RKI Instruments 30-0954RK-209D Owner's manual

Category
Carbon monoxide (CO) detectors
Type
Owner's manual
www.rkiinstruments.com
30-0954RK-209D Series Sample Draw
Aspirator Adapter
with M2As for CH
4
(%Volume) and H
2
S
Operator’s Manual
Part Number: 71-0339
Revision: D
Released: 8/28/20
2 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
WARNING
Read and understand this instruction manual before operating detec-
tor. Improper use of the detector could result in bodily harm or death.
Periodic calibration and maintenance of the detector is essential for
proper operation and correct readings. Please calibrate and maintain
this detector regularly! Frequency of calibration depends upon the
type of use you have and the sensor types.
Typical calibration frequencies for most applications are between 3
and 6 months, but can be required more often or less often based on
your usage.
30-0954RK-209D Series Sample Draw Aspirator Adapter • 3
Product Warranty
RKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials,
workmanship, and performance for a period of one year from the date of shipment from RKI
Instruments, Inc. Any parts found defective within that period will be repaired or replaced, at our
option, free of charge. Parts must be returned to RKI Instruments, Inc. for repair or replacement. This
warranty does not apply to those items which by their nature are subject to deterioration or
consumption in normal service, and which must be cleaned, repaired or replaced on a routine basis.
Examples of such items are:
Warranty is voided by abuse including mechanical damage, alteration, rough handling, or repair
procedures not in accordance with instruction manual. This warranty indicates the full extend of our
liability, and we are not responsible for removal or replacement costs, local repair costs, transportation
costs, or contingent expenses incurred without our prior approval.
THIS WARRANTY IS EXPRESSLY IN LIEU OF ANY AND ALL OTHER WARRANTIES
AND REPRESENTATIONS, EXPRESSED OR IMPLIED, AND ALL OTHER
OBLIGATIONS OR LIABILITIES ON THE PART OF RKI INSTRUMENTS, INC.
INCLUDING BUT NOT LIMITED TO, THE WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. IN NO EVENT SHALL RKI INSTRUMENTS,
INC. BE LIABLE FOR INDIRECT, INCIDENTAL OR CONSEQUENTIAL LOSS OR
DAMAGE OF ANY KIND CONNECTED WITH THE USE OF ITS PRODUCTS OR
FAILURE OF ITS PRODUCTS TO FUNCTION OR OPERATE PROPERLY.
This warranty covers instruments and parts sold to users only by authorized distributors, dealers and
representatives as appointed by RKI Instruments, Inc.
We do not assume indemnification for any accident or damage caused by the operation of this gas
monitor and our warranty is limited to the replacement of parts or our complete goods. Warranty
covers parts and labor performed at RKI Instruments, Inc. only, and does not cover field labor or
shipment of parts back to RKI.
a) Pump diaphragms and valves c) Batteries
b) Fuses d) Filter elements
4 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
Table of Contents
Chapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Chapter 2: Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Aspirator Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
M2A Series H
2
S and CH
4
Detector Heads . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
Chapter 3: Installation and Startup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
Making Tubing Connections. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Start Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
Chapter 4: Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Normal Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Information Screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
4-20 mA Signal Output Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Alarm Indications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
Chapter 5: Configuration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Viewing and Changing M2A Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
Chapter 6: Gas Type Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Selecting the Gas Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Chapter 7: Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Preventive Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
Calibration Frequency. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
Blowback Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Setting the Drain Fault Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
Replacing Components of the Aspirator Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
30-0954RK-209D Series Sample Draw Aspirator Adapter • 5
Chapter 8: RS-485 Modbus Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Wiring the M2A in a Modbus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Using the M2A in a 4-Wire Modbus System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Modbus Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
Supported Modbus Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 56
Parts List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
Appendix A: Control Button Quick Reference Guide. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
Appendix B: PLC and DCS Device Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Appendix C: Function Code 16 Registers. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
6 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
Chapter 1: Introduction
Overview
This manual describes the 30-0954RK-209D Series sample draw aspirator adapter supplied with an
M2A for 0 - 100 %volume CH
4
and an M2A for 0 - 100 ppm H
2
S. The 30-0954RK-209D may be used
with a variety of detector heads. This manual describes the %volume CH
4
/H
2
S M2A combination. It
also describes how to install, use, and maintain the adapter. A spare parts list at the end of this manual
lists replacement parts.
Specifications
Table 1 lists specifications for the Sample Draw Aspirator Adapter.
WARNING: When using the 30-0954RK-209D Series, you must follow the instructions and
warnings in this manual to assure proper and safe operation of the 30-0954RK-209D
Series and to minimize the risk of personal injury. Be sure to maintain and calibrate
the 30-0954RK-209D Series as described in this manual.
Table 1: Specifications
Target Gas/Detection Range •Methane (CH
4
): 0 - 100 % volume, 1% increments
Hydrogen Sulfide (H
2
S): 0 - 100 ppm, 1 ppm increments
Alarm Settings (Alarm 1/Alarm 2) Methane: 100 %volume / 100 %volume
Hydrogen Sulfide: 10 ppm / 50 ppm
Input Power 19 - 30 VDC
Signal Output 4 to 20 mA, 500 ohms impedance max
RS-485 Modbus
Maximum Compressed Air Supply
Pressure
140 PSI
Outlet Pressure to Aspirator 5 - 50 PSI adjustable (determined by required flow rate)
Recommended Sample Flow Rate 3 SCFH (standard cubic feet per hour)
Recommended H
2
S Detector Dilution
Flow Rate
0.6 SCFH (standard cubic feet per hour)
Low Flow Setpoint 0.6 SCFH (standard cubic feet per hour)
Accuracy IR CH
4
:
± 5% of reading or ± 2% of full scale (whichever is greater)
Hydrogen Sulfide
:
± 5% of reading or ± 2 ppm H
2
S (whichever is greater)
30-0954RK-209D Series Sample Draw Aspirator Adapter • 7
Chapter 2: Description
Overview
This version of the 30-0954RK-209D sample draw aspirator adapter consists of an aspirator panel with
a 65-2645RK-05 M2A series H
2
S detector and a 65-2658RK-CH4 M2A series methane detector
installed. Fresh air dilution is used for the sample going to the H
2
S detector. The sample draw aspirator
adapter uses compressed air flowing through a venturi to draw air into a sample chamber for each
detector. A flow diagram for the system is shown below.
Figure 1: Flow Diagram
Compressed
Air In
(Vacuum
Port)
Atmosphere
Air In
Particle
Filter
Dilution Flow Meter
H2S Detector
Sample
Inlet/Blowback
Valve
V
Drain
Fault
Switch
Flow
Fail
Switch
Sample Flow Meter
Calibration Gas
Fitting
Calibration
Valve
Water
Trap
CH4 Detector
Restrictor
AspiratorRegulator
Exhaust
8 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
Aspirator Panel
The sample draw adapter consists of fourteen major components (see Figure 2): the detector heads
(described in “M2A Series H
2
S and CH
4
Detector Heads” on page 11), regulator, aspirator, detector
adapter, detector chambers, sample flowmeter, H
2
S dilution flowmeter, particle filter, calibration
valve, inlet/blowback valve, water trap, low flow switch, drain fault switch, and low flow and drain
fault contact housing.
Figure 2: Component Location
Detector Heads
Figure 2 above shows the location of the two detector heads. Please see “M2A Series H
2
S and CH
4
Detector Heads” on page 11 for a complete description of the detector heads.
Adjustment
Screw
Calibration
Valve
Push Button
Calibration
Valve
Water Trap
Inlet/Blowback
Valve
H2S M2A
Detector
Head
H2S Dilution
FlowMeter
Low Flow
Switch
Sample Flow
Meter
Low Flow
and Drain
Fault
Contact
Housing
Particle Filter
(not factory
installed)
Calibration
Fitting
1/4 O.D. Tube
Compressed
Air Inlet Fitting
Inlet/Blowback Valve Button
1/4 O.D. TubeAspirator/
Sample Exhaust Fitting
IR CH4 M2A
Detector
Head
Drain
Fault
Switch
H2S Dilution
FlowMeter
Inlet Hose
Barb
Pressure AdjustmentKnob
Pressure Regulator
Aspirator
Pressure Gauge
1/4 O.D. TubeSample Inlet Fitting
30-0954RK-209D Series Sample Draw Aspirator Adapter • 9
Regulator
The regulator has an inlet port on its left side with a 1/4” tube fitting. The maximum allowable inlet
pressure is 140 PSI. A gauge at the bottom of the regulator indicates the output pressure. The output
pressure, and detector flow, can be adjusted using the knob on the front of the regulator. The detector
flow rises or falls as the output pressure is increased or decreased.
Aspirator
The aspirator inlet is connected to the output port on the right side of the regulator and the vacuum port
on top is connected to the detector chamber. It has a venturi orifice inside it which generates a vacuum
at its top port when compressed air flows through it. The compressed air and the air drawn from the
detector chamber into the top port of the aspirator both exhaust at the right side of the aspirator.
Detector Adapter
The infrared (IR) CH
4
detector and the H
2
S detector each have a detector adapter screwed onto the
bottom of the detector. The detector adapters are installed hand tight and have a gasket inside that seals
against the detector. When removing the detector adapter to change the detector, be sure not to lose this
gasket. The IR CH
4
and H
2
S detectors each require their own specific detector adapter. The IR CH
4
detector adapter cannot be installed on the H
2
S detector and vice versa.
Detector Chamber
Each detector has a detector chamber. The chamber has three thumbscrews which fasten it to the
detector adapter. An O-ring at the top of the chamber seals the chamber/adapter interface. The inlet of
each chamber is on the side and the exhaust of each chamber is on the bottom. The IR CH4 detector’s
inlet is connected to the exhaust of the sample flowmeter. The IR CH
4
sensor’s exhaust is routed to the
inlet of the H
2
S detector’s chamber. The exhaust of the H
2
S detector’s chamber is connected to the
vacuum port of the aspirator.
Sample Flowmeter
The flowmeter indicates the flow to the IR CH
4
detector. It has a 1/4” OD tube fitting at its inlet and
exhaust port. The exhaust port of the flow meter is connected to the IR CH
4
detector’s detector
chamber. The flowmeter’s indication range is 0.5 - 5 SCFH. It has no flow adjustment valve because
the flowrate is controlled by the regulator setting.
H
2
S Dilution Flowmeter
The H
2
S dilution flowmeter indicates the flow rate of fresh air entering the sample stream. It has a 1/4
OD tube fitting at its exhaust port. The exhaust port of the flowmeter is connected to the sample line
before it enters the H
2
S detector’s detector chamber. The H
2
S dilution flowmeter’s inlet port has a hose
barb fitting installed. The H
2
S dilution flowmeter’s indication range is 0 - 1.0 SCFH and it has a flow
adjustment valve. The flow to the H
2
S detector is the sum of the sample flow and the dilution flow.
Particle Filter
A particle filter with a piece of gray tubing on each end is shipped with the aspirator adapter but is not
factory installed. The particle filter must be attached to the H
2
S dilution flowmeter’s inlet port using
the gray tubing during installation. If tubing from a fresh air area needs to be routed to the particle
filter, that fresh air line can be connected to the gray tubing on the open end of the particle filter or the
gray tubing can be removed and the fresh air line can be connected directly to the open end of the
particle filter. If the particle filter gets clogged, the flowrate through the flowmeter will decrease. The
10 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
flowrate should be monitored, as a change in flowrate will affect the accuracy of the H
2
S reading.
Calibration Valve
The calibration valve is a manual operation spring return valve with a push button actuator. This valve
is used to switch from sample flow to calibration gas during the calibration process. When the button is
pushed and held, the sample port is closed and the calibration port is opened.
Inlet/Blowback Valve
The inlet/blowback valve is a manual operation spring return valve. A push button actuator is located
on the front of the valve. This valve is used to switch from sample flow to blow back in the event of a
clogged sample line. When the push button is pressed and held, the valve diverts the compressed air
supply back through the sample line to clear obstructions.
WARNING: The blowback pressure can be as high as 140 PSI. Make sure that all personnel and
equipment are clear of the sample line inlet end to avoid personal injury or equipment
damage if a sample line obstruction is blown out of the sample line.
Water Trap
The water trap prevents water from entering the flow system and is self-draining.
Low Flow/Drain Fault Contact Housing
The low flow switch provides open contacts in normal operation in the low flow/drain fault contact
housing that close in the event of a low flow condition. Two contact wires protrude from the side of the
switch and enter the low flow contact housing through a plastic cable bushing on the left side of the
housing. Terminals are provided in the low flow/drain fault contact housing for field connection to the
switch contacts.
The drain fault switch senses a break or very low flow in the water drain line and provides open
contacts in normal operation in the low flow/drain fault contact housing which close when the switch
senses a break, a very low drain flow, or a shut down.
Connecting monitoring devices to the low flow and drain fault contact terminals provides the user with
a notification of a contact closure. Each set of contacts may have their own monitoring device or they
may share one. See “Installation” on page 19 for instructions to wire these terminals using the two
scenarios.
A second plastic cable bushing on the bottom of the low flow/drain fault contact housing allows for
cable entry to the housing. The size range of the cable which can be routed through the second cable
bushing is .064’ - .210’ OD.
30-0954RK-209D Series Sample Draw Aspirator Adapter • 11
M2A Series H
2
S and CH
4
Detector Heads
This section describes the external and internal components of the M2A series H
2
S and CH
4
detector
heads that are installed on this version of the 30-0954RK-209D aspirator panel. Other versions may
have different detectors installed.
External Description
This section describes the junction box and all external components of the M2A transmitter.
Figure 3: M2A External Components, IR CH
4
Detector
Magnetic
Wand
Junction Box Cover
3/4 NPT Conduit Port, Plugged
3/4 NPT
Conduit Port
Mounting Slot (2x)
Window
IR CH4 Detector
12 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
Figure 4: M2A External Components, H
2
S Detector
Junction Box
The M2A’s cast aluminum junction box is dust and weather resistant. The junction box also protects
the M2A and all connections made to it. Use the three 3/4 NPT conduit ports to mount the detector to
the junction box (factory installed in the bottom port) and connect wiring from an external device (left
port). The top port is shipped with a factory installed and sealed conduit plug to avoid leaks into the
junction box. If necessary, the conduit plug can be removed and the top port can be used for wiring.
WARNING: If the top conduit port is used for wiring, be sure to seal the threads to ensure water
does not enter the junction box. See “Wiring” on page 20 for complete wiring
instructions.
Use the junction box’s two mounting holes to mount the M2A to a vertical surface at the monitoring
site. The window in the cover on the front of the junction box allows you to view the OLED display
and use the magnetic wand to actuate the magnetic control switches so you can perform non-intrusive
calibration. Removing the cover allows you to access the interior of the junction box.
Magnetic Wand
The magnetic wand is a short plastic rod with a magnet in one end. It is used to actuate the magnetic
control switches on the control PCB while the junction box cover is still installed so that non-intrusive
calibration can be performed.
Magnetic
Wand
Junction Box Cover
3/4 NPT Conduit Port, Plugged
3/4 NPT
Conduit Port
Mounting Slot (2x)
Window
H2S Detector
30-0954RK-209D Series Sample Draw Aspirator Adapter • 13
Infrared (IR) CH
4
Detector
An IR CH
4
detector is generally used instead of the catalytic combustible detector in applications
where there may be catalyst poisons such as silicone present or where oxygen is not present in the
monitoring environment.
The IR CH
4
detector is made up of a miniature infrared methane %volume detector housed and
encapsulated in a pipe nipple. The pipe nipple has 3/4 NPT threads on each end and a 1 1/4” hex that
allows removal or installation of the detector with a wrench. A porous flame arrestor that is coated
with a hydrophobic film that repels liquids is on one end of the detector and allows sample gas to enter
the detector. Four color coded leads, red, white, green, and black, extend from the other end of the
detector. The leads allow you to connect the detector to the amplifier.
H
2
S Detector
The H
2
S detector has a replaceable plug-in sensor inside the detector housing. A rubber boot and a
spacer help ensure that the sensor remains plugged into the detector housing body.
Figure 5: H
2
S Detector Component Location
Spacer
Rubber Boot
H2S Plug-inSensor
Flame
Arrestor
Guard
Detector
Housing Cap
Detector HousingBody
Cap Gasket
14 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
Internal Description
This section describes the internal components of the M2A. The internal components of the M2A
include the terminal PCB which provides for all the wiring connections to the M2A and the control
PCB which displays the gas reading and has the control buttons.
Figure 6: M2A Internal Components
Terminal PCB
The terminal PCB is encapsulated in epoxy for protection against moisture and physical damage. It is
mounted into the rear of the junction box with three standoffs and rests on a thin layer of foam. A
banana jack is screwed into each of the standoffs and is used for mounting the control PCB. The
terminal PCB converts the electrical output from the detector to a signal which can be displayed by the
LCD display, a 4 - 20 mA signal (that is proportional to the detection range), and an RS-485 Modbus
output signal. The 4 - 20 mA signal may be used by a recording device, gas monitor controller, or
programmable controller. The Modbus output may be used to connect the M2A to a Modbus network.
The terminal PCB also controls three relays: one fail and two gas alarm.
Two columns of plug-in style terminal strips are used to make all wiring connections to the M2A. The
column on the left consists of the power/signal, detector, and Modbus terminal strips. The column on
the right consists of the relay terminal strips. A 20 position connector at the bottom of the terminal
PCB is used to connect the terminal PCB to the control PCB with a ribbon cable. A 5 position
connector on the left side of the terminal PCB is used by factory or field service personnel to program
the M2A. On the combustible and CO
2
versions of the M2A, a factory adjust pot just above the
NO
Control PCB
Junction Box
Terminal PCB
3/4 NPT Conduit
Opening for Wire Entry
M2A TRANSMITTER
RKIINSTRUMENTS
3/4 NPT Conduit Opening, Plugged
30-0954RK-209D Series Sample Draw Aspirator Adapter • 15
programming connector is used to set the detector current.
Figure 7: Terminal PCB Component Location
Power/Signal Terminal Strip
The power/signal terminal strip is a three position plug-in style terminal strip located at the top of the
left terminal column. It is used to connect 24 VDC power to the M2A and to connect the 4 - 20 mA
output signal to a device.
The signal output, the S terminal, does not have to be connected for the M2A to function. The S
terminal is used if you want to connect the 4 - 20 mA output signal to another device such as a gas
monitoring controller, chart recorder, or programmable controller (PLC).
Detector Terminal Strip
The detector terminal strip is a four position plug-in style terminal strip and is the middle terminal strip
in the left terminal column. For the IR CH
4
detector, all four terminals are used to connect the detector
to the M2A. For the H
2
S detector, only two of the terminals are used for connecting a detector.
Figure 8: Detector Terminal Strip
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Modbus Terminal Strip
Ground Jumper
Programming Connector
(Factory Use)
Power/Signal Terminal Strip
Termination Jumper
Ribbon Cable Connector
Detector Current Adjust Pot
(Comb/CO2 Type Only, Factory Adjust)
Detector Terminal Strip
(Comb/CO2 Version Shown)
Relay
Terminal
Strips
H2S
Detector
Terminal
Strip
IR CH4
Detector
Terminal
Strip
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
A B C
RS 485 PWR/SIG
R W G B
LEL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
16 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
NOTE: The detector is factory-wired to the M2A. See “Wiring” on page 20 for all wiring procedures
related to M2A.
Modbus Terminal Strip
The Modbus terminal strip is a three position plug-in style terminal strip and is the bottom terminal
strip in the left terminal column. It allows connection of the M2A into a Modbus network.
Relay Terminal Strips
The right column of terminal strips consists of, from top to bottom, the fail, alarm 1, and alarm 2 relay
terminal strips. They are three-position plug-in style terminal strips. The relay terminal strips are used
to connect devices such as lights and horns that are controlled by the relay contacts. The relay contacts
are rated at 115 VAC, 5 amps. The relay contacts may also be used to control higher rated relays.
Termination Jumper
A two pin header with a termination jumper installed is located below the Modbus terminal
strip. The jumper has no function unless the M2A is wired into a Modbus installation. See
“Chapter 8: RS-485 Modbus Output” on page 50 for a description of using the M2A in a
Modbus system.
Ground Jumper
A two pin header with a jumper installed is located along the left side of the terminal PCB. Leave this
jumper installed unless directed to remove it for a Modbus installation. See “Chapter 8: RS-485
Modbus Output” on page 50 for a description of using the M2A in a Modbus system.
Control PCB
The LCD display and control switches are located on the control PCB. It is installed on top of the
terminal PCB by lining up its three spacing standoffs with the banana jacks in the terminal PCB
mounting standoffs and pushing it onto the banana jacks. The jacks retain the control PCB.
Figure 9: Control PCB Component Location
ENTER
DOWN/NO
UP/YES
Push Button
Control
Switches
Magnetic Switches
ENTER
UP/YES
DOWN/NO
Alarm 1 LED
Alarm 2 LED
Fail LED
RX LED
TX LED
M2A TRANSMITTER
RKI INSTRUMENTS
OLED Display
30-0954RK-209D Series Sample Draw Aspirator Adapter • 17
LCD Display
The LCD display is located at the top of the control PCB. It indicates the current gas reading and
displays messages and parameters in the M2A’s programs.
Control Buttons
The M2A includes three push button switches that allow you to enter the M2A’s operating modes,
navigate through the modes, update settings, and save changes to the settings. The push button
switches are located along the bottom edge of the control PCB (see Figure 9). The UP/YES button is
on the left, the DOWN/NO button is in the middle, and the ENTER button is on the right.
Just above each push button switch is a magnetic switch with the same function as the push button
switch below it. The magnetic switches are for use in non-intrusive calibration. They are actuated by
bringing the magnetic wand close enough to them to actuate them. Although the magnetic switches
have the same functions as the push button switches, it is not practical to use them for operations other
than calibration because it is not possible to actuate two magnetic switches at once with only one
magnetic wand. Since displaying the Information Screen only requires the use of one switch, the wand
may be used to show the Information Screen (see “Information Screen” on page 27).
Status LEDs
The M2A includes five status LEDs that are located above the display (see Figure 9).
Fail LED
The fail LED turns on when the M2A is experiencing a fail condition. A fail condition can be
caused by a detector failure or low detector signal.
Alarm 1 LED
The alarm 1 LED is on when the M2A is experiencing an alarm 1 condition.
Alarm 2 LED
The alarm 2 LED is on when the M2A is experiencing an alarm 2 condition.
Table 2: M2A Control Button Functions
Switch Function
UP (YES) Saves settings
Changes the displayed setting
Enters the Calibration Program
Enters Gas Type Mode (press with DOWN/NO button)
Enters Configuration Mode (press with ENTER button)
DOWN (NO) Cancels setting changes
Changes the displayed setting
Enters Gas Type Mode (press with UP/YES button)
Enters Modbus Mode (press with ENTER button)
Displays the Information Screen
ENTER Initiates operations
Enters Configuration Mode (press with UP/YES button)
Enters Modbus Mode (press with DOWN/NO button)
Functions as an alarm reset switch
18 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
RX & TX LED’s
These LED’s indicate data being received (RX) and transmitted (TX) when the M2A’s Modbus
output is operating.
30-0954RK-209D Series Sample Draw Aspirator Adapter • 19
Chapter 3: Installation and Startup
Overview
This section describes installation and startup procedures for the sample draw aspirator adapter.
Installation
Figure 10: Outline & Mounting Dimensions
1. Install the mounting panel to a flat vertical surface using the four mounting holes (0.50’ diameter)
in the corners of the panel. Make sure that sufficient space is allowed to remove the detector head’s
cover, and to bring power/signal wiring to the detector head. Also provide sufficient clearance for
routing of sample, compressed air, and exhaust lines.
20 • 30-0954RK-209D Series Sample Draw Aspirator Adapter
Wiring
WARNING: Always verify that the power source is OFF before making any wiring connections.
1. Remove the junction box cover.
2. Grasp the control PCB by its edges.
3. Gently pull until the control PCB is pulled away from the banana jacks. Take care not to pull too
hard and damage the cable which connects the control and terminal PCB’s.
4. Let the control PCB hang by the cable. The terminal strips are now visible on the terminal PCB.
The control PCB may be left hanging while wiring is done. If desired, the control PCB may be
disconnected from the cable and set aside while wiring.
5. The detector leads are factory wired. Verify that the detector leads are wired to the detector
terminal strip as shown in the applicable figure below:
IR CH
4
Detector
Red wire to terminal labeled LEL R, white wire to terminal labeled LEL W, green wire to
terminal labeled LEL G, black wire to terminal labeled LEL B.
Figure 11: IR CH
4
Detector Wiring
•H
2
S Detector
Red wire to terminal labeled TOXIC +, black wire to terminal labeled TOXIC -.
Figure 12: H
2
S Detector Wiring
Green
Black
Red
White
- S + A B C
RS 485 PWR/SIG
R W G B
LEL
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
Black
Red
C NC NO
ALARM 1
C NC NO
ALARM 2
C NC NO
FAIL
- S +
PWR/SIG
TOXIC OXY
+ - + -
RS 485
A B C
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67

RKI Instruments 30-0954RK-209D Owner's manual

Category
Carbon monoxide (CO) detectors
Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI