Miller MH000000 Owner's manual

Type
Owner's manual

This manual is also suitable for

Miller MH000000 is a versatile, professional-grade welding machine designed to deliver exceptional performance for a wide range of welding applications. With its advanced features and user-friendly design, this machine empowers you to tackle even the most demanding welding tasks with precision and efficiency.

Miller MH000000 is a versatile, professional-grade welding machine designed to deliver exceptional performance for a wide range of welding applications. With its advanced features and user-friendly design, this machine empowers you to tackle even the most demanding welding tasks with precision and efficiency.

15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
9900 Software Calibration Guide
July, 2013
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
A. 9900 Software Calibration
Before you make any changes it is recommended that you make a copy of the configuration file.
To do this go into the Channels Configuration Editor and press the “Save As” button on the
bottom of the screen. Change the name by adding today’s date so you can quickly tell when it
was copied. After making a copy open the old file so that all the changes will be made in the
original file. If you need to bring back the old file, just open the saved file and use “Save As”
and change its name back to the original file.
For all channels except ARC, the Closed Loop Sensitivity must be changed to 0 during the
calibration procedure. Write down what the number is, so it can be restored when you are
finished.
CURRENT Channel Configuration Editor Screen
Press Blue Button to Show Auto Calibration Boxes
Blue Button
Set to Zero for Open Loop Mode
When Calibrating Weld Current
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Channel Configuration Editor, Calibration Term Definitions
RANGE
Units: Units of Measure for the channel. When doing calibration checks, make measurements
using the same units as selected here. Note that the Travel channel’s units of measure change
from Rotational (RPM) to Surface Speed (Linear IPM, CPM) in the weld program when there is
a non-zero value entered in the Diameter field.
Minimum Value: This is the minimum controlled value for the channel. No value for this
channel can be entered in below this value in the weld program. This value should not be
changed during a touch up calibration.
Maximum Value: This is the maximum controlled value for the channel. No value for this
channel can be entered in above this value in the weld program. This value should not be
changed during a touch up calibration.
EQUIVALENT DC REFERENCE VOLTS
Out Minimum Value: This is the analog voltage that is put out by the 9900-10 board when the
channel is operated at the minimum value. Changing this number changes the Actual (measured)
value of the controlled device when near the minimum setting.
Out Maximum Value: This is the analog voltage that is put out by the 9900-10 board when the
channel is operated at the maximum value. Changing this number changes the Actual (measured)
value of the controlled device when near the maximum setting.
In Minimum Value: This is the input voltage that comes into the module from the feedback
transducer for the channel, when operating at the minimum setting. Changing this number
changes the read back display on the 9900 screen. It does not change the Actual (measured)
value, only the Displayed value when in the open loop mode. Making this number higher makes
the display read lower and vice versa.
In Maximum Value: This is the input voltage that comes into the module from the feedback
transducer for the channel, when operating at the maximum setting. Changing this number
changes the read back display on the 9900 screen. It does not change the Actual (measured)
value, only the Displayed value when in the open loop mode. Making this number higher makes
the display read lower and vice versa.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Touch Up Calibration Procedure
Write a test program that will run each channel, but when you want to calibrate a channel, use
Dry Run and turn off every channel but the one that you are testing. For your test program,
make sure that the ALC is locked out in every segment. Also make sure to turn off Wait for Arc
and Touch Retract. Do not have any pulsing in your program for calibration. It is best to do the
calibration program in the Standard Mode.
To do a complete calibration you will need to run each channel near the high end of its range and
then near the low end of the range. You will need an independent test transducer for each of the
channels to measure parameter values and compare them to the 9900 program values:
For the Weld Current channel(s) you should have a calibrated current shunt or clamp-on
ammeter.
For Arc Voltage channel(s) you should have a 0 to 30 volt DC test power supply and a calibrated
DVM.
For Travel channel(s) you should have a calibrated tachometer with surface speed wheel capable
of measuring distance traveled.
Alternatively, depending on accuracy and traceability required, you could use the ammeter on
the welding power source for the weld current channel, a voltmeter and three 9 volt batteries for
the voltage channel, and do time/distance measurements for the travel and wire channels.
Touch Up Calibration Using Auto Calibration Routine (Analog Channels)
Weld Current Channel
1) Set up the system to weld. Make sure gas and water are on and set up some thick material or a
copper block to weld on during the calibration check. Use a tungsten size that will hold up to the
maximum current that will be used, for a few seconds. Set your test weld program to weld near
the high end of the weld current range (300 amps in the example screen). Call up the weld
program to run then de-select the unneeded channels from the Dry Run screen.
2) Connect the weld current test transducer or prepare to take readings from the welding power
supply if using that as the reference.
3) Run the program in the main weld segment for at least 8 seconds.
4) After running the test program, go into the Channel Configuration Editor and select the Current
channel that you are working with. First you will need to press the blue button. After you press
the Blue button, the calibration boxes will open up. Also, a green and a red button will appear.
The green button (marked “C”) is the “calculate” button, the red button is marked “U” and is the
“undo” button.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Weld Current Configuration Editor Showing Auto Calibration Boxes
The Program, Read, Diameter and Time boxes will be filled in with what was just run. Next,
enter in the Actual value that you measured. After you enter in the Actual value, press the green
button (calculate) once. When you press the green button you will see the In and Out numbers
change. You will also see that the green button turns grey, so you can not press it again. If you
have made a mistake, or would like to go back, press the Red button (undo). After the changes
you will need to press the Save button to keep what you have just done. You can now run the
calibration in the other half of the range.
5) Set the weld current in the test program to run near the low limit (15 amps in the example
screen). When ready, run the test program as in Step 3 above.
6) After running the test program at the lower current, go to the Current Channel in the Channel
Configuration Editor. Press the Blue button, to open the calibration boxes. Press the green
button to calculate the new calibration values. Save the values.
Green Button --
CALCULATE
Red Button --
UNDO
Auto Calibration Boxes
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
7) Go back and run the weld current at a couple of different points through the range to verify
accuracy and linearity to the desired tolerance. You can repeat the auto calibration sequence for
increased accuracy if needed.
8) When satisfied with calibration results, set the Closed Loop Sensitivity number back to the
original value. Record the final calibration values for future reference.
Arc Channel (With Arc Length Control)
1) There is no welding when calibrating the Arc Channel. Only the 9900 Displayed value is adjusted
in this step. You will need a calibrated DC test power supply capable of 0 to 30 VDC at 0.2A or
more. Alternatively you could use three 9 volt batteries and a calibrated DVM as a substitute
DC voltage source. Set up your test weld program to run near the high end of the arc voltage
range (30 volts is good). Call up the weld program to run then de-select the unneeded channels
from the Dry Run screen.
2) Turn off the welding power source and then connect the test power supply to the torch and the
work. Connect the negative output of the test power supply to the torch electrode (connecting
to the tungsten is ok) and connect the positive output to the Work ground. Turn on the test
power supply and set to 30.0 volts. If there is excessive current flowing from the test power
supply (20ma or more), disconnect the electrode weld cable (torch cable) from the welding
power supply.
If using the 9 volt batteries instead of the test power supply, connect all three in series then
connect the negative side to the torch electrode and positive to Work ground. Immediately
connect a DVM across the batteries. If the voltage is continually dropping, disconnect the
electrode weld cable from the welding power supply. Leave the voltmeter connected for
monitoring the voltage during the test. The voltage should be near 27 volts and should not drop
significantly during the test.
3) Run the weld program and let it run in the main weld segment for at least 8 seconds. Note the
voltage reading on the test power supply or on the batteries then press the Stop button.
4) Go into the Channel Configuration Editor and select the Arc channel that you are working with.
First you will need to press the blue button. After you press the Blue button, the calibration
boxes will open up. Also, a green and a red button will appear. The green button (marked “C”)
is the “calculate” button, the red button is marked “U” and is the “undo” button.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Arc Channel Configuration Editor Screen
Press Blue Button to Show Calibration Boxes
The Program, Read, Diameter and Time boxes will be filled in automatically with what was just
run. Next, enter in the Actual value that was displayed on the test power supply or the value you
measured across the batteries during the run. After you enter in the Actual value, press the green
button (calculate) once. When you press the green button you will see the In numbers change.
You will also see that the green button turns grey, so you can not press it again. If you have
made a mistake, or would like to go back, press the Red button (undo). After the changes you
will need to press the Save button to keep what you have just done. You can now run the
calibration in the other half of the range.
Do Not Change Closed Looping
Parameters When Calibrating Voltage
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Arc Channel Configuration Editor Showing Auto Calibration Boxes
5) Adjust the output of the test power supply to 6 volts. If using batteries, move the output leads
to use just one battery (9 volts).
6) Set the Arc voltage in the test program to run near the low limit (6 volts in the example screen).
When ready, run the test program as in Step 3 above.
7) After running the test program at the lower voltage, go to the Arc Channel in the Channel
Configuration Editor. Press the Blue button, to open the calibration boxes. Press the green
button to calculate the new calibration values. Save the values.
8) Go back and run the test weld program and vary the voltage on the test power supply to check
the voltage read back values at a few points in the range of 6 to 30 volts. If using batteries, try
one battery, then two, then three to verify accuracy. You can repeat the auto calibration
sequence for increased accuracy if needed.
9) When satisfied with calibration results, record the final calibration values for future reference.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Rotation Travel Channel
1) There is no welding when calibrating the Rotation Travel Channel. You will need a calibrated test
tachometer with surface speed wheel and distance readout. Alternatively you can do a
time/distance measurement using tape measure and stopwatch (or timed weld program). Set
up your test weld program to run near the high end of the rotation travel range (in the upper
10% of the speed range). In the weld program, enter the diameter of the faceplate or tooling
that you intend to use to measure surface speed. It is recommended to not use any Initial,
Upslope, Downslope or Final segments in the calibration test weld program (enter “0” for
times). Call up the test weld program to run then de-select the unneeded channels from the
Dry Run screen.
2) Connect the travel tachometer or prepare to take distance measurements on the tooling
diameter. Be sure this diameter is the same as what is entered in the test weld program.
3) Run the program in the main weld segment for at least 30 seconds.
4) After running the test program, go into the Channel Configuration Editor and select the Rotation
Travel channel that you are working with. First you will need to press the blue button. After
you press the Blue button, the calibration boxes will open up. Also, a green and a red button
will appear. The green button (marked “C”) is the “calculate” button, the red button is marked
“U” and is the “undo” button.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Analog Travel (Rotation) Configuration Editor Showing Auto Calibration Boxes.
There are Two Analog Travel Channels in this Example (Subindex 1 and 2).
Use Next and Prev to Toggle Between Them (Rotation and Boom).
5) The Program, Read, Diameter and Time boxes will be automatically filled in with what was just
run. Next, enter in the Actual value that you measured. Do not multiply or divide your
measurement to compensate for times less than or greater than 1 minute, this is automatically
compensated for in the application program. Enter only the Actual value, do not change the
other fields at this time. After you enter in the Actual value, press the green button (calculate)
once. When you press the green button you will see the In and Out numbers change. You will
also see that the green button turns grey, so you can not press it again. If you have made a
mistake, or would like to go back, press the Red button (undo). After the changes you will need
to press the Save button to keep what you have just done. You can now run the calibration in
the other half of the range.
6) Set the rotation travel speed in the test program to run near the low limit (lower 10% of the
speed range). When ready, run the test program as in Step 3 above.
Auto Calibration Boxes
Green Button --
CALCULATE
Red Button --
UNDO
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
7) After running the test program at the lower travel speed, go to the Rotation Travel Channel in
the Channel Configuration Editor. Press the Blue button, to open the auto calibration boxes.
Enter the new Actual (measured) value then press the green button to calculate the new
calibration values. Save the values.
8) Go back and run the rotation at a couple of different points through the range to verify accuracy
and linearity to the desired tolerance. You can repeat the auto calibration sequence for
increased accuracy if needed.
9) When satisfied with calibration results, set the Closed Loop Sensitivity number back to the
original value. Record the final calibration values for future reference.
Linear Travel Channel
1) There is no welding when calibrating the Linear Travel Channel. You will need a calibrated test
tachometer with surface speed wheel and distance readout. Alternatively you can do a
time/distance measurement using tape measure and stopwatch (or timed weld program). Set
up your test weld program to run near the high end of the rotation travel range (in the upper
10% of the speed range. It is recommended to not use any Initial, Upslope, Downslope or Final
segments in the calibration test weld program (enter “0” for times). Call up the test weld
program to run then de-select the unneeded channels from the Dry Run screen.
2) Connect the travel tachometer or prepare to take distance measurements on the carriage or
boom.
3) Run the program in the main weld segment for at least 30 seconds.
4) After running the test program, go into the Channel Configuration Editor and select the Linear
Travel channel that you are working with. First you will need to press the blue button. After
you press the Blue button, the auto calibration boxes will open up. Also, a green and a red
button will appear. The green button (marked “C”) is the “calculate” button, the red button is
marked “U” and is the “undo” button.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Boom Travel (Linear) Configuration Editor Showing Auto Calibration Boxes.
There are Two Analog Travel Channels in this Example (Subindex 1 and 2).
Use Next and Prev to Toggle Between Them (Rotation and Boom).
5) The Program, Read, Diameter and Time boxes will be automatically filled in with what was just
run. Next, enter in the Actual value that you measured. Do not multiply or divide your
measurement to compensate for times less than or greater than 1 minute, this is automatically
compensated for in the application program. Enter only the Actual value, do not change the
other fields at this time. After you enter in the Actual value, press the green button (calculate)
once. When you press the green button you will see the In and Out numbers change. You will
also see that the green button turns grey, so you can not press it again. If you have made a
mistake, or would like to go back, press the Red button (undo). After the changes you will need
to press the Save button to keep what you have just done. You can now run the calibration in
the other half of the range.
6) Set the Linear travel speed in the test program to run near the low limit (lower 10% of the speed
range). When ready, run the test program as in Step 3 above.
Green Button --
CALCULATE
Red Button --
UNDO
Auto Calibration Boxes
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
7) After running the test program at the lower travel speed, go to the Linear Travel Channel in the
Channel Configuration Editor. Press the Blue button, to open the auto calibration boxes. Enter
the new Actual (measured) value then press the green button to calculate the new calibration
values. Save the values.
8) Go back and run the linear travel at a couple of different points through the range to verify
accuracy and linearity to the desired tolerance. You can repeat the auto calibration sequence
for increased accuracy if needed.
9) When satisfied with calibration results, set the Closed Loop Sensitivity number back to the
original value. Record the final calibration values for future reference.
Wire Feed Channel
1) There is no welding when calibrating the Wire Feed Channel. You will need a calibrated test
tachometer with surface speed wheel and distance readout. Alternatively you can do a
time/distance measurement using tape measure and stopwatch (or timed weld program). Set
up your test weld program to run near the high end of the wire feeder range (in the upper 10%
of the speed range. It is recommended to not use any Initial, Upslope, Downslope or Final
segments in the calibration test weld program (enter “0” for times). Call up the test weld
program to run then de-select the unneeded channels from the Dry Run screen.
2) Connect the test tachometer wheel so that it can drive from the surface of the feed rolls or from
the driven wire directly. Alternatively you can make measurements of the wire fed out during
the run with a tape measure.
3) Run the program in the main weld segment for at least 15 seconds.
4) After running the test program, go into the Channel Configuration Editor and select the Wire
Feed channel that you are working with. First you will need to press the blue button. After you
press the Blue button, the auto calibration boxes will open up. Also, a green and a red button
will appear. The green button (marked “C”) is the “calculate” button, the red button is marked
“U” and is the “undo” button.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Wire Feed Configuration Editor Showing Auto Calibration Boxes.
5) The Program, Read, Diameter and Time boxes will be automatically filled in with what was just
run. Next, enter in the Actual value that you measured. Do not multiply or divide your
measurement to compensate for times less than or greater than 1 minute, this is automatically
compensated for in the application program. Enter only the Actual value, do not change the
other fields at this time. After you enter in the Actual value, press the green button (calculate)
once. When you press the green button you will see the In and Out numbers change. You will
also see that the green button turns grey, so you can not press it again. If you have made a
mistake, or would like to go back, press the Red button (undo). After the changes you will need
to press the Save button to keep what you have just done. You can now run the calibration in
the other half of the range.
6) Set the Wire Feed speed in the test program to run near the low limit (lower 10% of the speed
range). When ready, run the test program as in Step 3 above.
7) After running the test program at the lower wire feed speed, go to the Wire Feed Channel in the
Channel Configuration Editor. Press the Blue button, to open the auto calibration boxes. Enter
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
the new Actual (measured) value then press the green button to calculate the new calibration
values. Save the values.
8) Go back and run the wire feeder at a couple of different points through the range to verify
accuracy and linearity to the desired tolerance. You can repeat the auto calibration sequence
for increased accuracy if needed.
9) When satisfied with calibration results, set the Closed Loop Sensitivity number back to the
original value. Record the final calibration values for future reference.
Digital Channels
The calibration of digital channels is done primarily by checking the speed/distance of the
particular channel, there are no analog voltage adjustments or Auto Calibration routines except
for Input Only on certain channel types (similar to the Arc Voltage channel using the 9900-AV-
1). The min and max speeds are determined by entries in the calibration screen along with the
steps per revolution of the motor and gear ratio of the gearbox used on the fixture. Once these
values are established, changes should not be necessary, calibration is only checked or verified.
If there is a disparity in output numbers that were correct originally, it could be an indication of a
problem.
Mechanical Oscillator Channel
The Mechanical Oscillator Channel controls oscillation width as it’s primary parameter. The
Width can be checked in the calibration test program by setting up a width longer than what is
normally used, eg. 2 inches. Set the dwell times so there is enough time to mark the excursions
of the torch during each cycle of the oscillator (or take a reading on a dial indicator attached to
the slide). Set the oscillation speed at about 50 IPM. See the Oscillator Configuration Screen on
the next page with descriptions of the relevant fields:
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Mechanical Oscillator Configuration Editor Showing Set Up Parameters
Range Values
Units: Inch or Metric units for Oscillator Width
Min Value: Minimum width allowed to be programmed (in Units)
Max Value: Maximum width allowed to be programmed (in Units)
Equiv DC Reference Volts
Min Value: Not Used
Max Value: Not Used
Misc Info:
Stepper Counts/Rev: This is how many counts (cycles) to the stepper drive to rotate the motor
shaft one revolution
Drive Ratio: This is how many revolutions of the motor shaft will cause the slide to
travel one unit (Inch, in this case).
Acceleration/Deceleration: This is how many motor steps per second/second are allowed to
get to final (programmed) speed.
Maximum Speed: This is maximum slide speed in Units per Minute (IPM, in this case).
Closed Looping Parameters: Closed Looping parameters are Not Used.
15 Goodyear Street, Irvine, California 92618 U.S.A.
Tel: (949) 951-1515 Fax: (949) 951-9237
Calibration Tips
Calibration of Current Channel
Make sure that you have a part that can handle the current without burning through. Make sure
that you run the current for about 8 seconds after the program has gotten to the weld segment.
You can use the meter on the power supply to determine the actual reading.
Calibration of Arc Channel
You will need to supply a voltage between the work and the tungsten. In the program enter the
value of the voltage that is between the work and the tungsten. Run the program for at least 8
seconds in the weld program. Since you know the actual value, enter that number in the box and
then press the green button.
Calibration of Travel and Wire
Always do the calibration of the travel in the forward direction. For these channels you will have
to run the weld program to completion. For the faster speeds you only need the weld time to be
from 15 30 seconds. For the slower speeds you can go for a couple of minutes. If the delay
start for the travel and wire match the amount of time that the current is in initial and upslope,
and you have no stop delay, that is the time that each channel will be active. This will make it
easier when measuring to know if the channel parameter has moved the correct distance.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17

Miller MH000000 Owner's manual

Type
Owner's manual
This manual is also suitable for

Miller MH000000 is a versatile, professional-grade welding machine designed to deliver exceptional performance for a wide range of welding applications. With its advanced features and user-friendly design, this machine empowers you to tackle even the most demanding welding tasks with precision and efficiency.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI