Tektronix RSA5106B Declassification And Security Instructions

Type
Declassification And Security Instructions

This manual is also suitable for

RSA5100B Series
Real-Time Signal Analyzers
Declassication and Security
ZZZ
Instructions
xx
www.tektronix.com
*P077090200*
077-0902-00
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries or suppliers, and are protected by
national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supersedes that in all previously published
material. S pecications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
For product information, sales, service, and technical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to nd contacts in your area.
Table of Contents
Preface .............................................................................................................................................................. iii
Clear and Sanitize Procedures. .. .. ................... .. ..................... .. .. .. .. .. ................... .. .. ................... .. ................................ 1
Memory Devices............................................................................................................................................... 1
Media and Data Export Devices ............................................................................................................................. 7
Troubleshooting.................................................................................................................................................... 11
How to Clear or Sanitize a Non-Functional Instrument .................................................................................................. 11
Change Log ........................................................................................................................................................ 12
RSA5100B Series Declassication and Security Instructions i
Table of Contents
ii RSA5100B Series Declassication and Security Instructions
Preface
This document helps customers with data security concerns to sanitize or remove memory devices from the RSA5100B
Series Real-Time Signal Analyzers.
These products have data storage (memory) devices and data output devices (USB ports). These instructions tell how to
clear or sanitize the memory devices and disable the data output devices. The instructions also tell how to declassify an
instrument that is not functioning.
Reference
The procedures in this document are written to m eet the requirements specied in:
NISPOM, DoD 5220.22–M, Chapter 8
ISFO Process Manual for Certication & Accreditation of Classied Systems under NISPOM
Products
The following Tektronix products are covered by this document:
RSA5103B
RSA5106B
RSA5115B
RSA5126B
Terms
The following terms may be used in this document:
Clear.
This removes data on media/memory before reusing it in a secured area. All reusable memory is cleared to
deny access to previously stored information by standard means of access.
Erase.
This is equivalent to clear.
Instrument Declassication.
A term that refers to procedures that must be undertaken before an instrument is
removed from a secure environment. Declassication procedures include memory sanitization and memory removal,
and sometimes both.
RSA5100B Series Declassication and Security Instructions iii
Preface
Media storage/data export device.
Any of several devices that can be used to store or export d ata from the
instrument, such as a USB port.
Power off.
Some instruments have a “Standby” mode, in which power is still supplied to the instrument. For the
purpose of clearing data, putting the instrument in Standby mode does not qualify as powering off. For these products,
you will need to either press a rear-panel OFF switch or remove the power source from the instrument.
Remove.
This is a physical means to clear the data by removing the memory device from the instrument. Instructions
are available in the product Service Manual.
Sanitize.
This eradicates the data from media/memory so that the data cannot be recovered by other means or
technology. This is typically used when the device will be moved (temporarily or permanently) from a secured
area to a non-secured area.
Scrub.
This is equivalent to sanitize.
User Accessible.
User is able to directly retrieve the memory device contents.
User-modiable.
The user can write to the memory device during normal instrument operation, using the instrument
interface or remote control.
Vo latile memory.
Data is lost when the instrument is powered off.
Nonvolatile memory.
Data is retained when the instrument is powered off.
iv RSA5100B Series Declassication and Security Instructions
Clear and Sanitize Procedures
Memory Devices
The following tables list the volatile and nonvolatile memory devices in the standard instrument and listed options.
Detailed procedures to clear or sanitize these de vices, if any, are shown following each table.
Terminology
The following terms are used in the tables in this section:
User data Describes the type of information stored in the device. Refers to waveforms or other measurement data
representing signals connected to the instrument by users.
User settings Describes the type of information stored in the device. Refers to instrument settings that can be
changed by the user.
Both Describes the type of information stored in the device. It means that both user data and user settings are
stored in the device.
None Describes the type of information stored in the device. It means that neither user data nor user settings
are stored in the device.
Directly Describes how data is modied. It means that the user can modify the data.
Indirectly Describes how data is modied. It means that the instrument system resources modify the data and that
the user cannot modify the data.
RSA5100B Series Declassication and Security Instructions 1
Clear and Sanitize Procedures
Table1:Volatilememorydevices
Type and
min. size Function
Type of
user info
stored
Backed-up
by battery
Method of
modication
Data Input
method Location
User
accessible To clear Process to sanitize
SDRAM
1GB(std)
4GB
(Opt. 53)
Holds active
acquisition data
User data No Indirectly Firmware
operations
Digital
Interface
board
No
Remove power from the
instrument for at least
20 seconds.
Remove power from the
instrument for at least
20 seconds.
SDRAM
4GB)
Microprocessor
system memory
User data,
user settings
No Direct Written by
processor
system
Slot-PC
board
Yes
Remove power from the
instrument for at least
20 seconds.
Remove power from the
instrument for at least
20 seconds.
SRAM 2 MB Power PC
system memory
None No Indirect Firmware
operations
Digital
Interface
board
No
Remove power from the
instrument for at least
20 seconds.
Remove power from the
instrument for at least
20 seconds.
ZBT SRAM
2MB
Holds active
DPX acquisition
data
User data No Indirect Firmware
operations
DPX board No
Remove power from the
instrument for at least
20 seconds.
Remove power from the
instrument for at least
20 seconds.
SRAM, 1 MB
Unused None No None None DPX board No
Remove power from the
instrument for at least
20 seconds.
Remove power from the
instrument for at least
20 seconds.
QDR SRAM
20 MB
(Option 200)
Holds active
DPX acquisition
data
User data No Indirect Firmware
operations
DPX board No
Remove power from the
instrument for at least
20 seconds.
Remove power from the
instrument for at least
20 seconds.
2 RSA5100B Series Declassication and Security Instructions
Clear and Sanitize Procedures
Table 2: Nonvolatile memory devices
Type and
min. size Function
Type of
user info
stored
Method of
modication
Data Input
method Location
User
accessible To clear To sanitize
Fixed Hard
Drive (Part
of Option 59)
Holds instrument
operating system
and application
software. Holds all
user-storable data
such as waveforms,
measurement
results, and
instrument settings.
User data,
user settings
Indirect Firmware
operations, user
input
Internal Yes
Acquire a full memory length
acquisition of noise (no signal
attached to the input). Erase
the hard drive with commercial
erasure software. Reinstall
Microsoft Windows using the OS
Restore procedure in the Quick
Start User Manual. Reinstall
the instrument software using
the supplied recovery disk. See
Clearing Hard Disk Drives
.
Removetheharddrive.Store
the removed hard drive in
a secure area or destroy it.
(See page 5,
FixedHardDrive
Removal Procedure
.) When
the hard drive is removed,
no user data remains in the
instrument.
Removable
Hard Drive
(Option 56)
Holds instrument
operating system
and application
software. Holds all
user-storable data
such as waveforms,
measurement
results, and
instrument settings.
User data,
user settings
Indirect Firmware
operations, user
input
Front panel.
Additional
drives can be
purchased from
Tektronix, Inc.
Yes
Acquire a full memory length
acquisition of noise (no signal
attached to the input). Erase
the hard drive with commercial
erasure software. Reinstall
Microsoft Windows using the OS
Restore procedure in the Quick
Start User Manual. Reinstall
the instrument software using
the supplied recovery disk. See
Clearing Hard Disk Drives
.
Removetheharddrive.Store
the removed hard drive in a
secure area or destroy it. (See
page 6,
Removable Hard Drive
Procedure
.) When the hard
drive is removed, no user data
remains in the instrument.
Flash
16 Mbytes
Holds instrument
calibration data,
serial number, and
Option key.
None Indirect Firmware
operations
Digital Interface
board
No Not applicable, does not contain
user data or settings. Clearing
would disable instrument
functionality.
Not applicable, does not
contain user data or settings.
Sanitizing would disable
instrument functionality.
EEPROM
16Mbit
Holds data buffer
rmware
None None Factory
conguration
Acquisition
Board
No Not applicable, does not contain
user data or settings. Clearing
would disable instrument
functionality.
Not applicable, does not
contain user data or settings.
Clearing would disable
instrument functionality.
EEPROM
16Mbit
Holds board
interface rmware.
None None Factory
conguration
Acquisition
Board
No Not applicable, does not contain
user data or settings. Clearing
would disable instrument
functionality.
Not applicable, does not
contain user data or settings.
Clearing would disable
instrument functionality.
EEPROM,
8KB
Unused (for
future software
enhancements)
None None None
Digital Interface
board
No Not applicable. Does not contain
user data or settings. Not used
by the system.
Not applicable, does not
contain user data or settings.
Not used by the system.
RSA5100B Series Declassication and Security Instructions 3
Clear and Sanitize Procedures
Table 2: Nonvolatile memory devices (cont.)
Type and
min. size Function
Type of
user info
stored
Method of
modication
Data Input
method Location
User
accessible To clear To sanitize
EEPROM
16Mbit
(Option 65
& 66)
Holds option board
calibration data.
None None Factory
conguration
Zero-Span
Output/Digital
IQ Output
option board.
No Not applicable, does not contain
user data or settings. Clearing
would disable Zero Span Output
functionality
Not applicable, does not
contain user data or settings.
Clearing would disable Zero
Span Output functionality.
EEPROM,
1Kb
PCI conguration
PLX9056
None None Factory
conguration
Digital Interface
board
No Not applicable, does not contain
user data or settings. Clearing
would disable instrument
functionality.
Not applicable, does not
contain user data or settings.
Sanitizing would disable
instrument functionality.
EEPROM,
1Kb
PCIe conguration
–PEX8311
None None Factory
conguration
Digital Interface
board
No Not applicable, does not contain
user data or settings. Clearing
would disable instrument
functionality.
Not applicable, does not
contain user data or settings.
Sanitizing would disable
instrument functionality.
Flash, 1 Mb
Unused (for
future software
enhancements)
None None None RFI board No Not applicable, does not contain
user data or settings. Not used
by the system
Not applicable, does not
contain user data or settings.
Not used by the system
Flash,
32 Mb
Holds calibration
data
None None Factory
conguration
LO board
No Not applicable, does not contain
user data or settings.
Not applicable, does not
contain user data or settings.
RSA5115B
and
RSA5126B
only; Flash,
32 Mb
Holds calibration
data
None None Factory
conguration
MFC LO board
No Not applicable, does not contain
user data or settings. Clearing
would disable instrument
functionality.
Not applicable, does not
contain user data or settings.
Clearing would disable
instrument functionality.
RSA5126B
only; Flash,
32 Mb
Holds calibration
data
None None Factory
conguration
26 GHz module
No Not applicable, does not contain
user data or settings. Clearing
would disable instrument
functionality.
Not applicable, does not
contain user data or settings.
Clearing would disable
instrument functionality.
NOTE.
The following procedure covers only Tektronix installed software.
4 RSA5100B Series Declassication and Security Instructions
Clear and Sanitize Procedures
Clearing Hard Disk Drives
If your organization's security protocols allow the use of software to purge or sanitize hard drives, you can use commercial
software to erase free space on the hard drive before sending the instrument out for upgrades or repair. Follow the
instructions that come with the software to ensure that the erasure of sensitive data from the hard drive complies w ith your
organization's security protocols.
Sequence for clearing s ecure data from a hard drive:
1.
Delete all stored signals and instrument settings les (all les stored with a
tiq
,
csv
,or
Setup
extension).
2.
Acquire several full memory-length acquisitions of noise (no signal attached to the input).
Without Option 53, acquire a full-length acquisition 4 times.
With Option 53, acquire a full-length acquisition 16 times.
3.
Close the RSA5100B application and erase the free space on the hard disk using commercial software.
Fixed Hard Drive Removal
Procedure
The following procedure applies to instruments with Option 59 (Fixed Hard Disk Drive) installed. This procedure is a n
overview of the removal process. For detailed removal and replacement procedures, see the RSA5100B Series Real-Time
Signal Analyzers Service Manual, Tektronix part number 077-0903-XX.
WARNING.
Only qualied service personnel should perform this procedure. Read the Service Safety Summary and the
General Safety Summary in the RSA5100B Series Real-Time Signal Analyzers Service Manual before performing this
procedure.
WARNING.
Before performing this procedure, disconnect the power cord from the line voltage source. Failure to do so
could cause serious injury or death.
CAUTION.
To avoid damaging the hard drive, perform the following procedure in a static-safe environment with proper
electro-static discharge controls in place.
RSA5100B Series Declassication and Security Instructions 5
Clear and Sanitize Procedures
1.
Remove the front cover (if installed), the front panel trim, and the cabinet top.
2.
Remove the top EMI cover.
3.
Remove the hard drive and support bracket.
4.
Unplug the SATA cable from the hard drive.
5.
Remove the hard drive from the support bracket.
6.
Reinstall the support bracket.
7.
Reinstall the top EMI cover.
8.
Reinstall the cabinet top, the front panel trim, and front cover (if removed).
Removable Hard Drive
Procedure
The following procedure applies to instruments with Option 56 (Removable Hard Disk Drive) installed. For detailed
information on removing parts from the instrument, see the RSA5100B Series Real-Time Signal Analyzers Service Manual,
Tektronix part number 077-0903-XX.
WARNING.
Before doing this procedure, disconnect the power cord from the line voltage source. Failure to do so could
cause serious injury or death.
CAUTION.
To avoid damaging the hard drive, perform the following procedure in a static-safe environment with proper
electro-static discharge controls in place.
1.
Loosen the thumbscrews on the removable drive panel.
2.
Pull the removable drive out from the instrument.
6 RSA5100B Series Declassication and Security Instructions
Clear and Sanitize Procedures
Media and Data Export Devices
Table 3: Media and data export devices
Type and min.
size
Function
Method of
modication
Data Input method Location Process to disable
USB device ports Supports removable USB
ash drive. User storage of
reference waveforms, screen
images, and instrument
setups.
Directly Directly by system resources Front and rear panels
Remove all USB memory devices. USB devices can be
formatted, stored in a secure area, or destroyed.
The USB ports can be disabled. .(See page 7,
Disabling
the USB Ports
.)
You can set a BIOS password to prevent changes to the
BIOS. (See page 8,
Setting the BIOS Password
.)
LAN Ethernet
Connection
Supports remote control and
data transfer
Directly
System resources
Rear panel
Disconnect from Network cable
GPIB connector Supports remote control and
data transfer
Directly
System resources
Rear panel
N/A
Disabling the USB Ports
Use the following procedure to disable the front and rear panel USB ports.
CAUTION.
Disabling the USB ports will disable the front panel and touch screen. When the USB ports are disabled, no
mouse input is accepted. The PS2 connector on the rear panel does not support a mouse. When the USB ports are disabled,
the only way to control the instrument is through the GPIB/LAN programmatic interface.
CAUTION.
To avoid disabling the instrument, be sure to perform this procedure exactly as stated. A miscongured BIOS
can make the instrument unusable and require that it be returned to Tektronix for servicing.
1.
Reboot instrument and press the
Delete
key to access the BIOS settings screen.
2.
Select
Integrated Peripherals
.
3.
Select
USB Device Setting
.
RSA5100B Series Declassication and Security Instructions 7
Clear and Sanitize Procedures
4.
Change
USB 1.0 Controller
from
Enable
to
Disable
. Change
USB 2.0 Controller
from
Enable
to
Disable
.
5.
Save the BIOS settings and exit.
Setting the BIOS Password
Use the following procedure to set a BIOS password that will prevent others from changing the BIOS settings.
CAUTION.
To avoid disabling the instrument, be sure to perform this procedure exactly as stated. A miscongured BIOS
can make the instrument unusable and require that it be returned to Tektronix for servicing.
1.
Reboot the instrument and press the
Delete
key to access the BIOS settings screen.
2.
Select
Set Password
.
3.
Choose a BIOS password.
4.
Conrm the new BIOS password.
5.
Save the BIOS settings and exit.
Enabling the USB Ports
Use the following procedure to enable the front and rear panel USB ports.
CAUTION.
To avoid disabling the instrument, be sure to perform this procedure exactly as stated. A miscongured BIOS
can make the instrument unusable and require that it be returned to Tektronix for servicing.
1.
Reboot instrument and press the
Delete
key to access the BIOS settings screen.
2.
Select
Integrated Peripherals
.
3.
Select
USB Device Settings
.
8 RSA5100B Series Declassication and Security Instructions
Clear and Sanitize Procedures
4.
Change
USB 1.0 Controller
from
Disable
to
Enable
. Change
USB 2.0 Controller
from
Disable
to
Enable
.
5.
Save the BIOS settings and exit.
Disabling the BIOS
Password
Use the following procedure to disable the BIOS password.
CAUTION.
To avoid disabling the instrument, be sure to perform this procedure exactly as stated. A miscongured BIOS
can make the instrument unusable and require that it be returned to Tektronix for servicing.
1.
Reboot instrument and press the
Delete
key to access the BIOS settings screen.
2.
Enter the current BIOS password.
3.
Select
Set Password
.
4.
Press the
Enter
key.
5.
Press the
Enter
key again to conrm disabling the BIOS password.
6.
Save the BIOS settings and exit.
RSA5100B Series Declassication and Security Instructions 9
Clear and Sanitize Procedures
10 RSA5100B Series Declassication and Security Instructions
Troubleshooting
How to Clear or Sanitize a Non-Functional Instrument
If your instrument is not functioning, perform the following actions and return the instrument to Tektronix for repair.
Describe the initial problem with the product. Tektronix will install replacement parts and then repair and return the
instrument.
Hard Disk Drives
Remove the internal hard disk (Option 59) or removable hard disk (Option 56) using the procedures in this document.
External Memo ry Devices
Remove any USB ash drives or external hard drives from the instrument.
Refer to your company’s internal policies regarding handling or disposal of the external memory device.
Charges
Replacement of any missing hardware will be charged according to the rate at the time of replacement.
RSA5100B Series Declassication and Security Instructions 11
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17

Tektronix RSA5106B Declassification And Security Instructions

Type
Declassification And Security Instructions
This manual is also suitable for