UEi SSM1 User manual

  • Hello! I am an AI chatbot trained to assist you with the UEi SSM1 User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
SSM1
Instruction Manual
Measures
•Pressurefrom0~500psia
•Temperaturefrom-40˚~400˚F
Calculates
•Superheatforlowside
•Subcoolforhighside
(TVX)Systems
Refrigerants:
•R22
•R134a
•R404a
•R410a
Super Heat / Sub Cool Meter
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
SSM1:
Super Heat / Sub Cool Meter
ThankyouforpurchasingtheUEiSSM1
Superheat/SubcoolMeter.UsetheSSM1to
measuresuperheatandsubcoolbydirectconnec-
tionsfortemperatureandpressure.TheSSM1is
programmedwiththemostcommonfourrefrigerants,
withdatafromtheNIST-Prorefrigerationdatabase
fortemperature-pressurevalues.
Technical Note
Superheatandsubcoolarethemosteffective
measurestodetermineifasystemisproperly
chargedandoperatingwithinmanufacture
guidelines.
1.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
What is Super Heat / Sub Cool?
Refrigerantsuperheatreferstotherisein
temperatureofvaporaftertherefrigeranthascom-
pletelychangedfromaliquidtoagas.Thedewpoint
(temperatureatmeasuredpressurewhereallofthe
liquidevaporates)isdeterminedbypropertiesofthe
refrigerantandmeasuringthepressure.Whencom-
paringthistothemeasuredpipetemperatureofthe
returnlinewilldeterminesuperheat.TheSSM1will
calculatethisvaluedirectlyandupdateinrealtime.
Usesuperheatonxedoriceorcapillarysystems.
Refrigerantsubcoolingreferstothedropin
temperatureofliquidrefrigerantafterithascom-
pletelychangedfromagastoaliquid.Theboiling
point(temperatureatmeasuredpressurewherethe
liquidstopsboilingandstartstofurthercool)isdeter-
minedbypropertiesoftherefrigerantandmeasuring
pressure.Whencomparingthistothemeasuredpipe
temperatureattheexitofthecondenserdetermines
subcooling.TheSSM1willcalculatethisvalue
directly,andupdateinrealtime.UseSubcooling
onTXVsystems,astheyaredesignedtomaintaina
constantsuperheat.
2.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
OPERATION
1. Presstopowerthemeteron
2. ZeroPressure:Whiledisplayingpressureholdthe
“Pressure/TEMP”buttonfor3seconds.
3. Pressthe“Pressure/TEMP”buttontoselectthe
pressureortemperaturemode.
4. RefrigerantType:Pressthe“RefrigType”buttonto
scrollthroughtheavailablerefrigeranttypes
(R22,R134a,R404A,R40A).
5. Press“”buttontoswitchthebacklightonoroff.
6. Press“TempSH/SC”whiledisplayingtemperatureto
scrollthrougheithertheprobetemperature,calculated
superheatorcalculatedsubcool.
7. Press“Scale”toselectdesiredunitsforpressureor
temperature.Theavailableunitsforpressureinclude
PSI,BAR,KPA,Kg/cm2,InHgandcmHg.Theavailable
unitsfortemperatureinclude°Fand°C.
3.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
4.
CONNECTIONS
1. Connectastandardrefrigeranthosetothe¼”are
connectorontopoftheSSM1.(Fig.1)-OR-Connectthe
SSM1in-linetoarefrigerantsourceusingthetee
tting(Fig.2)
2. Connecttheincludedpipeclampprobetothek-type
inputjackontopoftheSSM1.(Fig.1)
Fig.2Fig.1
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
MEASURING SUPERHEAT
1.Connectthepressuretothesuctionline
serviceport
2.Attachthepipeclamptemperatureprobetothe
suctionline(mustbeatleast6”fromthe
compressor)
1.
2.
5.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
5.Add/removerefrigerantbyconnectingtheSSM1
in-linetoarefrigerantsourceandmonitorpressure
whileaddingorremovingrefrigerant.
NOTE: Refer to manufacturers
specification and procedures
when adding / removing
refrigerant.
6.
3.Press“RefrigType”to
selecttheproper
refrigerant.Theselected
refrigeranttypewillshow
onthedisplay.*
4.Press“TEMPSH/SC”to
selectthedesired
parametertoview.
*
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
MEASURING SUBCOOL
1.Connectpressuretotheliquidlineserviceport
2.Attachthepipeclamptemperatureprobeonthe
condenserexitline
7.
1.
2.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
3.Press“RefrigType”to
selecttheproper
refrigerant.Theselected
refrigeranttypewillshow
onthedisplay.*
4.Press“TEMPSH/SC”to
selectthedesired
parametertoview.
8.
*
5.Add/removerefrigerantbyconnectingtheSSM1
in-linetoarefrigerantsourceandmonitorpressure
whileaddingorremovingrefrigerant.
NOTE: Refer to manufacturers
specification and procedures
when adding / removing
refrigerant.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
TEMPERATURE CALIBRATION
Aprocesstosettemperaturecalibrationtoanice
pointisincludedtocompensatefordifferencesin
temperaturesensors,ortoresettheunittoproper
operation.Pleasefollowthesestepstoperforma
temperatureset;
1. Connecttemperatureprobe
2. Press“PressureTEMP”todisplayprobe
temperature(verifythat“TEMP”isshownon
theLCD.
3. Placethetemperatureprobesensingareaina
mixtureofcrushediceandwater.(Distilled
waterwillprovidethemostaccuratesource)
4. Stirtheicewatermixturewiththetemperature
probeandallowthereadingtostabilize
5. Pressandhold“TempSH/SC”buttonforfour
seconds
Thiswillsettheicepointto32°F.Themeterwill
compensateforchangesupto10°F.
9.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com
10.
SPECIFICATIONS
• OperatingEnvironment:32~122°F
• StorageEnvironment:0~140°F
• Over-rangeindication:OL
• Autopoweroff:After30minutesidle
NOTE: Turn OFF when not in use to extend
battery life.
• Battery:Standard9Vbattery,NEDA1604,
IEC6F22
Function Resolution Accuracy
0~500psia 0.1psi ±0.5%FS+1
(2kPa,0.02bars) digit
-40~400°F 0.1° ±1°F(-40~200°F)
(-40~204°C) or±0.5°C(-40~93°C)
±2°F(201~400°F)
or±1°C(94~204°C)
www.GlobalTestSupply.com