Beurer EM 38 Owner's manual

Type
Owner's manual
ENGLISH
TENS back belt
Instruction manual ................................................2
ESPAÑOL
Cinturón terapéutico lumbar TENS
Manual de instrucciones ....................................13
FRANÇAIS
Ceinture dorsale avec appareil TENS
Mode d’emploi ..................................................... 25
EM 38
READ THIS MANUAL COMPLETELY AND
CAREFULLY BEFORE USING THIS PRODUCT
Keep this manual in a safe location for future reference
LEA TODO ESTE MANUAL CON ATENCIÓN
ANTES DE USAR ESTE PRODUCTO
Conserve este manual en un lugar seguro para consultarlo en el futuro
LISEZ CE MODE D’EMPLOI COMPLÈTEMENT ET
ATTENTIVEMENT AVANT D’UTILISER CE PRODUIT
Conservez ce manuel en lieu sûr pour le consulter ultérieurement
Distributed by/ Distribuido por/
Distribué par :
Beurer North America LP
900 N Federal Highway, Suite 300
Hallandale Beach, FL 33009
www.beurer.com
¿Preguntas o comentarios? Llame gratis al
1-800-536-0366 o póngase en contacto info@beurer.com
Questions ou commentaires? Appeler gratuitement
1-800-536-0366 ou communiquez avec info@beurer.com
Questions or comments?
Call toll free 1-800-536-0366 or contact info@beurer.com
2
ENGLISH
1. Important safety notes
Signs and symbols
Whenever used, the following signs identify safety and property damage messages and designate a level of
hazard or seriousness.
READ AND UNDERSTAND THIS ENTIRE MANUAL, INCLUDING THE SAFETY SECTION AND ALL IN
-
STRUCTIONS AND WARNINGS, COMPLETELY AND CAREFULLY BEFORE USING THIS PRODUCT. FOL-
LOW ALL SAFETY INSTRUCTIONS AND WARNINGS TO AVOID HAZARDOUS SITUATIONS AND TO MAKE
CORRECT USE OF THIS PRODUCT.
WARNING WARNING indicates a hazardous situation which, if not avoided, could result in death or
serious injury.
CAUTION CAUTION indicates a hazardous situation which, if not avoided, may result in minor or
moderate injury.
NOTICE NOTICE addresses practices not related to personal injury, such as product and/or
property damage.
WARNING:
•
Use this device only on humans and according to its intended purpose as specified in this instruction manual.
Any improper use is dangerous and the warranty will be invalidated if the device is used for other than the
intended purpose.
•
The unit is intended for use in the lower back only.
•
The device should not be used by patients with:
– pacemakers, defibrillators or cardiac irregularities,
– metal implants, apnea monitors or other electronic auxiliary devices,
– diabetes or abnormal blood pressure,
– a tendency toward internal bleeding,
– pregnant women.
•
Therapy with the Beurer Back belt with TENS device is a proven and natural way of fighting pain, however,
it does not replace medical diagnosis or treatment. If you are not sure, consult your doctor before using this
device.
•
Consult your doctor always before use, if
– you suer from epilepsy or certain disabilities,
you suer from chronic pain conditions that have not been clarified, regardless of the region in the body.
•
Persons whose sensitivity is disturbed or impaired should only apply this device under supervision of your
physician.
•
Pain relief treatment following injury or an operation should only be carried out under the careful supervision
and instruction of a doctor or physical therapist.
•
Consult a doctor before use in the following situations:
– Simultaneous carried out medical treatments.
– Discomfort experienced as a result of the stimulation treatment.
– Persistent skin irritation under the electrodes.
•
In case of doubt, consult a doctor or medical professional for instructions and guidance.
Contents
1. Important safety notes ............................................2
2. Package Contents ...................................................4
3. About the unit ..........................................................4
4. Signs and symbols ..................................................4
5. Important Safeguards ..............................................5
6. Controls ...................................................................7
7. Before Use ...............................................................7
8. Operation .................................................................8
9. Cleaning and care of the unit and belt ....................9
10. Troubleshooting ...................................................10
11. Technical specifications ......................................10
12. FCC Information .................................................10
13. Warranty ..............................................................11
3
•
In acute emergencies, first aid takes priority.
•
Keep the unit out of the reach of children. Medical products are not toys.
•
Treatment should be monitored if used on or near children or persons who are handicapped and only have a
limited capability to use this device.
•
Packaging materials are a deadly hazard for children and can cause suocation. Remove all packaging materi-
als immediately and keep them away from children at all times.
•
Do not use the device while taking a bath or a shower.
•
Do not use this device while driving a car or another kind of vehicle or while operating machinery or heavy
equipment.
•
Never use the device in rooms where aerosols (sprays) are used or pure oxygen is being administered.
•
Do not use with AC Power source.
•
Use the unit only with the supplied accessory.
•
Do not use this unit when under influence of pain medication, alcohol or sleep aids.
CAUTION:
•
Do not use the device if it is not operating correctly, if it has been dropped, if it has fallen into water or if it has
been damaged.
•
If you feel pain or have an unpleasant feeling, stop treatment and consult your doctor.
•
Do not place the belt on:
– Injured or inflamed areas of skin, such as wounds, spots, rashes or reddening.
– Surgical scars in the process of healing.
– Burning or numb areas of skin.
– Swellings, peeling skin, or other sensitive points.
•
To prevent unpleasant electric shocks, never remove the device while it is still turned on.
•
Subjective perception of the electric current may alter with changing frequencies or pulse widths. Lower the
intensity as soon as the application becomes unpleasant or the pleasant prickling sensation is not felt for a
long period of time.
•
Ensure that during stimulation, no metallic objects can contact the electrodes, as otherwise local burns can
occur.
•
Do not attempt to disassemble, open and/or repair the device yourself in the event of a malfunction. You could
get injured. Moreover this will invalidate the warranty.
•
Do not switch the device on until the belt has been fixed on the body.
•
Do not use this device while sleeping.
•
Do not use with any creams, lotions or gels.
•
Do not use this device if suering from menstrual pains.
•
Do not use the control unit and belt again until they are completely dry.
NOTICE:
•
Do not operate this device near high-frequency transmitters, such as microwaves, short wave transmitters or
radio transmitters. These may impair the device’s ability to function.
•
Protect the device from moisture. If liquid should get into the device, remove the battery immediately and
discontinue use. Should this happen, contact our customer service.
•
Follow all laws and regulations regarding the proper disposal of batteries and electronic devices that are rele-
vant in your area. Consult your municipal authorities or your dealer for information about proper disposal.
•
Never use aggressive cleaning products or sti brushes to clean the device.
•
Do not allow water to penetrate the device.
•
Do not expose the device to direct sunlight and protect it from dirt and moisture.
•
Do not wash the belt in water. Do not wash the belt in the washing machine. Do not tumble dry.
•
Do not use bleach when washing the belt.
•
Do not dry clean your belt. Do not dry it over anything hot (e. g. a radiator) as it contains plastic parts.
•
The belt should never be ironed.
4
SAVE THESE INSTRUCTIONS FOR FUTURE REFERENCE
If the instruction manual is damaged or if you no longer have the instruction manual in your possession, please
contact Beurer customer service. Please see warranty for service contact.
The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use.
2. Package Contents
Back belt with control unit EM 38
Extension belt
3 x 1.5 V AAA Alkaline Batteries
This instruction manual
3. About the unit
Dear valued customer
Thank you for choosing one of our products.
Our name stands for high-quality, thoroughly tested products for applications in the areas of heat, weight, blood
pressure, body temperature, pulse, gentle therapy, massage and air.
Please read these instructions for use carefully and keep them for later use, be sure to make them accessible to
other users and observe the information they contain.
With kind regards
Your Beurer team
How does stimulation current work?
The back belt works on the basis of transcutaneous (meaning “through the skin“) electrical nerve stimulation
(TENS). TENS is a clinically proven, eective, medication-free method of treating certain types of pain which,
when used properly, is free of side eects. It is also approved as a means of simple self-treatment.
The pain-relieving eect is caused by suppressing the transmission of pain in the nerve fibers (particularly
through high-frequency impulses) and increasing the release of the body‘s own endorphins, which decrease the
sensation of pain.
Every condition that calls for the use of TENS must be clarified by your physician. He or she will also give you
information about the specific benefits of TENS self-treatment.
4. Signs and symbols
The following symbols appear in these instructions.
Note Important information to be noted.
Refer to instructions for use
Do not use this device with a pacemaker
Type BF applied part
SN Serial Number
Operating
Permissible operating temperature and humidity.
Storage
Permissible transport and storage temperature.
Permissible transport and storage humidity.
5
5. Important Safeguards
WARNING:
•
Use the belt only:
– on humans,
– for external use,
for the purpose for which it was developed and in the manner specified in these instructions for use.
•
This unit is not intended for commercial or clinical use, but only for individual, private household use.
•
Before use, ensure that there is no visible damage to the unit or accessories. When in doubt, do not use
the unit and contact your dealer or the customer service address provided.
•
If you have health concerns of any kind, contact your general practitioner.
•
Use the unit in the lower back area only. Using it on other body parts can cause severe health problems.
•
A slight reddening of the skin after use is normal and disappears after a short time. Refrain from using the
unit until the reddening has gone away.
•
If skin irritation results from a long therapy time, select a shorter application time.
•
In case of more severe skin irritation, discontinue treatment and consult a physician.
•
This unit is not intended for use by children or persons with physical, sensory (e.g. insensitivity to pain)
or mental impairments or those who do not know how to use it due to a lack of experience or knowledge,
unless they are supervised by a person responsible for their safety or have received instructions from them
on how to use the unit.
•
Keep children away from packaging materials (risk of suocation).
•
Do not use attachments that are not recommended by the manufacturer.
•
Carry out the first few minutes of treatment while sitting or lying down, so that if you experience a rare va-
sovagal reaction (i.e. you feel faint), you are not at unnecessary risk of injury. If you start to feel faint, shut
o the unit immediately and elevate your legs (for approx. 5 – 10 min.).
•
The treatment should be pleasant. If the unit does not function correctly or if you begin to feel unwell or
feel pain, stop using the unit immediately.
•
Take o the belt only when the unit is switched o.
•
Never apply the electrodes to injured skin areas.
WARNING:
To prevent harm to your health, we urgently recommend that you do not use the unit in the following
situations:
•
Do not use the unit if you use a pacemaker or other implants, such as an insulin pump.
•
If you have a high fever (e.g. > 102 °F / > 39 °C).
•
If you have known or acute arrythmias or other problems involving cardiac stimuli and their trans-
mission to the heart.
•
If you have seizure disorders (such as epilepsy).
•
If you are pregnant.
•
If you have cancer.
•
After operations in which increased muscle contractions could interfere with the healing process.
•
On acutely or chronically diseased (injured or inflamed) skin, e.g. painless or painful inflammation or redde-
ning.
•
If you have rashes (e.g. allergies), burns, bruises, swelling and open wounds or those in the process of
healing.
•
On surgical scars in the process of healing.
•
When simultaneously connected to a high-frequency surgical instrument. In this case, burns under the
stimulation current fields can result.
•
When under the influence of pain medication, alcohol or sleep aids.
•
Whenever doing anything in which an unforeseen reaction (such as increased muscle contraction despite
low intensity) can be hazardous, for example while driving a car or operating machinery.
•
On sleeping persons.
•
Do not use this unit simultaneously with other units that transmit electrical pulses to your body.
•
The unit is suitable for private, individual use.
•
For hygienic reasons, the belt may be used by one person only.
•
Ensure that during stimulation, no metallic objects can contact the electrodes, as otherwise local burns
can occur.
6
The unit must not be used in the following situations:
In the head area: Here, it can trigger cramping.
In the neck/carotid artery area: Here, it can trigger cardiac arrest.
In the throat and larynx area: Here, it can trigger muscle cramps that can cause suocation.
In the ribcage area: Here, it can increase the risk of ventricular fibrillation and cause cardiac arrest.
Safety precautions
Do not use the unit in close proximity to shortwave or microwave units (e.g. cell phones), as these can cause
fluctuations of the initial values of the unit.
Do not use while sleeping, driving a motor vehicle or simultaneously operating machinery.
Never immerse the unit in water or other liquids.
Do not use in close proximity to highly inflammable materials, gases or explosives.
Consult a doctor before use in the following situations:
Acute diseases, particularly in case of suspected or known blood coagulation disorders, susceptibility to
thromboembolitic disease or malignant growths.
If you have diabetes or other diseases.
Chronic pain conditions that have not been clarified, regardless of the region of the body.
Any sensitivity disorders with reduced sensation of pain (such as metabolic disorders).
Simultaneously carried out medical treatments.
Discomfort experienced as a result of the stimulation treatment.
Persistent skin irritation under the electrodes.
WARNING:
•
Using the unit is not a substitute for consultation or treatment from a physician.
•
Whenever you have any pain or illness, always contact your physician first.
Before using the unit for the first time
CAUTION:
•
Before you use the unit for the first time, remove all packaging materials.
•
Switch the unit o immediately if it malfunctions.
•
If the self-adhesive covers for the electrode connection are missing or become detached, we recommend
urgently that you ax the electrode covers provided.
•
Never wear the belt with bare metal electrodes. Using the belt without electrode covers can cause injuries.
•
The manufacturer shall not be held liable for damage or injuries caused by improper or incorrect use.
•
Protect the unit from dust, dirt and moisture.
•
If the unit has fallen or been dropped, exposed to extreme moisture or incurred other types of damage, it
must no longer be used. Do not expose the unit to high temperatures or direct sunlight.
•
Never open or attempt to repair the unit by yourself, as otherwise proper function is no longer guaranteed.
Failure to observe this regulation shall void the warranty.
•
If you need to have the unit repaired, contact customer service or an authorized dealer.
Battery Handling Safety Precautions
• Use only the size and type of batteries specied.
• Be sure to follow the correct polarity when installing the batteries. Reversed batteries may cause damage to
the device.
• Do not mix dierent types of batteries together (e.g. Alkaline and Carbon-zinc or rechargeable batteries) or old
batteries with fresh ones. Always replace batteries as a simultaneous set.
• If the batteries in the device are depleted or the device will not be used for a long period of time, remove the
batteries to prevent damage or injury from possible battery leakage.
• Do not try to recharge batteries not intended to be recharged; they can overheat and rupture (follow battery
manufacturer’s directions).
• Do not dispose of batteries in re, batteries may explode or leak.
• Clean the battery contacts and also those of the device prior to battery installation.
• Remove discharged batteries from the product and dispose/recycle in compliance with all applicable laws.
• Keep batteries away from children and pets. Batteries may be harmful if swallowed. Should a child or pet swal-
low a battery, seek medical assistance immediately.
7
6. Controls
Part No. Description Part No. Description
1 On/o/pause button 6 Battery status, appears when batteries are depleted
2 Intensity adjustment
increase
decrease
7 Intensity, level 0 – 20
8 Pause symbol, flashes when activated
3 Program/locking key 9 TIMER: Remaining time of the active program in
minutes
4 Battery compartment 10 Keylock active
5 Display with LCD display 11 Active program
7. Before Use
Putting on the belt
Before each use, ensure that your skin is clean and free of creams or oils.
1. Lay out the belt with the electrodes facing you.
2. Moisten the electrodes and your back with water. Ensure that the elec-
trode surfaces are moistened evenly to avoid localised voltage peaks
when they are applied to your skin.
3. Fit the belt around your waist so that the vertical seams A are at the
same place on each side of your body. The Beurer logo B is then in the
middle of the spinal column.
See the diagram
When positioned correctly, the electrodes C are slightly to the left
and right of your spinal column.
4. Then, close the fastener.
Ensure that the belt fits tightly enough to provide a good contact between the electrodes and your skin, but
does not constrict.
B
A
C
A
8
Extending the belt
If the belt is not adequate for your waistline, you can use the extension belt provided. Depending on your
waistline, the holder for the control unit may be away from the center of your body. It is important that the
belt is positioned so that the electrodes C are to the left and right of the spinal column.
Inserting the control unit
Insert the control unit into the holder on the belt. If inserted correctly, the
control unit clicks into the holder.
The control panel should be facing outwards and the display should be
facing upwards and towards you.
Automatic shuto
If the unit is not used after being switched on, it switches o after 3 minutes.
Changing the batteries
If the battery status indicator [6] flashes, change the batteries. To
insert or replace batteries, remove the control unit from the holder
of the belt.
1. Release the magnetic connection by gently pulling the control unit
away from the belt. You can feel the magnetic connection of the two
contact buttons release.
2. Pull the unit upwards and out of the holder.
3. Open the battery compartment [4] by pushing the cover downwards
at the arrow.
4. Insert 3 AAA Alkaline batteries. Ensure the correct polarity.
See the diagram inside the battery compartment.
NOTICE:
•
Do not use rechargeable batteries.
8. Operation
General information about use
The back belt is intended for individual treatment of pain in the lower back area.
NOTICE:
•
Do not switch on the unit until the belt is fitted correctly. See Section 5.
Switching on the unit
Press and hold down the on/o button [1] until the unit beeps and the LCD display [5] switches on.
When the unit is started for the first time, Program A is activated automatically.
Selecting the program
Press the program button P [3] to change programs. The following programs are available:
Program Frequency Time
A 4 Hz –110 Hz (3 phases) 30 min.
B 4 Hz 25 min.
C 2 Hz (burst) 25 min.
D 100 Hz 25 min.
9
Note:
For program A, the eect becomes stronger when switching from phase 1 to phase 2 (after approx. 10 min-
utes). If the intensity seems too high, decrease it using the intensity button [2].
Note:
If you change the program during treatment (for example from A to B), the initial intensity in the new
program increases incrementally to the previously set intensity. This can be stopped by pressing the
intensity button [2] for two seconds or by switching o the unit by pressing the on/o button [1] for two
seconds.
Stopping impulses that are too strong
You can decrease the intensity at any time or switch o the unit by holding the on/o button [1] (approx. 2 sec-
onds).
Adjusting the intensity
Press the intensity button [2] to increase the intensity incrementally or press the intensity button [2] to lower
the intensity. You can adjust the intensity in 20 Increments. Depending on the intensity level, you may feel a
prickling at first that can escalate to a muscle contraction. Select a setting that is within your comfort level.
Preventing unwanted pulse changes
To prevent an accidental increase in intensity during a treatment session, activate the keylock by pressing the
program selection button P [3] for approx. 2 seconds. The unit beeps and the “ ” symbol appears in the
display [5].
To cancel the keylock, press and hold the P [3] button for approx. 2 seconds.
Reacting to unpleasant sensations
If you feel a stinging or itching sensation, switch o the unit and proceed as follows:
Check whether the electrodes are working properly.
Check whether the round cover is still present over the electrode connections.
Take o the belt and moisten it completely once again.
When refitting the belt, ensure good skin contact and good moistening.
Inserting a pause
If you want to take a break or pause a program, press the Pause button [1] briefly. The unit beeps and the “II
pause symbol flashes in the display [5]. You can resume the program by pressing the Pause button [1] again.
Contact detection
If the electrodes have no contact with the body, the intensity is set automatically to zero. This prevents unwanted
stimulation. The intensity cannot be increased without contact with the body.
Memory function
The unit stores the most recently configured program.
After a battery change, the unit starts with the first program again.
9. Cleaning and care of the unit and belt
Cleaning the unit
Remove the batteries from the unit before each cleaning.
Clean the unit with a soft cloth moistened slightly with water after use. If the dirt is heavier, you can also mois-
ten the cloth with a mild soap solution.
Ensure that no water penetrates the device.
Do not use any chemical cleaners or abrasives for cleaning.
NOTICE:
•
The belt and control unit are not suitable for machine washing.
Cleaning the belt
Once you have removed the control unit from the belt, clean the belt using a soft cloth and a mild soap solution
or liquid detergent. Do not use bleach.
10
10. Troubleshooting
Problem/question Possible cause/remedy
The unit does not switch on when
you press the ON/OFF button.
1. Ensure that the batteries are inserted correctly and have contact. The
on/o button must be pressed and held down to switch the unit on.
2. If necessary, replace the batteries.
3. Please contact customer service.
No stimulation can be felt. 1. Increase the pulse intensity incrementally.
2. Press the on/o button to restart the program.
3. The batteries are almost empty. Replace them.
You feel an unpleasant sensation
from the electrodes.
The electrodes are poorly moistened. This can lead to skin irritation, as
the current is no longer distributed evenly over the entire surface. Switch
o the unit. Take o the belt, moisten the belt (electrodes) and your lower
back and restart the program according to the manual.
The skin in the treatment area
becomes red.
Discontinue the treatment immediately and wait until the skin returns to
normal. Temporary skin reddening under the electrode is harmless and
is a result of increased blood flow. However, if the skin irritation persists
and you feel itching or irritation, consult your doctor before continuing
use.
11. Technical specifications
Belt size for waist sizes from Approx. 30 in to 55 in (75 to 140 cm)
Weight (belt and unit) Approx. 0.9 lb (415 g) with batteries
Electrode size Approx. 3.2 in x 2.6 in (81 x 65 mm)
Parameters (500 Ohm load) Output voltage: max. 50 Vpp / 5.5 Vrms
Output current: max. 100 mApp / 11 mArms
Output frequency: 2 – 110 Hz
Pulse duration 60 – 220 μs per phase
Wave form Symmetric, biphasic rectangular pulses
Power supply 4.5 V (3 x 1.5 V AAA)
Operating conditions 32 °F to 104 °F (0 °C to 40 °C), 20 to 65% relative humidity
Storage 32 °F to 131 °F (0 °C to 55 °C), 10 to 90% relative humidity
Protection against ingress of water IPX0, ordinary protection
We reserve the right to make any technical alterations that are necessary in order to improve and develop the
product further. If the unit is used outside of the specifications, proper function is no longer guaranteed.
12. FCC Information
Changes or modifications to the product not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
Note:
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant
to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful
interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in ac-
cordance with the instructions, may cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interfe-
rence to radio or television reception, which can be determined by turning the equipment o and on, the user is
encouraged to try to correct the interference by one or more of the following measures:
– Reorient or relocate the receiving antenna.
– Increase the separation between the equipment and receiver.
– Connect the equipment into an outlet on a circuit dierent from that to which the receiver is connected.
11
– Consult the dealer or an experienced radio/TV technician for help.
13. Warranty
Limited Lifetime Warranty For Original Purchaser
Your Beurer back belt with TENS device, Model EM 38, excluding the battery, is warranted to be free from
defects in materials and workmanship for the life of the product under normal conditions of intended use and
service. This warranty extends only to the original retail purchaser and does not extend to retailers or subsequent
owners.
We will, at our option, repair or replace the Beurer back belt with TENS device, Model EM 38, without additional
charge, for any part or parts covered by these written warranties. No refunds will be given. Repair or replacement
is our only responsibility and your only remedy under this written warranty. If replacement parts for defective ma
-
terials are not available, Beurer reserves the right to make product substitutions in lieu of repair or replacement.
For warranty service contact our customer service department at 1-800-536-0366 or at info@beurer.com to
provide a description of the problem. If the problem is deemed to be within the scope of the limited lifetime
warranty, you will be asked to mail the product at your costs in its original package with proof of purchase, your
name, address and phone number. If the problem is not deemed to be within the scope of the limited lifetime
warranty, we will provide a quotation for repair respectively replacement and return shipping fee.
This warranty does not cover damage caused by misuse or abuse; accident; the attachment of unauthorized
accessory; alteration to the product; improper installation; misapplication; lack of reasonable care with respect
to the product; unauthorized repairs or modications; improper use of electrical/power supply; old worn battery;
normal wear; loss of power; dropped product; malfunction or damage of an operating part as a result of failure to
comply with instructions for use or to provide manufacturer’s recommended maintenance; transit damage; theft;
neglect; vandalism; or environmental conditions; loss of use during the period the product is at a repair facility or
otherwise awaiting parts or repair; or any other conditions whatsoever that are beyond the control of Beurer. This
warranty is void if the product is ever used in a commercial or business environment. The maximum liability of
Beurer under this warranty is limited to the purchase price actually paid by the customer for the product covered
by the warranty, as confirmed by proof of purchase, regardless of the amount of any other direct or indirect
damage suered by the customer.
This warranty is eective only if the product is purchased and operated in the country in which the product is
purchased. A product that requires modifications or adaptation to enable it to operate in any other country than
the country for which it was designed, manufactured, approved and/or authorized, or repair of products dama
-
ged by these modifications is not covered under this warranty.
THE WARRANTY PROVIDED HEREIN SHALL BE THE SOLE AND EXCLUSIVE WARRANTY. ANY IMPLIED
WARRANTIES, OBLIGATIONS, OR LIABILITES, INCLUDING BUT NOT LIMITED TO THE IMPLIED WAR
-
RANTY OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE, ARE LIMITED IN DURATI-
ON TO THE DURATION OF THIS APPLICABLE WRITTEN WARRANTY. Some states do not allow limitations
on how long an implied warranty lasts, so the above limitations may not apply to you.
IN NO EVENT SHALL BEURER BE LIABLE FOR ANY SPECIAL, INCIDENTAL, INDIRECT OR CONSE
-
QUENTIAL DAMAGES FOR BREACH OF THIS OR ANY OTHER WARRANTY, EXPRESS, IMPLIED OR ANY
OTHER THEORY OF LIABILITY, WHATSOEVER. Some states do not allow the exclusion or limitation of special,
incidental, or consequential damages, so the above limitation may not apply to you.
Beurer does not authorize anyone, including, but not limited to, retailers, the subsequent consumer purchaser of
the product from a retailer or remote purchaser, to obligate Beurer in any way beyond the terms set forth herein.
This warranty does not extend to the purchase of opened, used, repaired, repackaged and/or resealed products,
including but not limited to sale of such products on Internet auction sites and/or products by surplus or bulk
resellers. Any and all warranties or guarantees shall immediately cease and terminate in connection with any
products or parts thereof which are repaired, replaced, altered, or modified, without the prior explicitly written
consent of Beurer.
This warranty gives you specific legal rights, and you may also have other rights which may vary from state to
state.
For more information regarding our product line in the USA, please visit: www.beurer.com
12
Distributed by:
Beurer North America LP
900 N Federal Highway, Suite 300
Hallandale Beach, FL 33009. USA
www.beurer.com
Questions or comments? Call toll free 1-800-536-0366 or contact info@beurer.com
Made in China.
36
751.729_EM38-2018-03-14_02_IM_BEU-USA
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36

Beurer EM 38 Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI