• This device has been designed for single use only. Reusing this medical
device bears the risk of cross-patient contamination as medical devices
particularly those with long and small lumina, joints, and/or crevices
between components are difficult or impossible to clean once body fluids
or tissues with potential pyrogenic or microbial contamination have had
contact with the medical device for an indeterminable period of time.
The residue of biological material can promote the contamination of the
device with pyrogens or microorganisms which may lead to infectious
complications.
• For optimal balloon imaging, and to minimize potential radiation dose
attenuation, less than 10% contrast per fluid volume is recommended1.
• Verify indexed length of all treatment lumens prior to delivery of each
brachytherapy fraction. If treatment lumen lengths change by greater than
1mm, re-evaluate treatment planing prior to delivering a brachytherapy
fraction.
• The MammoSite ML should not be used to deliver radiation if the balloon
diameter is less than 3.5cm
• For patients who have or may have an allergic reaction to iodinated
materials, consider using a non-ionic contrast agent.
PRECAUTIONS
• The MammoSiteML RTS should be used only by physicians trained in
catheter placement, treatment planning and radiation delivery prior to the use
of the device.
• Do not use the contents if there is any evidence of damage to the package
or package seal that could compromise sterile integrity. Do not resterilize.
Single-use device.
• Exercise caution when handling the trocar.
• Keep catheter away from foreign materials at all times. Exercise extreme
caution when handling the MammoSiteML balloon prior to and during
implantation. Polyurethane materials are susceptible to damage by sharp
objects/instruments and excessive pulling or pushing.
• Avoid contact of the product with glove talc, lint, particulate matter, soaps,
oils, detergents or other surface contaminants. Caution should be used to
avoid contamination of the MammoSiteML device.
• Store product at ambient temperature.
• Use needle-free syringes (provided) when inflating or deflating the
MammoSiteML applicator. For proper inflation and deflation, ensure the
inflation lumen is not kinked or twisted.