Intended use of DURAtec Prophy
contra-angle
To support periodontopathy treatment. For
rotary polishing with rubber tips and polishing
brushes available from your KaVo agent.
Suitable for tips and brushes that are plugged
in or screwed on. Head housing is protected
against penetration of polishing paste by
special seals.
Insertion of rubber tips and polishing
brushes
Screw the desired tip into the head or plug in.
Check for secure retention/stop.
Note the manufacturer’s information
regarding the dental polisher and
dental polishing brushes.
Removal
Wait until the tip has come to a standstill, then
unscrew or pull off.
Care Instructions
External cleaning
The DURAtec contra-angle can be cleaned
on the outside by brushing under running
water or using an alcohol solution.
Do not place the DURAtec contra-
angle in an ultrasonic cleaner since
this may result in malfunction.
Disinfection
The DURAtec contra-angle can be disinfected on the outside with chemical disinfectants
(disinfection by spraying, wiping).
Do not place the DURAtec contra-angle in disinfectant baths since this may
result in malfunction.
Thermodisinfection
The DURAtec contra-angle can be washed in the thermodisinfector.
The DURAtec contra-angle must be sprayed through or oiled with KaVo
maintenance agent immediately after each cycle.
Maintenance with KAVOspray
Remove the bur.
Cover the DURAtec contra-angle with the
Cleanpac bag and plug onto the nozzle.
Press the spray button for 1 second.
Maintain the DURAtec contra-angle twice
daily, at midday and in the evening after
surgery, as well as before every sterilization.
Maintenance with the KaVo SPRAYrotor
Remove the bur.
Place the DURAtec contra-angle on the
appropriate SPRAYrotor coupling and cover
with the Cleanpac bag.
Maintain the DURAtec contra-angle twice
daily, at midday and in the evening after
surgery, as well as before every sterilization.
For further handling please refer to the
operating, maintenance and assembly
instructions of the SPRAYrotor.
Maintenance with QUATTROcare
Cleaning and maintenance unit with expansion
pressure for efficient cleaning and maintenance.
Maintain the DURAtec contra-angle twice
daily, at midday and in the evening after
surgery, as well as before every sterilization.
Refer to the QUATTROcare operating
instructions for further information on
application.