Andatech WM-PRO User manual

Type
User manual
BREATHALYSER
Congratulations on buying your new Wingmate Pro breathalyser.
We hope you enjoy owning your Wingmate Pro breathalyser. If you
have any questions regarding the unit or want to speak with us
further, please visit andatech.com.au or call 1300 800 200
(between 9am and 5pm Monday to Friday).
Specifications
Wingmate Pro Components
Sensor
Indication of B.A.C
Accuracy
Operation Temperature
Storage Temperature
Response Time
Display
Power Supply
Use Mouthpiece
Dimensions (mm)
Weight
Calibration
Fuel Cell Sensor
0.000 to 0.400 %BAC
± 0.005 %BAC at 0.050 %BAC
5˚C to 40˚C
0˚C to 40˚C
3 ~ 22 seconds
3-Digit LCD
Two 1.5V “AAA” size alkaline batteries
Compatible with Wingmate mouthpieces
114 (height) X 47 (width) X 24 (thickness)
61.5g (84g including batteries)
Recommended every 500 tests or
6 months, whichever comes first.
Australian Standards Requirement:
1 month
Contents:
Breathalyser,
mouthpieces (x5),
AAA batteries (x2),
and user’s manual
Warm Up Time
15 ~ 35 seconds
Mouthpiece Slot
Digital LCD Display
Power Button
Battery Cover (Back)
Breath out
(do not cover)
Operating Instructions
1. Insert a mouthpiece into the mouthpiece slot
2. Press the power button. “ON” will display and the breathalyser
will start warming up, which will take about 15 ~ 35 seconds.
3. The breathalyser will beep and display “BLO” on the screen.
Blow steadily into the mouthpiece and continue blowing until
you hear another beep.
(If you do not blow into the breathalyser within 20 seconds, it
will automatically switch off. Press the power button to
switch it back on.)
4. The breathalyser will display a circling line as it analyses the
breath sample. It will then beep and display the result for 20
seconds.
ǩ,IWKHWHVWUHVXOWLVRYHU%$&VHWE\GHIDXOWWKH/&'
backlight will change to red and display “WARN” and sound
an audible alert.
ǩ7HVWUHVXOWVDUHGLVSOD\HGIRUVHFRQGV/&'EDFNOLJKWZLOO
dim down in the last 5 seconds).
ǩ3UHVVWKHSRZHUEXWWRQGXULQJWKHODVWVHFRQGVWRWDNH
another test. The breathalyser will start warming up for a
second test.
5. To take another test, repeat from step 2.
ǩ7U\WRVSDFHFRQVHFXWLYHWHVWVDWOHDVWPLQXWHVDSDUW
6. Switch the breathalyser off by pressing the power button for 2
seconds.
Please note:
ǩ7KHEUHDWKDO\VHUZLOOWXUQRIIDXWRPDWLFDOO\ZKHQQRWLQXVH
Display
Device is ready to take a
test. Start blowing.
Battery low. Replace
with new batteries.
Displays test result Dual colour LCD backlight.
See Warnings Display section
in the manual.
Test result is displayed
as a 3 decimal place
%BAC reading
Result is
above
0.050% BAC
Warnings Display
How to Change Batteries
If the result is higher than 0.050% BAC, the
breathalyser will display “WARN” and sound
an audible alert.
If the result is above 0.400% BAC, “Hot” signal
is displayed along with alarm sound.
If you have tested over 500 times, CAL will
show up on the display screen for 3 seconds
before warming up for a test. Please send the
breathalyser back for calibration.
The battery voltage is too low to perform a
test. The device will turn off automatically
after showing this signal. Please replace with
new 1.5V alkaline batteries.
The breath sample was not strong enough for
a measurement. Please power on the unit
again and make sure the breath is strong and
steady during sampling.
When the ambient temperature is out of the
working temperature range (5°C~ 40°C) the
unit shows “OUT” and then turns off. Keep and
use the unit within the recommended
temperature range.
The breathalyser’s LCD Backlight will turn red when any of these
errors occur.
Apply a small amount of pressure on the battery cover
located on the back of the unit.
Push down and away from the unit to release the cover.
Replace with new alkaline batteries in the proper
positive(+) and negative(-) positions as indicated. Then,
replace the battery cover.
Step 1:
Step 2:
Step 3:
The Wingmate Pro uses a fuel cell sensor to measure breath alcohol
levels. Depending upon the amount of alcohol registered by the
sensor after a period of time, it will require recalibration. Andatech
recommends the Wingmate Pro is calibrated every 6 months or after
500
Standards, calibration every one (1) month is required.
uses, whichever comes first. To maintain the unit to Australian
unit
Pro
Version 1.0. Updated December 2015.
Wingmate Pro
Product Registration
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Andatech WM-PRO User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI