15WIKA operating instructions, models CPP700-H and CPP1000-H
EN
14209601.01 06/2017 EN/DE/FR/ES/IT
3. Unscrew the fine adjustment valve
4
by turning it counter-clockwise, until it slightly
stops.
4. Carefully screw the pressure relief valve
2
by turning it clockwise, until the valve
closes securely.
5. By means of the pump handle
5
, pump the system until an initial pressure is
generated.
⇒
The handles can now hardly be pressed together. Depending on the volume of the
calibration circuit, a pressure of 200 ... 400 bar (2,900 ... 5,800 psi) is generated in
the “initial pressure generation” position.
6. Hold the handles
5
pressed together and press the toggle control
6
.
⇒
The pump handles
5
are now in the “high-pressure generation” position.
If too high an initial pressure has been generated, the pump handles
5
can
no longer be completely pressed together and the toggle control
6
can no
longer be actuated
In this case, drain the built-up initial pressure by unscrewing the pressure
relief valve
2
.
7. Now, by means of the pump handles
5
, pump until the desired test pressure is
reached, but up to a maximum of about 600 bar (8,700 psi).
⇒
A test pressure higher than about 600 bar (8,700 psi) is achieved only by turning
the fine adjustment valve
4
clockwise.
After increasing the pressure, the value displayed can drop slightly again
for about 30 seconds. The causes can be thermodynamic effects, the hose
connection and the sealings.
Readjust the test pressure using the fine adjustment valve
4
. If pressure
continues to drop, check the measuring circuit for leaks.
5.8 Reducing pressure
Do not disassemble the reference instrument or the test item until the
pressure relief valve
2
is open and there is no pressure in the test pump.
1. To reduce the pressure, turn the fine adjustment valve counterclockwise.
2. Carefully open the pressure relief valve
2
by turning it clockwise no more than once.
⇒
Pressure is reduced and there is no pressure on the hand test pump.
3. Dismount the reference pressure measuring instrument and/or the test item.
4. After the pressure has been relieved, close the pressure relief valve
2
by turning it
clockwise.
5. Commissioning, operation