Power, On/Off icon
Cycle icon
Suction/Vacuum icon
Quick-Change icon
Power, On/Off button
Cycle increase/decrease button
LCD screen
Suction increase/decrease button
Quick-Change button
1.
2.
3.
4.
5.
Cycle indicator
Time/level indicator
Low battery indicator
Let-down/Stimulation Phase indicator
Expression Phase indicator
Suction indicator
Tubing Adapter Input
6.
7.
8.
9.
10.
11.
Bottle holder (2)
AC adapter input
Tubing adapter input
Battery compartment for 6 AA batteries (not included)
12.
13.
14.
15.
AC Adapter AA Battery Compartment (underneath pump)
Power Options (continued)
Ifbatteriesareinthepump,butACadapterisinuse,thepumpwillusetheadapterto
powerthepump,notthebatteries.Forbestperformance,werecommendusingyourpump
withtheACadapterprovided.TheACadapterprovidedhasauniversalACinputandcanbe
usedwith120Vacand230Vacsystemswith50Hzand60Hzmainspower.TheAfnitypump
isintendedtoonlybeusedwiththeprovidedexternaladapter(Lansinohpart#50455)by
BridgePowerCorporation,JMW110-KA-09-00-G-07,withelectricalrating110-230VAC,
50-60Hz.
Breastmilk Storage And awing Guidelines
For Healthy Term Babies
Checkwithyourmidwife,healthvisitororbreastfeedingcounsellorforspecicstorage
instructions.Whenyouarefreezingyourbreastmilk,donotllcontainersmorethan¾fullto
allowforexpansionfromfreezing.Datecontainerswiththedateyouexpressedyourmilkfor
future reference.
Source:LaLecheLeagueInternational,2009
Note:DONOTstorebreastmilkinfreezerdoor.Thedoortendstobethewarmestpartofthefreezer.
Dateallbreastmilkandusetheoldestbreastmilkrst.
awing
Thawfrozenmilkintherefrigeratororplacethecontainersundercoolrunningwater,
graduallywarmingthewateruntilthemilkisthawed.Thawedmilkcanbekeptinarefrigerator
forupto24hours.
NOTE: DONOTrefreezemilkonceitisthawed.
Gentlyshakethawedmilkbeforefeedingtobabytomixthelayersthathaveseparated.
Breastmilkisnothomogenizedandthe“cream”willrisetothetopofthebottle.Breastmilk
canvaryincolourandconsistencydependinguponthetimeofthedaythatthemilkwas
expressed,whatyouhaveeaten,andtheageofthebabyatthetimeofpumping.
NOTE: DO NOT MICROWAVE HUMAN MILK! Microwaving alters the composition of the milk.
Microwaving also may create “hot spots” in the milk that could burn the infant.
NOTE: DO NOT thaw breastmilk in hot or boiling water.
11
Breastmilk Storage Guidelines
FreshBreastmilk
ThawedBreastmilk
Self-contained freezer
approx.-18°C
Upto6months
Donotrefreeze
Freezer section
in refrigerator
Upto2weeks
Donotrefreeze
Room Temperature
19°Cto26°C
Upto6hours
1 hour
Refrigerator
<4°C
Upto72hours
24hours
Breastmilk Storage Bags
12
Lansinoh Breastmilk Storage Bags are specially
designed to be used with your pump
•Pre-sterilisedforsafety
•Madeoffood-safepolyethylene
•Gussetedbottomsobagcanstand
•Writeontaboutsidethellareaforadding
customised use by date
•Doublezipperforsecureseal
•Bagscomeinatissuedispenserforconvenience
13
Limited Consumer Product Warranty
Thisproductiswarrantedtotheoriginalconsumerpurchaser(the“purchaser”)tobefreefrom
defectsinmaterialandworkmanshipwhicharenotcommerciallyacceptablefortheperiodof
oneyearfromthedateofpurchaseonthepumpmechanismandtheperiodof90dayson
allotherproductcomponents.Thiswarrantygivesyouspeciclegalrightsandyoumayalso
haveotherrights,whichvary,bylocation.Yourstatutoryrightsareunaffectedbythiswarranty.
LANSINOHMAKESNOEXPRESSWARRANTYOFANYKINDREGARDINGTHEPRODUCT
OTHERTHANTHOSEWARRANTIESEXPLICITLYSETFORTHHEREIN.ANYIMPLIED
WARRANTY,INCLUDINGANYIMPLIEDWARRANTYOFMERCHANTABILITYORANY
IMPLIEDWARRANTYOFFITNESSFORAPARTICULARPURPOSE,TOTHEEXTENT
PERMITTEDBYLAW,SHALLBELIMITEDINDURATIONTOAPERIODOF90DAYSFROM
THEDATEOFPURCHASEBYTHEORIGINALPURCHASER.
IntheeventtheproductisfoundbyLansinohtonotmeettheabovelimitedwarranty,as
purchaser’ssoleandexclusiveremedyLansinohwillrepairor,attheoptionofLansinoh,
replacethisproductwithoutchargeforsuchreplacementorpartsorlabour.Thiswarranty
doesnotapplytoanyproductthathasbeensubjectedtomisuse,abuse,alterationorfor
normalwearandtear.OnlyLansinohbrandreplacementpartsarerecommendedforusewith
Lansinohbreastfeedingproducts.Improperorincorrectlyperformedmaintenanceorrepair,or
useofnon-Lansinohbrandreplacementpartsmaylimitorinvalidatethiswarranty.
TO THE EXTENT PERMITTED BY LAW, LANSINOH SHALL NOT BE LIABLE FOR ANY
INCIDENTAL OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO,
REPLACEMENT COSTS RESULTING FROM THE BREACH OF ANY WRITTEN OR
IMPLIED WARRANTY.
If you wish to make a claim under this warranty, please telephone 0113 259 1425.
You will be given a Freepost address to return the product, if necessary, together
with your original dated bill of sale or other dated proof of purchase and a brief
statement of the problem.
Warranty
Troubleshooting e Lansinoh Anity Breast Pump
Interruption of Electrical Service
Ifyouexperienceaninterruptioninthenormaloperationoftheelectricaland/ormechanical
functioningofthepump,simplydothefollowing:
•UnplugtheACadaptercordfromtheelectricaloutlet.
•Waitafewseconds,andthenplugtheACadaptercordbackintotheelectricaloutlet.
Discomfort While Pumping
Ifyoufeelanydiscomfort,youmayneedadifferentsizeange.
Visitwww.lansinoh.co.uk/customerserviceorcall01132591425.
Decreased Suction
Ifitfeelslikethesuctionhasdecreased,trythefollowing:
• Set cycle to its highest level (L6). The highest cycle setting creates a slower
pumping rhythm, which combined with the highest suction level (L8), allows the
Affinity breast pump to operate at its greatest level for milk expression.
• Removethetubingandtubingconnectorfromthebackofthepumpandplaceanger
overthepumpoutlet.Ifyoufeelsuction,thepumpisoperatingproperly,butthe
BreastmilkExpressionSetmaynotbecorrectlyassembledorthewhitevalvemaybe
damaged.Ifnosuctionisfeltonyournger,tryunpluggingtheadapterfromtheelectrical
outlet,waitafewseconds,andplugitbackintotheelectricaloutlet.
• MakesuretheBreastmilkExpressionSethasbeenproperlyassembled.Checkthe
assemblyinformationinthesection“AssemblyInstructions”,page7.
• Inspectthewhitevalveforatearorpinhole.Thiscomponentiscriticaltoachievingproper
suction.Replacethewhitevalvewithoneofthesparevalvesprovidedwiththeproduct.
• Whensinglepumping,checkthetubingadapteronthebackofthepumptomakesureit
isclosedforsinglepumping.Ifopen,closetheadapter.
• Ifdoublepumping,ensurethatthebreastmilkexpressionsetisonbothbreastsbefore
starting the pump.
Increased suction level for single pumping:
Ifitfeelslikethesuctionishigherthanusualwhensinglepumping,checkthattheplugforthe
pipeiscleanandthattheairwayintheplugisnotblocked.
Breastmilk Build-up in Breast cushion or Breast cushion Body
Ifbreastmilkdoesnotappeartodrainfastenoughintothebottle,reducethesuctionsetting
onthepumptoallowthemilktodrain.Oncemilkhasdrainedintothebottle,youmay
gradually increase the suction level.
14
Troubleshooting
Recommended separation distances between portable and mobile RF Communications Equipment and the Affinity Double
Electric Breast Pump - Table 206 according to EN 60601-1-2
Recommended separation distances
Recommended separation distances between portable and mobile RF communications and the Afnity Double Electric Breast Pump
TheAfnityDoubleElectricBreastPumpisintendedforuseinanelectromagneticenvironmentinwhichradiatedRFdisturbances
arecontrolled.ThecustomerortheuseroftheAfnityDoubleElectricBreastPumpcanhelppreventelectromagneticinterferenceby
maintainingaminimumdistancebetweenportableandmobileRFcommunicationsequipment(transmitters)andtheAfnityDouble
ElectricBreastPumpasrecommendedbelow,accordingtothemaximumoutputpowerofthecommunicationsequipment.
ForTransmittersratedatamaximumoutputpowernotlistedabove,therecommendedseparationdistancedinmetres(m)can
beestimatedusingtheequationapplicabletothefrequencyofthetransmitter.WherePisthemaximumoutputpowerratingofthe
transmitterinwatts(W)accordingtothetransmittermanufacturer.
NOTE1At80MHzand800Mhz,theseparationdistanceforthehigherfrequencyrangeapplies.
NOTE2Theseguidelinesmaynotapplyinallsituations.Electromagneticpropagationisaffectedbyabsorptionandreectionfrom
structures,objectsandpeople.
Rated maximum output power
of transmitter
W
0,01
0,1
1
10
100
800 MHZ to 2,5 GHz
P
d =
7
E
1
7cm
22cm
70cm
2,2m
7m
Separation distance according to frequency of transmitter m
3,5cm
11cm
35cm
1,1m
3,5m
80 MHZ to 800 MHz
P
d =
3,5
E
1
11,7cm
36,9cm
1,2m
3,7m
11,7m
150 kHZ to 80 MHz
P
d =
3,5
V
1
15 16
Declarations for Electromagnetic Compatibility (EMC)
Electromagnetic Emissions - Table 201 according to EN 60601-1-2
Guidance and manufacturer’s declaration - electromagnetic emissions
TheAfnityDoubleElectricBreastPumpisintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomerorthe
useroftheAfnityDoubleElectricBreastPumpshouldassurethatitisusedinsuchanenvironment.
Compliance
Group1
ClassB
ClassA
Complies
Emissions test
RFemissions
CISPR11
RFemissions
CISPR11
Harmonic emissions
IEC61000-3-2
Voltageuctuations/
ickeremissions
IEC61000-3-3
Electromagnetic environment - guidance
TheAfnityDoubleElectricBreastPumpusesRFenergyonlyforitsinternal
function.Therefore,itsRFemissionsareverylowandarenotlikelytocauseany
interference in nearby electronic equipment.
TheAfnityDoubleElectricBreastPumpissuitableforuseinallestablishments,
includingdomesticestablishmentsandthosedirectlyconnectedtothepubliclow-
voltagepowersupplynetworkthatsuppliesbuildingsusedfordomesticpurposes.
Electromagnetic Immunity (all devices) - Table 202 according to EN 60601-1-2
Guidance and manufacturer’s declaration - electromagnetic immunity
TheAfnityDoubleElectricBreastPumpisintendedforusetheelectromagneticenvironmentspeciedbelow.Thecustomerortheuser
oftheAfnityDoubleElectricBreastPumpshouldassurethatitisusedinsuchanenvironment.
IEC 60601 test level
±6kVcontact
±8kVair
±2kVforpowersupplylines
±1kVforinput/outputlines
±1kVline(s)toline(s)
±2kVline(s)toearth
<5%UT(0,5cycle)
40%UT(5cycles)
70%UT(25cycles)
<5%UTfor5s
3A/m
IMMUNITY test
Electrostatic
discharge(ESD)
IEC61000-4-2
Electricalfast
transient/burst
IEC61000-4-4
Surge
IEC61000-4-5
Voltagedips,short
interruptions and voltage
variationsonpower
supply input lines
IEC61000-4-11
Powerfrequency
(50/60Hz)magnetic
eldIEC61000-4-8
Compliance level
±6kVcontact
±8kVair
±2kVforpowersupplylines
±1kVforinput/outputlines
±1kVline(s)toline(s)
±2kVline(s)toearth
<5%UT(0,5cycle)
40%UT(5cycles)
70%UT(25cycles)
<5%UTfor5s
3A/m
Electromagnetic environment -
guidance
Floorsshouldbewood,concreteor
ceramictile.Ifoorsarecoveredwith
syntheticmaterial,therelativehumidity
should be at least 30%.
Mainspowerqualityshouldbethatofa
typical commercial or hospital environment.
Mainspowerqualityshouldbethatofa
typical commercial or hospital environment.
Mainspowerqualityshouldbethatofa
typical commercial or hospital environment.
Powerfrequencymagneticeldsshould
be at levels characteristic of a typical
locations in a typical commercial or
hospital environment.
NOTE:U
T
is the a.c mains voltage prior to application of the test level.
Technical Data
MEDICALEQUIPMENTWITHRESPECTTOELECTRICAL
SHOCK,FIREANDMECHANICALHAZARDSONLYIN
ACCORDANCEWITH60601-1ABDCSAC22:2No:60950-1
CONTROLNUMBER:44TS
ULFileNoE323414
CSAFileNoE323414
-
+
-
+
TypeBFAppliedPart
1007
Must not be disposed of together with household refuse.
UK
Enko Elektronik Kontrol Sistemleri
10006 Sokak No:64 A.O.S.B Çiğli İZMİR - TURKEY
17
Technical Data
Vacuum (approx.)
AC adapter
BridgePower Corporation
(JMW110-KA-09-00-G-07)
Protection Class II
ISO 13485
ISO 9001
CE 1007
EN 980, EN1041
EN14350 (for part numbers
50414, 50406, 50413)
Operation
Transport / Storage
+ 5 °C
+ 40 °C
- 20 °C
+ 50 °C
Operation
Transport / Storage
Electromagnetic Compatibility
EN60601-1-2 (2007)
20% - 95%
Humidity
20% - 80%
Humidity
100 - 230 V
50 / 60 HZ
0,4 A
-50 … -250 mmHG
-7 … -33 kPa
30 … 115 cpm
6 x AA
Batteries (not included) - Rechargeable batteries must be
fully charged before inserting into the pump unit.
18
Ifyouhaveanyquestions,pleasecallourcustomerservicedepartment:
Lansinoh Laboratories Inc.
JasonHouse,KerryHill,Horsforth
Leeds,YorkshireLS184JR,UnitedKingdom
Tel:+44(0)1132591425
www.lansinoh.co.uk
Quick Start Instructions
www.lansinoh.co.uk
If you are using the Lansinoh® Affinity Breast Pump for the first time it is
important to read the full instruction manual before you begin.
Quick-Start Instructions
Start with Phase I
Turnonpowerbypressingtheon/offbutton.
ThiswilltakeyouintoPhaseI-theLet-Down
(Stimulation)Phase.
Youwillhearahigherpitchedsoundinshortcycles.
Phase II (Expression Phase)
ThepumpwillautomaticallyswitchintophaseIIafter
twominutes.
IfyouwanttobypassorreducephaseIpress
Quick-Changebutton.
ThepumpwillswitchtoPhaseIIimmediately.
Thisischaracterisedbyalowerpitchsoundandlongercycles.
Adjust cycle
WhileyouareinphaseIIyoumayadjustcyclelevelby
pressing(+)or(-)nexttocycleicon.
Thehighestcyclesetting(L6)actuallycreatestheslowestpumping
rhythm,whilethelowestcyclesetting(L1)createsthefastestpumping
rhythm.Thehighestcycle settingcombinedwiththehighestsuction
levelprovidesthegreatestlevelformilkexpression.
Adjust suction
WhileyouareinphaseIorphaseIIyoumayadjust
suctionlevelbypressing(+)or(-)nexttosuctionicon.
Ifthesuctiononyourbreastreachesuncomfortablelevels,reduce
thesuctionlevelbypushingthe(-)Suctionbutton,ordisconnectthe
pumpconnectorfromitssocketontheelectricpump.
Change to Phase I (Let Down Phase)
AtanytimeduringExpressionphaseyoucanswitch
backtophaseI(LetDownPhase)bypressingthe
Quick-ChangeButton.
Stop pumping
Turnthepumpoffwithon/offbutton.
Indicator on
LCD screen
L1 up to L8
L1 up to L6
DE52035IF0410
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11

Lansinoh affinity User manual

Category
Breast pumps
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI