  • Hello! I am an AI chatbot trained to assist you with the Extech Instruments Extech HD755 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
UserManual
HeavyDuty
DifferentialPressureManometer
ModelHD755
HD755-en-GB_V1.8 9/16
2
Introduction
CongratulationsonyourpurchaseoftheExtechManometerModelHD755.Thisdevicemeasures
gaugeanddifferentialpressureintherangeof±0.500psi.TheHD755offerseleven(11)unitsof
measure(userselectablebyfrontpanelpushbutton)andhasadifferentialinputthatuses
convenientquickdisconnectfittings.AdvancedfeaturesincludeDATAHOLD,MINMAXAVG
RECORDmode,ZERO/OFFSET,AUTOPOWEROFF,andUSBPCinterface.Thismeterisshippedfully
testedandcalibratedand,withproperuse,willprovideyearsofreliableservice.Pleasevisitour
website(www.extech.com)tocheckforthelatestversionofthisUserGuide,ProductUpdates,
and
CustomerSupport.
MeterDescription
1. P1input
2. ACadaptorjack
3. P2Input
4. LCDDisplayIndicatesthemeasurementdata,unitofmeasure,
elapsedtimer,andfunctionsymbolsasdescribedinthismanual
5. USBPCinterfacejack(3.5mm)Forusewithdataacquisition
software
6. UNITbuttonPresstoselect
theunitofmeasure
7. DIFbutton–Presstodisplaydifferentialpressure.Alsousedto
OFFSETthedisplayedreadings.
8. RECORDbutton–PresstoaccesstheMIN/MAX/AVGrecording
mode
9. HOLD/ZERObutton–Presstofreezethedisplayedreading.Also
usedtozerothedisplay(pressandholduntil
displayshowsallzeroes).
10. BACKLIGHT
buttonPresstoswitchthe displaybacklightON.Thebacklightwill
automaticallyturnOFFafter40seconds.
11. POWERbuttonPresstoturnthemeterONorOFF
12. Batterycompartment(onrear)
HD755-en-GB_V1.8 9/16
3
Display
1. USBdataoutputactive
2. DifferentialPressureMode
3. DataHoldMode
4. Primarydisplay
5. RecordMode
6. MAX/MIN/AVGindicators
7. LowBatteryIndicator
8. Pressureunitofmeasure
9. ElapsedTimer
HD755-en-GB_V1.8 9/16
4
BasicOperation
MeterPower
TheHD755usesa9Vbatteryforpower.IfthemeterdoesnotswitchonwhenthePOWER
buttonispressedorifthedisplayshowsBAT,replacethebattery.
WhenthePOWERbuttonsispressed,themeterperformsashortselftestandrevertstothe
settingsinplace
whenthemeterwaslastturnedoff.
AutomaticPoweroffFeature
Inordertoconservebatterylife,themeterhasanautomaticpowerofffeature(after20
minutes).Todefeatthisfeature,pressandholdtheHOLDbuttonwhileturningthemeteron.
ReleasetheHOLDbuttonwhen‘n’appearsonthedisplay.Themeterwillnowremainonuntil
theuserturnsitofforuntilthebatteryweakens.NotethattheAutoPowerOFFfeaturewillbe
reactivatedthenexttimethemeteristurnedoffandthenon.
UnitofMeasure
PresstheUNITbuttontoselecttheunitofmeasure.Thismeterofferseleven(11)unitsof
measureaslistedbelow.
UnitofMeasure Displayicon
Poundspersquareinch psi
InchesofMercury inHg
Millibar mbar
MillimetersofMercury mmHg
KiloPascals kPa
Kilogramspercentimetersquared Kgcm
2
Inchesofwater inH
2
O
Feetofwater ftH
2
O
Centimetersofwater cmH
2
O
Bars Bar
Ouncespersquareinch ozin
2
FittingsConnection
ConnecttubingtotheP1,P2(orboth)inputport(s).Ifbothinputsareused(differentialmode),
themeterdisplaysapositivepressurereadingiftheP1pressureisgreaterthanP2anda
negativereadingifP2isgreaterthanP1.
HD755-en-GB_V1.8 9/16
5
ZeroAdjustandOffsetfunction
TheunitmustbekeptintheuprightpositionwhenZEROINGorwhentakingmeasurements.To
ZEROthemeter:Beforeuseandwithoutfittingsattachedtothemeter,pressandholdtheHOLD
buttonfor3secondsoruntilthedisplayshowsallzeroes.
ToOFFSETmeterreadings:Pressthe
DIFbuttonwhiletakingameasurement.Subsequent
readingswillrepresentthedifferencebetweenthecurrentmeasurementandthemeasurement
thatwasonthedisplaywhentheDIFbuttonwaspressed..PresstheDIFbuttonagaintoexitthe
OFFSETfunction.
MINMAXAVGModewithElapsedTimer
TheRECORDfunctionrecordsasinglemaximum(MAX),minimum(MIN),andaverage(AVG)
readingswithatimestamp.Theelapsedtimerisdisplayedonthelowerrighthandcornerofthe
LCDandinformstheuserofthetimeaMAXorMINreadingwasrecorded.
1. PresstheRECORDbutton once.TheRECindicatorwillappearonthedisplayandthe
displayedelapsedtimerwillstart(HH:MM:SSformat[hours:minutes:seconds]).
2. PresstheRECORDbutton againtoviewthehighest(MAX)reading
recordedsincethe
RECORDbuttonwasfirstpressed.TheMAXindicatorandmaximumreadingwillappearon
thedisplay.TheelapsedtimerwillshowthetimeoftheMAXevent.
3. PresstheRECORDbutton again,theMINindicatorandtheminimumvaluewillappearon
thedisplay.Theelapsed
timerwillshowthetimeoftheMINevent.
4. PresstheRECORDbutton again,theAVGindicatorandtheaveragevaluewillappearonthe
display.Theelapsedtimerwillcontinuetorun.
5. Toreturntonormaloperation,pressandholdtheRECORDbuttonuntilthedisplay
indicatorsREC,MAX,etc.switchoff.
NotethatintheRECORDmode,thebacklightandthepowerbuttonsaretheonlyavailable
functions.TheMin,Max,andAVGdatacannotbeexportedtothePC.
PCInterface
ThemeterhasabuiltinUSBportforusewiththeManometerDataAcquisitionsoftware.The
softwareallowstheusertoview,recordandprintreadingsfromtheHD755whileitisconnected
directlytoaPC.Refertothedocumentationincludedwiththesoftwareforcompletedetails.
DisplayedErrorCodes
Thefollowingcodesaredisplayedwhenthemeterencountersanerror.
ERR1:Pressureinputexceedsspecifiedlimits
ERR2:Pressureinputistoolowtoread
ERR3:TheDifferentialPressureinputexceedsspecifiedlimits
ERR4:TheDifferentialPressureinputistoolowtoread
HD755-en-GB_V1.8 9/16
6
Specifications
GeneralSpecifications
Display FourdigitLCDwithfunctionindicators
UnitsofMeasure BAR,mBAR,psi,kg/cm
2
,mmHg,inchesHg,kPa,feetH
2
O,inches
H
2
O,cmH
2
O,andozin
2
Accuracy ±0.3%FS+1d(@25°C)
Repeatability ±0.2%to0.5%FS
Linearity/Hysteresis ±0.3%FS
CombinedAccuracy ±1.1%FS
MaximumPressure 2psi
Responsetime 0.5stypical
Inputcircuit Differentialinputs(P1andP2)
InputPorts Twometal4.8mm,barbedfor1/8"(3.17mm)IDtubing
DataHold Freezesdisplayedreading
Sensor Builtinpiezoelectricsensors
Zero/Offsetfunction Pushbuttonactivation
Samplerate 0.5sec.(approx.)
Compatibility Airornoncorrosiveandnonionizedgas (notintended forusewith
liquids)
AutoPowerOFF Meterturnsoffafterapprox.20minutestoconserveenergy(can
bedeactivated)
PCInterface BuiltinUSBserialinterface
Storage
conditions 0to60
o
C(14to140
o
F);<80%RH
Operatingconditions 0to50
o
C(32to122
o
F);<80%RH
Batterypower 9Valkalinebattery(BATlowbatteryindicator)
Batterylife 125hourstypical
Dimensions 210x75x50mm(8.2x2.9x1.9”)
Weight 280g(9.8oz)
HD755-en-GB_V1.8 9/16
7
RangeSpecifications
Range/Units(Max.) Resolution
±0.034bar 0.001
±34.47mbar 0.01
±0.500psi 0.001
±0.035Kg/cm
2
0.001
±25.86mmHg 0.01
±1.018inchesHg 0.001
±3.447kPa 0.001
±1.154feetH
2
O 0.001
±13.85inchesH
2
O 0.01
±35.15cmH
2
O 0.01
±8.00ozin
2
0.01
BatteryReplacement
WhenthelowbatteryindicatorBATappearsonthedisplayorifthemeterdoesnotswitchon
whenthePOWERbuttonispressed,itisnecessarytoreplacethebattery.
Toreplacethebattery:
1. Openthebatterycompartmentbyslidingthemeter’srearpanelinthedirectionofthe
embossedarrow.
2. Replacethe9Valkalinebattery
3. Replacethebatterycompartmentcover.
Neverdisposeofusedbatteriesorrechargeablebatteriesinhouseholdwaste.
Asconsumers,usersarelegallyrequiredtotakeusedbatteriestoappropriatecollection
sites,theretailstorewherethebatterieswerepurchased,orwhereverbatteries
aresold.
Disposal:Donotdisposeofthisinstrumentinhouseholdwaste.Theuserisobligatedto
takeendoflifedevicestoadesignatedcollectionpointforthedisposalofelectricaland
electronicequipment.
OtherBatterySafetyReminders
Neverdisposeofbatteriesinafire.Batteriesmayexplodeorleak.
Nevermixbatterytypes.Alwaysinstallnewbatteriesofthesametype.
Warning: Pressure inputs of 10 psi or higher will damage the instrument. Do not
apply pressure higher than the maximum pressure listed in the specification table
HD755-en-GB_V1.8 9/16
8
Conversions
UsefulconversionFactors
From To Multiplier
inofH
2
O inofHg 0.07355
inofH
2
O cmofH
2
O 2.54
mmofHg inofH
2
O 0.03937