Omega RD200 and RD2800 Series Owner's manual

Type
Owner's manual
Basic Manual for
RD200
,
RD2800 Series Recorder
1 Back Terminal connection
Power Su
pp
l
y
2. Sensor Connection
Thermocou
p
le
DC Volta
g
e and Current**
***4/20mA connect 250ohm
resister to the terminal.
RTD
(
PT100ohm
)
***Make sure all terminals are ti
g
htl
y
connected.
3. Date
,
time
,
Year
(
Clock Settin
g)
(
See manual under Time settin
g)
3A. Push SHIFT + CLOC
K
ke
at same time
You will see Dis
p
la
y
-Ori
g
inal set at factor
y
time
)
C
.1.998 01.01 00:00
Yea
r
Month Date Time
3B. Push SET ke
y
You will see curser under first di
g
it of the month
Use Arrow ke
y
to move curse
r
3C. Chan
g
e each di
g
it to toda
y
s month
,
date
,
and time
Exam
p
le Jul
y
4
,
2007 at 4:15PM
C
.2.007 07.04 16:15
3D. Push ENTR
Y
Ke
y
4. In
p
ut
(
Ran
g
e
)
Settin
g
(
See manual under Ran
g
e/
p
rintin
g
ran
g
e
)
4A. Choose In
p
ut and ran
g
e close to
y
our in
p
ut t
yp
e
Exam
p
le:K thermocou
p
le for 0/1000 F. Choose No. 22
4B. Push SHIFT + RANGE Ke
y
at same time
You will see Dis
p
la
y
of ori
g
inal factor
y
settin
g
R
.01:0700.00 5.00
Ran
g
e
Tem
p
erature com
p
ensation
(
CJ
)
Thermocou
p
le 1 other 0
Ran
g
e Number form above chart
In
p
ut channel number: RD2800= 2 di
g
it RD200= 1 di
g
it
4C. Push SET Ke
y
You will see Curser under Channel number.
Use Arrow ke
y
to move curse
r
4CA. Exam
p
le 1. settin
g
of Thermocou
p
le
(
K t
yp
e 0 to 2000 F
)
R
.01:2210 200.0
Push SPAC ke
y
to remove decimal
Thermocou
p
le alwa
y
s 1
Now Push ENTR
Y
ke
y
Dis
p
la
y
will chan
g
e to channel 02 and factor
y
settin
g
.
To chan
g
e the in
p
ut
,
re
p
eat of ste
p
1.
(
Re
p
eat 4CA
)
Alwa
y
s
p
ush ENTR
Y
ke
y
after each channel
p
ro
g
ram.
If all channel in
p
ut is same as Channel 1
,
then
,
after
p
ro
g
ramin
g
Channel 1 and
p
ush ENTR
Y
ke
y,
p
erform COPY
4CB. Push SHIFT + COPY ke
y
at same time
Curser under referance channel: move curser b
y
Arrow ke
y
S
.C o:P
Y
01=02. 12
Co
p
ied chnnel 1 settin
g
to channel 2 to 12
Push: Entr
y
Ke
y
dis
p
la
y
g
oes back to:
R
.12:2210 200.0
Show last channel
p
ro
g
ramed.
When all channel are
p
ro
g
ramed: Push SHIFT + END ke
y
***ENTRY Ke
y
onl
y
tos tore the chan
g
e
,
Shift + End ke
y
to chan
g
e
the
p
ro
g
ram and store.
4CD: Defult tem
p
erature is alwa
y
s de
g
rees C. If De
g
rees F is needed:
Push SHIFT + F ke
y
for lon
g
er then 3 second
.d EG.C d EG.F d EG.C
Push SE
T
Ke
y
Youw will see curser under dEG.C
To chan
g
e to De
g
. F
,
Push U
p
/down Arrow ke
y
.d EG.C d EG.F d EG.F
Push ENTR
Y
ke
y
(
no need to
p
ush SHIFT + END ke
y)
Push Dis
p
la
y
ke
y
to
g
o back to
p
ro
g
am dis
p
la
y
.
4CA
A
. mV
,
Volta
g
e and Current
(
4/20mA
)
in
p
ut
(
4/20mA in
p
ut
,
make sure 250 ohm resister is connected to the terminal
)
Push SHIFT + RANGE Ke
y
at same time
R
.01:0700.00 5.00
This is DC volta
g
e in
p
ut
Exam
p
le 4/20mA in
p
ut
Push Set Ke
y
You will see curser under channel nummber
,
move curser b
y
Arrow ke
y
. 4/20mA with 250ohm Resiter is 1-5V
R
.01:0701.00 5.00
Ran
g
e of 1 to 5V onl
y
use
CJ is alwa
y
s 0 for volta
g
e
,
current and RTD
Ran
g
e number form the chart -5 to 5 V
Push ENTR
Y
Ke
y
**After Entr
y
ke
y,
Channel chan
g
e to next channel number
**If not same in
p
ut
,
then enter the in
p
ut number and ran
g
e
(
4CA
)
**If all channel to be same in
p
ut
,
p
erform Co
py
function as above.
(
4CB
)
After all channel are
p
ro
g
ram
,
Push SHIFT + END ke
y
5. Scale
(
Min. and Max readin
g
of in
p
ut ran
g
e
)
(
see manual under Scale
)
**Thermocou
p
le Scale is noramll
y
same as Ran
g
e so no need to
p
erform
Scale
p
ro
g
rammin
g
.
5A. Exam
p
le 4/20mA in
p
ut with scale of 0 to 100
Push SHIFT + SCAL
E
ke
at same time
(
4/20mA was set as 1 to 5V
)
You will see dis
p
la
y
:
S
.1 5
Push SET ke
y
(
You will see curser under 1
)
Chan
g
e to followin
g,
movin
g
curser usin
g
Arrow ke
y
S
.0 100
Push ENTRY Ke
y
Channel number chan
g
e to next channel
**If next channel is different then
p
revious
,
p
erform chan
g
e then
p
ush ENTRY
***If all cahnnel Scale to be same
,
Perform Co
py
(
4CB
)
function
When all channela re
p
ro
g
ramed
,
Push SHIFT + END ke
at same time.
6. Chart S
p
eed
Factor
y
set chart S
p
eed is 20mm
p
er hour
To chan
g
e chart S
p
eed: Push SHIFT + CHAR
T
ke
at same time
H
CS 0020
Push SET ke
y
Chart S
p
eed
You will see curser under chart s
p
eed
Exam
p
le: chan
g
e to 2 inches
(
50mm
)
p
er hour
H
CS 0050
Push Entr
y
Ke
y
. This will store the chan
g
e
(
no need to
p
ush END ke
y
basic
p
ro
g
rammin