GASTEC 13TP Operating instructions

  • Hello! I've reviewed the instructions for the Gastec 13TP Carbon Disulphide Detector Tube. This document provides detailed information on its specifications, safe operation, measurement procedures, and interference considerations. I'm ready to assist with any questions you might have about using this detection tube.
  • What is the measuring range of the tube?
    What color change indicates detection?
    What is the recommended sampling rate?
    What interferences can affect the results?
MEASUREMENT PROCEDURE :
If the Automatic Air Sampling Pump Model GSP-300FT-2 or GSP-501FT is used.
1. Priortooperationpleaseconfirmifblackcolourinletrubbertubeholderisequippedwiththepump.
2. Breaktipsoffafreshprimarytubeandsecondarytubebybendingeachtubeendinthetubetip
breakerofthepump.
3. Connect©markedendswithrubbertubingafterbreakingeachend.
4. Insertthetubeintothepumpinletwitharrow()onthetubepointingtowardpump.
5. Settheflowrateto100mL/minandtimerto10minutesatthepump.Pressthestartswitchofthe
pumptostartthesampling.
6. Aftersamplingiscompleted,removethedetectortubefromthepump.
7. Readtheconcentrationlevelattheinterfacewherethestainedreagentmeetstheunstainedreagent.
Formeasurementshigherthan1.0ppm,prepareafreshtube.Settheflowmetreat50mL/minand
timerto“10minutes”ofthepumpandstartthesamplingagain.
8. Ifnecessary,multiplythereadingsbythecorrectionfactorsofsamplingrateandatmosphericpressure
respectively.
INTERFERENCES :
Substance Concentration Interference Change colour by itself to
Hydrogensulphide 0.1ppm Yellowishgreen(0.1ppm)
Carbonylsulphide 0.8ppm Yellowishgreen(0.8ppm)
Dimethyldisulphide 1.0ppm Yellowishgreen(1.0ppm)
MethyleneChloride 50ppm No Nodiscolouration
Acetone 500ppm No Nodiscolouration
Methanol 200ppm No Nodiscolouration
Sulphurdioxide 0.1ppm Yellowishgreen(0.1ppm)
 Thistableofinterferencegasesprimarilyexpressestheinterferenceofeachcoexistinggasinthe
gasconcentrationrange,thatisequivalenttothegasconcentration.Therefore,thetestresultmay
giveapositiveresultfromothersubstancesnotlistedinthetable.Ifmoreinformationisneeded,
pleasecontactusorGastecrepresentatives.
WHEN MEASURING OVER AN INTERVAL OF 15 MINUTES :
 ThisDetectorTubecanmeasurecontinuouslyfor15minutes,soitispossibletomeasurethe
averageconcentrationfor15minutesinonemeasurement.InthecasewherethisDetectorTubeis
usedtomeasureover15minutes,followtheMEASUREMENTPROCEDUREandusevaluesfrom
thetablebelowforsetSamplingRate,SamplingTime,andCorrectionFactor.
MeasuringRange (0.033)-0.66ppm 0.66-1.59ppm
SamplingRate 100mL/min 50mL/min
CorrectionFactor 0.66 1.59
SamplingTime 15min 15min
DANGEROUS AND HAZARDOUS PROPERTIES :
 ThresholdLimitValue-TimeWeightedAveragebyACGIH(2022):1ppm
INSTRUCTIONS ON DISPOSAL :
 Thereagentofthepretreatmenttubeusesasmallamountofhexavalentchromium.Thereagentof
thedetectortubedoesnotusetoxicsubstance.Whendisposingthetuberegardlessofwhetherithas
beenusedornot,followtherulesandregulationsofyourlocalgovernment.
WARRANTY :
 Ifyouhaveanyquestionsregardinggasdetectionandqualityofthetubes,pleasefeelfreetocontact
usoryourGastecrepresentatives.
FOR SAFE OPERATION :
 CarefullyreadthismanualandtheinstructionmanualofyourAirSamplingPumpbeforeuse.
CAUTION : If you do not observe the following precautions, you may suffer
injuries or damage the product.
1. Whenbreakingthetubeends,keepawayfromtheeyes.
2. Donottouchthebrokenglasstubes,piecesorreagentwithbarehands.
NOTES : For maintaining performance and reliability of the test results,
observe the following.
1. RecommendtouseGastecGasSamplingdeviceModelGSP-300FT-2orGSP-501FTtogetherwith
GastecDetectorTubesonlyforthepurposesspecifiedintheinstructionmanualofthedetectortube.
2. Usethistubewithinthetemperaturerangeof0-35(32-95).
3. Usethistubewithintherelativehumidityrangeof0-90%.
4. Thistubemaybeinterferedwithbythecoexistinggases.Pleaserefertothetable"INTERFERENCES"
below.
5. Shelflifeandstorageconditionofthetubesaremarkedonthelabeloftheboxoftubes.
APPLICATION OF THE TUBE : UsethistubeforthemeasurementofCarbondisulphideinair
orintheindustrialareasandenvironmentalatmosphericcondition.
SPECIFICATION : (BecauseofGastec’scommitmenttocontinuedimprovement,specificationsare
subjecttochangewithoutnotice.)
 TheMinimumScale(0.05ppm)isnotprintedonthetubeandisindicatedasaScaleLineonly.
MeasuringRange (0.05)-1.0ppm 1.0-2.4ppm
SamplingRate 100mL/min 50mL/min
CorrectionFactor 1 2.4
SamplingTime 10min 10min
DetectingLimit 0.04ppm(1000mL)
ColourChange BlueYellowishgreen
ReactionPrinciple CS2+I2O5+H2S2O7SO2+CO2
SO2+BaseChloride
Coefcient of Variation: 5% (for 0.05 to 1.0 ppm)
** Shelf Life: Please refer to the validity date printed on the box of tubes.
** Store the tubes in a cool and dark place.
CORRECTION FOR ENVIRONMENTAL CONDITIONS :
Temperature : Nocorrectionisrequiredbetween0-35°C(32-95°F).
Humidity : Nocorrectionisrequiredbetween0-90%R.H.
Pressure : Tocorrectforpressure,usetheformulabelow
  TubeReading(ppm)×1013(hPa)
  AtmosphericPressure(hPa)
 *
Thisvalueisafterothercorrection(s),ifany,areapplied.
No.13TP
Instructions for
Carbon Disulphide Detector Tube
Manufacturer:GastecCorporation
8-8-6Fukayanaka,Ayase-City,Kanagawa252-1195,Japan
https://www.gastec.co.jp/
Telephone+81-467-79-3910Facsimile+81-467-79-3979
IM0113TPE1
PrintedinJapan
22J1Z
Humidity scrubber
Detecting Layer
Oxidising agent Pretreatment layer
/