LM-900 Cryosurgery System • Directions for Use
4. During rinsing, use a clean, soft bristle brush to loosen and remove as much visible soil as
possible from device. Focus brushing in challenging features such as mating surfaces and
crevices.
5. Continue brushing and rinsing until no visible soil remains on the device.
6. Prepare a neutral pH enzymatic cleaner solution according to the manufacturer’s instructions.
Immerse a clean, disposable, absorbent, non-shedding cloth into the cleaner solution and
wring thoroughly.
7. Commencing with the gun shaft/casing joint, wipe thoroughly, especially in hard to
reach areas.
8. During wiping, pour enzymatic solution down the shaft so that it runs the length of the shaft to
the probe, but does not contact the shaft/casing joint. After pouring the solution, use a clean,
soft bristle brush to reach challenging features with the cleaner solution.
9. Continue wiping and brushing for a minimum of one (1) minute. Repeat pouring the enzymatic
solution during the brushing and wiping activities. Rinse the cloth in clean water and repeat
steps 7 & 8 two (2) or more times (for a total of three (3) iterations) to fully remove soils.
10. Thoroughly rinse the shaft and probe with warm (35°C-40°C) water until no visible detergent
remains. While rinsing, use soft bristle brushes to clean out mating surfaces and crevices
and other challenging features.
11. Using a soft, lint-free, clean cloth, pat the probe and shaft dry. Hold the gun facing downwards
and carefully remove the probe.
12. Wipe the protected areas and inside threads of the shaft and probe with a clean, disposable,
absorbent, non-shedding cloth that has been immersed in cleaner and appropriately wrung
out. Continue wiping until all visible soil is removed.
13. Wipe the protected areas and threads of the shaft and probe repeatedly with a clean,
disposable, absorbent, non-shedding cloth that has been immersed in clean water and
appropriately wrung out. Rinse towel after wiping and repeat until no visible detergent
remains.
14. Pat the probe and shaft dry using a soft, lint-free, clean cloth. Visually inspect the device for
soil. Repeat the above cleaning procedure until no visible soil remains on the components.
15. Apply isopropyl alcohol (70% or greater) to a soft, lint-free, clean cloth (Do not saturate). Wipe
down the entire cryosurgical system (probe, shaft, gun, handle, tubing, console, etc.) for a
minimum of one (1) minute.
16. Dry the components using clean compressed air or a soft, lint-free, clean cloth.
DISINFECTION INSTRUCTIONS
Intermediate Level Disinfection of Cryosurgical Shaft
1. Instruments should be cleaned as soon as possible. Do not allow blood and debris to dry
on the instrument. If cleaning cannot be performed within a reasonable amount of time,
keep the shaft moist.
2. Prior to performing Intermediate Level Disinfection of the cryosurgical gun shaft, cleaning
must be performed following Cleaning Instructions.
3. Wearing appropriate protective equipment, hold the cryosurgical gun by the handle with
the shaft pointing downwards at an angle.
4. Obtain a clean, disposable, absorbent, non-shedding cloth that has been submerged in a
prepared high level disinfectant solution (Cidex OPA or equivalent) and wrung out so that
dripping has ceased.
5. Begin wiping the shaft surface up and down the entire length of the shaft. Wipe the inner
edge where the probe and shaft meet.
6. Continue wiping the surface of the shaft for one (1) minute. Re-submerge the cloth into the
prepared high level disinfectant solution, wring out until dripping has ceased and repeat
Step 5.
5