Benchmark Scientific P7700-1, P7700-10, P7700-100, P7700-1000, P7700-10M, P7700-20, P7700-200, P7700-5M, P7700-CAR, P7700-T10M, P7700-T5M Owner's manual

  • Hello! I am an AI chatbot trained to assist you with the Benchmark Scientific P7700-1 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
InstructionManual
adivisionof
1
Introduction……………………………………………………………...1
Parts&Features………………………………………………………..2
TipCompatibility……………………………………………………….
3
Operation………………………………………………………………….
4
TipsforProperPipetting…………………………………………...5
CleaningandSterilization………………………………….........
5
Specifications…………………………………………………………….6
CheckingAccuracyandPrecision……………………………….
6
Calibration…………………………………………………………………8
Troubleshooting………………………………………………………..
9
Introduction:
AccurisNextPettePipettesaredesignedforlaboratoryresearchuse,andspecificallyforthe
measurementandtransferofaqueoussolutions.TheNextPettepipettesoffervariablevolume
setting,andareavailableindifferentrangestocovervolumemeasurementsfrom0.1ulto
10ml.
NextPettepipettorsutilizetheprincipleofairdisplacement.Avacuumis
createdbythe
plungermovement,andaprecisevolumeofsampleliquidisdrawnintotheattachedpipette
tip.Disposablepipettipsmustalwaysbeusedforliquidaspiration,andtheshaftsofthe
NextPettepipettesaredesignedtoacceptawiderangeofdifferentbrandsandstylesof
disposable
pipettetips.
Properoperationandproperuseofdisposabletipswillpreventliquidsamplesfromentering
thepipetteandcontaminatingtheinternalparts.
EachvariablevolumeNextPetteincorporatesa4digitcounterdisplaytoshowthesetvolume.
Thevolumesettingcanbechangedusingthevolumeadjustmentdialbelowthepushbutton.A
volumeadjustmentlockislocatedjustbelowthetipejectorbutton,andthisallowstheuserto
securelylocktheadjustmentdialattheappropriatelysetvolume.
2
Thetopofthefingerhookhasaplaceundertheremovablewindowfor
anidentificationlabel.Eachpipettecomeswithasetofdifferent
coloredlabelsforcolorcoding.Underthewindowandlabelistheserial
numberlabelandcalibrationswitch.Seethecalibrationsectionofthis
manualfor
details.
3
TheNextPettePipettesfeatureauniversaltipfittingshaftthatisdesignedtoworkwithmost
manufacturer’sbrandsofpipettetips.Beforepipetting,alwaysensurethatapipettetipofthe
appropriatesizehasbeenattached.
SeebelowforachartidentifyingtheappropriatetipsizeforeachNextPettePipette:
P777001: 0.11μl 10ultip
P770010 0.510μl 10ultip
P770020 220μl 200ultip
P7700100 10100μl 200ultip
P7700200 20200μl 200ultip
P77001000 1001000μl 1000ultip
P77005M 5005000μl 5000ultip
P770010M 100010000μl 10,000ultip
AttachingtheTip:
Attachthetipbyfirmlypressingthepipetteconeintothetip.Apropersealisnecessaryfor
optimalpipettingaccuracyandprecision.
NOTE:For5mland10mlpipettes(P77005MandP770010M)alwaysensurethattheincluded
filterisproperlyinstalledontothepipetteshaftpriortotipattachment.Failuretodosocan
resultinharmtotheinternalcomponentsofthepipette.
4
(Beforeoperation,pleaserefertothe“attachingthetip”instructionsabove.)
1. Volumesetting:
Ensurethatthevolumeadjustmentlockisintheunlockedposition
(pushedup).Turnthevolumeadjustmentdialandmonitorthe
volumedisplaytoselectthepipettingvolume.Whenthevolume
isselected,pushthevolumelocktothelockedposition.
2. Aspiratetheliquidsample:
Pushtheplungerbuttondownwiththethumbtothefirststop.
Holdingthepipettevertically,immersethepipettipintotheliquidsample2mmto3mm.
Slowlyreleasepressureontheplungerbuttonsothattheliquidsampleisdrawnintothe
pipettip.NOTE:Itisveryimportantthatthesampleliquidisaspiratedslowly.Leavethe
tipimmersedinthesamplefor1to2secondsthenremoveitfromthesample.
NOTE:Alwayskeepthepipetteinanuprightverticalpositionwhenthereisliquidinthe
pipettetip.Layingthepipettehorizontallymaycauseliquidtoflowfromthetipinto the
pipetteshaft,causingcontamination.
3. Dispensing
Placethepipettipsoitistouchingthewallofthecontainerortubeintowhichthesample
willbedispensed.Thepipetteshouldbeheldata30to40degreeanglerelativetothewall
ofthecontainer.
Presstheplungerbuttonslowlytothefirststoptodispensethesample.Increasepressure
ontheplungerbuttontopassthefirststopandreachthesecondstop(thisistheblowout
stepwhichhelpstodispensethefullamountofsamplefromthetip).
Duringtheblowoutstep,slowlymovethepipettipagainstthe wallofthevessel.Thiswill
helptoremoveanyremainingliquidfromthepipettip.
Removethepipettetipsawayfromthevesselwallandallowtheplungertoreturntothe
startingposition.
4. Ejectingthetip
Holdthepipetteoverasuitablewastecontainerandpressthetipejectorbuttontoremove
thetip.
1.2.3.4.

SecondStop
VolumeSetting
Aspirating
FirstStop
Dispensing Ejectingtip
Example:
5
Accuratepipettingrequiresskillsthatimprovewithpractice.Itisalsonecessarytofollow
recommendedbestpracticesandunderstandproperpipettingtechnique.Thefollowingare
recommendationsonhowtogetthebestperformancefromyourNextPettepipette:
Alwaysoperatetheplungerbuttonslowlyandsmoothlywhenaspiratinganddispensing
samples.
Theendofthepipettetipsshouldbeimmersed2to3mmbelowthesurfaceofthe
sampletobeaspirated,andthisdepthorimmersionshouldbekeptconstantduringthe
aspiration.
Alwayskeepthepipetteinaverticalpositionduringthepipettingprocess.
Useanewpipettetipwheneverswitchingfromonetypeofsampletoanother.
Carefullycheckthepipettetipsused.Anyinconsistenciesinmanufacturewillresultin
inaccurateandimpreciseresults.
Carefullycheckthepipettetipsforleftoversampleliquidstayinginsidethetipduring
thepipettingprocess.
Alwaysusetheprerinsingtechniqueonanewtipbeforeaspiratingthemeasured
volume.
Sampleliquidshouldnotenterthepipetteshaft.
Neverlaythepipettehorizontallywhenthereissampleliquidinapipettetip.
Donotaspirateliquidsabove70C.
Afterpipettingacidsoranyotheraggressiveliquids,itisrecommendedtodisassemble
thepipetteandrinsetheplunger,shaft,sealandotherinternalelementswithdistilled
water.
Cleaning:Externalsurfacesofthepipettecanbecleanedwithwaterorisopropylalcohol.
Sterilization:TheNextPetteisfullyautoclavableat121°Cfor20minutes.Afterautoclaving,itis
importanttoletthepipettecompletelycooltoroomtemperaturepriortouse.
Important:Alwayschecktheautoclavetoconfirmthemaximumtemperaturesettingof121°C
priortoautoclavingtheNextPette.Temperatureshigherthan121°Corasterilizationprocess
longerthan20minutesmaydamagetheNextPette.
6
Item No. Volume Increment
A*≤±%
Min volume
A*±%
Max volume
CV*≤±%
Min volume
CV*≤±%
Max volume
P7700-1
0.1-1μl 0.001ul)
20.00 2.00 12.00 1.20
P7700-10
0.5-10μl 0.01ul)
3.80 0.50 4.00 0.50
P7700-20
2-20μl 0.02ul)
3.00 0.80 3.00 0.80
P7700-100
10-100μl 0.1ul)
1.50 0.80 1.00 0.20
P7700-200
20-200μl 0.2ul)
1.20 0.60 1.50 0.80
P7700-1000
100-1000μl 1ul)
1.50 0.60 1.00 0.60
P7700-5M
500-5000μl 5ul)
0.60 0.50 0.60 0.50
P7700-10M
1000-10000μl 10ul)
2.50 0.50 2.00 0.50
ItisrecommendedthateachNextPettepipetteischeckedonaregularbasistoconfirmthatitis
withinitscalibrationspecifications.
Thegravimetrictestmethod(accordingtoDINENISO8655)isrecommendedforchecking
accuracyandprecision.
Itisrecommendedtouseafullycalibratedelectronicbalanceforthegravimetrictesting.A4
digitbalancecanbeusedtotestvolumesgreaterthan10ul.For10ulandlessa5digitbalance
isrecommended.For1.0ulandless,a6digitbalanceisrecommended.
1. Setthemaximumvolumeonthepipette.
2. Conditionthepipettebeforetesting.Attachanappropriatepipettetip,aspirateand
dispensethetestliquid(distilledwater,at20°C,isrecommended)5times.Discardthe
pipettetip.
3. Performagravimetrictest:
a. Attachanewpipettetip.Note:thepipettetipsshouldbeofhighquality,and
shouldbethesametipsthat
willbeusedonaregularbasiswiththepipette.
b. Prerinsethetipsafewtimesbyaspiratinganddispensingthetestliquid.This
willcoattheinnersurfaceofthetipwithliquidandenableanaccuratetest.
c. Weighthemeasuredvolumeofliquid fromthepipette
byaspiratingtoa
weighingvesselonthebalance.
d. Calculatetheliquidvolumebyconvertingusingthecorrectionfactorconsidering
thetemperatureofthewater.
e. Atleast10aspirationsandweighingsin3volumeranges:100%,50%,10%of
nominalvolume,arerecommendedinordertocalculatethe
accuracyand
precisionresults
7
Calculation(fornominalvolume)
X
1
=Weighingresults Z=Correctionfactor
N=Numberofweighings (e.g.1.0029ul/mgat20°C,1013hPa)
 
Accuracy CoefficientofVariation


StandardDeviation
NOTE:TheaccuracyandprecisionspecificationshavebeendevelopedusingBenchmark
pipettetips,usingthegravimetricmethodandperforming10measurementsofdistilled
wateratatemperatureof20°C+/‐1°CaccordingtotheENISO8655standard.
8
TheNextPetteisfactorycalibratedforusewithaqueoussolutions.Ifthepipetteoperationis
inaccuratewhenpipettingaqueoussolutions,oritisdesiredtopipetteliquidsofadensityor
viscositydifferentthanwater,orspecialtypesofpipettetipsaretobeused,itmaybe
necessarytoadjust
thecalibration.
1. Settheadjustmentdialofthepipettortoavolumefortesting(ie.Thenominalvolume
ofthepipette).
2. Pipetteameasuredvolumeofsampleliquidontoananalyticalbalance,andusingthe
gravimetricmethod(withtheappropriatecorrectionfactor),calculatetheactual
volumeofthepipettedsample.Iftheactualvolumediffersfromthevolumesetonthe
NextPettedisplay,itisnecessarytoadjustthecalibration.
3. Removetheclearwindowandlabelonthetopsurfaceofthefingerhook.
4. Pushtheredcalibrationlockawayfromtheadjustmentdialandsimultaneouslypull
upwardonthevolumesettingdialuntilitclicksintotheraisedposition.Releasethered
calibrationlock.
5. Turnthevolumeadjustmentdialtothevolumethathasbeencalculatedforthesample
thatwasmeasured(theactualvolume).
6. Pushthecalibrationlockagainawayfromtheadjustmentdial,simultaneouslypushthe
volumesettingdialdownuntilitlocksintoposition.Releasethecalibrationlock.
7. AdjustthevolumeoftheNextPettebacktothenominalvolume,andtestthecalibration
againusingthegravimetricmethod.Ifthevalueisstilloff,repeatthecalibration
procedure.
3.Removethe
clearwindow
4.Pullupvolume
adjustmentdial
5.Turnvolume
adjustmentdial
6.Pushdown
volumesettingdial
9
Dropletsofliquidremaininthetipafterdispensing:
Thetipisemptiedtoofast
Solution1.)Decreasethespeedandforcewhenpressingtheplungerbutton
Thetiphasbeenoverused
Solution1.)Replacethepipettetipwithanewone
Airpocketsappearintheliquid
ofthepipettetip
Thesampleisaspiratedtoofast
Solution1.)Decreasethespeedofaspiration(raisethepipettingbuttonsmoothly
duringsampleuptake)
Thepipettetipisdefective
Solution2.)Checkthepipettetipforqualityissues,replacethepipettetip
Plungerbuttonisdifficulttodepress
(ordoesnotmovesmoothly):
Thepistonmaybecontaminatedoroutofshape.
Solution1.)Cleanthepistonandapplysiliconelubricant
Solution2.)Replacethepistonorlowerportionofthepipette
Thetipisleakingordrippingliquid:
Thepipettetipisdefectiveornotcompatible
Solution
1.)Replacethepipettetipwithanewone
Solution2.)Replacethepipettetipwithanewonefromadifferentbrand/style
Thetipisnotproperlysealedonthepipetteshaft
Solution1.)Pressthetipfirmlyonthepipetteshaft
Theinternalseal,Oring,pistonorpipetteshaftiscontaminatedofdamaged
Solution1.)Disassembleandcleanthepistonandsealwithdistilledwater
Solution2.)Checkthepartsfordamage/wear.Replacepartasrequired
Thedispensedvolumeismoreorlessthanthesetvolume:
Plunger
buttonhasbeenpressedtoofar(pasttheblowoutphase)beforesampleuptake
Solution1.)Operatecarefully,usingproperpipettingtechnique
Pipetteisoutofcalibration
Solution1.)Recalibratethepipettefollowingtheinstructionsinthemanual
10

11
adivisionof
/