SALINITYSENSORML66M
USERSGUIDE
CENTRE FOR MICROCOMPUTER APPLICATIONS
http://www.cma-science.nl
2|ML66mSalinitySensorUser’sGuide
Shortdescription
TheSalinitysensorML66mmeasures thesalinityofasolutioninthe rangefrom 0to
50ppt.TheSalinitysensorconsistsofaSalinityelectrodeandanamplifier.
Salinityisthemeasureofallthesaltsdissolvedinwater.Salinityisoftenexpressedas
parts per thousand (ppt), which is
approximately equal to grams of salt per liter of
solution. The average ocean salinity is 35 ppt
1
and the averageriver water salinity is
0.5pptorless.
The Salinity sensor is an I2C digital sensor, which gives calibrated values of the
measured quantity. This sensor can only be connected to special interfaces that
supportI2CdigitalsensorsliketheCMAMoLabinterface.Thesensorcableneededto
connect the sensor to MoLab is not supplied with the sensor,
it is supplied with
MoLab.
Sensorspecifications
TheSalinitysensorML66misadigitalsensorthatconvertsthemeasuredsalinityvalue
toadigitalvaluevia16bitanalogtodigitalconversion.Themaximalsamplingrateof
thesensoris10Hz.
Howthesensorworks
The sensor determines salinity based on electrical conductivity. This Salinity probe
measures how much electrical current flows through the water. Water that has
dissolvedsaltinitwillconductelectricitybetterthanwaterwithnodissolvedsalt.The
more salt thatis dissolvedin thewater, the betterthe waterwill
conductelectricity.
Salinityisthencalculatedfromthemeasuredconductivityvalue.
The sensor has a built in compensation for changes in conductivity due to
temperature.
Calibration
The Salinity sensor is supplied with a factory calibration and gives values in ppt. The
Coach 6 program allows shifting the predefined calibration or creating a new two
point calibration if needed. The user calibration is stored in nonvolatile user sensor
memory.Formoreaccuratemeasurementssensorcanbe
calibrated.
Zerocalibrationpoint
Keepthesalinityprobeoutofanysolutione.g.,inair.Callthisvalue0ppt.
Standardsolutioncalibrationpoint
Placethesalinityprobeintoastandardsolutione.g.,35pptandwaitforthevalue
tostabilize.Enterthevalueofthestandard
solution35ppt.

1
35gramsofsaltper1literoceanwater
ML66mSalinitySensorUser’sGuide| 3
Toprepareasolutionwithsalinityvalueof35pptat25°C:
Pour500mLofdistilledwaterintoacontainer.
Add33.03gsodiumchloride(NaCl)and stirthemixtureuntilallsalthasdissolved.
Addenoughdistilledwatertomake1Lofsolution(1000mL).

Takingmeasurements
This Salinity sensor works only with specific interfaces. The sensor will be
automaticallydetectedwhenconnectedtosuchaninterface.Fordetailedinformation
aboutmeasurementswithsensorsconsulttheUserManualsoftheinterfaceandthe
Coach6software.
Totakemeasurements:
SoakthetipoftheSalinityelectrodein
distilledwaterforabout10minutes.Ifthis
isnotpossible,rinsethetipthoroughlywithdistilledwater.
Wipe the outer part of the electrode body with a clean paper towel. Shake
vigorouslytoremoveanydropletsfromthecellchamber.
PlacetheSalinityelectrodeinthesample
tobetested.Thesamplemustbeatleast
3cmdeeptoensurethecellchamberisfullysubmerged.
Stirthesolutiongently,waitfor10secondstoallowthereadingstostabilize.
Ifyouaretakingreadings inasolutionthathasatemperaturebelow10
o
Corabove
35
o
C,allowmoretimeforthereadingstostabilize.
Cleanthoroughlyafterthemeasurementiscompletedtoavoidanycontamination
fortheelectrode’snextuse.Theprobecanthenbestoreddry.
Warning: Do not place the electrode in viscous, organic liquids, such as heavy oils,
glycerin(glycerol)orethyleneglycol.Do notplacetheprobeinacetoneornonpolar
solvents,suchaspentaneorhexane.
UsingtheSalinitysensorwithMLsensors
It is very important to know that the Salinity sensor will interact with other ML
sensors, if they are placed in the same solution and are connected to the same
interface (e.g., the same MoLab). This situation arises because the Salinity sensor
outputs a
signal in the solution, and this signal can affect the reading of another
sensor.Thefollowingsensors cannotbeconnected tothesameinterfaceand placed
inthesamesolutions:
Dissolvedoxygensensor,
pHsensor,
Conductivitysensor,and
More sensors can be connected at thesame time
to the interface but only one at a
timecanbeplacedinsidethesolutiontotakereadings.
4|ML66mSalinitySensorUser’sGuide
Suggestedexperiments
TheSalinitysensorcanbeusedinthefollowingexperiments:
Measurementsofsalinityofwatersources.
Measurethechangeinsalinityofsaltwaterasthewaterevaporates.
Monitortherateofreactioninachemicalreactioninwhichdissolvedsaltionsand
solution conductivity varies with time due
to an ionic specie being consumed or
produced.
TechnicalSpecifications
Sensorkind Digital:onsensordigitalconversion
(16bitsresolution,communicationviaI2C)
Measuringrange 0..50ppt
Resolution ±0.00076ppt
Uncertainty(accuracy) Typical±2%aftercalibrationat25°C
Maximalsamplingrate 10Hz
Responsetime 98%offullscalereadingin5seconds
Temperaturerange 0°Cto80°C
Temperature
compensation
Automaticbetween5°Cand35°C
Cellconstant 10cm
1
Dimensions Sensorhousing:62x18x16mm
Connection 5pinsminijackplug
Warranty:
The Salinity sensor ML66m is warrante d to be free from defects in materials and
workmanshipfor a periodof12monthsfromthedateofpurchaseprovidedthatithas
been used under normal laboratory conditions. This warranty does not apply if the
sensorhasbeendamagedbyaccidentormisuse.

Rev. 10/06/2012
Note:Thisproductistobeusedforeducationalpurposesonly.Itisnotappropriatefor
industrial,medical,research,orcommercialapplications.