  • Hello! I am an AI chatbot trained to assist you with the Daniel 370XA Gas Chromatograph System Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Revision A
7P00370-H01
April 2014
370XA Gas Chromatograph
System Reference Manual
NOTICE
ROSEMOUNT ANALYTICAL, INC. (“SELLER”) SHALL NOT BE LIABLE FOR TECHNICAL OR EDITORIAL ERRORS
IN THIS MANUAL OR OMISSIONS FROM THIS MANUAL. SELLER MAKES NO WARRANTIES, EXPRESSED OR
IMPLIED, INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR
PURPOSE WITH RESPECT TO THIS MANUAL AND, IN NO EVENT, SHALL SELLER BE LIABLE FOR ANY SPECIAL
OR CONSEQUENTIAL DAMAGES INCLUDING, BUT NOT LIMITED TO, LOSS OF PRODUCTION, LOSS OF
PROFITS, ETC.
PRODUCT NAMES USED HEREIN ARE FOR MANUFACTURER OR SUPPLIER IDENTIFICATION ONLY AND MAY
BE TRADEMARKS/REGISTERED TRADEMARKS OF THESE COMPANIES.
THE CONTENTS OF THIS PUBLICATION ARE PRESENTED FOR INFORMATIONAL PURPOSES ONLY, AND
WHILE EVERY EFFORT HAS BEEN MADE TO ENSURE THEIR ACCURACY, THEY ARE NOT TO BE CONSTRUED
AS WARRANTIES OR GUARANTEES, EXPRESSED OR IMPLIED, REGARDING THE PRODUCTS OR SERVICES
DESCRIBED HEREIN OR THEIR USE OR APPLICABILITY. WE RESERVE THE RIGHT TO MODIFY OR IMPROVE
THE DESIGNS OR SPECIFICATIONS OF SUCH PRODUCTS AT ANY TIME.
SELLER DOES NOT ASSUME RESPONSIBILITY FOR THE SELECTION, USE OR MAINTENANCE OF ANY
PRODUCT. RESPONSIBILITY FOR PROPER SELECTION, USE AND MAINTENANCE OF ANY SELLER PRODUCT
REMAINS SOLELY WITH THE PURCHASER AND END-USER.
ROSEMOUNT AND THE ROSEMOUNT ANALYTICAL LOGO ARE REGISTERED TRADEMARKS OF ROSEMOUNT
ANALYTICAL. THE EMERSON LOGO IS A TRADEMARK AND SERVICE MARK OF EMERSON ELECTRIC CO.
©2014
ROSEMOUNT ANALYTICAL, INC.
HOUSTON, TX
USA
All rights reserved. No part of this work may be reproduced or copied in any form or by any means–graphic,
electronic, or mechanical–without first receiving the written permission of Rosemount Analytical Inc. Houston,
Texas, U.S.A.
WARRANTY
1. LIMITED WARRANTY: Subject to the limitations contained in Section 2 herein and except as otherwise
expressly provided herein, Rosemount Analytical (“Seller”) warrants that the firmware will execute the
programming instructions provided by Seller, and that the Goods manufactured or Services provided by
Seller will be free from defects in materials or workmanship under normal use and care until the expiration
of the applicable warranty period. Goods are warranted for twelve (12) months from the date of initial
installation or eighteen (18) months from the date of shipment by Seller, whichever period expires first.
Consumables and Services are warranted for a period of 90 days from the date of shipment or completion
of the Services. Products purchased by Seller from a third party for resale to Buyer (“Resale Products”) shall
carry only the warranty extended by the original manufacturer. Buyer agrees that Seller has no liability for
Resale Products beyond making a reasonable commercial effort to arrange for procurement and shipping
of the Resale Products. If Buyer discovers any warranty defects and notifies Seller thereof in writing during
the applicable warranty period, Seller shall, at its option, promptly correct any errors that are found by
Seller in the firmware or Services, or repair or replace F.O.B. point of manufacture that portion of the Goods
or firmware found by Seller to be defective, or refund the purchase price of the defective portion of the
Goods/Services. All replacements or repairs necessitated by inadequate maintenance, normal wear and
usage, unsuitable power sources, unsuitable environmental conditions, accident, misuse, improper
installation, modification, repair, storage or handling, or any other cause not the fault of Seller are not
covered by this limited warranty, and shall be at Buyer's expense. Seller shall not be obligated to pay any
costs or charges incurred by Buyer or any other party except as may be agreed upon in writing in advance
by an authorized Seller representative. All costs of dismantling, re-installation and freight and the time and
expenses of Seller's personnel for site travel and diagnosis under this warranty clause shall be borne by
Buyer unless accepted in writing by Seller. Goods repaired and parts replaced during the warranty period
shall be in warranty for the remainder of the original warranty period or ninety (90) days, whichever is
longer. This limited warranty is the only warranty made by Seller and can be amended only in a writing
signed by an authorized representative of Seller. Except as otherwise expressly provided in the Agreement,
THERE ARE NO REPRESENTATIONS OR WARRANTIES OF ANY KIND, EXPRESSED OR IMPLIED, AS TO MER-
CHANTABILITY, FITNESS FOR PARTICULAR PURPOSE, OR ANY OTHER MATTER WITH RESPECT TO ANY OF
THE GOODS OR SERVICES. It is understood that corrosion or erosion of materials is not covered by our
guarantee.
2.
LIMITATION OF REMEDY AND LIABILITY: SELLER SHALL NOT BE LIABLE FOR DAMAGES CAUSED BY DELAY
IN PERFORMANCE. THE SOLE AND EXCLUSIVE REMEDY FOR BREACH OF WARRANTY HEREUNDER SHALL BE
LIMITED TO REPAIR, CORRECTION, REPLACEMENT OR REFUND OF PURCHASE PRICE UNDER THE LIMITED
WARRANTY CLAUSE IN SECTION 1 HEREIN. IN NO EVENT, REGARDLESS OF THE FORM OF THE CLAIM OR
CAUSE OF ACTION (WHETHER BASED IN CONTRACT, INFRINGEMENT, NEGLIGENCE, STRICT LIABILITY,
OTHER TORT OR OTHERWISE), SHALL SELLER'S LIABILITY TO BUYER AND/OR ITS CUSTOMERS EXCEED THE
PRICE TO BUYER OF THE SPECIFIC GOODS MANUFACTURED OR SERVICES PROVIDED BY SELLER GIVING
RISE TO THE CLAIM OR CAUSE OF ACTION. BUYER AGREES THAT IN NO EVENT SHALL SELLER'S LIABILITY TO
BUYER AND/OR ITS CUSTOMERS EXTEND TO INCLUDE INCIDENTAL, CONSEQUENTIAL OR PUNITIVE
DAMAGES. THE TERM “CONSEQUENTIAL DAMAGES” SHALL INCLUDE, BUT NOT BE LIMITED TO, LOSS OF
ANTICIPATED PROFITS, LOSS OF USE, LOSS OF REVENUE AND COST OF CAPITAL.
i
TABLE OF CONTENTS
Section 1 Introducing the 370XA . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Section 2 Anatomy of the 370XA . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Exterior: Front View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Exterior: Left View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Exterior: Right View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Exterior: Rear View . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Exterior: Local Operator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6
Interior: Back plane . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Interior: Maintainable Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Section 3 Installing the 370XA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Site Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Site requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Mounting requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Actions Upon Recent of GC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Inspection and verification of received equipment . . . . . . . . . . . . 3-2
Mount the GC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Minimum installation clearances . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Pole mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Optional gas bottle accessories . . . . . . . . . . . . . . . . . . . . 3-6
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Wall mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Secure the 370XA to the mounting bracket . . . . . . . . . . . . . . . . . 3-14
Mount the sample conditioning system . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Attach the single-stream, side-mounted sample
conditioning system to the GC . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17
Connect sample conditioning system tubing to the GC . . . . . . . . 3-21
Connect to the carrier gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-23
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
ii
Actuation gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Helium actuation gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Alternative actuation gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-25
Connect to the calibration gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Connect to the sample gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-28
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Terminal wiring diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
Connect to serial ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
RS-232 wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-35
RS-485 wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-36
Connect to Ethernet ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-37
Ethernet port 1 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-38
Ethernet port 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-39
Connect to external devices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40
Digital inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-40
Digital output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-41
Analog input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-42
Analog outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-42
Connect to power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-44
Power source wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-44
Electrical and signal ground . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-46
Start up and configure the gas chromatograph . . . . . . . . . . . . . . . . . . . . 3-46
Apply carrier and actuation gas . . . . . . . . . . . . . . . . . . . . . . . . . . 3-46
Apply calibration gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-47
Apply sample gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-47
Turning on power for the first time . . . . . . . . . . . . . . . . . . . . . . . . 3-48
Set the time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-50
Configure the gas chromatograph’s Ethernet ports . . . . . . . . . . . 3-50
Connect a computer directly to the 370XA . . . . . . . . . . . . . . . . . 3-51
Connect a local computer to the 370XA . . . . . . . . . . . . . . . . . . . 3-51
Configure the serial communications settings from the 370XA . . 3-53
Set the calibration gas concentration values . . . . . . . . . . . . . . . . 3-54
Calibrating the GC for the first time . . . . . . . . . . . . . . . . . . . . . . . 3-55
Section 4 Using the 370XA. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Interacting with the LOI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Menu operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Screen operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Enter numeric data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Enter alphanumeric data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
How to perform common tasks with the LOI . . . . . . . . . . . . . . . . . . . . . . . 4-4
Acknowledge an alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Acknowledge all alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Acknowledge and clear all alarms . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
View the maintenance log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
View the event log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
370XA Gas Chromatograph System Reference Manual
APRIL 2014 7P00370-H01
iii
View a live chromatogram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-6
View an archived chromatogram . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Start a single stream analysis run . . . . . . . . . . . . . . . . . . . . . . . . . 4-8
Start a calibration analysis run . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
LOI screen descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
The View menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
The Current Alarms screen . . . . . . . . . . . . . . . . . . . . . . 4-11
The Reports screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-12
The Reports Viewer screen . . . . . . . . . . . . . . . . . . . . . . 4-12
The Chromatogram screens . . . . . . . . . . . . . . . . . . . . . 4-12
The Hardware menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
The Heaters screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-13
The Valves screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-14
The EPC screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
The Discrete Input screen . . . . . . . . . . . . . . . . . . . . . . . 4-16
The Discrete Output screen . . . . . . . . . . . . . . . . . . . . . . 4-17
The Analog Input screen . . . . . . . . . . . . . . . . . . . . . . . . 4-18
The Analog Output screen . . . . . . . . . . . . . . . . . . . . . . . 4-19
The Application menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
The System screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-20
The Component Data Table screen . . . . . . . . . . . . . . . . 4-21
The Timed Events Table screen . . . . . . . . . . . . . . . . . . 4-22
The Streams screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
The Status Display screen . . . . . . . . . . . . . . . . . . . . . . . 4-23
The Communications screen . . . . . . . . . . . . . . . . . . . . . 4-23
TCP/IP Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-25
The Logs menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
The Alarm Log screen . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
The Event Log screen . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
The Maintenance Log screen . . . . . . . . . . . . . . . . . . . . 4-27
The Add a New Maintenance Log screen . . . . . . . . . . . 4-27
The GC Control menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-27
The Auto Sequence screen . . . . . . . . . . . . . . . . . . . . . . 4-27
The Single Stream screen . . . . . . . . . . . . . . . . . . . . . . . 4-28
The Halt screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-28
The Calibration screen . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
The Validation screen . . . . . . . . . . . . . . . . . . . . . . . . . . 4-29
The Auto Valve Timing screen . . . . . . . . . . . . . . . . . . . . 4-30
The Tools menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
The Screen Control screen . . . . . . . . . . . . . . . . . . . . . . 4-30
The Set GC Time screen . . . . . . . . . . . . . . . . . . . . . . . . 4-31
The Module Validation screens . . . . . . . . . . . . . . . . . . . 4-31
The Login screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
iv
Advanced configuration and operation topics . . . . . . . . . . . . . . . . . . . . . 4-32
Validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Prepare for a validation . . . . . . . . . . . . . . . . . . . . . . . . . 4-33
Run a validation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-35
Change the calibration gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-36
Auto valve timing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-37
Warm-start mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Maintenance mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-39
Conserve calibration gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-40
New Module Software Assistant . . . . . . . . . . . . . . . . . . . . . . . . . 4-42
Section 5 Replacing the maintainable module. . . . . . . . . . . . . . . . 5-1
Remove the maintainable module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Install the maintainable module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Section 6 Overhauling the maintainable module. . . . . . . . . . . . . . 6-1
Replace a stream selection solenoid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Replace an analytical valve solenoid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Overhaul an analytical valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
Overhaul the sample shut-off valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-22
Replace the carrier pressure control valve . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Replace the carrier gas pressure sensor . . . . . . . . . . . . . . . . . . . . . . . . . 6-30
Replace the detectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-31
Replace the chromatograph columns . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-37
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-37
Column assignments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-39
Column drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-39
Replace the o-rings in a tubing connector . . . . . . . . . . . . . . . . . . . . . . . . 6-48
Section 7 Troubleshooting the 370XA . . . . . . . . . . . . . . . . . . . . . . 7-1
Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Flowcharts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
The GC won’t power up. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Un-normalized total error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
Carrier pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
Heater won’t reach set point . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
Module validation fails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
Auto valve timing fails . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
1 - 1
Section 1: Introducing the 370XA
This manual contains information pertaining to the 370XA gas
chromatograph from Rosemount Analytical. The purpose of this
manual is to provide detailed installation, operation, maintenance, and
troubleshooting procedures.
1.1 General description
The 370XA gas chromatograph from Rosemount Analytical is the
latest analyzer to join the XA series of Emerson gas chromatographs.
Designed to simplify natural gas measurement analysis, the 370XA
provides greater ease of use and increased measurement
performance for your C6+ BTU/CV analysis.
Another unique benefit of the 370XA is its Maintainable Module™
technology, which allows you to easily replace the GC module in the
field in approximately two hours, including warm-up and purge, greatly
reducing downtime and overall operating costs.
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
1 - 2
Structurally, the 370XA has two major sections. The upper section (1)
consists of the analysis module (3). The lower section (2) of the gas
chromatograph houses the electronics (4) and the LCD display.
370XA Gas Chromatograph System Reference Manual
APRIL 2014 7P00370-H01
1 - 3
1.2 Specifications
Electronics
Power 24V DC (standard) at the unit.
21-30V DC (operating range) at the unit.
Class 2 power supply as required for
CSA approval.
Note: Provide the GC with one 5-amp
circuit breaker for protection.
Power consumption at 22° C (72° F) 50 Watts (startup)
20 Watts (steady state)
Construction
Environmental temperature -20° C (-4 °F) to 60 °C (140 °F)
Enclosure protection rating IP65 and Type 4X
Dimensions
(without sample system or mounts)
460 mm (H) x 305 mm (W) x 280 mm (D)
18” (H) x 12” (W) x 11” (D)
Mounting options Pipe, wall
Weight
(without sample system or mounts)
22 kg (50 lbs)
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
1 - 4
Performance
Application four-minute C6+ analysis
Repeatability Controlled environment
+/-0.0125% calorific value
+/- 0.125 BTU/scf per 1000 BTU/scf
-20 °C (-4 °F) to 60 °C (140 °F)
+/-0.025% calorific value
+/- 0.25 BTU/scf per 1000 BTU/scf
Metrology approvals Measurement Canada, NMi (OIML)
Calculations ISO 6976, AGA-8, GPA 2172 (using the
GPA 2145 physical properties table)
Carrier gas Zero-grade helium at 6.2 BarG (90
PSIG).
Zero grade hydrogen available as an
option.
Actuation gas Helium, nitrogen, or clean dry air at 6.2
BarG (90 PSIG).
Sample input pressure range 0.7 to 1.7 BarG (10 to 25 PSIG)
Valves Three 6-port diaphragm chromatograph
valves.
Oven Airless iso-thermal
Detector Thermal conductivity detector (TCD)
Streams One sample stream and one calibration
stream
Number of chromatograms stored
internally
Stores up to 30 days of analysis report
data and up to 2500 individual
chromatograms.
370XA Gas Chromatograph System Reference Manual
APRIL 2014 7P00370-H01
1 - 5
Standard communications
Ethernet Two available connections: one RJ-45
plug-in port and one 4-wire termination.
Both with 10/100 Mbps.
Analog input One standard input filtered with transient
protection, 4–20 mA that is user scalable
and assignable.
Analog outputs Two isolated outputs, 4–20 mA.
Digital inputs One input that is user assignable,
optically isolated, and rated to 30 VDC @
0.5 A.
Digital output One user-assignable output, Form C and
electro-mechanically isolated, 24 VDC.
Serial ports Two termination blocks, configurable as
RS-232 or RS-485.
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
1 - 6
1.3 Certifications
ATEX
Manufacturer Rosemount Analytical, Inc.
Houston, TX, USA
Product Danalyzer 370XA Gas Chromatograph
Certificate number Sira 13ATEX1030
Certification code Ex d IIB+H2 T6 Gb
Ambient range Ta = -20 °C to +60 °C
Serial number device dependent
Year of manufacture device dependent
Warnings DO NOT OPEN WHEN AN EXPLOSIVE
ATMOSPHERE MAY BE PRESENT.
DO NOT OPEN WHILE ENERGIZED.
USE SUPPLY CABLES OR WIRES
SUITABLE FOR AT LEAST 80 °C.
Other markings
Electrical ratings
DC: 21 - 30 V, , 55 W MAX
Number and size of the conduit
entries
3 conduit entries: M32 X 1.5
EN 60079-0:2009 Explosive atmospheres - Part 0:
Equipment - General requirements
EN 60079-1:2007 Explosive atmospheres - Part 1:
Equipment protection by flameproof
enclosures "d"
IECEx Ex d IIB+H2 T6 Gb
- 20 °C to +60 °C
IP65 IECEx CSA 13.0005
370XA Gas Chromatograph System Reference Manual
APRIL 2014 7P00370-H01
1 - 7
CSA Class I, Div. 1; Groups B, C, D; T6; Type 4X
Class I, Zone 1; Ex/AEx d IIB + H
2
; T6; IP65
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
1 - 8
This page is intentionally left blank.
2 - 1
Section 2: Anatomy of the 370XA
2.1 Exterior: Front View
1. Explosion-proof dome
Unscrew the explosion-proof dome to gain access to the
analytical oven.
2. Explosion-proof body
The explosion-proof body houses the electronics.
3. Local operator interface
The local operator interface (LOI) contains keys and an LCD
display for interacting with the gas chromatograph.
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
2 - 2
2.2 Exterior: Left View
1. Locking bolt
Secures the dome to the enclosure. To engage the lock, screw
in the hex screw with a 2 mm hex wrench; to release the lock,
unscrew the hex screw with a 2mm hex wrench.
2. Cable entry
M32 conduit entry. This is the most convenient entry point for
the power cable. If this entry is not used for cables then a
certified plug must be used.
Certification type Additionally supplied certified parts
CSA M32-to-¾-inch adapter and a ¾-inch sealing plug
ATEX/IECEx M32 plug
370XA Gas Chromatograph System Reference Manual
APRIL 2014 7P00370-H01
2 - 3
3. Grounding lug
Connects to an external grounding system, as required for
ATEX-certified installations.
4. Tubing bulkhead
The central hub for the sample, carrier, and other gas inlets.
2.3 Exterior: Right View
1. Upper cable entry
Certification type Additionally supplied certified parts
CSA M32-to-¾-inch adapter and a ¾-inch sealing plug
ATEX/IECEx M32 plug
System Reference Manual 370XA Gas Chromatograph
7P00370-H01 APRIL 2014
2 - 4
2. Port hole
Unscrew and remove to gain access to the field wiring
connections and the circuit boards.
3. Lower cable entry
Supplied with an M32 certified plug and an M32-to-¾ NPT
adapter. This is the most convenient entry point for connecting
cables to the Ethernet port, communications ports, and
external device terminals. If this entry is not used for cables
then a certified plug must be used.
370XA Gas Chromatograph System Reference Manual
APRIL 2014 7P00370-H01
2 - 5
2.4 Exterior: Rear View
1. Mounting bolts
These M8 bolts are used to mount the GC to its final
destination, depending upon the mounting configuration
chosen. Each bolt is 1.8 centimeters long.
/