Tektronix TPS2014B User manual

Type
User manual

This manual is also suitable for

xx
TPS2000B Series
Digital Storage Oscilloscopes
ZZZ
Service Manual
*P077044701*
077-0447-01
TPS2000B Series
Digital Storage Oscilloscopes
ZZZ
Service Manual
xx
This document supports rmware version 1.00 and above.
Warning
The servicing instructions are for use by qualied personnel
only. To avoid personal injury, do not perform any servicing
unless you are qualied to do so. Refer to all safety summaries
prior to performing service.
Revision A
www.tektronix.com
077-0447-01
Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its subsidiaries
or suppliers, and are protected by national copyright laws and international treaty provisions.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication
supersedes that in all previously published material. Specications and price change privileges reserved.
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Contacting Tektronix
Tektronix, Inc.
14150 SW Karl Braun Drive
P.O. Box 500
Beaverto
n, OR 97077
USA
For product information, sales, service, and te chnical support:
In North America, call 1-800-833-9200.
Worldwide, visit www.tektronix.com to nd contacts in your area.
Table of Contents
General Safety Summary .......................................................................................... v
Service Safety Summary......................................................................................... vii
Preface .............................................................................................................. ix
Products........................................................................................................ ix
Related Manuals .............................................................................................. ix
Probe Safety .................................................................................................... x
Specications ....................................................................................................... 1
Signal Acquisition System......... ........................ ........................ ........................ ... 1
Time Base System ............................................................................................. 3
Triggering System ............................................................................................. 4
Display.......................................................................................................... 7
Batteries......................................................................................................... 7
External DC Input ............................................................................................. 8
External Interfaces............................................................................................. 8
Mechanical Characteristics ................................................................................... 9
Environmental Performance.................................................................................. 9
Operating Information ............................................................................................ 11
Theory of Operation............................................................................................... 13
Acquisition Board ............................................................................................ 14
Power Supply and Battery System.......................................................................... 16
Display Module ............................................................................................... 16
Front Panel .................................................................................................... 16
Performance Verication ......................................................................................... 19
Required Equipment.......................................................................................... 19
Test Record .................................................................................................... 20
Self Test................................................................................................... 23
Self Calibration........................................................................................... 23
Check DC Gain Accuracy............................................................................... 24
Check DC Measurement Accuracy .................................................................... 25
Check Bandwidth ........................................................................................ 27
Check Sample Rate and Delay Time Accuracy....................................................... 28
Check Edge Trigger Sensitivity ................ .................. ...................................... 29
Check External Edge Trigger Sensitivity................ ................................ .............. 30
Adjustment Procedures ........................................................................................... 33
Required Equipment.......................................................................................... 33
Adjustment Procedure........................................................................................ 34
Maintenance........................................................................................................ 41
Preparation..................................................................................................... 41
Preventing ESD ............................................................................................... 41
TPS2000B Series Digital Storage Oscilloscopes Service Manual i
Table of Contents
Inspection and
Cleaning...................................................................................... 42
Removal and Installation Procedures....................................................................... 45
Rear Case ................................................................................................. 47
Fan......................................................................................................... 49
I/O Board ................................................................................................. 50
Acquisition Board...... .......... .......................... .......................... .......... .......... 51
Display Ada
pter Board .................................................................................. 52
Front-Panel Knob ........................................................................................ 54
Front Case ................................................................................................ 54
Front-Panel Board ....................................................................................... 56
Display Module .......................................................................................... 57
Compact Flash Board.................................................................................... 58
Keypad .................................................................................................... 5
8
Troubleshooting............................................................................................... 60
Troubleshooting Tree.................................................................................... 61
Lithium-Ion Battery Maintenance .......................................................................... 66
Repackaging Instructions .................................................................................... 69
Diagrams ........................................................................................................... 71
Repl
aceable Parts .................................................................................................. 73
Parts Ordering Information .................................................................................. 73
Using the Replaceable Parts List............................................................................ 74
ii TPS2000B Series Digital Storage Oscilloscopes Service Manual
List of Figures
Figure 1: TPS2000B Series module-level block diagram. Remove CH3 and CH4 for a 2-channel
version ......................................................................................................... 13
Figure 2: Adjustment setups ..................................................................................... 35
Figure 3: Re
moving and installing the rear case ............. ................................ .................. 48
Figure 4: M odules accessed from the rear of the oscilloscope ............. .............. .............. ...... 50
Figure 5: Using a screwdriver to remove the Acquisition board ... ...................... .................... 52
Figure 6: Removing the BNC cover, shell, and retaining nuts ............................................... 54
Figure 7: Removing the front case .............................................................................. 55
Figure 8: Removing the Front-Panel module................................................................... 56
Figure 9
: Removing the Display module ....................................................................... 57
Figure 10: Removing and installing the keypad ............................................................... 59
Figure 11: Oscilloscope troubleshooting tree (page 1 of 4)....................... ............................ 61
Figure 12: Oscilloscope troubleshooting tree (page 2 of 4)................................................... 62
Figure 13: Oscilloscope troubleshooting tree (page 3 of 4)................................................... 63
Figure 14: Oscilloscope troubleshooting tree (page 4 of 4)................................................... 64
Figu
re 15: TPS2000B Series block diagram ................................................................... 71
Figure 16: Exploded view diagram.............................................................................. 78
TPS2000B Series Digital Storage Oscilloscopes Service Manual iii
Table of Contents
List of Tables
Table 1 : Signal acquisition system ...................... .................. .................. ..................... 1
Table 2: Tim
e base system ........................................................................................ 3
Table 3: Triggering system ........................................................................................ 4
Table 4: Display .................................................................................................... 7
Table 5: Batteries ................................................................................................... 7
Table 6: External DC input........................................................................................ 8
Table 7: External interfaces ....................................................................................... 8
Table 8: M
echanical characteristics .............................................................................. 9
Table 9: Environmental performance ............................................................................ 9
Table 10: Required equipment ................................................................................... 19
Table 11: Test record.............................................................................................. 20
Table 12: Required equipment ................................................................................... 33
Table 13: Adjustments (steps 73-108 are for 4-CH models only)............................................ 37
Table
14: Internal inspection checklist.......................................................................... 43
Table 15: List of error codes ..................................................................................... 65
Table 16: Parts list column descriptions ........................................................................ 74
Table 17: Replaceable parts list.................................................................................. 75
iv TPS2000B Series Digital Storage Oscilloscopes Service Manual
General S afety Summary
General Safet
ySummary
Review the fo
llowing safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
To avoid pot
ential hazards, use this product only as specied.
Only qualied personnel should perform service procedures.
While using this product, you may need to access other parts of a larger system.
Read the s afety sections of the other component manuals for warnings and
cautions r
elated to operating the system.
To Avoid Fire or Persona l
Injury
Use proper power cord. Use only the power cord specied for this product and
certied for the country of use.
Connect and disconnect properly. Do not connect or disconnect probes or test
leads while they are connected to a voltage source.
Connect and disconnect properly. Connect the probe output to the measurement
instrument before connecting the probe to the circuit under test. Connect the
probe reference lead to the circuit under test before connecting the probe input.
Disconnect the probe input and the probe reference lead from the circuit under test
before disconnecting the probe from the measurement instrument.
Observe all terminal ratings. To avoid re or shock hazard, observe all ratings
and markings on the product. Consult the product manual for further ratings
inf
ormation before making connections to the p roduct.
Do not apply a potential to any terminal, including the common terminal, that
ex
ceeds the maximum rating of that terminal.
Do not operate without covers. Do not operate this product with covers or panels
re
moved.
Do not operate with suspected failures. If you suspect that there is damage to this
p
roduct, have it inspected by qualied service personnel.
Avoid exposed circuitry. Do not touch exposed connections and components when
power is present.
Replace batteries properly. Replace batteries only with the specied type and
rating.
Recharge batteries properly. Recharge batteries for the recommended charge cycle
only.
Use proper AC adapter. Use only the AC adapter specied for this product.
Wear eye protection. Wear eye protection if exposure to high-intensity rays or
laser radiation exists.
TPS2000B Series Digital Storage Oscilloscopes Service Manual v
General Safety Summary
Do not operate i
n wet/damp conditions.
Do not operate in an explosive atmosphere.
Keep product surfaces clean and dry.
Provide prop
er ventilation. Refer to the manual's installation instructions for details
on installing the product so it has proper ventilation.
TermsinThisManual
These terms may appear in this manual:
WARNING.
Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION
. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Symbols and Terms on the
Product
These t
erms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you rea d
the ma
rking.
WARNING indicates an injury hazard not immediately accessible as you
read
the marking.
CAUTION indicates a hazard to property including the product.
The following symbol(s) may appear on the product:
vi TPS2000B Series Digital Storage Oscilloscopes Service Manual
Service Safety Summary
Service Safet
y Summary
Only qualie
d personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering rst aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, switch off the instrument power, then
disconnect the power cord from the mains power.
UseCareWhenServicingWithPowerOn. Dangerousvoltagesorcurrentsmay
exist in
this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing p rote ctive panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
TPS2000B Series Digital Storage Oscilloscopes Service Manual vii
Service Safety Summary
viii TPS2000B Series Digital Storage Oscilloscopes Service Manual
Preface
This service manual for the TPS2000B Series Digital Storage Oscilloscopes
provides instructions to verify the performance of, calibrate, troubleshoot, and
repair the os
cilloscopes to the module level.
Unless noted otherwise, the term "oscilloscope" refers to all of the models in
the TPS2000
BSeries.
Products
This manu
al covers the following products:
TPS2012B Digital Storage Oscilloscope
TPS2014B Digital Storage Oscilloscope
TPS2024B Digital Storage Oscilloscope
Related Manuals
These manuals contain additional documentation for the oscilloscopes. They are
available at www.tektronix.com/manuals.
Language User manual p art number
Progr
ammer manual part
number
Engl
ish
071-
2733-XX
French 071-2734-XX
1
Italian 071-2735-XX
1
German
071
-2736-XX
1
Spanish
071
-2737-XX
1
Ja
panese
07
1-2738-XX
1
Portuguese 071-2739-XX
1
Simplied Chinese
071-2740-XX
1
Traditional Chinese
071-2741-XX
1
Korean 071-1450-XX
1
Russian 071-2743-XX
1
077-
0444-XX
1
These manuals contain a language overlay for the front-panel controls.
TPS2000B Series Digital Storage Oscilloscopes Service Manual ix
Preface
Probe Safety
A guard around the probe body provides a nger barrier for protection from
electric shock.
WARNING. To avoid electric shock when using the probe, keep ngers behind the
guard on the probe body, and do not touch metallic portions of the probe head
while it i
s connected to a voltage source.
Connect the probe output to the oscilloscope before connecting the probe to
the cir
cuit under test. Disconnect the probe input and the probe reference lead
from the circuit under test before disconnecting the probe from the measurement
instrument.
x TPS2000B Series Digital Storage Oscilloscopes Service Manual
Specications
These specications apply to all TPS2000B Series oscilloscopes. To verify that an
oscilloscope meets specications, it must rst meet the following conditions:
The oscilloscope must have been operating continuously for twenty minutes
within the specied operating temperature range. (See Table 9.)
You must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C.
The oscilloscope must be within the factory calibration interval of one year.
The V/div specications apply with the probe calibration set to 1X probe
mode.
Specications begin in the following table. All specications are guaranteed
unless noted "typical." Specications that are marked with the
symbol are
checked in the Performance Verication section.
S
ignal Acquisition System
Table 1: Signal acquisition system
Characteristic Description
Number of Input Channels Two (TPS2012B) or four (TPS2014B, TPS2024B) identical channels, digitized simultaneously
Input Coupling DC, AC, or GND
AC coupling connects a capacitor in series with the input circuitry. The DC input impedance
becomes very high, since capacitance is in series w ith all paths to ground.
Ground coupling mode provides a reference waveform derived from the values identied during
SPC. This reference waveform shows visually where signal reference (ground) is expected
to be.
Input Impedance, DC Coupled 1 M ±2%inparallelwith20p3pF
Overvoltage Category
Maximum Voltage
CAT I and CAT II
300 V
RMS
CAT III
150 V
RMS
Maximum Input Voltage (between
Signal and Reference at input
BNC)
1
At front panel connector, 300 V
RMS
; derate at 20 dB/decade above 100 kHz to 13 V peak
AC at 3 MHz and above.
All inputs are rated 300 V CAT II, signal to oating common.
Derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and above. For
nonsinusoidal waveforms, peak value must be less than 450 V. Excursion above 300 V should
be less than 100 ms duration. RMS signal level, including any DC component removed through
AC coupling, must be limited to 300 V. If these values are exceeded, damage to the oscilloscope
may result. Refer to the Overvoltage Category description. (See Table 9 on page 9.)
Maximum Float Voltage
Rated 600 V
RMS
CAT II.
Between instrument common (chassis) and BNC shell. Using rated connectors or accessories.
Number of Digitized Bits 8-bit resolution (except when set to 2 mV/div), each channel sampled simultaneously.
TPS2000B Series Digital Storage Oscilloscopes Service Manual 1
Specications
Table 1: Signal acquisition system (cont.)
Characteristic Description
Supported Voltage Probe
Attenuation Factors
1X, 10X, 20X, 50X, 100X, 500X, 1000X
This adjusts the display scale factor of the instrument to accommodate various probe types.
Accuracy of the probe used must be added to the accuracy specications of instrument.
The Tektronix probe Interface is not provided, so the user m ust adjust the setting to match
the probe characteristics.
Supported Current Probe Scales 1 mV/mA, 5 mV/mA, 1 mV/A, 10 mV/A, 20 mV/A, 50 mV/A, 100 mV/A, 200 mV/A
This adjusts the display scale factor of the instrument to accommodate various probe types.
Accuracy of the probe used must be added to the accuracy specications of instrument.
The Tektronix probe Interface is not provided, so the user m ust adjust the setting to match
the probe characteristics.
Acquisition Modes
Sample, Peak Detect, and Average
Envelope mode not provided
±3% for Sample or Average acquisition mode, 5 V/div to 10 mV/div
DC Gain Accuracy
±4% for Sample or Average acquisition mode, 5 mV/div and 2 mV/div
Measurement Type Accuracy
Average of 16 waveforms with
vertical position at zero
±(3% × |reading| + 0.1 div + 1 mV)
Average of 16 waveforms with
vertical position not at zero for 2 mV
to 200 mV/div.
±[3% × (|reading + vertical position|) + 1 % of |vertical
position| + 0.2 div + 2 mV]
DC Measurement Accuracy,
Avera
ge Acquisition Mode
Average of 16 waveforms with
vertical position not at zero for >200
mVto5V/div.
±[3% × |reading + vertical position| + 1% of |vertical
position| + 0.2 div + 50mV]
Change in measured volts between any two averages of 16 waveforms acquired under the
same setup and ambient conditions.
DC Voltage Measurement
Repeatability, Average Acquisition
Mode
(3% of |reading| + 0.05 div)
Analog Bandwidth
Dened in Section 4.6 of IEEE std 1057. The difference between the upper and lower
frequencies, at which the amplitude response, as seen in the data record, is 0.707 (-3 dB) of
the response seen in the data record at the specied reference frequency. Specied only –3 dB
point, does not include in-band response.
TPS2012B and TPS2014B TPS2024B
Analog Bandwidth in Sample
and Average modes at BNC or with
probe set to 10X, DC Coupled,
typical
DC to 100 MHz for all V/div settings
with bandwidth limit at full.
DC to 200 MHz for all V/div settings 5 mV and above,
for operating temperatures from 0°C to 40 °C.
DC to 180 MHz for all V/div settings 5 mV/div and
above, for operating temperatures from 40 °C to
50 °C
Analog Bandwidth in Peak Detect
mode (50 s/div to 5 m s/div), DC
coupled, typical
DC to 75 MHz for all V/div settings with bandwidth limit at full.
Lower Frequency Limit, AC
Coupled
10 Hz at BNC
1 Hz when using a 10X passive probe
2 TPS2000B Series Digital Storage Oscilloscopes Service Manual
Specications
Table 1: Signal acquisition system (cont.)
Characteristic Description
TPS2012B and TPS2014B TPS2024B
<3.5ns <2.1ns
Rise Time at BNC, typical
Rise time is generally calculated from the following formula: Rise time in ns = 350 / Bandwidth
in MHz
Peak Detect Response
Captures 50% or greater amplitude of pulses 12 ns wide typical (50 s/div to 5 μs/div) in the
center 8 vertical divisions
Vertical Position Range
2 mV/div to 200 mV/div, ±1.8 V
>200 mV/div to 5 V/div, ± 45 V
Vertical Position Accuracy
2 mV/div to 200 mV/div, ±(1% × selected value +0.1 div + 2 mV) within the range ±1.8 V
>200 mV/div to 5 V/div, ±(1% × selected value + 0.1 div + 50 mV) within the range ±45 V
Greater than 1000:1 up to 50 MHz, derated to 400:1 at 200 MHzSingle Channel Common Mode
Rejection, typical
Sine wave with VOLTS/DIV setting at 5 mV. Signal applied between the channel (signal and
signal reference) and chassis. Ratio of the acquired signal amplitude to the amplitude of
the signal.
100:1 at 60 Hz, reducing to 10:1 with 50 MHz sine wave, w ith equal volts/div and coupling
settings on each channel.
Channel-Channel Common Mode
Rejection Ratio, typ ical
With the same signal applied to each channel, this is the ratio o f the acquired signal amplitude
to the amplitude of the MATH difference waveform, CH1–CH2, CH2–CH1, CH3–CH4, or
CH4–CH3.
TPS2012B and TPS2014B TPS2024B
50 MHz:100:1 100 MHz:100:1
Channel-to-Channel Crosstalk
Measured on one channel, with test signal applied between signal and common of the other
channel, and with the same VOLTS/DIV and coupling settings on each channel.
Waveform acquisition Rate, typical Up to 180 waveforms per second, per channel (Sample acquisition mode, no measurements)
1
The maximum peak-to-peak voltage that can be displayed is 40 V
p-p
with 1x attenuation.
Time Base System
Table 2: Time base system
Characteristic Description
TPS2012B and TPS2014B TPS2024BSample Rate Range
5S/sto1GS/s 5S/sto2GS/s
Waveform Interpolation (sin x)/x
Waveform interpolation is activated for sweep speeds of 100 ns/div and faster.
Record Length
2500 samples for each channel
TPS2012B and TPS2014B TPS2024BSeconds/Division R ange
5 ns/div to 50 s/div, in a 1, 2.5, 5 sequence 2.5 ns/div to 50 s/div, in a 1, 2.5, 5 sequence
Sample Rate and Delay Time
Accuracy
±50 ppm over any 1mstimeinterval
TPS2000B Series Digital Storage Oscilloscopes Service Manual 3
Specications
Table2: Timebasesystem(cont.)
Characteristic Description
2.5ns/divto10ns/div:(-4div×s/div)to20ms
25 ns/div to 100 ms/div: (-4 div × s/div) to 50 ms
250 μs/div to 10 s/div: (-4 div × s/div) to 50 s
Position Range
25 s/div to 50 s/div: (-4 div× s/div) to 250 s
Conditions
Accuracy
Single-shot, sample mode ±(1 sample interval + 100 ppm × |reading|
+0.6ns)
>16 averages
±(1 sample interval + 100 ppm × |reading|
+0.4ns)
Delta Time Measurement Accuracy
(Full Bandwidth)
The limits apply for signals having amplitude > 5 divisions, slew rate at the measurement
points of > 2.0 divisions/ns, and acquired > 10 mV/Div.
Sample interval is the time between the samples in the waveform record.
Triggering System
Table 3: Triggering system
Characteristic Description
Holdoff Range
500 ns to 10 s
The ability to set large values of Holdoff is limited by the difculty in adjusting the Holdoff
at Sec/Div settings less than 100 ms/div. This is because Holdoff cannot be set in Scan
Mode, which begins at 100 ms/div when Trigger Mode is AUTO. By adjusting Trigger Mode
to NORMAL, the Scan Mode operation is turned off, and Holdoff can be adjusted at larger
Sec/Div settings.
External Trigger Input Impedance
1M±2%inparallelwith20p3pF
External Trigger Maximum Input
Voltage
300 V
RMS
CAT 1 and CAT II
150 V
RMS
CAT III
At front panel connector, signal input to BNC shell, 300 V
RMS
; derate at 20 dB/decade above
100 kHz to 13 V peak AC at 3 MHz and above.
Based upon sinusoidal or DC input signal. For nonsinusoidal waveforms, peak value must be
less than 420 V. Excursion above 300 V should be less than 100 ms duration. RMS s ignal level
must be limited to 300 V. If these values are exceeded, damage to the instrument may result.
Input is rated 300 V CAT II, signal to oating common.
External Trigger Maximum Float
Voltage
Rated 600 V CAT II.
Instrument chassis is normally considered to be at earth ground.
Between instrument chassis and BNC shell
4 TPS2000B Series Digital Storage Oscilloscopes Service Manual
Specications
Table 3: Triggering system (cont.)
Characteristic Description
Trigger Source Sensitivity
All products
1 div from DC to 10 MHz
TSP2012B
TPS2014B
1.5 div from 10 MHz to 100 MHz
TSP2024B 1.5 div from 10 MHz to 100 MHz
2 div from 100 MHz to 200 MHz
EXT
1V
p-p
from 50 Hz to 100 MHz
2V
p-p
from 100 MHz to 200 MHz
EXT/5 5X the value of EXT
EXT/10 10X the value of EXT
Trigger Sens
itivity, Edge Trigger
Type. DC Coupled
The minimum signal levels required for stable edge triggering of an acquisition when the trigger
Source is DC coupled. A stable trigger is one that results in a uniform, r egular visual display,
triggered on the selected slope. The trigger point must not switch between opposite slopes on
the waveform, and the display must not “roll” across the screen on successive acquisitions.
Trigger Frequency Readout typically stabilizes at 25% more signal than generates a stable
visual display. At 5 mV/div, 50% additional signal is typically required.
Coupling Sensitivity
AC Same as DC-coupled limits from 50 Hz and above
NOISE REJ Reduces the DC-coupled trigger sensitivity by 2 times for greater
than 10 mv/div to 5 V/div
HF REJ
Same as the DC-coupled limit from DC to 7 kHz, attenuates signals
above 80 kHz
Trigger Sensitivity, Edge Trigger
Type, typical
LF REJ
Same as the DC-coupled limits for frequencies above 300 kHz,
attenuates signals below 300 kHz
SET LEVEL TO 50%, typical Operates with input signals equal to or greater than 50 Hz
Using a 10 M probe will not affect the operation of this function.
Source
Range
CH1, CH2, CH3
2
,
CH4
2
±8 divisions from center of screen
EXT ±4 V
EXT/5
±20 V
EXT/10
±35 V
Trigger Level Range
The adjustable resolution for Trigger Level is 0.02 division for Ch1 or Ch2 source, (and Ch3,
Ch4 for the TPS2014B, TPS2024B)
20 mV for Ext source
100 mV for Ext/5 source
200 mV for Ext/10 source
TPS2000B Series Digital Storage Oscilloscopes Service Manual 5
Specications
Table 3: Triggering system (cont.)
Characteristic Description
Accuracies are for signals having rise and fall times 20 ns
Source
Accuracy
Internal
±(0.2div+5mV)withi4divisions from center screen having rise
and fall times of >20 ns.
EXT
±(6% of setting + 250 mV) for signals less than ±2 V
EXT/5 ±(6% of setting + 500 mV) for signals less than ±10 V
EXT/10 ±(6% of setting + 1 V) for signals less than ±20 V
Trigger Level Accuracy, typical
A sine wave with 20 ns rise time corresponds to about 18 MHz.
Default Settings, Video Trigger Coupling is AC and Auto except for a single sequence acquisition
Signal Formats and Field Rates,
Video Trigger Type
Supports NTSC, PA L, and SECAM broadcast systems for any eld or any line
Composite video signal
Source
Range
Internal
Peak-to-peak amplitude of 2 div
EXT
1 V of composite video
EXT/5 5 V of composite video
EXT/10 10 V of composite video
Sensitivity, Video Trigger Type,
typical
A 2 division composite video signal will have a 0.6 division sync tip.
Pulse Width Trigger
Pulse Width Tr igger modes
Trigger when < (Less than), > (Greater than), = (Equal), or (Not Equal)
Positive pulse or Negative pulse
Pulse Width Trigger Point
Equal: The oscilloscope triggers when the trailing edge of the pulse crosses the trigger level.
Not Equal: If the pulse is narrower than the specied width, the trigger point is the trailing
edge. Otherwise, the oscilloscope triggers when a pulse continues longer than the time
specied as the Pulse Width.
Less than: The trigger point is the trailing edge.
Greater than (also called time-out trigger): The oscilloscope triggers when a pulse continues
longer than the time specied as the Pulse Width.
Pulse Width Range
Selectable from 33 ns to 10 s
Pulse Width Resolution 16.5 ns or 1 part per thousand, whichever is larger
Equal Guard Band t > 330 ns: ±5% guard band < ± (5.1% + 16.5 ns)
t 330 ns: guard band = ±16.5 ns
Not Equal Guard Band t > 330 ns: ±5% guard band < ±(5.1% + 16.5 ns)
165 ns < t 330 ns: guard band = -16.5 ns ±33 ns
t 165 ns: guard band = ±16.5 n s
Trigger Frequency Counter
Readout Resolution 6 digits
Accuracy, typical
±51 ppm including all frequency reference errors and ±1 count errors
6 TPS2000B Series Digital Storage Oscilloscopes Service Manual
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92

Tektronix TPS2014B User manual

Type
User manual
This manual is also suitable for

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI