KaVo DURAtec 2731 Operating instructions

Type
Operating instructions
Dear dentist:
KaVo hopes that you will benefit greatly from working with your new quality product, the
DURAtec Prophy contra-angle 2731
Please observe the following notes to assure continued trouble-free, economical and safe product
operation:
Technical Data
DURAtec Prophy contra-angle 2731
for prophylaxis
Drive speed: max. 10,000 rpm
Reduction: 5.4:1
Usable with rubber tips and polishing brushes
Recommended working speed: 400 - 600 rpm
Important Remarks
The DURAtec contra-angle is:
- only to be used by a qualified dental practitioner, in connection with the provision of
dental treatment to dental patients, any other use may prove dangerous and is
strictly forbidden,
- a medical product according to the applicable national provisions of law.
In conformity with these regulations, the DURAtec contra-angle may only be used by
a competent person for the described application, and under consideration of
- the valid regulations on work safety,
- the valid accident prevention measures,
- these operating instructions.
In conformity with these regulations it is the duty of the user,
- to use only faultless tools,
- to ensure that it is used for the correct purpose,
- to protect patients and third parties from all dangers,
- to avoid any contamination by the product.
Putting Into Operation
The contra-angle should be sterilised before
being operated for the first time.
Connection to the drive motor
Insertion
Push the DURAtec contra-angle onto the motor
coupling until latched in.
Pull on the contra-angle before treatment
to verify that it is firmly seated on the motor
coupling.
Removal
Pull the DURAtec contra-angle off the motor while
turning it slightly.
Do not mount or remove the contra-angle
while the coupling is rotating as this may
damage the driver.
Vertrieb/distribution: Hersteller/manufacturer:
KaVo Dental GmbH Kaltenbach & Voigt GmbH
Bismarckring 39 Bismarckring 39
D-88400 Biberach D-88400 Biberach
DURAtec Prophy
contra-angle 2731
Mat.-No. 0.551.0330
en INSTRUCTIONS FOR USE
0.589.1192 - 11.02 - 3
0
1
2
4
05891192_2731_1102ur.qxd 18.10.2006 11:38 Uhr Seite 1
Intended use of DURAtec Prophy
contra-angle
To support periodontopathy treatment. For
rotary polishing with rubber tips and polishing
brushes available from your KaVo agent.
Suitable for tips and brushes that are plugged
in or screwed on. Head housing is protected
against penetration of polishing paste by
special seals.
Insertion of rubber tips and polishing
brushes
Screw the desired tip into the head or plug in.
Check for secure retention/stop.
Note the manufacturers information
regarding the dental polisher and
dental polishing brushes.
Removal
Wait until the tip has come to a standstill, then
unscrew or pull off.
Care Instructions
External cleaning
The DURAtec contra-angle can be cleaned
on the outside by brushing under running
water or using an alcohol solution.
Do not place the DURAtec contra-
angle in an ultrasonic cleaner since
this may result in malfunction.
Disinfection
The DURAtec contra-angle can be disinfected on the outside with chemical disinfectants
(disinfection by spraying, wiping).
Do not place the DURAtec contra-angle in disinfectant baths since this may
result in malfunction.
Thermodisinfection
The DURAtec contra-angle can be washed in the thermodisinfector.
The DURAtec contra-angle must be sprayed through or oiled with KaVo
maintenance agent immediately after each cycle.
Maintenance with KAVOspray
Remove the bur.
Cover the DURAtec contra-angle with the
Cleanpac bag and plug onto the nozzle.
Press the spray button for 1 second.
Maintain the DURAtec contra-angle twice
daily, at midday and in the evening after
surgery, as well as before every sterilization.
Maintenance with the KaVo SPRAYrotor
Remove the bur.
Place the DURAtec contra-angle on the
appropriate SPRAYrotor coupling and cover
with the Cleanpac bag.
Maintain the DURAtec contra-angle twice
daily, at midday and in the evening after
surgery, as well as before every sterilization.
For further handling please refer to the
operating, maintenance and assembly
instructions of the SPRAYrotor.
Maintenance with QUATTROcare
Cleaning and maintenance unit with expansion
pressure for efficient cleaning and maintenance.
Maintain the DURAtec contra-angle twice
daily, at midday and in the evening after
surgery, as well as before every sterilization.
Refer to the QUATTROcare operating
instructions for further information on
application.
05891192_2731_1102ur.qxd 18.10.2006 11:38 Uhr Seite 2
Maintenance with LIFEtime
Cleaning, maintenance and disinfection unit for
automatic and efficient external and internal cleaning.
Maintain the DURAtec contra-angle twice daily, at
midday and in the evening after surgery, as well as
before every sterilization.
Refer to the LIFEtime operating instructions for
further information on application.
Improper maintenance and care can lead
to premature wear and malfunction.
We recommend using original KaVo care
products.
Sterilization
The DURAtec contra-angle can be sterilized in the autoclave at 135 °C (275 °F).
Note the manufacturers information regarding the dental polisher and
dental polishing brushes.
Before each sterilization cycle the DURAtec contra-angle must be
maintained with KaVo maintenance agent.
Remove the DURAtec contra-angle from the sterilizer immediately after
each sterilization cycle to prevent any risk of contact corrosion.
Accessories
Are supplied by your Dental Depot
Description Mat.-No.
Instrument stand 2151 0.411.9501
Cellulose pad, 100 pcs. 0.411.9862
Cleanpac, 10 pcs. 0.411.9691
KAVOspray 2112 A 0.411.9640
ROTAspray 2142 A 0.411.7520
QUATTROcare Spray 2108 A 0.411.7720
LIFEtime Spray 2118 A 0.411.9700
Further advice for use and safety
When the DURAtec contra-angle 2731 will not be used for an extended
period of time, it should be cleaned and lubricated according to the
instructions and stored in a dry place.
Work must stop immediately if there are damages, unnatural running
noises or too strong vibrations.
To avoid further damage, please contact the KaVo Customer Service Center at
Warthausen, or a KaVo authorised office or your Dental Depot.
KaVo LIFEtime
®
93°C
10min
93°C
10min
93°C
10min
93°C
10min
93°
min
min
93°
F
05891192_2731_1102ur.qxd 18.10.2006 11:38 Uhr Seite 3
KaVo Dental GmbH. D-88400 Biberach/Riss
Telefon +49 7351 56-0 Fax +49 7351 56-1488
Internet: www.kavo.com