12. Stop the enzyme reaction by quickly pipeing 100 µL of Stop
Solution into each well. It is important that the Stop Solution is
spread quickly and uniformly throughout the microwells to
completely inactivate the enzyme. Results must be read
immediately after the Stop Solution is added, or within one hour
if the microwell strips are stored at 2°C to 8°C in the dark.
13. Read absorbance of each microwell on a spectro-photometer
using 450 nm as primary wave length (you can use 620 nm as
reference wave length; 610 nm to 650 nm is acceptable). Blank the
plate reader according to the manufacturer's instructions by using
the blank wells. Determine the absorbance of both the samples
and the human IL-13 standards.
Calculation of results
• Calculate the average absorbance values for each set of duplicate
standards and samples. Duplicates should be within 20% of the
mean value.
• Create a standard curve by ploing the mean absorbance for each
standard concentration on the y-axis, against the human IL-13
concentration on the x-axis. Draw a best t curve through the
points of the graph (a 5-parameter curve t is recommended).
• To determine the concentration of human IL-13 for each sample,
rst calculate the mean absorbance value for the duplicate wells
of the sample, then extend a horizontal line from this point on the
y-axis to the standard curve. At the point of intersection, extend a
vertical line to the x-axis and read the corresponding human IL-13
concentration.
• If instructions in this protocol have been followed samples have
been diluted 1:2, the concentration read from the standard curve
must be multiplied by the dilution factor (x 2).
• Calculation of samples with a concentration exceeding that of
standard 1 may result in inaccurate, low human IL-13 levels
(Hook Eect). Such samples require further external predilution
according to expected human IL-13 values with Assay Buer (1x)
in order to precisely quantitate the actual human IL-13 level.
• Each testing facility should establish a control sample of known
human IL-13 concentration and run this additional control with
each assay. If the values obtained are not within the expected
range of this control, the assay results may be invalid.
A basic understanding of immunoassay development and technical
experience in ELISA performance are conditional for the successful
use of this Module Set.
The protocol provided is just a guideline. The type of substrate as well
as all other reagents not included in the Module Set may inuence the
test characteristics.
Human IL-13 module set characteristics
Specificity
The assay detects both natural and recombinant human IL-13. The
interference of circulating factors of the immune system was evaluated
by spiking these proteins at physiologically relevant concentrations
into a IL-13 positive serum. No cross-reactivity was detected.
Expected values
A panel of 24 sera samples from randomly selected apparently healthy
donors was were tested for human IL-13. The detected human IL-13
levels ranged between 0 and 44.4 pg/mL with a mean level of
8.2 pg/mL and a standard deviation of 12.1 pg/mL
Customer and technical support
Visit thermosher.com/support for the latest service and support
information.
• Worldwide contact telephone numbers
• Product support information
– Product FAQs
– Software, patches, and updates
– Training for many applications and instruments
• Order and web support
• Product documentation
– User guides, manuals, and protocols
–Certicates of Analysis
– Safety Data Sheets (SDSs; also known as MSDSs)
Note: For SDSs for reagents and chemicals from other
manufacturers, contact the manufacturer.
Limited product warranty
Life Technologies Corporation and/or its aliate(s) warrant their
products as set forth in the Life Technologies' General Terms and
Conditions of Sale at www.thermosher.com/us/en/home/global/
terms-and-conditions.html. If you have any questions, please contact
Life Technologies at www.thermosher.com/support.
Bender MedSystems GmbH | Campus Vienna Biocenter 2 | 1030 Vienna, Austria
For descriptions of symbols on product labels or product documents, go to thermofisher.com/symbols-definition.
The information in this guide is subject to change without notice.
DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, THERMO FISHER SCIENTIFIC INC. AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT,
PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Important Licensing Information: These products may be covered by one or more Limited Use Label Licenses. By use of these products, you accept the terms and conditions of all
applicable Limited Use Label Licenses.
©2019 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries unless otherwise specified. All other
trademarks are the property of their respective owners.
thermofisher.com/support | thermofisher.com/askaquestion
thermofisher.com
16 July 2019