P5200A

Tektronix P5200A, P5202A, P5205A, P5210A User manual

  • Hello! I am an AI chatbot trained to assist you with the Tektronix P5200A User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
x
P5200A Series
High Voltage Differential Probes
ZZZ
Instruction Manual
*P077053602*
077-0536-02
P5200A Series
High Voltage Differential Probes
ZZZ
Instruction Manual
xx
Revision B
www.tektronix.com
077-0536-02
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are
protected by na
tional copyright laws and international treaty provisions.
Tektronix pro
ducts are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all
previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl B raun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit ww w.tek.com to nd contacts in your area.
Warranty
Tektronix warrants that this product will be free from defects in materials and workmanship for a period o f one (1) year from the date of
shipment. If any such product proves defective during this warranty period, Tektronix, at its option, either will repair the defective
product without charge for parts and labor, or will provide a replacement in exchange for the defective product. Parts, modules and
replacement products used by Tektronix for warranty work may be new or reconditioned to like new performance. All replaced
parts, modules and products become the property of Tektronix.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the warranty period
and make suitable arrangements for the performance of service. Customer shall be responsible for packaging and shipping the
defective product to the service center designated by Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the
product to Customer if the shipment is to a location w ithin the country in which the Tektronix service center is located. Customer shall
be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate maintenance and
care. Tektronix shall not be obligated to furnish service under this warranty a ) to repair damage resulting from attempts by personnel
other than Tektronix representatives to install, repair o r service the product; b) to repair damage resulting from improper use or
connection to incompatible equipment; c) to repair any damage or malfunction caused by the use of non-Tektronix supplies; or
d) to service a product that has been modied or integrated with other products when the effect of such modication or integration
increases the time or difculty of servicing the product.
THIS WARRANTY IS GIVEN BY TE KTR ONIX W ITH RESPEC T TO THE PROD UC T IN LIEU OF A NY OTHE R WARRANTIES,
EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDO RS DISCLAIM ANY IMPLIED WARRANTIES O F MERCHANTAB ILITY OR
FITNESS FOR A PARTICULAR PURPOSE . TE KTRONIX' RES P ONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS
IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO T HE CUSTOMER FOR BREACH OF THIS WARRANTY. TEKTRONIX
AND ITS VENDORS W ILL NOT BE LIABLE FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIA L DAMAGES
IRRESPECTIVE O F WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH
DAMAGES.
[W2 15AUG04]
Table of Content
s
Table of Contents
General safety summary .......................................................................................... ... ... ... ... ... .... . iii
Compliance I
nformation ............................................................................................................... v
EMC Compliance................................................................................................................. v
Safety Compliance............................................................................................................... vi
Environmen
tal Considerations................................................................................................. viii
Preface................................................................................................................................. ix
Probe Operating Information.......................................................................................................... 1
Connectin
g to the Instrument.................................................................................................... 2
Probe Controls ...................................................................................................................5
Functional Check.... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .......................................... .7
Accessori
es............................................................................................................................ 8
Connecting to the Circuit... ... ....... ....... ... ... . ... ... ... ....... ....... ... ... ... . ... ... ....... ....... ... ... ... . ... ... ....... .. 8
Options................................................................................................................................ 23
Operatin
g Basics ..................................................................................................................... 24
Operating Characteristics and Probing Techniques . ... ... ... ... ... ... ... .. . .. ... ... ... ... ... ... ... ... ....................... . 24
Specications .........................................................................................................................28
Warrant
ed Speci cations ....................................................................................................... 28
Typical Specications ........................................................................................................... 29
Nominal Specications.......................................................................................................... 31
Perfor
mance Graphs.. ........... ... ... ... .. ... ... ... ........... ........... ... ... ... ..... ... ... ... ........... ........... ... ... .. 32
Performance Verication ............................................................................................................. 36
Required Equipment . ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... ... .. 36
Test Pr
ocedures .................. ... ... ... ... ... ... ... ... .. . .. ... ... ... ... ... ... ... ... .................... ... ... ................ 37
Adjustments........................................................................................................................... 41
Equipment Required ........................................................................................................... 43
Adjus
tment Procedures ... .............. ... .............. ... ... ... ... .. ... ... ... ... ... .............. ... .............. ... ... ... .. 44
Troubleshooting..... ... . .. . ... ... ... ... ... ... ... ... ... ... ... ... ... ............................ ... ... ... ... ... ......................... 55
Host Instrument Firmware ...................................................................................................... 55
Erro
r Conditions ... ............................ ... ... ... ... ............................ ... ... ... ... ... ... ... ... ... ... ... ... ... . .. 55
Cleaning .. ... ... ... ... ... ... ... ... ................... ... ... ... ................... ... ... ... ... ... ... ... ... ... . ... ... ... ... ... ..... 56
Service........................................................................................................................... 56
Ind
ex
P5200A Series H igh Voltage Differential Probes Instruction Manual i
Table of C ontent
s
ii P5200A Series High Voltage Differential Probes Instruction M anual
General safety s
ummary
General safet
y summary
Review the following safety precautions to avoid injury and prevent damage to this product or any products connected to it.
To avoid potential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
While using this product, you may need to access other parts of a l arger system. Read the safety s ections of the other
component manuals for warnings and cautions related to operating the system.
To avoid re or personal injury
Use proper power cord. Use only the power cord specied for this product and certied for the country of use.
Connect and disconnect properly. Do not connect or disconnect probes or test leads while they are connected
to a voltag
e source.
Connect a
nd disconnect properly.
Connect the probe output to the measurement instrument before connecting the
probe to the circuit under test. Connect the probe reference lead to the circuit under test before connecting the probe
input. D isconnect the probe input and the probe reference lead from the circuit under test before disconnecting the probe
from the m
easurement instrument.
Ground t
he product.
This product is indirectly grounded through the grounding conductor of the mainframe power cord.
To avoid electric shock, the grounding conductor m ust be connected to earth ground. Before making connections to the input
or output terminals of the product, ensure that the product is properly grounded.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings and markings on the product. Consult the
produc
t manual for further ratings information before making connections to the product.
Do not a
pply a potential to any terminal, including the common terminal, that exceeds the maximum rating of that terminal.
Power d
isconnect.
The power cord disconnects the product from the power source. Do not block the power cord; it
must remain accessible to the user at all times.
Do not operate without covers. Do not operate this product with covers or panels removed.
Do not operate with suspected failures. If you suspect that there is damage to this product, have it inspected by
qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when power is present.
Use proper AC adapter. Use only the AC adapter specied for this product.
Do not operate in wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
P5200A Series H igh Voltage Differential Probes Instruction Manual iii
General safety s
ummary
Termsinthismanual
These terms may appear in this manual:
WARNING. Warning s tatements identify conditions or practices that could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property.
Symbols and terms on the product
These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the marking.
WARNING indicates an injury hazard not immediately accessible as you r ead the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
iv P5200A Series High Voltage Differential Probes Instruction M anual
Compliance Info
rmation
Compliance In
formation
This section lists the EMC (electromagnetic compliance), safety, and environmental standards with which the instrument
complies.
EMC Compliance
EC Declaration of Conformity EMC (Applies Only to the P5200A Probe)
Meets intent of Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following
specications as listed in the Ofcial Journal of the European Communities:
EN 61326-1:2006, EN 61326-2-1:2006. EMC requirements for electrical equipment for measurement, control, and
laboratory use.
123
CISPR 11:2003. Radiated and conducted emissions, Group 1, Class A
IEC 61000-4-2:2001. Electrostatic discharge immunity
IEC 61000-4-3:2002. RF electromagnetic eld immunity
IEC 61000-4-4:2004. Electrical fast transient/burst immunity
IEC 61000-4-5:2001. Power line surge immunity
IEC 61000-4-6:2003. Conducted RF immunity
IEC 61000
-4-11:2004. Voltage dips and interruptions immunity
4
EN 61000-3-2:2006. AC power line harmonic emissions
EN 61000-3-3:1995. Voltage changes, uctuations, and icker
European contact.
Tektronix UK, Ltd.
Western Peninsula
Western Road
Bracknell, RG12 1RF
United Kingdom
1
This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
2
Emissions which exceed the levels required by this standard may occur whe
n this equipment is connected to a test object.
3
To ensure compliance with the EMC standards listed here, high quality shielded interface cables should be used.
4
Performance Criterion C applied at the 70%/25 cycle Voltage-Dip and the 0%/250 cycle Voltage-Interruption test levels
(IEC 61000-4-11).
Australia / New Zealand Declaration of Conformity EMC
Complies with the EMC provision of the Radiocommunications Act per the following standard, in accordance with ACMA:
CISPR 11:2003. Radiated and Conducted Emissions, Group 1, Class A, in accordance with EN 61326-1:2006 and
EN 61326-2-1:2006.
P5200A Series H igh Voltage Differential Probes Instruction Manual v
Compliance Info
rmation
Safety Compliance
Equipment Type
Differential Voltage Probe
EC Declarati
on of Conformity Low Voltage
Compliance was demonstrated to the following specication as listed in the Ofcial Journal of the European Communities:
Low Voltage Directive 2006/95/EC.
EN 61010-031/A1:2008. Safety requirements for electrical equipment for measurement, control and laboratory use - Part
031: Safety requirements for handheld probe assemblies for electrical measurement and test.
Canadian Certication
CAN/CSA-C22.2 No. 61010-031-07/A1:2010, 1st Edition. Safety requirements for handheld probe assemblies for electrical
measurem
ent and test.
Additional Compliances
IEC 61010-031/A1:2008. Safety requirements for electrical equipment for measurement, control and laboratory use - Part
031: Safety requirements for handheld probe assemblies for electrical measurement and test.
Pollution Degree Description
A measure of the contaminants that could occur in the envir onment around and within a product. Typically the internal
envir
onment inside a product is considered to be the same as the external. Products should b e used only in the environment
for which they are rated.
Pollution Degree 1. No pollution or only dry, nonconductive pollution occurs. Products in this category a re generally
encapsulated, hermetically sealed, or located in clean rooms.
Pollution Degree 2. Normally only dry, nonconductive pollution occurs. Occasionally a temporary conductivity that is
caus
ed by condensation must be expected. This location is a typical ofce/home environment. Temporary condensation
occurs only when the product is out of service.
Pollution Degree 3. Conductive pollution, or dry, nonconductive pollution that becomes conductive due to condensation.
These are sheltered locations where neither temperature nor humidity is controlled. The area is protected from direct
sun
shine, rain, or direct wind.
Pol
lution Degree 4. Pollution that generates persistent conductivity through conductive dust, rain, or snow. Typical
outdoor locations.
Pollution Degree
Pollution Degree 2 (as dened in IEC 61010-1). Note: Rated for indoor use only.
vi P5200A Series High Voltage Differential Probes Instruction Manual
Compliance Info
rmation
Installation & Measurement (Overvoltage) Category Descriptions
Terminals on this product may have different installation or measurement (overvoltage) category designations. The
installation and measurement categories are:
Measurement Category IV. For measurements performed at the source of low-voltage installation.
Measurement Category III. For measurements performed in the building installation.
Measurement Category II. For measurements performed on circuits directly connected to the low-voltage installation.
Measurement Category I. For measurements perform ed on circuits not directly connected to MAINS.
Overvoltage Category (AC Adap ter)
Overvolta
ge Category II (as dened in IEC 61010-1)
P5200A Series H igh Voltage Differential Probes Instruction Manual vii
Compliance Info
rmation
Environmental Considerations
This section provides information about the environmental impact of the product.
Product End-of-Life Handling
Observe the following guidelines when recycling an instrument or component:
Equipment recycling. Production of this equipment required the extraction and use of natural resources. The equipment
may contain s ubstances that could be harmful to the environment or human health if improperly handled at the product’s
end of life. To avoid release of such substances into the environment and to reduce the use of natural resources, we
encourage you to recycle this product in an appropriate system that will ensure that most of the materials are reused
or recycled appropriately.
This symbol indicates that this product complies with the applicable European Union r equirements according
to Directives 2002/96/EC and 2006/66/EC on waste electrical and electronic equipment (WE EE) and
batteries. For information about recycling options, c heck the Support/Service section of the Tektronix Web
site (www.tek.com).
Restriction of Hazardous Substances
This product has been classied as Monitoring and Control equipment, and is outside the scope of the 2002/95/EC RoHS
Directi
ve.
viii P5200A Series High Voltage Differential Probes Instruction Manual
Preface
Preface
This document provides operating information and specications for the Tektronix P5200A Series high voltage differential
probes. The probes share similar functions, properties, and operating procedures, and are discussed in the rstpartofthe
manual. The specications and performance verication procedures for the probes follow.
WARNING. Only use the accessories that are designed for your probe and that are rated at or above the voltages you are
measuring. (See Table i on page x.) (See Table ii on page xiii.)
Name Bandwidth Attenuation Oscilloscope interface
P5200A 50 MHz
50X/500X BNC
P5202A 100 MHz
20X/200X TekProbe BNC-Level 2
P5205A 100 MHz
50X/500X TekProbe BNC-Level 2
P5210A 50 MHz
100X/1000X TekProbe BNC-Level 2
P5200A Series H igh Voltage Differential Probes Instruction Manual ix
Preface
Figure i: P5200A High Voltage Differential Probe with accessories
Table i: P5200A, P5202A & P5205A probe standard accessories derating table
Combined probe and accessory common-mode voltage
and in
put voltag e -t o- earth ratings
Acce
ssory
P520
2A
P520
0A & P5205A
P521
0A
1
Extender leads
450
VCATI
300 V C AT II
100
0VCATII
600 V CAT III
230
0VCATI
1000 V CAT III
Hook clips (AC280-FL) 450 V CAT I
300 V C AT II
1000 V C AT II
600 V CAT III
1000 V CAT I
1000 V CAT III
Pincer clips (AC283-FL) 450 V C AT I
30
0VCATII
1000 V C AT II
60
0 V CAT III
1000 V CAT I
10
00 V CAT III
A
lligator clips (AC285-FL)
4
50VCATI
300 V C AT II
1
000VCATII
600 V CAT III
1
000VCATI
1000 V CAT III
1
The P5200A, P5202A & P5205A standard accessories can also be used with the P5210A probe, but only at the reduced voltage levels listed here.
x P5200A Series High Voltage Differential Probes Instruction Manual
Preface
Figure ii: P5202A High Voltage Differential Probe with accessories
P5200A Series H igh Voltage Differential Probes Instruction Manual xi
Preface
Figure iii: P5205A High Voltage Differential P robe with accessories
xii P5200A Series High Voltage Differential Probes Instruction Manual
Preface
Figure iv: P5210A High Voltage Differential Probe with accessories
Table ii: P5210A probe standard accessories derating table
Combined probe and accessory common-mode voltage
and in
put voltage-to-earth ratings
Acce
ssory
P520
2A
1
P520
0A & P5205A
1
P521
0A
Extender leads
450
VCATI
300 V C AT II
100
0VCATII
600 V CAT II
230
0VCATI
1000 V CAT III
Test probe (TAT P) 450 V C AT I
300 V C AT II
1000 V C AT II
600 V CAT II
1000 V CAT I
1000 V CAT II
Small hook tip (TASH) 450 V CAT I
30
0VCATII
1000 V C AT II
60
0VCATII
2300 V CAT I
10
00 V CAT II
L
arge hook tip (TALH)
4
50VCATI
300 V C AT II
1
000VCATII
600 V CAT II
2
300 V CAT I
1000 V CAT II
1
The P5210A standard accessories can be used with these probes at the reduce d voltage levels listed in this table.
P5200A Series H igh Voltage Differential Probes Instruction Manual xiii
Probe Operating
Information
Probe Operati
ng Information
The P5200A Series probes share many common features, including connections to the circuit, compensation box buttons,
and operating basics. The probes connect to the host oscilloscope through one of two probe-to-oscilloscope interfaces:
BNC this connection is a shielded 50 coaxial cable with an outer ground connection and center signal pin. The BNC
interface is used on the P5200A probe, which allows you to c onnect directly to most g round-re ferenced oscilloscopes.
TekProbe
Level 2 this interface adds probe communications with the oscilloscope to accomplish calibrated offset at
the probe tip. This interface is used on the P5202A, P5205A, and P5210A probes, and many Tektronix oscilloscopes.
These probes can also be used with Tektronix oscilloscopes that use the TekVP I-interface, by connecting through an
optional
TPA-BNC Adapter.
TPA-BNC Adapter
The TPA-BNC Adapter is an optional accessory that enables you to use existing TekProbe-interface probes with oscilloscopes
that feature the TekVPI probe interface. The adapter recognizes TekProbe-interface probes and supplies the necessary
power, serial c ommunication, and offset control as used by the connected TekProbe product accessory.
WARNING. To reduce risk of shock or re, do not exceed the ratings of the TPA-BNC adapter; it is not intended to be
connected to voltages above 30 VAC, 42 Vpk, or 60 V DC. For BNC probes, connect the probe directly to the oscilloscope.
1 P5200A Series High Voltage Differential Probes Instruction Manual
/