  • Hello! I am an AI chatbot trained to assist you with the Taylor Technologies K-1805 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
DROP TEST
MOLYBDENUM (1 drop = 2, 5, 20, or 50 ppm)
COMPONENTS:
2 x 4030 Pipets, Calibrated 0.5 & 1.0 mL, plastic w/cap
1 x 4078 Pipet, Graduated, 3 mL (0.5 mL div.), plastic
1 x 5375 Instruction
2 x 9198 Sample Tubes, Graduated, 25 mL, plastic w/cap
1 x R-0887 Molybdenum Standard 20 ppm
1 x R-0890 Molybdenum Buffer Solution
1 x R-0891 Molybdenum Indicator Solution
1 x R-0892 Molybdenum Titrating Solution, DB
TO ORDER REPLACEMENT PARTS & REAGENTS CALL TOLL-FREE 800-837-8548.
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
For 1 drop = 2 or 5 ppm Mo
1.Rinseandll25mLsampletube(#9198)tomarkwithwatertobetested.
NOTE: For1drop=2ppm,llto25mLmark.
For1drop=5ppm,llto10mLmark.
2.Rinse and ll second 25 mL sample tube to same mark with distilled, deionized, or
molybdenum-freetapwater.
3.For25mLsample,usingseparate1.0mLpipets(#4030),add1.0mLR-0890Molybdenum
BufferSolutionand1.0mLR-0891MolybdenumIndicatorSolutiontoeach25mLsample
tube.For10mLsample,add0.5mLofeachreagenttoeachsampletube.Swirltomix.
4.Add R-0892 Molybdenum Titrating Solution, dropwise,swirlingandcountingaftereach
drop,to25mLsampletubecontainingmolybdenum,untilcolormatches25mLsampletube
containingnomolybdenum,oruntilnofurtherchangeincoloroccurs.Alwaysholdbottlein
verticalposition.
5.For25mLsample,multiplydropsofR-0892MolybdenumTitratingSolutionby2.Record
as parts per million (ppm) molybdenum. For 10 mL sample, multiply drops of R-0892
MolybdenumTitratingSolutionby5.Recordasppmmolybdenum.
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548) • 410-472-4340 8/00
NOTE: Toconvertmolybdenum(Mo)readingstomolybdate(MoO
4
2-
),multiplyMoreadings
by1.7;toconverttosodiummolybdatedihydrate(Na
2
MoO
4
·
2H
2
O),multiplyby2.52.
For 1 drop = 20 or 50 ppm Mo
1.Rinseandllone25mLsampletube(#9198)to25mLmarkwithdistilled,deionized,or
molybdenum-freetapwater.
2.Usinga3mLpipet(#4078),placewatertobetestedinsecond25mLsampletube.
NOTE: For1drop=20ppm,llpipetto2.5mLmark.
For1drop=50ppm,llpipetto1.0mLmark.
3.Diluteto25mLmarkwithdistilled,deionized,ormolybdenum-freetapwater.
4.Usingseparate1.0mLpipets(#4030),add1.0mLR-0890MolybdenumBufferSolutionand
1.0mLR-0891MolybdenumIndicatorSolutiontoeach25mLsampletube.Swirltomix.
5.AddR-0892MolybdenumTitratingSolutiondropwise,swirlingandcountingaftereachdrop,
to25mL sampletube containingmolybdenum,until colormatches 25mLsample tube
containingnomolybdenumoruntilnofurtherchangeincoloroccurs.Alwaysholdbottlein
verticalposition.
6.For2.5mLsample,multiplydropsofR-0892MolybdenumTitratingSolutionby20.Record
as parts per million (ppm) molybdenum. For 1.0 mL sample, multiply drops of R-0892
MolybdenumTitratingSolutionby50.Recordasppmmolybdenum.
NOTE: Toconvertmolybdenum(Mo)readingstomolybdate(MoO
4
2-
),multiplyMoreadings
by1.7;toconverttosodiummolybdatedihydrate(Na
2
MoO
4
·
2H
2
O),multiplyby2.52.
NOTE: R-0891MolybdenumIndicatorSolutionhasalimitedshelf-life.Occasionallycheck
indicator performance by following procedure for 1 drop = 2 ppm using R-0887
Molybdenum Standard 20 ppm as the sample. Titration should require 10 drops
toendpoint.(An alternate kit, K-1805P,isavailable, which replacesR-0891 witha
stable,two-partreagentsystem.)
Instr. #5375
/