SEW 6420 DM User manual

Category
Multimeters
Type
User manual
INSTRUCTION MANUAL
DIGITAL MULTIMETER
with bargraph
6420 DM
NOTE
This meter has been designed and tested According
to CE Safety Requirements for Electronic Measuring
Apparatus, IEC / EN 61010-1 and other safety
standards. Follow all warnings to ensure safe
operation.
WARNING
READ "SAFETY NOTES" (NEXT PAGE) BEFORE
USING THE METER.
1. INTRODUCTION
INDEX PAGE
1.INTRODUCTION........................ 1
2.SAFETY NOTES......................... 2
3.FEATURES................................ 3
4.SPECIFICATIONS...................... 4-5
5.GENERAL.................................. 6
6.INSTRUMENT LAYOUT............... 7-11
7.MEASUREMENT......................... 12-13
8.MAINTENANCE......................... 14
9.FUSE REPLACEMENT................ 15
- 1 -
NOTE
This meter has been designed and tested According
to CE Safety Requirements for Electronic Measuring
Apparatus, IEC / EN 61010-1 and other safety
standards. Follow all warnings to ensure safe
operation.
WARNING
READ "SAFETY NOTES" (NEXT PAGE) BEFORE
USING THE METER.
1. INTRODUCTION
INDEX PAGE
1.INTRODUCTION........................ 1
2.SAFETY NOTES......................... 2
3.FEATURES................................ 3
4.SPECIFICATIONS...................... 4-5
5.GENERAL.................................. 6
6.INSTRUMENT LAYOUT............... 7-11
7.MEASUREMENT......................... 12-13
8.MAINTENANCE......................... 14
9.FUSE REPLACEMENT................ 15
- 1 -
3.FEATURES
Safety Standard :
EN 61010-1 CAT 600V
EN 61326-1
To disable the auto-off function, press the "Range"
key and turn on the 6420 DM multimeter at the
same time.
4000 - count LCD.
Full automatic measurement.
Voltage measurement.
Current measurement.
Resistor measurement.
Frequency measurement.
Range change function.
Bar graph display.
Relative mode.
Maximum / Minimum modes.
Peak hold mode.
VAHz mode measures frequency in voltage or
current mode.
Data Hold function.
Continuity check.
Diode measurement.
Low battery indication.
40M Hz frequency counter.
3V DC power supply.
Input impedance 10 M
- 3 -
2.SAFETY NOTES
Read the following safety information carefully
Before attempting to operate or service the meter.
Use the meter only as specified in this manual.
Otherwise, the protection provided by the meter
may be impaired.
Rated environmental conditions
(1) Indoor Use.
(2) Installation Category .
(3) Pollution Degree 2.
(4) Altitude up to 2000 meters.
(5) Relative humidity 80% max.
(6) Ambient temperature 0~40°C
Observe the International Electrical Symbols
listed below
Meter is protected throughout by double
insulation or reinforced insulation.
Warning ! Risk of electric shock.
Caution ! Refer to this manual before using
the meter.
AC... Alternating current.
DC... Direct current.
.
- 2 -
3.FEATURES
Safety Standard :
EN 61010-1 CAT 600V
EN 61326-1
To disable the auto-off function, press the "Range"
key and turn on the 6420 DM multimeter at the
same time.
4000 - count LCD.
Full automatic measurement.
Voltage measurement.
Current measurement.
Resistor measurement.
Frequency measurement.
Range change function.
Bar graph display.
Relative mode.
Maximum / Minimum modes.
Peak hold mode.
VAHz mode measures frequency in voltage or
current mode.
Data Hold function.
Continuity check.
Diode measurement.
Low battery indication.
40M Hz frequency counter.
3V DC power supply.
Input impedance 10 M
- 3 -
2.SAFETY NOTES
Read the following safety information carefully
Before attempting to operate or service the meter.
Use the meter only as specified in this manual.
Otherwise, the protection provided by the meter
may be impaired.
Rated environmental conditions
(1) Indoor Use.
(2) Installation Category .
(3) Pollution Degree 2.
(4) Altitude up to 2000 meters.
(5) Relative humidity 80% max.
(6) Ambient temperature 0~40°C
Observe the International Electrical Symbols
listed below
Meter is protected throughout by double
insulation or reinforced insulation.
Warning ! Risk of electric shock.
Caution ! Refer to this manual before using
the meter.
AC... Alternating current.
DC... Direct current.
.
- 2 -
- 4 -
AC Voltage
Input impedance 10 M
DC Voltage
Input impedance 10 M
AC uA, mA & DC uA, mA
Overload Protection 0.5 A(250 V) fast blow fuse
4.SPECIFICATIONS
(All at 23°C±5°C 80%R.H)
Range
400 mV
4 V
40 V
400 V
750 V
Accuracy
± (1% rdg+5dgt)
± (1% rdg+5dgt)
± (1% rdg+5dgt)
± (1% rdg+5dgt)
± (1% rdg+5dgt)
Resolution
0.1 mV
1 mV
10 mV
100 mV
1 V
Range
400 mV
4 V
40 V
400 V
1000 V
Accuracy
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
Resolution
0.1 mV
1 mV
10 mV
100 mV
1 V
Range
400 uA
4000 uA
40 mA
400 mA
Accuracy
± (1% rdg+5dgt)
± (1% rdg+5dgt)
Resolution
0.1 uA
1 uA
0.01 mA
0.1 mA
- 5 -
DC AC 10A
Overload Protection 12.5 A(500 V) fast blow fuse
Resistance
Overload Protection 500V DC
Continuity Test
Frequency counter
Overload protection 500V RMS
Range
400
Audible threshold
Less than 35
Range
400
4 K
40 K
400 K
4 M
40 M
Accuracy
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (2.0% rdg+4dgt)
Resolution
0.1
1
10
100
1 K
10 K
Range
4K Hz
40K Hz
400K Hz
4M Hz
40M Hz
Accuracy
± (1% rdg+2dgt)
± (1% rdg+2dgt)
± (1% rdg+2dgt)
± (1% rdg+2dgt)
± (1% rdg+2dgt)
Resolution
1 Hz
10 Hz
100 Hz
1K Hz
10K Hz
Range
10 A
Accuracy
± (2% rdg+3dgt)
Resolution
0.1 A
- 4 -
AC Voltage
Input impedance 10 M
DC Voltage
Input impedance 10 M
AC uA, mA & DC uA, mA
Overload Protection 0.5 A(250 V) fast blow fuse
4.SPECIFICATIONS
(All at 23°C±5°C 80%R.H)
Range
400 mV
4 V
40 V
400 V
750 V
Accuracy
± (1% rdg+5dgt)
± (1% rdg+5dgt)
± (1% rdg+5dgt)
± (1% rdg+5dgt)
± (1% rdg+5dgt)
Resolution
0.1 mV
1 mV
10 mV
100 mV
1 V
Range
400 mV
4 V
40 V
400 V
1000 V
Accuracy
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
± (0.5% rdg+3dgt)
Resolution
0.1 mV
1 mV
10 mV
100 mV
1 V
Range
400 uA
4000 uA
40 mA
400 mA
Accuracy
± (1% rdg+5dgt)
± (1% rdg+5dgt)
Resolution
0.1 uA
1 uA
0.01 mA
0.1 mA
- 5 -
DC AC 10A
Overload Protection 12.5 A(500 V) fast blow fuse
Resistance
Overload Protection 500V DC
Continuity Test
Frequency counter
Overload protection 500V RMS
Range
400
Audible threshold
Less than 35
Range
400
4 K
40 K
400 K
4 M
40 M
Accuracy
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (1.5% rdg+3dgt)
± (2.0% rdg+4dgt)
Resolution
0.1
1
10
100
1 K
10 K
Range
4K Hz
40K Hz
400K Hz
4M Hz
40M Hz
Accuracy
± (1% rdg+2dgt)
± (1% rdg+2dgt)
± (1% rdg+2dgt)
± (1% rdg+2dgt)
± (1% rdg+2dgt)
Resolution
1 Hz
10 Hz
100 Hz
1K Hz
10K Hz
Range
10 A
Accuracy
± (2% rdg+3dgt)
Resolution
0.1 A
- 7 -
6.INSTRUMENT LAYOUT
(1) Function Switch (8) Range Button
(2) LCD Display (9) HOLD Button
(3) V. Hz. Terminal (10) REL Button
(4) mA. uA Terminal (11) PEAK Button
(5) 10A Terminal (12) VAHz Button
(6) COM Terminal (13) MAX/MIN Button
(7) Select Button
5.GENERAL
Dimensions
192mm(L) x 88.6mm(W) x 45.2mm(D)
Weight
Approx. 350g (battery included)
Power source
AAA battery 1.5V x 2.
Low Battery Indication
"BATT" sign appears on the display when the
battery voltage drops below accurate operating
level.
Accessories
Test leads,
Batteries,
Instruction Manual.
- 6 -
(1)
(2)
(9)
(13)
(4)
(7)
(6)
(3)(5)
(10)
(11)
(8)
(12)
- 7 -
6.INSTRUMENT LAYOUT
(1) Function Switch (8) Range Button
(2) LCD Display (9) HOLD Button
(3) V. Hz. Terminal (10) REL Button
(4) mA. uA Terminal (11) PEAK Button
(5) 10A Terminal (12) VAHz Button
(6) COM Terminal (13) MAX/MIN Button
(7) Select Button
5.GENERAL
Dimensions
192mm(L) x 88.6mm(W) x 45.2mm(D)
Weight
Approx. 350g (battery included)
Power source
AAA battery 1.5V x 2.
Low Battery Indication
"BATT" sign appears on the display when the
battery voltage drops below accurate operating
level.
Accessories
Test leads,
Batteries,
Instruction Manual.
- 6 -
(1)
(2)
(9)
(13)
(4)
(7)
(6)
(3)(5)
(10)
(11)
(8)
(12)
B
B
B
B
- 9 -- 8-
(1) Function selection rotary switch
The rotary switch selects the function.
(2) LCD Display
3999 count LCD with LOW BATTERY indication.
(3) Volt / Ohm / Frequency Terminal
This is positive input terminal for voltage / ohm /
frequency measurement. Use the RED test lead
to connect.
(4) uA / mA Terminal
This is positive input terminal for current. The
Maximum is 4000uA for uA measurement. The
Maximum is 400mA for mA measurement.
(5) 10A Terminal
This is positive input terminal for 10A current.
(6) COM Terminal
This is the ground input terminal. Use the BLACK
test lead to connect.
(7) Select Button
For AC / DC function selection.
In the resistance + continuity + diode function,
press the Select button to select resistance,
continuity or diode function.
(8) Range Button
Press the Range button to selects the manual
range mode and turns off the AUTO annunciator
and turns on the manual annunciator and
changes the full scale range. In manual range
mode, each time press Range button (less than
one second), the range increments and a new
value is displayed. To exit the manual range
mode and return to auto mode, press the RANGE
button (More than one second).
(9) HOLD Button
Press the HOLD button (HOLD annunciator turns
on) makes the meter stop updating the LCD
display. This mode can be nested in most of the
special modes. Enabling HOLD function in
automatic mode makes the meter switch to
manual mode, but the full scale range remains
the same. HOLD function can be cancelled by
changing the measurement mode, pressing
range, or push HOLD again.
(10)REL Button
The "REL" mode means relative mode. Press the
REL button, the REL annunciator turns on, zero
the display, and store the displayed reading as a
reference value. Press the REL button more than 1
second to exit the REL mode. In the REL mode, the
value shown on the display is always difference
between the stored reference value and the
present reading. If the new reading is the same as
The reference value, the display will be zero.
B
B
B
B
- 9 -- 8-
(1) Function selection rotary switch
The rotary switch selects the function.
(2) LCD Display
3999 count LCD with LOW BATTERY indication.
(3) Volt / Ohm / Frequency Terminal
This is positive input terminal for voltage / ohm /
frequency measurement. Use the RED test lead
to connect.
(4) uA / mA Terminal
This is positive input terminal for current. The
Maximum is 4000uA for uA measurement. The
Maximum is 400mA for mA measurement.
(5) 10A Terminal
This is positive input terminal for 10A current.
(6) COM Terminal
This is the ground input terminal. Use the BLACK
test lead to connect.
(7) Select Button
For AC / DC function selection.
In the resistance + continuity + diode function,
press the Select button to select resistance,
continuity or diode function.
(8) Range Button
Press the Range button to selects the manual
range mode and turns off the AUTO annunciator
and turns on the manual annunciator and
changes the full scale range. In manual range
mode, each time press Range button (less than
one second), the range increments and a new
value is displayed. To exit the manual range
mode and return to auto mode, press the RANGE
button (More than one second).
(9) HOLD Button
Press the HOLD button (HOLD annunciator turns
on) makes the meter stop updating the LCD
display. This mode can be nested in most of the
special modes. Enabling HOLD function in
automatic mode makes the meter switch to
manual mode, but the full scale range remains
the same. HOLD function can be cancelled by
changing the measurement mode, pressing
range, or push HOLD again.
(10)REL Button
The "REL" mode means relative mode. Press the
REL button, the REL annunciator turns on, zero
the display, and store the displayed reading as a
reference value. Press the REL button more than 1
second to exit the REL mode. In the REL mode, the
value shown on the display is always difference
between the stored reference value and the
present reading. If the new reading is the same as
The reference value, the display will be zero.
(11)PEAK Button
The PEAK mode is used to measure the peak
value of a signal. It is useable with AC Current
measurements. Press the PEAK button, the
PMAX annunciator turns on and the display
shows the maximum peak value. When the
PEAK button is pressed again (less than one
second), the PMIN annunciator turns on, PMAX
annunciator turns off and the display shows the
minimum peak value. When the PEAK button is
pressed for the third time (less than one second),
the PMAX and PMIN annunciators are blinking,
and display shows current input value. The meter
returns to normal operation if the PEAK button is
pressed and held for longer than one second.
(12)VAHZ Button
If VAHZ button pressed in ACV or ACA measure-
ment mode, the meter enters frequency counter
mode with automatic range selection. Therefore,
pressing RANGE button in VAHZ mode doesn't
change the frequency range. However, range
changes the sensitivity of frequency detection.
If the input signal has a small amplitude, the
user shall increase the sensitivity.
(13)MAX / MIN Button
Press button to enter the MAX/MIN recording
mode. When the MAX/MIN button is pressed for
the first time, the "MAX" annunciator turns on, and
the display shows the maximum value. When the
MAX/MIN button is pressed again (less than one
second), the "MIN" annunciator turns on, and the
display shows the minimum value. When the
MAX/MIN button is pressed for the third time (less
than one second), the MAX and MIN annunciators
are blinking, and display shows current input
value. The meter returns to normal operation if the
MAX/MIN button is pressed and held for longer
than one second. Pressing the HOLD button in
MAX/MIN mode makes the meter stop updating
the maximum or the minimum value.
-10 -
-11 -
(11)PEAK Button
The PEAK mode is used to measure the peak
value of a signal. It is useable with AC Current
measurements. Press the PEAK button, the
PMAX annunciator turns on and the display
shows the maximum peak value. When the
PEAK button is pressed again (less than one
second), the PMIN annunciator turns on, PMAX
annunciator turns off and the display shows the
minimum peak value. When the PEAK button is
pressed for the third time (less than one second),
the PMAX and PMIN annunciators are blinking,
and display shows current input value. The meter
returns to normal operation if the PEAK button is
pressed and held for longer than one second.
(12)VAHZ Button
If VAHZ button pressed in ACV or ACA measure-
ment mode, the meter enters frequency counter
mode with automatic range selection. Therefore,
pressing RANGE button in VAHZ mode doesn't
change the frequency range. However, range
changes the sensitivity of frequency detection.
If the input signal has a small amplitude, the
user shall increase the sensitivity.
(13)MAX / MIN Button
Press button to enter the MAX/MIN recording
mode. When the MAX/MIN button is pressed for
the first time, the "MAX" annunciator turns on, and
the display shows the maximum value. When the
MAX/MIN button is pressed again (less than one
second), the "MIN" annunciator turns on, and the
display shows the minimum value. When the
MAX/MIN button is pressed for the third time (less
than one second), the MAX and MIN annunciators
are blinking, and display shows current input
value. The meter returns to normal operation if the
MAX/MIN button is pressed and held for longer
than one second. Pressing the HOLD button in
MAX/MIN mode makes the meter stop updating
the maximum or the minimum value.
-10 -
-11 -
time of measurement is only 50ms, the least
significant digit will not display.
(5)Diode Measurement
Diode measurement mode shares the same
configuration with 4.000 V manual voltage
measurement mode.
If the test circuit is open or the voltage drop
between the two ports of the device (diode) under
test are larger than 2 V, the LCD panel will show
"OL".
The buzzer generates a 2K Hz sound whenever
the digit number is less than 0.25V. Because the
cycle time of measurement is only 50ms, the least
significant digit will not display.
(6)Frequency measurement
Switch to frequency measurement mode.
Insert the BLACK test lead to COM and the RED
one to the Hz terminal.
Apply the test leads to the points across which the
frequency is to be measured, and read the result
directly from the display.
7. MEASUREMENT
Before proceeding with measurement , read the
safety notes.
(1)Voltage measurement
Insert the BLACK test lead to COM and the RED
One to the other terminal.
Switch to V function for ACV or DCV
function.
Use the test lead tip to the circuit and read the
reading of display directly.
If the readings exceed AC 600V(DC 600V), maybe
the reading value is wrong and it is dangerous.
(Refer to the safety notes)
(2)Current measurement
Switch to mA / uA or 10A function for AC current
or DC current measurement.
(3)Resistance Measurement
Switch to OHM range and make sure there is no
power in the circuit being measured. Insert the
BLACK lead to the COM and the RED one to
the OHM terminal.
Connect the test leads to the circuit or device
under test and read the display directly.
(4)Continuity Check
Continuity check shares the same configuration
with 400.0 manual resistance measurement
mode, but with buzzer output to indicate continuity.
The buzzer generates a 2KHz sound whenever
the digit number less than 35 . Because the cycle
-13 -
-12 -
time of measurement is only 50ms, the least
significant digit will not display.
(5)Diode Measurement
Diode measurement mode shares the same
configuration with 4.000 V manual voltage
measurement mode.
If the test circuit is open or the voltage drop
between the two ports of the device (diode) under
test are larger than 2 V, the LCD panel will show
"OL".
The buzzer generates a 2K Hz sound whenever
the digit number is less than 0.25V. Because the
cycle time of measurement is only 50ms, the least
significant digit will not display.
(6)Frequency measurement
Switch to frequency measurement mode.
Insert the BLACK test lead to COM and the RED
one to the Hz terminal.
Apply the test leads to the points across which the
frequency is to be measured, and read the result
directly from the display.
7. MEASUREMENT
Before proceeding with measurement , read the
safety notes.
(1)Voltage measurement
Insert the BLACK test lead to COM and the RED
One to the other terminal.
Switch to V function for ACV or DCV
function.
Use the test lead tip to the circuit and read the
reading of display directly.
If the readings exceed AC 600V(DC 600V), maybe
the reading value is wrong and it is dangerous.
(Refer to the safety notes)
(2)Current measurement
Switch to mA / uA or 10A function for AC current
or DC current measurement.
(3)Resistance Measurement
Switch to OHM range and make sure there is no
power in the circuit being measured. Insert the
BLACK lead to the COM and the RED one to
the OHM terminal.
Connect the test leads to the circuit or device
under test and read the display directly.
(4)Continuity Check
Continuity check shares the same configuration
with 400.0 manual resistance measurement
mode, but with buzzer output to indicate continuity.
The buzzer generates a 2KHz sound whenever
the digit number less than 35 . Because the cycle
-13 -
-12 -
9. FUSE REPLACEMENT
There are 3 fuses on the PCB.
0.5A for mA. Range and 12.5A for 10A range.
Another 0.5A for Volt / OHM / frequency range.
Open the meter case, then remove and replace the
fuse.
Only replace with the same specification fuse.
8. MAINTENANCE
Battery replacement
When low battery warning appears, change new
batteries as follows
(1) Disconnect the test leads from the instrument and
turn off power.
(2) Unscrew the case and replace new batteries.
(3) Close the case and lock the screw.
Cleaning and Storage
Periodically wipe the case with a damp cloth and
detergent. Do not use abrasives or solvents.
If the meter is not used for over 60 days, remove the
battery for storage.
WARNING
To avoid electrical shock or damage to the meter,
do not get water inside the case.
-15 -
-14 -
FUSE 0.5A
FUSE 12.5A
BATTERY AAA 1.5V X2
FUSE 0.5A
Due to our policy of constant improvement and development, we reserve
the right to change specifications without notice.
(5 X 20 mm)
(6 X 32 mm)
(5 X 20 mm)
/ 250V
/ 250V
/ 500V
9. FUSE REPLACEMENT
There are 3 fuses on the PCB.
0.5A for mA. Range and 12.5A for 10A range.
Another 0.5A for Volt / OHM / frequency range.
Open the meter case, then remove and replace the
fuse.
Only replace with the same specification fuse.
8. MAINTENANCE
Battery replacement
When low battery warning appears, change new
batteries as follows
(1) Disconnect the test leads from the instrument and
turn off power.
(2) Unscrew the case and replace new batteries.
(3) Close the case and lock the screw.
Cleaning and Storage
Periodically wipe the case with a damp cloth and
detergent. Do not use abrasives or solvents.
If the meter is not used for over 60 days, remove the
battery for storage.
WARNING
To avoid electrical shock or damage to the meter,
do not get water inside the case.
-15 -
-14 -
FUSE 0.5A
FUSE 12.5A
BATTERY AAA 1.5V X2
FUSE 0.5A
Due to our policy of constant improvement and development, we reserve
the right to change specifications without notice.
(5 X 20 mm)
(6 X 32 mm)
(5 X 20 mm)
/ 250V
/ 250V
/ 500V
CAT - measurements performed at the source of
the low voltage installations.
CAT - measurements performed in building
installations.
CAT - measurements performed on circuits
directly connected to the low voltage installations.
For
For
For
CAT - For measurements performed on circuits not
directly connected to mains.
CAT - measurements performed at the source of
the low voltage installations.
CAT - measurements performed in building
installations.
CAT - measurements performed on circuits
directly connected to the low voltage installations.
For
For
For
CAT - For measurements performed on circuits not
directly connected to mains.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19

SEW 6420 DM User manual

Category
Multimeters
Type
User manual

Ask a question and I''ll find the answer in the document