Roche cobas e 602 Short Guide

Type
Short Guide
cobas
®
8000 modular analyzer series
Short Guide
Version 2.0
25549 Titel und Rück. ShortGuide.indd 1 30.06.10 15:04
1
Ta ble of Contents
Table of Contents 1
System Overview - Workflow Guide 2
Starting Inspection 2
How to select calibrations and control
s3
Requesting calibrations and controls using Overview screen 3
Requesting calibrations in Calibration Status screen 4
QC after calibration 4
Requesting controls in QC Status screen 4
Start Routine 5
End of Routine 5
How to install a new application 6
Loading new application parameters 6
Test Parameter Setting 7
Test Assignment 7
Test Key Assignment 8
Configuration in Utility 8
Set up Calibrators 9
Set up Calibrators manually (for chemistry only) 10
Set up Roche Controls 10
Edit target values and ranges for Roche controls 11
Edit target values and ranges for non Roche controls 11
Set up non Roche Controls 12
Rack Assignment for non barcoded calibrators and controls 13
2
System Overview - Workflow Guide
Starting Inspection
Load reagents and auxillary reagents onto the analyzer according to reagent load list.
Remove all empty packs from the system.
After the Reagent Prime for ISE, recalibrate the ISE unit before you resume routine operation.
Check system alarms.
Select beginning of the day maintenance.
Select Clear or Backup and Clear.
Activate the Preventive Action check box, if not selected.
Print load list.
Perform Prime for ISE. Select all.
3
How to select calibrations and controls
Requesting calibrations and controls using Overview screen
Calibration area
QC area
Print load list.
Load calibrator and QC racks on the analyzer according to calibrator and QC load list.
4
Requesting calibrations in Calibration Status screen
QC after calibration
o Insert the QC rack directly after the calibrator racks.
Requesting controls in QC Status screen
Select test.
Select calibration method e.g. .
Calibration request is active when bar is blue .
Print load list.
Select test
or select for all active Reagent Packs.
Select for Stand by QC.
Select test
Print load list.
5
Start Routine
End of Routine
Deselect rack reception mode, if necessary.
Load green rack:
Pos. 1 MultiClean
Pos. 2 SysClean
Pos. 3 Activator
Load racks with samples.
Area Rack Reception Mode
Deselect Rack Reception Mode.
6
How to install a new application
Loading new application parameters
Activate the check box of parameter to be downloaded.
Check if the parameters are downloaded.
Select one of the available search conditions.
Select the unit, if necessary:
c 701 / c 502: For first installation parameters for dilution and cleaner reagent packs have to be downloaded.
7
Test Parameter Setting
Test Assignment
Select
Optional settings Define decimal places, select Automatic
Rerun, define Control Interval time,
AutoQC on board stability, Repeat Limit,
Qualitative fields.
Select ISE module. Select Na, K, Cl or Na, K.
Select c 701 module.
Select checkbox ,
Select c 502 module.
Select checkbox
Select c 602 module.
Select checkbox
8
Test Key Assignment
Configuration in Utility
Select
Select a test.
Select option for Routine/Stat, Calibrator and Control.
Select checkmark and Hours.
Define cycles for wash rack: photometric: 5, ISE: 15.
9
Set up Calibrators
Load reagent packs on the module.
Select Expired Reagent Flag.
Select Auto Masking. Select Reagent Information Printout.
Select Print Dilution Ratio.
Select one of the available search conditions.
Activate the check box of the parameter to be downloaded.
10
Set up Calibrators manually (for chemistry only)
Set up Roche Controls
Load reagent packs on the module.
Select an empty line.
Type in calibrator name, code, lot no., and expiration date.
Type in the concentration values.
Select one of the available search conditions.
Activate the check box of the parameter to be downloaded.
11
Edit target values and ranges for Roche controls
Log on to data manager.
Select Control Material.
Edit target values and ranges for non Roche controls
Log on to data manager.
Select Control Material.
Select
Highlight the control material you want to edit.
Select
Highlight the test you want to edit.
Select Select .
Type in the value.
Select
Highlight the control material you want to edit.
Select
Highlight the test you want to edit.
Select
Type in the value.
12
Set up non Roche Controls
Log on to data manager.
Select Control Material.
Select
and right-click in the work area.
Select
on the shortcut menu.
Complete the fields according to the information from the control material manufacturer.
o Control name
o Control Code (enter a number between 25000 - 32000)
o Sample type
o
Lot No.
o Expiry date
o Instrument
o
Test
to save the test without closing the window, or
to save the test and close the window.
13
Rack Assignment for non barcoded calibrators and controls
To remove the assignment
To remove the assignment
Do not mix barcoded and non-barcoded calibrators/ controls on the same rack.
COBAS and LIFE NEEDS ANSWERS
are trademarks of Roche.
©2010 Roche
Roche Diagnostics GmbH
D-68298 Mannheim
Germany
www.roche.com
06473792018
25549 Titel und Rück. ShortGuide.indd 2 30.06.10 15:04
- Ch. 0010 V .02
2897
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15

Roche cobas e 602 Short Guide

Type
Short Guide

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI