6. Integrated Pneumatic Backup Test:
a. Verifypre-useassemblyowmeter
setto10L/min
b. TurnINOmaxDSIRbackupswitchON
c. Verify measured values
NO 14-26ppm
NO2≤<≤ 1.0ppm
d. TurnbackupswitchOFF
7. Performance Test:
a.VerifyO2owmeterissetto10L/min
b. SetINOMAX
doseto
40ppm
c. Verify
monitored
values
d. SetINOMAXdoseto0ppm
• “SetDoseisZero,PleaseCloseCylinderValve”
reminderwillappear-DONOTclosecylindervalveat
thistime
e. TurnoxygenowmeterOFF
8. INOblender Test:
a. Removeinjectormodulefrompre-useassembly
b. RemoveO2tubingfromowmeterandattachto
INOblenderoutlet
c. SetINOblenderowto10L/min,INOMAXdoseto
40ppm
d. ConrmacceptableNOvalue
NO 32-48ppm
e. SetINOblenderdoseandowto0
Pre-Use Checkout
WARNING:ThisreferencecardisnotasubstitutefortheINOmaxDSIRandINOblenderOperation
andMaintenancemanuals.Refertothesemanualsforallapplicablecautionsandwarnings.
AcceptableNOValue 35-45ppm
AcceptableNO2 Value <1.5ppm
FiO295% ± 3%
1. TurndeviceON,lowcalibrationwillbeginandcomplete
2. Initial Connections:
Conrmattachmentofthefollowing:
a. Waterseparatorcartridge,watertrapbottle,and
patientgassamplelineinplace
b. Injectormodulecableandtubingareconnected
c. PluginpowercordandverifyACpowerlightisON
d. Regulatorto800ppmINOMAXcylinder
e. RegulatorhosetoINOmaxDSIRinlet
f. INOblenderhoseconnectedandwhitelockinplace
g. Oxygensource(50psig)tobackofINOblender
h. IRcableinplace
3.Assemblepre-useset-upconnectors(diagramonback
ofcard).DonotturnO2owmeterONyet.
4. High Pressure Leak Test:
Open/closeINOMAX800ppmcylindervalve
a. Verify,atleast500psigcylinderpressure
b. Verify,nodecreaseincylinderpressurefor
30seconds
5. Manual Purge/Alarm Verication:
a. PressCANCELtoexitpre-usewizard.
b. VerifytheINOMAXcylindervalveisclosed.
c. SettheO2owmeterto10L/min
d. PurgeINOmaxDSIR.
•SettheINOMAXdoseto40ppm.
•“CylinderValveClosed”alarmwilloccur.
•Cylindergaugepressureshoulddropto0psig.
•MeasuredNO2willincreaseandthendecreaseas
NO2ispurgedfromthesystem.
•“LowCylinderPressure”alarmwilloccur.
e. OpentheINOMAXcylindervalve.
f. TurntheINOMAXdosetozero.The“SetDoseis
Zero,CloseCylinderValve”indicatorwillappear.
Thisindicatorwilldisplayanytimethesetdose
isreturnedtozero;however,duringthispre-use
procedure,leavethecylinderopenandtouchthe
screentoresettheindicator.
End of Procedure
If device is not used within 10 minutes, depressurize cylinder regulator
(Thefollowinginstructionsareprovidedforwhentheon-screenpre-usewizardisnotused.)
2e 2f
2g
2h
2a
8d
8e
8c
8c
8a
8a
8e
8b
6b
6d
Series 3 software