Example 5: An antigen is added to well #7 and its corresponding antibody is added to well
#1. A low concentration of antibody is added to well #2. Evaluating the end of the precipitin
line, which is turned away from well #2, allows determination of low antibody
concentrations.
Example 6: To determine immunologic specificity, a mixture of two different antigens is
added to well #7. The respective antibody samples are added to well #1 and well #2. After
incubation, an intersection where the lines meet implies that the reactions have unrelated
specificities, which are known as nonidentity reactions.
Example 7: Similar to Example 6, two different antigens are added to well #7. An antibody
sample is added to well #2, and a mixture of two antibodies is added to well #1. The result is
a continuous arc between wells #1 and #7 and between wells #2 and #7. This is known as the
identity arc, which indicates that the reactants in well #1 and well #7 and in wells #2 and #7
have the same immunological specificity. Also note that the arc does not affect the second
band between wells #1 and #7.
Example 8: Similar to Example 7, the relative concentrations of two different antibodies used
in well #1 are such that the resulting bands are superimposed and indistinguishable. The
antibody in well #2 forms an arc with the antigen in well #7, but a spur projects from the arc.
This result is known as a partial identity and indicates that the specificities in the reactants in
wells #1 and #7 are contained in the precipitin line between wells #2 and #7; however, the
exact number of specificities in a single band cannot be determined.
Cited References
1. Oudin, J.; L’analyse, (1948). immunochemique qualitative. Methode par diffusion des antigens au sein de l’immunoserum precipitant gelose. Premiere
Parte Inst Pasteur 75:30-52.
2. Ouchterlony, O. (1949). Antigen-antibody reactions in gels and the practical applications of this phenomenon in laboratory diagnosis of diphtheria.
Med Diss Stockholm.
This product (“Product”) is warranted to operate or perform substantially in conformance with published Product specifications in effect at the time of sale,
as set forth in the Product documentation, specifications and/or accompanying package inserts (“Documentation”) and to be free from defects in material and
workmanship. Unless otherwise expressly authorized in writing, Products are supplied for research use only. No claim of suitability for use in applications
regulated by FDA is made. The warranty provided herein is valid only when used by properly trained individuals. Unless otherwise stated in the
Documentation, this warranty is limited to one year from date of shipment when the Product is subjected to normal, proper and intended usage. This
warranty does not extend to anyone other than the original purchaser of the Product (“Buyer”).
No other warranties, express or implied, are granted, including without limitation, implied warranties of merchantability, fitness for any particular
purpose, or non infringement. Buyer’s exclusive remedy for non-conforming Products during the warranty period is limited to replacement of or
refund for the non-conforming Product(s).
There is no obligation to replace Products as the result of (i) accident, disaster or event of force majeure, (ii) misuse, fault or negligence of or by Buyer, (iii)
use of the Products in a manner for which they were not designed, or (iv) improper storage and handling of the Products.
Current product instructions are available at www.thermoscientific.com/pierce. For a faxed copy, call 800-874-3723 or contact your local distributor.
© 2011 Thermo Fisher Scientific Inc. All rights reserved. Unless otherwise indicated, all trademarks are property of Thermo Fisher Scientific Inc. and its
subsidiaries. Printed in the USA.