DustTrak
Aerosol Monitor
Environmental
Enclosure
Model 8535
(For DustTrak™ II and DRX Aerosol Monitor
Models 8530 and 8533)
Operation and Maintenance Manual
P/N 6002097, Revision R
May 2023
Start Seeing the Benefits
of Registering Today!
Thank you for your TSI® instrument purchase. Occasionally, TSI®
releases information on software updates, product enhancements and
new products. By registering your instrument, TSI® will be able to
send this important information to you.
http://register.tsi.com
As part of the registration process, you will be asked for your
comments on TSI products and services. TSI’s customer feedback
program gives customers like you a way to tell us how we are doing.
i
Copyright©
TSI Incorporated / 2008-2023 / All rights reserved.
Part Number
6002097 / Revision R / May 2023
Address
TSI Incorporated / 500 Cardigan Road / Shoreview, MN 55126 / USA
Fax No.
(651) 490-3824
LIMITATION OF WARRANTY AND LIABILITY (effective April 2014)
(For country-specific terms and conditions outside of the USA, please visit www.tsi.com.)
Seller warrants the goods, excluding software sold hereunder, under normal use and
service as described in the operator's manual, shall be free from defects in workmanship
and material for twenty-four (24) months, or if less, the length of time specified in the
operator's manual, from the date of shipment to the customer. This warranty period is
inclusive of any statutory warranty. This limited warranty is subject to the following
exclusions and exceptions:
a. Hot-wire or hot-film sensors used with research anemometers, and certain other
components when indicated in specifications, are warranted for 90 days from the
date of shipment;
b. DustTrak internal pump for Models 8530 and 8533 is warranted for two (2) years or
4000 hours, whichever comes first;
c. DustTrak external pump for Models 8530EP and 8533EP is warranted for two (2)
years or 8760 hours, whichever comes first;
d. DustTrak internal pump for Models 8530 and 8533 is warranted for operation within
ambient temperatures between 545°C. Warranty is void when the internal pump is
operating outside of this temperature range;
e. Parts repaired or replaced as a result of repair services are warranted to be free from
defects in workmanship and material, under normal use, for 90 days from the date of
shipment;
f. Seller does not provide any warranty on finished goods manufactured by others or on
any fuses, batteries or other consumable materials. Only the original manufacturer's
warranty applies;
g. This warranty does not cover calibration requirements, and seller warrants only that
the instrument or product is properly calibrated at the time of its manufacture.
Instruments returned for calibration are not covered by this warranty;
h. This warranty is VOID if the instrument is opened by anyone other than a factory
authorized service center with the one exception where requirements set forth in the
manual allow an operator to replace consumables or perform recommended
cleaning;
i. This warranty is VOID if the product has been misused, neglected, subjected to
accidental or intentional damage, or is not properly installed, maintained, or cleaned
according to the requirements of the manual. Unless specifically authorized in a
separate writing by Seller, Seller makes no warranty with respect to, and shall have
no liability in connection with, goods which are incorporated into other products or
equipment, or which are modified by any person other than Seller.
The foregoing is IN LIEU OF all other warranties and is subject to the LIMITATIONS
stated herein. NO OTHER EXPRESS OR IMPLIED WARRANTY OF FITNESS FOR
PARTICULAR PURPOSE OR MERCHANTABILITY IS MADE. WITH RESPECT TO
SELLER’S BREACH OF THE IMPLIED WARRANTY AGAINST INFRINGEMENT, SAID
WARRANTY IS LIMITED TO CLAIMS OF DIRECT INFRINGEMENT AND EXCLUDES
CLAIMS OF CONTRIBUTORY OR INDUCED INFRINGEMENTS. BUYER’S
EXCLUSIVE REMEDY SHALL BE THE RETURN OF THE PURCHASE PRICE
DISCOUNTED FOR REASONABLE WEAR AND TEAR OR AT SELLER’S OPTION
REPLACEMENT OF THE GOODS WITH NON-INFRINGING GOODS.
ii Model 8535
TO THE EXTENT PERMITTED BY LAW, THE EXCLUSIVE REMEDY OF THE USER OR
BUYER, AND THE LIMIT OF SELLER'S LIABILITY FOR ANY AND ALL LOSSES,
INJURIES, OR DAMAGES CONCERNING THE GOODS (INCLUDING CLAIMS BASED
ON CONTRACT, NEGLIGENCE, TORT, STRICT LIABILITY OR OTHERWISE) SHALL
BE THE RETURN OF GOODS TO SELLER AND THE REFUND OF THE PURCHASE
PRICE, OR, AT THE OPTION OF SELLER, THE REPAIR OR REPLACEMENT OF THE
GOODS. IN THE CASE OF SOFTWARE, SELLER WILL REPAIR OR REPLACE
DEFECTIVE SOFTWARE OR IF UNABLE TO DO SO, WILL REFUND THE PURCHASE
PRICE OF THE SOFTWARE. IN NO EVENT SHALL SELLER BE LIABLE FOR LOST
PROFITS, BUSINESS INTERRUPTION, OR ANY SPECIAL, INDIRECT,
CONSEQUENTIAL OR INCIDENTAL DAMAGES. SELLER SHALL NOT BE
RESPONSIBLE FOR INSTALLATION, DISMANTLING OR REINSTALLATION COSTS
OR CHARGES. No Action, regardless of form, may be brought against Seller more than
12 months after a cause of action has accrued. The goods returned under warranty to
Seller's factory shall be at Buyer's risk of loss, and will be returned, if at all, at Seller's risk
of loss.
Buyer and all users are deemed to have accepted this LIMITATION OF WARRANTY AND
LIABILITY, which contains the complete and exclusive limited warranty of Seller. This
LIMITATION OF WARRANTY AND LIABILITY may not be amended, modified or its terms
waived, except by writing signed by an Officer of Seller.
Service Policy
Knowing that inoperative or defective instruments are as detrimental to TSI as they are
to our customers, our service policy is designed to give prompt attention to any
problems. If any malfunction is discovered, please contact your nearest sales office or
representative, or call TSI's Customer Service department at (800) 680-1220 (USA) or
(001 651) 490-2860 (International) or visit www.tsi.com.
Trademarks
TSI and the TSI logo are registered trademarks of TSI Incorporated in the United
States and may be protected under other country’s trademark registrations. VELCRO is
a registered trademark of Velcro Industries B.V.
Operation and Maintenance Manual iii
Contents
Contents ............................................................................................. iii
Safety Information ............................................................................. v
Laser Safety ............................................................................................ vi
Labels ..................................................................................................... vii
Description of Caution/Warning Symbols .............................................. viii
Caution and Warning Symbols .............................................................. viii
Reusing and Recycling .......................................................................... viii
Chapter 1 Parts Identification and Unpacking ............................... 1
Parts Identification: Environmental Enclosure ......................................... 1
Unpacking the Model 8535 DustTrak Aerosol Monitor
Environmental Enclosure ..................................................................... 3
Model 8535 DustTrak Aerosol Monitor Environmental Enclosure
Optional Accessories ........................................................................... 4
Chapter 2 Setting Up ........................................................................ 9
Install the Aerosol or Respirable Aerosol Inlet ......................................... 9
Install Heat Shield Assembly (optional accessory) ................................ 10
Install Water Trap Bottle ........................................................................ 12
Install Sample Inlet Tubing .................................................................... 13
Install Impactor in the Sample Flow (optional accessory) ...................... 14
Install Thiamis 1000 GSM/GPS Communication Modem
(optional accessory) ........................................................................... 14
Install Heated Inlet Sample Conditioning Module (optional accessory) . 14
Install Battery Pack System (optional accessory) .................................. 15
Install External Weatherproof AC Adapter (if desired) ........................... 17
Install DustTrak™ Monitor in the Environmental Enclosure ................... 19
Connecting Internal USB Cables ........................................................... 20
Connecting Sample Tubing ................................................................... 22
Supplying Power to the DustTrak™ II/DRX Aerosol Monitor
(and optional accessories) ................................................................. 23
Mounting the Environmental Enclosure to a Surveyor Tripod ................ 25
Transporting the Environmental Enclosure ............................................ 26
Setting-up the Solar Cell Power System (optional accessory) ............... 27
Setting-up TrakPro Data Analysis Software ........................................ 38
Chapter 3 Operation ....................................................................... 39
Overview ............................................................................................... 39
How to Properly Orient the Environmental Enclosure ............................ 39
Changing and Re-charging the Battery Packs ....................................... 40
Zeroing the DustTrak™ II/DRX Aerosol Monitor ..................................... 44
Setting the Flow Rate for the Respirable Aerosol Inlet
(optional accessory) ........................................................................... 44
Locking the Environmental Enclosure ................................................... 44
Checklist for Sampling with the Environmental Enclosure ..................... 45
Communication and Data Acquisition .................................................... 45
iv Model 8535
Chapter 4 Maintenance ................................................................... 47
Daily Maintenance Checks ..................................................................... 47
Cleaning the Aerosol Inlet ...................................................................... 47
Cleaning the Respirable Aerosol Inlet .................................................... 47
When to Change the Battery .................................................................. 47
Replacing the Battery Pack .................................................................... 48
Recharging the Battery Packs ................................................................ 48
Battery Pack Life .................................................................................... 50
Emptying the Water Trap ....................................................................... 52
Storage Precautions ............................................................................... 52
Chapter 5 Troubleshooting the Environmental Enclosure ......... 53
Appendix A Specifications ............................................................. 55
Environment Enclosure .......................................................................... 55
Solar Power System ............................................................................... 56
Operation and Maintenance Manual v
Safety Information
WARNING
The DustTrak™ II/DRX aerosol monitor is not rated
for intrinsic safety. The DustTrak™ monitor, with
the Environmental Enclosure, must NEVER be
operated under conditions where there is a risk of
fire or explosion.
WARNING
Use of components other than those specified by
TSI® may impair the safety features provided by
the equipment.
WARNING
The instrument has been designed to be used with
batteries supplied by TSI®. DO NOT use a substitute.
The TSI® charger (P/N 801809) has been designed
to be used with the battery packs supplied by TSI®.
DO NOT use a substitute charger to charge TSI®
battery packs.
Old batteries must be properly recycled in
accordance with the local environmental regulations.
WARNING
DO NOT use non-rechargeable batteries in this
instrument. Fire, explosions, or other hazards
may result.
C A U T I O N
The enclosure is designed to be water resistant to
rain or spray. It is not designed to be waterproof
when immersed. Setting it in a pool of water will
result in flooding the inner compartment with water.
This will severely damage both your DustTrak II /
DRX aerosol monitor and battery pack. DO NOT set
the Environmental Enclosure in water!
The enclosure has no IP rating or NEMA rating for
dust or water ingress.
vi Model 8535
NOTICE
Prior to using the Battery Pack for the first time, a full recharge is
recommended. Recharging Battery Pack(s) immediately after
use (within one hour maximum) is critical to obtaining optimal
recharge time, battery health, and battery life.
NOTICE
This product must not be used in residential areas or in
establishments directly connected to a low voltage power supply
network which supplies buildings used for domestic purposes.
Laser Safety
The Model 8533/8534 DustTrak DRX monitor is a Class I laser-
based instrument
During normal operation, you will not be exposed to laser radiation
Precaution should be taken to avoid exposure to hazardous
radiation in the form of intense, focused, visible light
Exposure to this light may cause blindness
Take these precautions:
DO NOT remove any parts from the DustTrak DRX monitor
unless you are specifically told to do so in this manual
DO NOT remove the housing or covers. There are no serviceable
components inside the housing
WARNING
The use of controls, adjustments, or procedures other
than those specified in this manual may result in
exposure to hazardous optical radiation.
WARNING
There are no user-serviceable parts inside this
instrument. The instrument should only be opened by
TSI® or a TSI® approved service technician.
WARNING
If the DustTrak monitor is used in a manner not
specified by the manufacturer, the protection
provided by the equipment may be impaired.
Operation and Maintenance Manual vii
When operated according to the manufacturer’s instruction, this
device is a Class I laser product as defined by U.S. Department of
Health and Human Services standards under the Radiation Control
for Health and Safety Act of 1968. A certification and identification
label like the one shown below is affixed to each instrument.
Labels
Advisory labels and identification labels are attached to the
instrument.
1. Serial Number Label
(bottom)
2. Laser Radiation Label
(internal)
DANGER!
VISIBLE LASER RADIATION
WHEN
OPEN. AVOID DIRECT
EXPOSURE TO BEAM
WARNING: NO USER
SERVICABLE
PARTS INSIDE. REFER
SERVICING
TO QUALIFIED PERSONNEL
or
4. European symbol for non-disposable item. Item
must be recycled.
viii Model 8535
Description of Caution/Warning Symbols
Appropriate caution/warning statements are used throughout the
manual and on the instrument that require you to take cautionary
measures when working with the instrument.
Caution
C A U T I O N
Failure to follow the procedures prescribed in this
manual might result in irreparable equipment
damage. Important information about the operation
and maintenance of this instrument is included in
this manual.
Warning
WARNING
Warning means that unsafe use of the instrument
could result in serious injury to you or cause damage
to the instrument. Follow the procedures prescribed.
Caution and Warning Symbols
The following symbols may accompany cautions and warnings to
indicate the nature and consequences of hazards:
Warns that the instrument contains a laser and that
important information about its safe operation and
maintenance is included in the manual.
Warns that the instrument is susceptible to electro-
static discharge (ESD) and ESD protection should be
followed to avoid damage.
Indicates the connector is connected to earth ground
and cabinet ground.
Reusing and Recycling
As part of TSI® Incorporated’s effort to have a minimal
negative impact on the communities in which its
products are manufactured and used:
DO NOT dispose of used batteries in the trash.
Follow local environmental requirements for
battery recycling.
If instrument becomes obsolete, return to TSI® for
disassembly and recycling.
1
Chapter 1
Parts Identification and Unpacking
Carefully unpack the Model 8535 DustTrak™ Aerosol Monitor
Environmental Enclosure from the shipping container. Use the tables
and illustrations below to make certain that there are no missing
components. Contact TSI® Incorporated (TSI®) immediately if
anything is missing or damaged.
Parts Identification: Environmental Enclosure
Identify the parts of the Model 8535 DustTrak Aerosol Monitor
Environmental Enclosure (Figure 1 and Figure 2). Please become
familiar with these components before proceeding.
Figure 1: Inside View of Environmental Enclosure
Water Trap
Bottle
Battery
VELCRO
Strap
Inlet
Fitting
DustTrak
VELCRO
Strap
DC Power
Cable
Tripod
Mount
Screw Plug
DustTrak
Battery
Bracket
2 Chapter 1
Figure 2: Outside Views of Environmental Enclosure
Antenna Cable
Strain Relief and
Plug
USB Pass
Through
Connection
Antenna Mount
Location
DC Power
Connector
Heat Shield
Mount
Location
Inlet Assembly
(sampling)
Inlet Ring
AC Adapter Mount
Location
Parts Identification and Unpacking 3
Unpacking the Model 8535 DustTrak Aerosol
Monitor Environmental Enclosure
Compare all the components you received with those listed in the
table below. If any parts are missing, contact TSI® Incorporated.
Item
Qty
Part Number
Description
1
801567
Environmental
Enclosure Omni-
Direct Inlet with
Extended Rain Cap
1
Refer to
801830
accessory kit.
Sample Inlet Tubing
1
Refer to
801830
accessory kit.
Inlet Fitting for
Sample Tubing
1
Refer to
801830
accessory kit.
Water Trap Bottle
4
Refer to
801830
accessory kit.
VELCRO® Straps
1
Refer to
801830
accessory kit.
Internal DC Power
Cable
1
801842-XX
12 VDC
Weatherproof
Power Adapter
(100 to 250 VAC,
CE Compliant)
4 Chapter 1
Item
Qty
Part Number
Description
1
Refer to
801830
accessory kit.
Transport Inlet Plug
1
Refer to
801830
accessory kit.
Tripod Mount
Screw Plug
1
Refer to
801830
accessory kit.
Tripod Mount
Screw Plug Wrench
1
Refer to
801830
accessory kit.
Antenna Cable
Fitting Plug
1
6002097
Model 8535
DustTrak™ Aerosol
Monitor
Environmental
Enclosure
Operation and
Maintenance
Manual
Model 8535 DustTrak Aerosol Monitor
Environmental Enclosure Optional Accessories
Listed below are optional accessory items for the Model 8535
DustTrak Aerosol Monitor Environmental Enclosure. Contact TSI®
for purchase information.
Item
Part Number
Description
801807
Internal 12 VDC
Battery System
(includes 2
batteries and
1 charger)
Parts Identification and Unpacking 5
Item
Part Number
Description
801817
Dual Battery Wiring
Harness (For use
with 2 battery
packs, See
Chapter 3 for
details)
801808
Replacement 12
VDC Battery Pack
801809
Replacement 12
VDC Battery Pack
Charger
801810
Heat Shield
Assembly
854057
Tripod
6 Chapter 1
Item
Part Number
Description
801811
Solar Cell Power
System
801901
GSM/GPS
Communication
Modem with 8535
mounting kit.
801850
Heated Inlet
Accessory
801850 with
Auto Zero Module
801851
Heated Inlet
Accessory
801851 without
Auto Zero (for users
who currently own
an Auto Zero
Module)
801830
Environmental
Enclosure Spare
Parts
801903
Router Kit
Parts Identification and Unpacking 7
Item
Part Number
Description
801906
Router Mounting Kit
8 Chapter 1
(This page intentionally left blank)
9
Chapter 2
Setting Up
The setup of the Environmental Enclosure is an important part in
allowing reliable and accurate sampling of aerosols in a wide range of
conditions. TSI® cannot ensure accurate measurements if any of the
components are set up incorrectly. Failure to follow these procedures
could result in damage to the enclosure or its components.
Install the Aerosol or Respirable Aerosol Inlet
The inlet assembly allows ambient aerosols to be sampled efficiently
in a wide range of conditions. Proper installation into its sampling
position ensures that the DustTrak II/DRX aerosol monitor
measures an accurate ambient aerosol mass concentration. Follow
these instructions to install either of the aerosol inlets:
1. Remove the Inlet Transport Plug from bubble pack.
C A U T I O N
Always hold the inlet assembly by the knurled portion
of the tube. DO NOT twist on the inlet assembly cap.
2. Thread the inlet assembly into upper inlet ring found on the top of
the Environmental Enclosure (see Figure 3).
Figure 3: Thread the Inlet into the Inlet Ring
10 Chapter 2
NOTICE
Make sure that a thin film of vacuum grease is coating both O-rings
on the inlet tube to ensure a good seal before installing it.
DO NOT use the Respirable Aerosol Inlet in the rain. As it will
aspirate rainwater into the cyclone body.
If a Heat Shield is used, you need to install the inlet after you install
the Heat Shield (see Figure 6).
Install Heat Shield Assembly (optional accessory)
The optional Heat Shield Assembly is designed to protect the
DustTrak monitor from excessive heating inside the Environmental
Enclosure due to direct sunlight.
NOTICE
Be sure to remove the Inlet Transport Plug from the top of the
Enclosure prior to attaching the Heat Shield Assembly.
1. Remove the 4 waterproof bolt assemblies from the Heat Shield
mount locations (Figure 4). Save these nuts and bolts for future
usage when the Heat Shield is not used.
Figure 4: Remove Waterproof Bolt Assemblies from
Heat Shield Mount Locations
2. Remove the Inlet Transport Plug from the top of the Enclosure at
the Inlet Ring.
/