17
F09: Display the temperature in CO
2
sensor box
<Direction for
use>
Input “F09” on temperature display and press ENT key to show the
temperature in CO
2
sensor box. If you press ENT key or leave for 90seconds,
automatically returns to PV display.
In SV is +45°C, CO
2
valve turns OFF when the temperature is out of
SV±2.0°C.
Once the temperature retrieves in SV±2.0°C, Auto Zero is forcibly executed
since the apparatus falls into the condition that the main power is supplied.
When CO
2
is not injected, the temperature in sensor box may be out of
±2.0°C, or it may be within 60minutes after the main power is supplied, or
Auto Zero may be being executed.
F10: Display a coefficient of adjustment value for CO
2
zero point (initial: 500)
<Direction for
use>
Input “F10” on temperature display and press ENT key to display a coefficient
of adjustment value for CO
2
zero point on CO
2
display. If you press ENT key
or leave the device for 90seconds, automatically returns to PV display.
<Description> The condition is that CO
2
sensor voltage should be adjusted in 4.0V when
internal CO
2
level is 0%. Perform Auto Zero and ensure the sensor output
voltage should be approx. 4.0V when ambient air is got properly by the pump.
At that time the value should be approx. “500”.
If CO
2
in the ambient air is several %, the value would be lower than “500”
since CO
2
sensor output voltage is lower than 4.0V.
<Calculation>
Note)
(CO
2
sensor output voltage during Auto Zero pump performs / 4.0V) x 5000
Only bigger three digits are displayed. e.g “5000” → displayed as “500”
F11: Display a coefficient of adjustment value for CO
2
span point (initial: 500)
<Direction for
use>
Input “F11” on temperature display and press ENT key to display an
adjustment value for CO
2
zero point on CO
2
display. If you press ENT key or
leave the device for 90seconds, automatically returns to PV display.
<Description> The value is adjusted when CO
2
is calibrated. For example, calibrate CO
2
with
SV 4.5% when the CO
2
level is displayed as 5.0%, however the actual level is
4.5%. At that time the value in F11 should be lower than “500”, ex “460”. If the
value is abnormal, ex “999”, the actual internal level can be 1.0%, however
the CO
2
is displayed as 5.0%.
F12: Display a temperature in ambient temperature sensor
<Direction for
use>
Input “F12” on temperature display and press ENT key to display a
temperature of ambient temperature sensor on CO
2
display. If you press ENT
key or leave the device for 90seconds, automatically returns to PV display.
Note) This temperature is not synchronized with actual ambient temperature since it
is affected by waste heat in electrical parts. (indicated approx. 4°C higher than
the actual temperature)
F13: Initialize the non-volatile memory (initial: 000)
This function should be used only when data in the non-volatile memory
cannot be restored owing to unexpected events (ex: noise).
Do not use in normal condition.
<Direction for
use>
Input “F13” on temperature display and press ENT key to display “000” on
CO
2
display, then change the 1
st
digit to “1” and press ENT key to initialize all
the data in non-volatile memory.