Extech Instruments GRT350 User manual

Category
Measuring, testing & control
Type
User manual

Extech Instruments GRT350 is a top-of-the-line Earth Resistance and Resistivity Tester, designed to provide accurate and reliable measurements. With its advanced safety features and user-friendly interface, the GRT350 is ideal for a wide range of applications, including testing earthing systems, ground rods, and soil resistivity. Key features include earth resistance testing with three ranges (0-20Ω/0-200Ω/0-2kΩ), earth voltage measuring range of 0-300Vac, and earth resistivity testing using the Wenner 4-terminal method.

Extech Instruments GRT350 is a top-of-the-line Earth Resistance and Resistivity Tester, designed to provide accurate and reliable measurements. With its advanced safety features and user-friendly interface, the GRT350 is ideal for a wide range of applications, including testing earthing systems, ground rods, and soil resistivity. Key features include earth resistance testing with three ranges (0-20Ω/0-200Ω/0-2kΩ), earth voltage measuring range of 0-300Vac, and earth resistivity testing using the Wenner 4-terminal method.

User Manual
Earth Resistance and Resistivity Tester
Model GRT350
Additional User Manual Translations available at www.extech.com
GRT350-en-GB_v1.4 12/17
2
Introduction
Thank you for selecting the Extech Instruments Model GRT350 Earth Resistance and Resistivity
Tester. The GRT350 has been designed and tested according to the IEC Publication 348, safety
requirements for Electronic Measuring Apparatus, EN 61010-1, EN 61326-1, EN 61557-1, EN
61557-5 and other safety standards. This device is shipped fully tested and calibrated and, with
proper use, will provide years of reliable service. Please visit our website (www.extech.com) to
check for the latest version of this User Guide, Product Updates, Product Registration, and
Customer Support.
Features
Microprocessor controlled with advanced safety features
Two line LCD display
Auto-Ranging
Earth resistance testing with three ranges: 0-20Ω/0-200Ω/0-2kΩ
Earth voltage measuring range of 0-300Vac
Automatic C (current) spike check.
Automatic P (potential) spike check.
2-wire test
3-wire test
4-wire test
Earth Resistivity (Wenner 4 terminal)
Auto power off
Low battery indicator
Data hold
200 memory locations to save test data
Safety standard: EN 61010-1 CATIV 300V, EN 61326-1
GRT350-en-GB_v1.4 12/17
3
Safety Information
Read the all safety information carefully before attempting to operate the meter.
Use the meter only as specified in this manual. Otherwise, the protection provided by the
meter may be impaired.
Rated environmental conditions:
Indoor & outdoor use
Installation Category IV 300V
Pollution Degree 2
Altitude up to 2000m
Relative Humidity 80% maximum
Ambient temperature 0-40°C
Observe the International Electrical Symbols listed below:
Double insulation
This symbol, adjacent to a terminal, indicates that, under normal use, hazardous
voltages may be present
This symbol, adjacent to another symbol or terminal, indicates the user must refer
to the manual for further information.
Earth (ground) terminal
Equipment complies with current EU directives
WARNING
To avoid electrical shock, do not touch the terminals during tests
Never apply Voltage higher than 300V across P1 and P2 terminals
GRT350-en-GB_v1.4 12/17
4
Meter Description
1. C1 terminal (Black test lead connection)
2. P1 terminal (Green test lead connection)
3. P2 terminal (Yellow test lead connection)
4. C2 terminal (Red test lead connection)
5 Display
6. Rc LED
7. Rp LED
8. Save button
9. Cursor buttons
10. Cursor buttons
11. ESC button
12. Power button
13. TEST/STOP button
GRT350-en-GB_v1.4 12/17
5
Operation
Battery Voltage check
Press the "ON/OFF" button, if "Battery Low" appears on the display, replace the batteries.
Earth Voltage measurement
1. Connect the test leads as shown below.
(1) Earth electrode (rod) under test (2) Test spike
2. Press the "ON/OFF" button and wait for “Select Function” to appear on the display.
3. Press the "" button 3 times and wait for "ACV" to appear on the display.
4. Press the "TEST/STOP" button.
5. The earth voltage will appear on the display if it is present.
6. Press SAVE to save the measurement in Log memory.
7. Press ESC to return to the “Select Function” display.
Measurement Considerations
Two-terminal testing of earth resistance is appropriate for most general purpose testing in
normally conductive soil. However, two-terminal measurements include test lead and contact
resistance in the measurement and the result will be a reading slightly higher than the true earth
resistance.
When measured results are higher than desired or if measurement directives require multi-
terminal techniques, switch to the 3 or 4 terminal techniques as needed.
GRT350-en-GB_v1.4 12/17
6
Earth Resistance measurement
Note: The measured results may be influenced by inductive or capacitive coupling if the test
leads are twisted or adjacent to each other.
When connecting the Probes, keep the leads separated.
Setup
Insert the Potential electrodes (P) and the Current electrodes (C) as deep as possible into the
soil. The distance between electrodes must be 5 to 10m (16 to 32’).
Four-terminal earth resistance measurements
(1) Earth electrode (rod) under test (2) Potential spike (3) Current spike
Three-terminal earth resistance measurements
(1) Earth electrode (rod) under test (2) Potential spike (3) Current spike
Two-terminal earth resistance measurements
(1) Earth electrode (rod) under test (2) Potential spike
GRT350-en-GB_v1.4 12/17
7
Testing
1. Connect the test leads for 2, 3 or 4 terminal Earth Resistance testing.
P1 = Green lead, C1 = Black lead, P2 = Yellow lead, C2 = Red lead
2. Press the ON/OFF button and wait for the “Select Function” screen to appear.
3. Press the "" button to select "2P", "3P" or "4P" depending on the test.
4. Press "TEST/STOP" button to begin the test.
5. The meter will beep while the test is in progress (for approximately 10 seconds) and then
the reading will appear on the lower line of the display.
6. Press SAVE to save the results in Log memory.
7. Press ESC to return to the “Select Function” display.
Notes for "Rc" & "Rp" LED indications:
Rc: No test current output. Check the connections and settings.
Rp: If Rp is on and the display indicates “> 2 kΩ”, the earth resistance is greater than 2000Ω.
Earth Resistivity Testing (ρ) (Wenner 4-point method)
Insert the four electrodes at equal distances (and in a straight line) at an interval of 1 to 30m (3.3
to 32.3’) apart. The depth should be 5% (or less) the distance between the electrodes.
For example: if the distance between the electrodes is 5m (16’), the electrodes should be
inserted to a depth of 25cm (10”). (500cm x 5% = 25cm)
NOTE: The length can be set between 1 and 50m (3.3’ and 164’).
1. Connect the test leads for 4 terminal testing.
2. Press the ON/OFF button and wait for the “Select Function” screen to appear.
3. Press the "" button and select "Resistivity".
4. Press "TEST/STOP" button. The setting length will display on the screen.
5. Press the "" or "" arrow buttons to set the count up or down.
6. Press "TEST/STOP" button to begin the test.
7. Press SAVE to save the results in Log memory.
8. Press ESC to return to the “Select Function” display.
GRT350-en-GB_v1.4 12/17
8
Save Function
The meter permits saving measurements in up to 200 memory locations.
When a test has completed, press the SAVE button to store the measured data.
The display will show the location number of the saved data. SAVE ###
LOG Display
1. Press the ON/OFF button and wait for the “Select Function” screen to appear.
2. Press the button and select LOG Display.
3. Press the TEST/STOP button. The log number will display on the screen.
4. Press the "" or "" arrow button to scroll through the saved data.
5. Press ESC to get back to the “Select Function” display.
LOG Clear
1. Press the ON/OFF button and wait for the “Select Function” screen to appear.
2. Press the button and select LOG Clear
3. Press the TEST/STOP button. Are you sure? will display on the screen.
4. Press the TEST/STOP button to clear the log data. Successful! will display on the
screen.
5. Press the TEST/STOP button to verify that the data log is clear. “No Log data” will
display on the screen.
6. Press ESC to get back to the “Select Function” display.
GRT350-en-GB_v1.4 12/17
9
Maintenance
1. Rear cover
2. Battery compartment
3. Fuse
Fuse Replacement
1. Disconnect the test leads from the instrument.
2. Remove the rear cover by removing the two screws.
3. Remove and replace the fuse with a new one of the same value and size
0.5A/500V, (5 x 20mm).
4. Replace and secure the rear cover.
Battery replacement
When Battery Low appears on the display, replace the batteries.
1. Disconnect the test leads from the instrument and remove the rear cover and the
batteries.
2. Replace with eight 1.5V AA batteries, taking care to observe correct polarity.
3. Reinstall battery holder and battery cover.
Never dispose of used batteries or rechargeable batteries in household waste.
As consumers, users are legally required to take used batteries to appropriate collection
sites, the retail store where the batteries were purchased, or wherever batteries are sold.
Disposal: Do not dispose of this instrument in household waste. The user is obligated to
take end-of-life devices to a designated collection point for the disposal of electrical and
electronic equipment.
Cleaning
WARNING: To avoid electrical shock or damage to the meter, do not allow water inside the
case.
Periodically wipe the case with a damp cloth and detergent, do not use abrasives or solvents.
GRT350-en-GB_v1.4 12/17
10
Specifications
General Specifications
Earth Resistivity Ranges (ρ)
(ρ = 2 πLR)
L = distance between probes
R = resistance
0.06-6.28 kΩ.m
0.62-62.8 kΩ.m
6.28-628 kΩ.m
Test Frequency
820Hz
Test Current
2mA
Data Memory
200 manual locations
Test lead lengths
Green lead 5 m (16 ft)
Black lead 10m (32 ft)
Yellow lead 10m (32ft)
Red lead 15m (49ft)
Temperature & Humidity
Operating: 0 to 50°C (32 to 122°F); 80%R.H.
Storage: -10 to 60°C (14 to 140°F); 80%R.H.
Altitude
2000m (6500’) max
Power Source
1.5V(AA) x 8 batteries
Dimensions
250(L) x 190(W) x 110(D)mm (9.84 x 7.5 x 4.33”)
Weight
Approx. 1430g (batteries installed)
Fuse
500mA/500V (5 x 20mm) fast blow, ceramic type
Safety Standards
CAT IV-300V, EN 61010-1, EN 61326-1, EN 61557-1, EN
61557-5
Range Specifications
Earth Resistance
Range
Resolution
0 to 20 Ω
0.01 Ω
0 to 200 Ω
0.1 Ω
0 to 2 kΩ
1 Ω
Earth Voltage
0 to 300 VAC
(50/60 Hz)
1VAC
Copyright © 2014-2017 FLIR Systems, Inc.
All rights reserved including the right of reproduction in whole or in part in any form
ISO-9001 Certified
www.extech.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10

Extech Instruments GRT350 User manual

Category
Measuring, testing & control
Type
User manual

Extech Instruments GRT350 is a top-of-the-line Earth Resistance and Resistivity Tester, designed to provide accurate and reliable measurements. With its advanced safety features and user-friendly interface, the GRT350 is ideal for a wide range of applications, including testing earthing systems, ground rods, and soil resistivity. Key features include earth resistance testing with three ranges (0-20Ω/0-200Ω/0-2kΩ), earth voltage measuring range of 0-300Vac, and earth resistivity testing using the Wenner 4-terminal method.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI