Weinmann SOMNObalance e Instructions For Making Settings

Type
Instructions For Making Settings

This manual is also suitable for

SOMNObalance (e)
autoCPAP device with exhalation relief
SOMNOsoft 2 (e)
CPAP device with exhalation relief
Instructions for making settings
Contents
SOMNObalance (e) / SOMNOsoft 2 (e) EN 3
WM 67441 07/10
Contents
1 Introduction 4
2 Overview 4
2.1 Control panel ....................................................................... 4
2.2 Menu structure ..................................................................... 6
3 Physician menu 7
3.1 Navigate in the Physician menu ............................................ 7
3.2 Parameters in the Physician menu ......................................... 8
3.3 Disable softstart generally .................................................... 9
3.4 Activate exchange mode .................................................... 10
4 Therapy data 11
4.1 Save therapy data .............................................................. 11
4.2 Read out therapy data ........................................................ 11
4.3 Clear therapy data .............................................................. 11
Introduction
4 EN SOMNObalance (e) / SOMNOsoft 2 (e)
WM 67441 07/10
1 Introduction
This document explains how to set therapy parameters directly on
the therapy device. Alternatively, you can set the parameters using
the SOMNOadjust remote setting system or WEINMANNsupport
software. This document is not a complete set of instructions for
use. Follow the instructions for use for the therapy device.
2Overview
2.1 Control panel
1
2
3
4
Overview
SOMNObalance (e) / SOMNOsoft 2 (e) EN 5
WM 67441 07/10
Key
No. Element Function
1 On/off key
Switch the therapy device on and off.
Increase an adjustable value (+).
2 Softstart key
Activate/deactivate softstart
Call up softstart setting during therapy (keep
key depressed).
Open the Patient menu.
Access to Physician menu (keep key
depressed, then also keep humidifier key
depressed for 4 seconds)
Reduce an adjustable value (-).
3 Humidifier key
Activate/deactivate humidifier.
Set humidifier stage during therapy (keep key
depressed).
Scroll to next parameter of menu currently
open.
4 Info key
Open the info menu (with device on
standby).
Display current leakage (during therapy).
Save therapy data manually on the card.
(Keep key depressed).
Exit a menu.
Overview
6 EN SOMNObalance (e) / SOMNOsoft 2 (e)
WM 67441 07/10
2.2 Menu structure
P
CPAP
CPAP
APAP
APAP
Pmin
Pmax
Mode
SOMNObalance (e)
SOMNOsoft 2(e)
Pressure setting
softPAP lock: SLOC
Softstart pressure: P
Lock info menu ILOC
Time
Date
Clear therapy data:
CLEAR
Device
softPAP stage: soft
Speed of
pressure rise
Exchange mode CHAN
Parameter
Physician menu
SOMNObalance (e) / SOMNOsoft 2 (e) EN 7
WM 67441 07/10
3 Physician menu
3.1 Navigate in the Physician menu
1. Connect the device to the electricity supply.
The device is on standby.
2. Press the softstart key and keep it depressed.
The first parameter in the patient menu appears first (SoFT).
3. To reach the Physician menu, continue to keep the sofstart key
depressed and then also press the humidifier key .
After about 4 seconds, the first parameter starts flashing and
can be set.
4. Press the on/off key (
+) to increase the value.
5. Press the softstart key (-) to reduce the value.
Tip If you keep the relevant key depressed, the speed of adjustment
accelerates.
6. To scroll through the menu and switch to the next parameter,
press the humidifier key .
The therapy device adopts the setting.
7. To exit the Physician menu, press the info key.
Alternatively,
wait 15 seconds. If you do not press any key for 15 seconds, the
device automatically exits the Physician menu.
Physician menu
8 EN SOMNObalance (e) / SOMNOsoft 2 (e)
WM 67441 07/10
3.2 Parameters in the Physician menu
Parameter Display
on
device
Value range
Therapy mode
APAP
CPAP
CPAP, APAP (SOMNObalance (e) only)
If CPAP: CPAP
pressure
P
SOMNObalance / SOMNOsoft 2:
4 hPa - 18 hPa
SOMNObalance e / SOMNOsoft 2e :
4 hPa - 20 hPa
If APAP: lower
pressure limit
P
min
4 hPa-P
max
If APAP: upper
pressure limit
P
max
SOMNObalance / SOMNOsoft 2:
P
min
-18 hPa
SOMNObalance e / SOMNOsoft 2e :
P
min
-20 hPa
Rate of pressure
rise (APAP)
0. hPa - 0.4 hPa/s
softPAP level
Soft
0 (none/off)
1 (slight drop)
2 (normal drop)
softPAP lock
SLOC
0 (lock disabled, patient can change
softPAP setting)
1 (lock enabled, patient cannot
change softPAP setting)
Physician menu
SOMNObalance (e) / SOMNOsoft 2 (e) EN 9
WM 67441 07/10
3.3 Disable softstart generally
If the therapy requires it, you can also disable softstart. To do so, set
the softstart pressure of P to --.- hPa.
Softstart pressure P
APAP: 4 hPa - P
max
CPAP: 4 hPa - CPAP
Lock info menu
ILOC
0 (off)
1 (lock enabled; expanded info menu
is not displayed)
Activate
exchange mode
CHAN
0 (device not configured as an
exchange device)
1 (device configured as an exchange
device)
Time hh:mm
Date DD.MM.YYYY
Clear therapy
data
CLEAr
Parameter Display
on
device
Value range
Physician menu
10 EN SOMNObalance (e) / SOMNOsoft 2 (e)
WM 67441 07/10
3.4 Activate exchange mode
In the CHAN menu entry, you can configure a therapy device as an
exchange device. Once you have activated CHAn 1, CArd flashes in
the display and the therapy device cannot be switched on. The
patient needs to put in the SD card from his previous therapy
device. The therapy device automatically adopts the settings of the
previous therapy device.
If you are unable to activate exchange mode, take the
SD card out of the device and call up the Physician
menu again. The setting can now be adjusted.
In exchange mode, the therapy device automatically
creates a backup file containing the previous therapy
data on the SD card from the previous therapy device.
This file can be read by WEINMANNsupport.
If exchange mode is activated, the therapy device
cannot be switched on.
Exchange mode can be deactivated again via the
Physician menu if it has been activated inadvertently.
Therapy data
SOMNObalance (e) / SOMNOsoft 2 (e) EN 11
WM 67441 07/10
4 Therapy data
4.1 Save therapy data
In WEINMANNadjust you can set which therapy data are stored on
the SD card. Further details are described in the online help for
WEINMANNadjust.
4.2 Read out therapy data
You can read out and evaluate the data using WEINMANNsupport.
The instructions for use for the therapy device describe how to read
out the therapy data directly on the device.
To be able to read out the extended info menu (RDI, percentage
proportion of therapy with impermissibly high leakage and 90 %
pressure percentile), the lock on the info menu must be deactivated
(ILOC 0).
4.3 Clear therapy data
1. Call up the Physician menu.
2. Press the humidifier key until CLeAr appears in the display.
3. Press the on/off key and then also the humidifier key until
CLeAr flashes.
The data are deleted as soon as the display stops flashing. This
clearing step cannot be cancelled.
Weinmann Geräte für Medizin
GmbH+Co. KG
Postfach 540268
D-22502 Hamburg
Kronsaalsweg 40
D-22525 Hamburg
T: +49-(0)40-5 47 02-0
F: +49-(0)40-5 47 02-461
www.weinmann.de
Center for Production, Logistics,
Service
Weinmann Geräte für Medizin
GmbH+Co. KG
Siebenstücken 14
D-24558 Henstedt-Ulzburg
T: +49-(0)4193-88 91-0
F: +49-(0)4193-88 91-450
WM 67441 07/10
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12

Weinmann SOMNObalance e Instructions For Making Settings

Type
Instructions For Making Settings
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI