Vaisala HMP60 Owner's manual

Category
Measuring, testing & control
Type
Owner's manual

This manual is also suitable for

USER'S GUIDE
Vaisala Humidity and Temperature Probe
HMP60 and HMP110
M211060EN-A
PUBLISHED BY
Vaisala Oyj Phone (int.): +358 9 8949 1
P.O. Box 26 Fax: +358 9 8949 2227
FI-00421 Helsinki
Finland
Visit our Internet pages at www.vaisala.com
© Vaisala 2010
No part of this manual may be reproduced in any form or by any means, electronic or
mechanical (including photocopying), nor may its contents be communicated to a third
party without prior written permission of the copyright holder.
The contents are subject to change without prior notice.
Please observe that this manual does not create any legally binding obligations for
Vaisala towards the customer or end user. All legally binding commitments and
agreements are included exclusively in the applicable supply contract or Conditions of
Sale.
________________________________________________________________________________
VAISALA________________________________________________________________________ 1
Table of Contents
CHAPTER 1
GENERAL INFORMATION............................................................................7
About This Manual...................................................................7
Contents of This Manual.......................................................7
Version Information...............................................................8
Related Manuals...................................................................8
Documentation Conventions.................................................8
Safety.........................................................................................9
ESD Protection.........................................................................9
Recycling ................................................................................10
Regulatory Compliances.......................................................10
Patent Notice ..........................................................................10
Trademarks.............................................................................11
License Agreement................................................................11
Warranty..................................................................................11
CHAPTER 2
PRODUCT OVERVIEW................................................................................13
Introduction to HMP60 and HMP110 ....................................13
Basic Features and Options..................................................14
Filter Options.......................................................................15
Installation Accessories (Optional)......................................16
Probe Mounting Clamp .......................................................16
Probe Mounting Flange.......................................................16
Loop Power Converter........................................................17
Cables.................................................................................18
CHAPTER 3
INSTALLATION............................................................................................19
Dimensions.............................................................................19
Mounting the Probe................................................................20
Wiring ......................................................................................21
Wiring with the Loop Power Converter ...............................22
Power Supply Requirements...............................................23
Recommendations.........................................................23
CHAPTER 4
OPERATION.................................................................................................25
Getting Started .......................................................................25
Serial Line Communication...................................................25
Connecting to the Serial Interface.......................................25
Installing the Driver for the USB Cable...............................27
Terminal Application Settings for Probes in RS-485 Mode.28
User's Guide ______________________________________________________________________
2 ___________________________________________________________________ M211060EN-A
Terminal Application Settings for Probes in Analog Mode..30
List of Serial Commands .......................................................31
Device Information and Status..............................................32
View Device Information......................................................32
View Calibration Information ...............................................33
View Order Code.................................................................33
View Serial Number.............................................................33
View Software Version........................................................34
Serial Line Output Commands..............................................34
Start Measurement Output..................................................34
Stop Measurement Output..................................................34
Output the Measurement Message Once ...........................35
Configuring Serial Line Operation........................................35
Set Serial Line Settings.......................................................35
Set Serial Interface Mode....................................................36
Set Output Interval...............................................................37
Set Measurement Filtering..................................................37
Set Probe Address ..............................................................38
Set Serial Interface Delay....................................................38
Set Measurement Units.......................................................39
Calibration Commands..........................................................39
Calibrate Humidity Measurement........................................39
Clear Adjustment of RH Measurement................................40
Calibrate Temperature Measurement .................................40
Clear Adjustment of T Measurement...................................41
View User Adjustment Parameters .....................................41
Other Commands ...................................................................42
Display Command List ........................................................42
Display the Currently Active Errors .....................................42
Connect to the Probe in POLL Mode ..................................42
Close the Connection in POLL Mode..................................42
Reset the Probe...................................................................43
Restore Factory Settings.....................................................43
CHAPTER 5
MAINTENANCE............................................................................................45
Periodic Maintenance.............................................................45
Cleaning ..............................................................................45
Changing the Filter..............................................................46
Calibration Procedure............................................................46
Adjustment Procedure (HMP110)..........................................47
One-point Adjustment of RH Measurement (HMP110).......47
Two-point Adjustment of RH Measurement (HMP110).......48
One-point Adjustment of T Measurement (HMP110)..........50
Repair Maintenance................................................................52
Changing the INTERCAP Sensor (HMP60)........................52
Changing the HUMICAP 180R Sensor (HMP110)..............53
CHAPTER 6
TROUBLESHOOTING..................................................................................55
Solving Typical Problems......................................................55
Technical Support ..................................................................56
________________________________________________________________________________
VAISALA________________________________________________________________________ 3
CHAPTER 7
TECHNICAL DATA......................................................................................57
Specifications.........................................................................57
Performance (HMP60)........................................................57
Relative Humidity...........................................................57
Temperature...................................................................57
Dewpoint........................................................................57
Performance (HMP110)......................................................58
Relative Humidity...........................................................58
Temperature...................................................................58
Dewpoint........................................................................58
Operating Environment.......................................................59
Inputs and Outputs..............................................................59
Mechanics...........................................................................59
Options and Accessories......................................................60
User's Guide ______________________________________________________________________
4 ___________________________________________________________________ M211060EN-A
This page intentionally left blank.
________________________________________________________________________________
VAISALA________________________________________________________________________ 5
List of Figures
Figure 1 HMP60 (left) and HMP110 (right).............................................13
Figure 2 Filter Options............................................................................15
Figure 3 Probe Mounting Clamp in Use.................................................16
Figure 4 Probe Mounting Flange............................................................16
Figure 5 Loop Power Converter .............................................................17
Figure 6 Cable with Threaded Connector ..............................................18
Figure 7 USB Serial Interface Cable ......................................................18
Figure 8 HMP60 and HMP110 Dimensions ...........................................19
Figure 9 Installation with Plastic M12 Nuts, Dimensions........................19
Figure 10 Wiring of Analog Output...........................................................21
Figure 11 Wiring of Digital Output ............................................................21
Figure 12 Wiring with the Loop Power Converter Module........................22
Figure 13 PuTTY Terminal Application ....................................................29
Figure 14 Changing the Sensor................................................................52
List of Tables
Table 1 Manual Revisions.......................................................................8
Table 2 Related Manuals ........................................................................8
Table 3 Quantities Measured by the HMP60 and HMP110..................14
Table 4 Filter Properties........................................................................15
Table 5 Pinout of the Probe Connector.................................................21
Table 6 Default Serial Communication Settings....................................26
Table 7 List of Serial Commands..........................................................31
Table 8 Additional Commands for Probes with RS-485 Output............31
Table 9 Troubleshooting Table..............................................................55
Table 10 HMP60 Options and Accessories ............................................60
Table 11 HMP110 Options and Accessories ..........................................60
User's Guide ______________________________________________________________________
6 ___________________________________________________________________ M211060EN-A
This page intentionally left blank.
Chapter 1 ________________________________________________________ General Information
VAISALA________________________________________________________________________ 7
CHAPTER 1
GENERAL INFORMATION
This chapter provides general notes for the manual and the product.
About This Manual
This manual provides information for installing, operating, and
maintaining Vaisala Humidity and Temperature Probe HMP60 and
HMP110.
Contents of This Manual
This manual consists of the following chapters:
- Chapter 1, General Information, provides general notes for the manual
and the product.
- Chapter 2, Product Overview, introduces the product features and
options.
- Chapter 3, Installation, provides you with information that is intended
to help you install HMP60 and HMP110.
- Chapter 4, Operation, contains information that is needed to operate
HMP60 and HMP110.
- Chapter 5, Maintenance, provides information that is needed in basic
maintenance of HMP60 and HMP110.
- Chapter 6, Troubleshooting, describes common problems, their
probable causes and remedies, and contact information for technical
support.
- Chapter 7, Technical Data, provides the technical data of the HMP60
and HMP110.
User's Guide ______________________________________________________________________
8 ___________________________________________________________________ M211060EN-A
Version Information
Table 1 Manual Revisions
Manual Code Description
M211060EN-A March 2010. This manual. First version.
Related Manuals
Table 2 Related Manuals
Manual Code Manual Name
M211059EN HMP60 and HMP110 Quick Guide
M211106EN Loop Power Converter Quick Reference guide
Documentation Conventions
Throughout the manual, important safety considerations are highlighted
as follows:
WARNING
Warning alerts you to a serious hazard. If you do not read and follow
instructions very carefully at this point, there is a risk of injury or even
death.
CAUTION
Caution warns you of a potential hazard. If you do not read and follow
instructions carefully at this point, the product could be damaged or
important data could be lost.
NOTE
Note highlights important information on using the product.
Chapter 1 ________________________________________________________ General Information
VAISALA________________________________________________________________________ 9
Safety
The product delivered to you has been tested for safety and approved as
shipped from the factory. Note the following precautions:
WARNING
Ground the product, and verify outdoor installation grounding
periodically to minimize shock hazard.
CAUTION
Do not modify the unit. Improper modification can damage the product
or lead to malfunction.
ESD Protection
Electrostatic Discharge (ESD) can cause immediate or latent damage to
electronic circuits. Vaisala products are adequately protected against
ESD for their intended use. It is possible to damage the product,
however, by delivering electrostatic discharges when touching,
removing, or inserting any objects inside the equipment housing.
To make sure you are not delivering high static voltages yourself:
- Handle ESD sensitive components on a properly grounded and
protected ESD workbench.
- When an ESD workbench is not available, ground yourself to the
equipment chassis with a wrist strap and a resistive connection cord.
- If you are unable to take either of the above precautions, touch a
conductive part of the equipment chassis with your other hand before
touching ESD sensitive components.
- Always hold component boards by the edges and avoid touching the
component contacts.
User's Guide ______________________________________________________________________
10 __________________________________________________________________ M211060EN-A
Recycling
Recycle all applicable material.
Dispose of batteries and the unit according to statutory regulations. Do
not dispose of with regular household refuse.
Regulatory Compliances
The HMP60 and HMP110 are in conformity with the provisions of the
following EU directive(s):
ROHS Directive (2002/95/EEC)
WEEE Directive (2002/96/EEC)
EMC Directive (2004/108/EC)
The electromagnetic compatibility of the HMP60 and HMP110 has been
tested according to the following product family standards:
EN 61326-1:2006 Electrical equipment for measurement, control and
laboratory use - EMC requirements - Basic immunity test requirements.
EN 55022:2006 + Am 1:2007 Information technology equipment - Radio
disturbance characteristics - Limits and methods of measurement.
Class B.
Patent Notice
The HMP60 and HMP110 are protected by the following patents and
patent applications and their corresponding national rights:
Finnish patent 98861, French patent 6650303, German patent 69418174,
Japanese patent 3585973, UK patent 0665303, U.S. patent 5607564.
Chapter 1 ________________________________________________________ General Information
VAISALA_______________________________________________________________________ 11
Trademarks
Vaisala INTERCAP® and Vaisala HUMICAP® are registered
trademarks of Vaisala Oyj.
Windows® is a registered trademark of Microsoft Corporation in the
United States and/or other countries.
License Agreement
All rights to any software are held by Vaisala or third parties. The
customer is allowed to use the software only to the extent that is provided
by the applicable supply contract or Software License Agreement.
Warranty
For certain products Vaisala normally gives a limited one-year warranty.
Visit our Internet pages for more information and our standard warranty
terms and conditions: www.vaisala.com/services/warranty.html.
Please observe that any such warranty may not be valid in case of
damage due to normal wear and tear, exceptional operating conditions,
negligent handling or installation, or unauthorized modifications. Please
see the applicable supply contract or Conditions of Sale for details of the
warranty for each product.
User's Guide ______________________________________________________________________
12 __________________________________________________________________ M211060EN-A
This page intentionally left blank.
Chapter 2 __________________________________________________________ Product Overview
VAISALA_______________________________________________________________________ 13
CHAPTER 2
PRODUCT OVERVIEW
This chapter introduces the product features and options.
Introduction to HMP60 and HMP110
Vaisala Humidity and Temperature Probes HMP60 and HMP110 are
simple and cost-effective humidity transmitters suitable for volume
applications, integration into other manufacturers’ equipment, incubators,
glove boxes, greenhouses, fermentation chambers, and data loggers.
HMP60 uses the interchangeable Vaisala INTERCAP® sensor. No
recalibration is required after sensor replacement.
HMP110 uses the Vaisala HUMICAP 180R® sensor for increased
accuracy. The HMP110 requires calibration after sensor replacement.
This can be done on the serial line using the optional Vaisala USB cable.
0912-101
Figure 1 HMP60 (left) and HMP110 (right)
User's Guide ______________________________________________________________________
14 __________________________________________________________________ M211060EN-A
Table 3 Quantities Measured by the HMP60 and HMP110
Quantity Abbreviation Metric Unit Non Metric Unit
Relative humidity RH %RH %RH
Dewpoint/Frostpoint
temperature*
T
d/f
ºC ºF
Temperature T ºC ºF
* When the dewpoint is below 0 °C, the probe outputs frostpoint for T
d
Basic Features and Options
- Analog output mode: two analog output channels, selectable from
0 ... 1 V / 0 ... 2.5 V / 0 ... 5 V / 1 ... 5 V
- RS-485 output mode
- Small size
- Low power consumption
- IP65 aluminum body
- Options and accessories:
- Filter options: plastic grid filter, membrane filter, and stainless
steel sintered filter
- One channel loop power converter 4 ... 20 mA (separate module,
compatible with humidity accuracy only)
- Shielded 0.3 m and 3.0 m connection cables with threaded
connector for probe connection, open end wires on the other end
- Plastic M12 installation nuts
Chapter 2 __________________________________________________________ Product Overview
VAISALA_______________________________________________________________________ 15
Filter Options
The available filters for HMP60 and HMP110 are shown in Figure 2
below. For order codes, see section Options and Accessories on page 60.
1001-008
Figure 2 Filter Options
Table 4 Filter Properties
No. Filter Diameter Pore Size
1 Plastic grid filter (fastest response time) 12 mm -
2 Membrane filter 12 mm 0.2 µm
3 Stainless steel sintered filter 12 mm 38 µm
User's Guide ______________________________________________________________________
16 __________________________________________________________________ M211060EN-A
Installation Accessories (Optional)
Probe Mounting Clamp
The optional mounting clamp makes it easy to install the probe on the
wall of the measurement environment. The probe can be detached for
calibration simply by loosening the lower screw.
1001-138
Figure 3 Probe Mounting Clamp in Use
Probe Mounting Flange
The probe mounting flange is a silicone flange can be used to hold the
probe in a through-wall installation. The flange is a general purpose
mounting accessory for Ø 12mm probes, and comes with a sealing plug
for coaxial cables that is not needed when the flange is used with HMP60
and HMP110.
0911-109
Figure 4 Probe Mounting Flange
Chapter 2 __________________________________________________________ Product Overview
VAISALA_______________________________________________________________________ 17
Loop Power Converter
The loop power converter is an open frame module that converts one
0 ... 2.5 VDC voltage output to a 4 ... 20 mA current output. To use the
loop power converter module, the probe:
- must be in the analog output mode
- the desired quantity must be on channel 1
- channel 1 must be scaled to 0 ... 2.5 V
Wiring instructions are provided in section Wiring with the Loop Power
Converter on page 22.
1001-001
Figure 5 Loop Power Converter
User's Guide ______________________________________________________________________
18 __________________________________________________________________ M211060EN-A
Cables
The probe cables have a standard M8 connector. Also other compatible
M8 series cables can be used.
0910-135
Figure 6 Cable with Threaded Connector
The USB Serial Interface Cable uses a snap-on connector. The USB
cable is intended for maintenance purposes only, not permanent
installation.
0809-002
Figure 7 USB Serial Interface Cable
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64

Vaisala HMP60 Owner's manual

Category
Measuring, testing & control
Type
Owner's manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI