Kam OWD Water Cut Meter User manual

Type
User manual

This manual is also suitable for

KAM
®
OWD
®
WATER CUT METER
PTB 08 ATEX 1026
TEL +1 713 784 0000
FAX +1 713 784 0001
KAM CONTROLS, INC.
3939 Ann Arbor Drive
Houston, Texas 77063 USA
www.KAM.com
User Manual
OWDMANUAL 1114
An ISO 9001 certified company
OWDMANUAL 1114
KAM CONTROLS, INC.
1
TABLE OF CONTENTS
SECTION TITLE PAGE
1 Introduction 2
•AvailableModelsandMountingOptions 2
•TheoryofOperation 3
•Features3
•Applications 3
2 Specifications 4
•Specications 4
•DimensionalDrawings 5
3 Installation 10
•InstallationFlowRequirements 10
•LACTunitinstallation 11
•GeneralDo'sandDon't's 12
•MainLine 13
•FastLoop 16
•Wiring 17
4 OWD Operation 21
•HyperterminalSoftware  21
•Calibration 25
•ChangingtheRange 26
•ModbusInterface 26
5 Maintenance 27
•Cleaning/Inspection 27
•Troubleshooting 27
•AntennaReplacement 28
APPENDIX A: ModbusRegisters 35
CAUTION:
WheninstallingtheOWD
®
sensorinapipelinecontainingpetroleumproducts,
petrochemicals,wastewaterswiththepresenceofpressure&temperature,andhigh-pressure
steamrefertothePipelineOperators’"Health,SafetyandEnvironmentalPolicyProcedures"to
ensuresafeinstallation.
KAMCONTROLS,INC.reservestherighttomakechangestothisdocumentwithoutnotice.
OWDMANUAL 1114
KAM CONTROLS, INC.
2
P
INTRODUCTION
AVAILABLE MODELS and MOUNTING OPTIONS
FIG. 1-1
FIG. 1-2
FIG. 1-3
Option3:Fixed-mountOWD
®
onadensitometerloop,with
1/2",3/4",1",or2"FNPT
Option 2: Retractable OWD
®
on a main pipe
with2"MNPTsealhousing
Option 1: Retractable OWD
®
onamainpipewith
2",3",or4"angedsealhousing
Full-opening
Ball Valve
Full-opening
Ball Valve
Q
Q
Q
Q
Q
FIG. 1-4
Option4:2"OWD
®
owthrough
sensorwithintegrated
KAM
®
SMS™StaticMixingSpool
Recommended KAM
®
SMS™
StaticMixingSpool
Recommended KAM
®
SMS™
StaticMixingSpool
Recommended
KAM
®
SMS™
StaticMixingSpool
Included
KAM
®
SMS™
StaticMixing
Spool
OWDMANUAL 1114
KAM CONTROLS, INC.
3
INTRODUCTION CONTINUED
Rugged,easytouseandextremelyaccurate,theKAM
®
OWD
®
OilWaterDetectoristheidealinstrumentforcontinu-
ouslymonitoringwaterconcentrationinyourpipeline.ItisdesignedinaccordancewithAPI,ASTM,ISO,EI,UL,and
DINstandardsamongstothers.Especiallyvitalinproductionmanagement,theOWD
®
sensor lets you maximize oil
productionversusproducedwater.ThesimplicityofdesignandqualityofengineeringemployedintheOWD
®
sensor
meantherearenomovingparts.Patentedmicrowavesensorsmeasuretheconductivity,dielectric,andboththereal
andimaginarypartofpermittivityoftheuidwithanextremelyhighdegreeofaccuracy,andmeasurementisfully
automaticwithouttheneedforoperatorinterventionorsupervision.TheoutputsignalcanbesenttoFlow
Computers,SCADA,PLC’sortoaCentralControlRoomforloggingordisplayonchartrecordersormonitors.
TheKAM
®
OWD
®
sensoralsousesinternalreferencestoautocalibratefordriftcausedbytemperaturechangesof
theelectronics,theagingoftheelectronicscomponents,uidpressure,anduidtemperature.
TheKAM
®
OWD
®
owthroughmodelcanbeusedinananalyzer/densitometerloop,forprocessoptimizationwhere
anaccuratedeterminationofwaterconcentrationisimportant,anditisvitaltooptimizingthe
desalinizationprocess.Placedonthedesaltersampleline,oroneachsampleline,theKAM
®
OWD
®
owthrough
modelprovidesreal-timeinformationaboutyourdesalterperformance.
Toensurethehighestdegreeofaccuracy,theowmustbehomogenous.InstalledupstreamofyourOWD
®
sensor,thepatentedKAM
®
SMP™StaticMixingPlateorKAM
®
SMS™StaticMixingSpoolcreateafully
homogenousmixtureinyourpipeline.Inlowvelocitysituations,theuseofaKAMMLMeasurementLoopmay
berequiredinordertocreateahomogenousowformeasurement.Propercalibration,alsokeytocomplete
accuracy,canbeachievedintheeldwiththeKAM
®
PKFPortableKarlFischerMoistureAnalyzer.
Becauseitcanbeinsertedintoyourpipeortankthrougha2",3",or4"hottap,theOWD
®
sensorhelpsyouavoid
costlydrainage,theneedforabypassloop,orhavingtocutasectioninthepipe.Allwettedpartsaremachinedfrom
316stainlesssteel.Shaftlengthsfrom1to3feetareavailablewithoff-the-shelflengthscomingin12",24",and36".
Metricandcustomlengthsareavailable.
Range(waterinoil) 0-1% 0–5% 0–10% 0–20% 0–40% 0–100%
Accuracy(atlistedrange)0.01% 0.05% 0.10% 0.20% 0.40% 1.00%
TABLE 1-5 MEASUREMENTCAPABILITIES:CALIBRATEDRANGEANDACCURACIES
NOTE:TheKAMOWDcanbecalibratedtoanyrangebetween0-1%and0-100%.
THEORY OF OPERATION
OWDMANUAL 1114
KAM CONTROLS, INC.
4
SPECIFICATIONS
Media: Crudeoil,renedproductsandchemicals
Material: Wettedparts-316stainlesssteel/titanium(Othermaterialsavailable)
Fluidtemp: To300ºF(149ºC)Hightempmodelavailableto600ºF(315ºC)
Electronicstemp: -40ºFto158ºF(-40ºc to +70ºC)Tempsbelow0ºrequireheattracing
Powerrequirements: 24VDC/1ampat24watts
Accuracy: 1%offullrange**
Repeatability: +/-0.01%
Resolution: +/-0.01%
Minimum water detection: 100 PPM
Outputs: 4–20mA
4-20mAor0–5VDC
Alarm relay
RS232/RS485
HART
Mounting: ½",¾",1"or2"FNPTowthrough
(Othersizes,includingmetric,areavailable)
2"MNPTsealhousing
2",3",or4"angedsealhousing
Pressureratings: ANSI150,300,600,900,1500
Flowconditions: WellmixedinaccordancewithAPIMPMSChapter8,Section2,Table1
Sensordimensions: Ø1.5"x4.5"to14.5"(38mmhx114mmto368mm)dependingonrange
EXenclosures: Sensorelectronics-3"x6"x3"
(76mmx152mmx76mm)
Shaftlength: 12"to36"
(305mmto914.4mm)
Off-the-shelflengthsare12",24",and36"
(609.6mm,762mm,914.4mm,)
PipeSize: ½"to48"(15mmto1200mm)
Weight: from20lbs.(9kg)
*TheKAM
®
OWD
®
mustbeinstalledinaccordancewithAPIMPMSChapter8,Section2,Table1.
**Ifentrainedgasisconstant,itseffectisfactoredout.IfentrainedgasisintroducedorremovedafterOWD
®
calibrationitwillshiftwatercutmeasurementbyapproximately1-2%forevery1%changeingaslevels.
OWDMANUAL 1114
KAM CONTROLS, INC.
5
SPECIFICATIONS CONTINUED
DIMENSIONAL DRAWINGS 0-100% MODELS
CALCULATING SHAFT LENGTH 0-100% MODELS
D1
Valve
Length
Main Pipe
Pipe OD
Determineminimumshaftlengthforproperinsertioninthepipeline.
Off-the-shelflengthsare12",24",and36".
TABLE 2-2 DIMENSIONS INSERTABLE OWD
®
SENSOR 0-100%
C
D
E
CustomShaft
Lengths
A
B
F
G
H
FIG. 2-1 INSERTABLE OWD
®
SENSOR 0–100%
A
B
C
D
E
F
G
H
INCHES
1.48
2.4
4.62
max11.65
2.3
5.7
7
4.6
MM
38
61
117
296
58
145
178
117
ShaftLengthsareavailablein.5"
(12.7mm)increments.
Standardsizesare24",30",36",48",
and60"(609.6mm,762mm,914.4mm,
1219mm,1524mm).
*Sealhousingdimensionsvaryaccordingtoangesizeandpressurerating.11.65"isthemaximumlength
requiredincalculatingshaftlength.
D1
+ ValveLength:
+ PipeODx.5:
+ Sealhousinglength
– Probefactor
= Total/minimumshaftlengthforproper
insertion distance
13"
1"
OWDMANUAL 1114
KAM CONTROLS, INC.
6
SPECIFICATIONS CONTINUED
TABLE 2-4 DIMENSIONS INSERTABLE OWD
®
SENSOR 0-20%
C
D
E
Electronics
Enclosure
CustomShaft
Lengths
A
B
F
G
H
FIG. 2-3 INSERTABLE OWD
®
SENSOR 0–20%
A
B
C
D
E
F
G
H
INCHES
1.48
4.4
6.62
max 13
2.3
5.7
7
4.6
MM
38
112
168
330
58
145
178
117
ShaftLengthsareavailablein.5"
(12.7mm)increments.
Standardsizesare24",30",36",48",
and60"(609.6mm,762mm,914.4mm,
1219mm,1524mm).
DIMENSIONAL DRAWINGS 0-20% MODELS
CALCULATING SHAFT LENGTH 0-20% MODELS
*Sealhousingdimensionsvaryaccordingtoangesizeandpressurerating.13"isthemaximumlength
requiredincalculatingshaftlength.
D1
Valve
Length
Main Pipe
Pipe OD
Determineminimumshaftlengthforproperinsertioninthepipeline.
Off-the-shelflengthsare12",24",and36".
D1
+ ValveLength:
+ PipeODx.5:
+ Sealhousinglength
– Probefactor
= Total/minimumshaftlengthforproper
insertion distance
13"
2"
OWDMANUAL 1114
KAM CONTROLS, INC.
7
SPECIFICATIONS CONTINUED
D1
+ ValveLength:
+ PipeODx.5:
+ Sealhousinglength
– Probefactor
= Total/minimumshaftlengthforproper
insertion distance
C
D
E
Electronics
Enclosure
CustomShaft
Lengths
A
B
F
G
H
TABLE 2-6 DIMENSIONS INSERTABLE OWD
®
SENSOR 0-5%
FIG. 2-5 INSERTABLE OWD
®
SENSOR 0–5%
A
B
C
D
E
F
G
H
INCHES
1.48
6.4
8.62
max 13
2.3
5.7
7
4.6
MM
38
163
219
330
58
145
178
117
ShaftLengthsareavailablein.5"
(12.7mm)increments.
Standardsizesare24",30",36",48",
and60"(609.6mm,762mm,914.4mm,
1219mm,1524mm).
DIMENSIONAL DRAWINGS 0-5% MODELS
CALCULATING SHAFT LENGTH 0-5% MODELS
D1
Valve
Length
Main Pipe
Pipe OD
13"
3"
Determineminimumshaftlengthforproperinsertioninthepipeline.
Off-the-shelflengthsare12",24",and36".
*Sealhousingdimensionsvaryaccordingtoangesizeandpressurerating.13"isthemaximumlength
requiredincalculatingshaftlength.
OWDMANUAL 1114
KAM CONTROLS, INC.
8
SPECIFICATIONS CONTINUED
FIG 2-7 2" OWD
®
FLOW THROUGH SENSOR WITH INTEGRATED KAM
®
SMS™ STATIC MIXING SPOOL
TABLE 2-8 DIMENSIONS
A
B
C
D
FLANGESIZE
N/A
N/A
N/A
2"150#
2"300#
2"600#
MM
119
184
419 ± 13
533
559
559
INCHES
4.7
7.25
16.5±0.5
21
22
22
A
E
E
B
C
D
2"150#,300#,or600#weld-
neckanges
SS316bothends
2"sch.seamlesspipeSS316
FIG 2-9
SS316½"samplevalvewith
½"pitotprobe
OWDMANUAL 1114
KAM CONTROLS, INC.
9
SPECIFICATIONS CONTINUED
TABLE 2-11 DIMENSIONS
A
B
C
D
E
MM
282
Min.152
Min254
406
691
INCHES
11. 1
Min.6"
Min.10"
Min.16"
27.2
A
E
B
C
D
FIG 2-10 1" OWD
®
0-1% FLOW THROUGH
1"WNRFFlanges
OWDMANUAL 1114
KAM CONTROLS, INC.
10
KAM SMS
STATIC MIXING
SPOOL
INSTALLATION
PLEASENOTE:InallKAMOWDInstallations,theusershouldinsurethattheKAMOWDisinstalledinaturbulent
owwiththeReynoldsNumberabove2000.Additionally,allKAMOWD'sshouldbeinstalledinaccordancewithAPI
MPMSChapter8,Section2,Table1.
ForKAMOWD'soperatinginoil-continuousow,thesensormustbeinstalledintheverticaldownowwithamini-
mumowvelocityof4feetpersecond.AKAMSMSStaticMixingSpoolisrequirediftheowvelocityisbetween4
and7feetpersecond.FIG.3-1.Lowrangemodelsalsorequireacorrecteddensityinput.
ForKAMOWD'soperatingintheOilContinuousPhase,thesensormustbeinstalledintheverticaldownow.A
KAMSMSStaticMixingSpoolisrequirediftheowvelocityisbetween4and7feetpersecond.FIG.3-1.
ForKAMOWD'soperatingintheWaterContinuousPhase,thesensormustbeinstalledintheverticalupow.A
KAMSMSStaticMixingSpoolisrequirediftheowvelocityisbetween4and7feetpersecond.FIG.3-2.
Insituationswheretheowvelocityislessthat4feetpersecond,KAMCONTROLSrecommendstheinstallation
ofaKAMMLMeasurementLoop,incorporatingsuctionandinjectionnozzles,apump,andtheOWDonaseparate
loop,ensuringahomogenous,high-velocityowacrossthemeasurementsensor.FIG.3-3.
TheKAMOWDmaybeinstalledhorizontallywhentheminimumowvelocityisabove10feetpersecond.
KAM SMS
STATIC MIXING
SPOOL
Q
LOWRANGEAND
OIL CONTINUOUS
PHASE OPERATION
Q
WATER CONTINUOUS
PHASE OPERATION
FIG. 3-3 INSTALLED ON A
KAM ML MEASUREMENT LOOP
INSTALLATION FLOW REQUIREMENTS
Q
INJECTION
SUCTION
TheKAMMLMeasurementLooputilizesa
pumptodrawarepresentativeowfromthemain
lineintoacirculationloop,incorporatingaKAM
OWD.Theloopowisinjectedbackintothemain
pipelineupstreamofthesuctionsite,
creatingmixingandhomogeneitypriorto
suction.Ahomogenousowof10-13fpsis
maintainedatthesensorheadforthemost
accurate measurement at all times.
FIG. 3-1 INSTALLATION VERTICAL FLOW DOWN
FIG. 3-2 INSTALLATION VERTICAL FLOW UP
OWDMANUAL 1114
KAM CONTROLS, INC.
11
INSTALLATION CONTINUED
LACT UNIT INSTALLATION
TheKAMOWDshouldbeinstalled
onaTwiththeowperpendicularto
theshaft.AKAMSMPStaticMixing
Plateshouldbeinstalled3to5pipe-
linediametersbeforethesensor.For
allotherinstallations,pleaseconsult
withfactory.
CAUTION:NeverinstalltheOWDon
aTwiththetopofthesensorfacing
intotheow.Thiswillminimizeow
acrosstheantennasandresultin
inaccurate measurement.
0-5%modelsrequireacorrected
densityinputinordertoachieve
stated accuracy.
Flow
KAMSMPStaticMixingPlate
installed3to5pipeline
diameters prior to sensor
FIG. 3-4 LACT UNIT
OWDMANUAL 1114
KAM CONTROLS, INC.
12
INSTALLATION CONTINUED
GENERAL INSTALLATION DO'S AND
DON'TS
Always install OWD
®
sensorswiththe
electronics enclosure
shadedfromdirect
sunlight.
DONOTuseTeon
tapeonthreadscon-
nectingtotheOWD
®
owthroughsensor.
DOuseliquidthread
sealant.
OWDMANUAL 1114
KAM CONTROLS, INC.
13
INSTALLATION CONTINUED
Removealltheprotectivepackagingmaterials,andensurethattheOWD
®
sensorwasnotdamaged
duringtransit.
REMINDER:PleaserefertotheInstallationFlowRequirementsonP.9ofthismanualtoensurepropersensor
placementwhereatallpossible.FlowconditionsmustsatisfyAPIMPMSChapter8.2requirementsinorderto
achieveaccurateOWDperformance.
Incoldweather,ifOWDisexposedtoanopenenvironment,KAMCONTROLSrecommendsoperatorsinsulate
theOWD,andifthepipelineisheatedthattheheatingtracebeextendedtoincludetheOWD.
IfthepipelineisnotgoingtoowforextendedamountoftimeandthepipeisnotheatedthenOWDshouldbe
takenouttoavoiddamagetothesensorprobebyfreezingwater.
PRIOR TO INSTALLATION
MAIN LINE INSTALLATION
INITIAL CALIBRATION
TheKAM
®
OWD
®
sensorshouldbeinstalled
accordingtoFIG.3-13.Afull-openingballvalveis
usedtoisolatetheOWD
®
sensorfromthepipeline
duringinstallationorremoval.Thesealhousingof
theOWD
®
sensorallowstheprobetobeinserted
andremovedfromthepipeunderpressureandow
conditions.Itistheusersresponsibilityto
ensurethattheOWD
®
sensorbeplacedatthemost
representativepointwithintheow
prole(seelocationrecommendationsabove).
TheOWD
®
sensorshouldbeinsertedsothatthe
windowoftheprobeislocatedinthecenterofthe
diameterofthepipeline.
Note:Iflinepressureexceeds100psi,useaKAM
®
ITInsertionToolwheninstalling/removing
theKAM
®
OWD
®
sensor.
LockingCollar
SealHousing
Full-opening
Ball Valve
SocketCapScrew
FIG. 3-5 KAM
®
OWD
®
INSTALLED ON A MAIN PIPE
ThoughtheOWDhasbeencalibratedinthefactory,operatorsshouldconductaninitialcalibrationinprocess
conditions.Thiscanbedoneintwoways:
Off-line:Priortoinstallation,operatorscangothroughtheproceduresforoff-linecalibrationoutlinedonpage25ofthis
manual.Thismethodrequiressamplesof100%producedwateranddryoiloroilwithaknownpercentageofwater.
On-line:Afterinstallation,operatorscanfollowcalibrationproceduresforin-linecalibrationoutlinedonpage25ofthis
manual.Thismethodrequiresaccuratesamplingandsampleprocessing.
OWDMANUAL 1114
KAM CONTROLS, INC.
14
Priortomountingverifythatthetipofthesensor
isallthewayinsidethesealhousing.
FIGS.3-14,3-15.
Ifsensorisnotfullyenclosedinsidetheseal
housing,pulltheshaftbackuntiltheprobeis
allthewayinthesealhousingandtighten
theSocketCapScrewsonthelockingcollar.
ThiswillpreventtheOWD
®
shaftfromsliding
andtheprobefromgettingdamagedduring
mounting.
1.
2.
FIG. 3-6
FIG. 3-7
INSTALLATION CONTINUED
Measurethedistance(D1)fromtheoutsidediameterofmainpipetotheendofthe
connectionwheretheOWD
®
sensorisgoingtobeinstalled.FIG.3-16.
D1
D1
CalculatetheminimuminsertiondistancefortheOWD
®
.
Minimuminsertiondistance(MID)=
D1 +
PipeWallThickness(WT)+
GasketThickness+
A(SeeTABLE3-16)
ExampleforD1=16",WT=1/4",
GasketThickness=1/8"anda0-100%OWD
®
sensor:
MID=16+1/4+1/8+3
MID=193/8"
4.
3.
FIG. 3-8
A/INCHES
7"
5"
3"
OWD
®
RANGE
0-5%SENSOR
0-20%SENSOR
0-100%SENSOR
TABLE 3-9
OWDMANUAL 1114
KAM CONTROLS, INC.
15
INSTALLATION CONTINUED
MeasurethecalculatedMIDfromthetopofthe
LockingCollarandplaceamarkwithapermanent
markerortapeontheShaft.FIG.3-18.
5.
MID
6.
BoltorscrewtheOWD
®
sensortothevalveordesignatedinstallationlocation.
(KAMCONTROLSrecommendsusingthreadsealantandnotTeontapeforthethreadedOWD
®
).
Openfullopeningvalve.7.
LoosenSocketCapScrewsonthelockingcollar.
8.
FIG. 3-10
PushOWD
®
sensorinuntilthemarkisatthetopedgeofthelockingcollar.EnsurethatOWD
®
owindicatoris
alignedwithpipelineowdirection.FIG.3-19.
9.
Mark
Re-tightentheSocketCapScrews.
TightenthehexnutsholdingdowntheLockingCollaronehalfturn.(Fig.3-19)Theseshouldneverbeover-tight-
ened.Theirmajorfunctionistoapplylightpressureonthechevronpackingtoensureasealbetweentheseal
housingbodyandtheinsertionshaft.
10.
11.
REMOVING THE OWD
®
SENSOR
ToremovetheOWD
®
sensor,rstshutoffpowertotheinstrument.LoosentheSocketCapScrewontheLockDown
Collar.SlidetheOWD
®
sensorupwarduntiltheproberestsinsidethesealhousing.Next,closetheFull-openingBall
Valvetightly.Drainoilfromvalveifpossible.TheOWD
®
sensormaynowbeunboltedfromthesystem.
Note:Iflinepressureexceeds100psi,useaKAM
®
ITInsertionToolwheninstalling/removing
theKAM
®
OWD
®
sensor.
RemovalshouldbeconductedinaccordancewithallregionalandClassrequirements.
FIG. 3-11
Hex Nuts
OWDMANUAL 1114
KAM CONTROLS, INC.
16
INSTALLATION CONTINUED
REMOVING THE OWD
®
SENSOR FROM FLOW THROUGH SPOOLS WITH FIXED INSERTION
ToremovetheOWD
®
sensor,rstshutoffpowertothe
instrument.Discontinueowinloopfromthemainline
anddrainuidfromloop.Theprobecanremovedfromthe
housingbyremovingscrewsoncollar.Fig.3-23.Theprobe
canthenbeliftedfromthecellfortesting/inspection/cali-
bration purposes.
Removalshouldbeconductedinaccordancewithallre-
gionalandClassrequirements.
FIG. 3-12
Hex screws
OWDMANUAL 1114
KAM CONTROLS, INC.
17
INSTALLATION CONTINUED
WIRING
Kam
Controls
Incorporated
3939 Ann Arbor Drive
Houston, TX 77063 USA
Tel + 1 713 784 0000
Fax + 1 713 784 0001
www.Kam.com
OWD Rev. 4
KAM OWD
Made in USA
24V (+) in
4-20mA (-)
24V (-) in
4-20mA (+)
ANA OUT (-)
ANA OUT (+)
DIG OUT (-)
DIG OUT (+)
AIN2
RS485 (+)
RS485 (-)
GND
RS232 RX
GND
RS232 TX
DIG IN (-)
DIG IN (+)
GND
FIG. 3-13 WIRING DIAGRAM 0-100% MODELS
D1
R1
+
OWDMANUAL 1114
KAM CONTROLS, INC.
18
INSTALLATION CONTINUED
INPUTS
24V(–)IN GND
24V(+)IN 24-30VDC
DIGIN(–) Pulseinput,discreteinputfordifferentmodesofoperation(0or5volt)
DIGIN(+)
AIN2 DensityIn–Notrequiredon0-100%models
OUTPUTS
4-20mA(–) Currentoutput,sourcepowered.Settorequisitepercentwateratfactory.
4-20mA(+)
ANOUT(–) Canbe4-20mAoranalogvoltage(usedforlocaldisplayortopassdensitytoPLC)
ANOUT(+)
DIGOUT(–) Alarmorrelay(digitalcontactclosure)
DIGOUT(+)
INPUT/OUTPUT
RS232 Consoleport–communicationinterfaceforcalibration,connectiontoPLC
RS485 Modbusinterface
LED INDICATORS
D1 Power
SERIAL PORT CONNECTIONS
DB9(female)
5 GND
3 RS232RX
2 RS232TX
AllwiringandmaintenanceontheKAMOWDmustbedoneinaccordancewithregionalandclassicationre-
quirements.Itistheuser'sresponsibilitytounderstandtheserequirements.
ItisalsorecommendedthattheOWDbewiredwithexiblewiring/conduitwithadditionalslack/lengthinthe
wiretoaccommodateinsertion,removal,andtesting.
Operator'sshouldtakeallpossibleprecautionstoavoidanymoisturefromenteringtheelectronicsenclosure.
Theenclosureshouldnotbeleftopenininclementweatherorforlongperiodsoftime,especiallyduringopera-
tionascondensationwillaccumulate.Lidshouldbetightlyscrewedshut,allconduitsshouldbesealedand
securedinaccordancewithregionalandclassicationrequirements,andunused3/4"NPTopeningsshouldbe
sealwithprovidedplug.Donotpowerwashtheunit.
OWDMANUAL 1114
KAM CONTROLS, INC.
19
FIG. 3-14 WIRING DIAGRAM OWD MODELS WITH RANGES OTHER THAN 0-100%
INSTALLATION CONTINUED
3939 Ann Arbor Drive
Houston, TX 77063 USA
Tel + 1 713 784 0000
Fax + 1 713 784 0001
www.Kam.com
Made in USA
V (+) in
AGND
V (-) in
AOUT0
AGND
AOUT1
DENSITY
DOUT
GND
GND
RS485 (+)
RS485 (-)
GND
DIN
GND
GND
RXDO
TXDO
TEC+
D2
24-30VDCPower
4-20mAOutput
4-20mAOutputor
analogvoltage
Alarm or Relay
Density Input
FactoryUse
RS485
RS232
FactoryUse
Main power
indicator
Sharedground
(RS485andRS232)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41