  • Hello! I am an AI chatbot trained to assist you with the Taylor Technologies K-1683 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
COMPONENT SHEET
K-1683 BOILER/COOLING SYSTEM WATER TESTS
COMPONENTS:
P Alkalinity
1 x 5228 Instruction
*1 x R-0638G-C Phenolphthalein Indicator, 2 oz w/ green cap, DB
*1 x R-0687G-C Sulfuric Acid .12N, 2 oz w/ green cap, DB
Free Chelant (NTA)
2 x 2269 Cap, Dispenser, 18 mm, white
1 x 4027 Funnel, 58 mm, plastic
1 x 4030 Pipet, Calibrated (0.5 & 1.0 mL) w/ cap, plastic
1 x 5529 Instruction
1 x 6009 Filter Paper, 9.0 cm, #610, 100/box
*1 x R-0620B-I Hardness Indicator Powder, 10 g w/ blue dot
1 x R-0877-C Magnesium Chloride Reagent, 2 oz, DB
1 x R-0878-C NTA Buffer Solution, 2 oz
2 x R-0879-C NTA Cosolvent Solution, 2 oz
Chloride
1 x 5212 Instruction
1 x R-0630-C Chromate Indicator, 2 oz, DB
*1 x R-0638G-C Phenolphthalein Indicator, 2 oz w/ green cap, DB
*1 x R-0687G-C Sulfuric Acid .12N, 2 oz w/ green cap, DB
1 x R-0706C Silver Nitrate Reagent, 2 oz, DB
Total Hardness
1 x 5202 Instruction
1 x R-0619B-C Hardness Buffer, 2 oz w/ blue cap, DB
*1 x R-0620B-I Hardness Indicator Powder, 10 g w/ blue dot
1 x R-0683-C Hardness Reagent, 2 oz, DB
31 Loveton Circle, Sparks, MD 21152 U.S.A.
800-TESTKIT(837-8548) • 410-472-4340 7/16
Sodium Sulfite
1 x 5203 Instruction
*1 x R-0638G-C Phenolphthalein Indicator, 2 oz w/ green cap, DB
1 x R-0699-C Iodide Iodate Reagent, 2 oz, DB
1 x R-0725-I Acid Starch Indicator Powder, 10g
Misc.
1 x 5518 Instruction (Component Sheet)
2 x 9198 Sample Tube, Graduated (25 mL) w/ cap, plastic
*Kit includes one.
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
Instr. #5518
DROP TEST
FREE CHELANT (1 drop = 10 ppm NTA)
COMPONENTS:
2 x 2269 Cap, Dispenser, 18 mm, white
1 x 4027 Funnel, 58 mm, plastic
1 x 4030 Pipet, Calibrated (0.5 & 1.0 mL) w/ cap, plastic
1 x 5529 Instruction
1 x 6009 Filter Paper, 9.0 cm, #610, 100/box
1 x 9198 Sample Tube, Graduated (25 mL) w/ cap, plastic
1 x R-0620B-I Hardness Indicator Powder, 10 g w/ blue dot
1 x R-0877-C Magnesium Chloride Reagent, 2 oz, DB
1 x R-0878-C NTA Buffer Solution, 2 oz
2 x R-0879-C NTA Cosolvent Solution, 2 oz
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles or pipets, always hold bottle/pipet
in a vertical position.
Free Chelant Test
1. Filter water to be tested to clarify.
2.Rinseandll25mLsampletube(#9198)to10mLmarkwithlteredsample.
3.Usinga1.0mLpipet(#4030),add0.5mLR-0878NTABuerSolution.Swirltomix.
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548) • 410-472-4340 7/16
Instr. #5529
4.Cap R-0879 NTA Cosolvent Solution with dispenser cap (#2269). Add R-0879 NTA
CosolventSolutionto20.5mLmark.Swirltomix.
5.Add1leveldipperR-0620BHardnessIndicatorPowder.Swirltomix.Samplewillturnblue
if chelant is present.
6.AddR-0877MagnesiumChlorideReagentdropwise,swirlingandcountingaftereachdrop,
untilcolorchangesfrombluetored.
7.MultiplydropsofR-0877MagnesiumChlorideby10.Recordaspartspermillion(ppm)free
chelantasNTA(nitrilotriaceticacid).
COMPONENTS:
1 x 5202 Instruction
1 x 9198B Sample Tube, Graduated (25 mL) w/ cap & blue dot, plastic
1 x R-0619B-C Hardness Buffer, 2 oz w/ blue cap, DB
1 x R-0620B-I Hardness Indicator Powder, 10 g w/ blue dot
1 x R-0683-C Hardness Reagent, 2 oz, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Total Hardness Test
1.Rinseandll25mLsampletube(#9198B)to25mLmarkwithwatertobetested
(Fig.1).
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add5dropsR-0619BHardnessBuer.Swirltomix.
3.Add1dipperR-0620BHardnessIndicatorPowder.Swirluntildissolved.Sample
willturnred(Fig.2)ifhardnessispresent.
4.AddR-0683HardnessReagentdropwise,swirlingandcountingaftereachdrop,
untilcolorchangesfromredtoblue(Fig.3).
DROP TEST
TOTAL HARDNESS (1 drop = 10 ppm)
5.MultiplydropsofR-0683HardnessReagentby10.Recordaspartspermillion
(ppm)totalhardnessascalciumcarbonate.
NOTE: For14.6mLsample,recorddropsasgrainspergallon(gpg)totalhardness
ascalciumcarbonate.
Fig. 1
Fig. 2
Fig. 3
Instr. #5202
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 7/16
COMPONENTS:
1 x 5203 Instruction
1 x 9198W Sample Tube, Graduated (25 mL) w/ cap & white dot, plastic
1 x R-0638W-C Phenolphthalein Indicator, 2 oz w/ white cap, DB
1 x R-0699-C Iodide Iodate Reagent, 2 oz, DB
1 x R-0725-I Acid Starch Indicator Powder, 10 g
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE
800-TEST KIT (800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: When dispensing reagents from dropper bottles, always hold bottle in a
vertical position.
Sodium Sulfite Test
NOTE: Sample must be cooled to less than 100ºF (38ºC) to prevent high test
results. Sample must be protected from air contact while cooling to prevent
low test results.
1.Collectwatertobetestedinaclean,preferablylarge-mouthed,bottletooverowing.
Immediately cap and cool to room temperature.
2.Rinseandll25mLsampletube(#9198W)to25mLmarkwithcooled(room
temperature) water to be tested (Fig. 1).
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
DROP TEST
SODIUM SULFITE (1 drop = 10 ppm)
3.Add1dropR-0638WPhenolphthaleinIndicator.Swirltomix.Samplewillturn
pink(Fig.2).
4.AddR-0725AcidStarchIndicatorPowderadipperatatime,swirlingaftereach
dipper, untilcolorchanges from pinktocolorless. Add2 moredippers. Swirl
until dissolved.
5.AddR-0699IodideIodateReagentdropwise,swirlingandcountingaftereach
drop, until sample changes from colorless to a faint but permanent blue (Fig. 3).
6.Multiply drops of R-0699 Iodide Iodate Reagent by 10. Record as parts per
million(ppm)sodiumsulte.
NOTE: For14.6mLsample,recorddropsasgrainspergallon(gpg)sodiumsulte.
NOTE:Forresultsassulte,multiplysodiumsulteconcentrationby0.64.
Fig. 1
Fig. 2
Fig. 3
Instr. #5203
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548)•410-472-4340 9/16
DROP TEST
CHLORIDE (1 drop = 10 ppm)
COMPONENTS:
1 x 5212 Instruction
1 x 9198 Sample Tube, Graduated, 25 mL, plastic w/cap
1 x R-0630 Chromate Indicator, DB
1 x R-0638 Phenolphthalein Indicator, DB
* 1 x R-0686 Sulfuric Acid N, DB
* 1 x R-0687 Sulfuric Acid .12N, DB
1 x R-0706 Silver Nitrate Reagent, DB
* 1 x R-0736 Sulfuric Acid .6N, DB
* Kit may only contain one of these reagents.
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE 800-TEST KIT
(800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
NOTE: Whensultecontentofsamplewatertobetestedexceeds10ppm,thesulteshould
beoxidizedtopreventinterferenceintest.A25mLsampleisrstadjustedtotheap-
propriatepH,then1mL(or24drops)ofR-06493%HydrogenPeroxideSolution(sold
separately)isaddedandthoroughlymixed.Continuewiththerestoftheprocedure.
1.Rinseandll25mLsampletube(#9198)to25mLmarkwithwatertobetested.
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add2dropsR-0638PhenolphthaleinIndicator.Ifcolorless—proceedtoStep3.Ifred,add
R-0686SulfuricAcidN,R-0687SulfuricAcid.12N,orR-0736SulfuricAcid.6Ndropwise,
swirlingaftereachdrop,untilcolorchangesfromredtocolorless.
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548) • 410-472-4340 5/08
3.Add5dropsR-0630ChromateIndicator.Swirltomix.Sampleshouldturnyellow.
4.AddR-0706SilverNitrateReagentdropwise,swirlingandcountingaftereachdrop,untilcolor
changesfromyellowtoamilkysalmon(brick)red.Alwaysholdbottleinverticalposition.
NOTE: DonotaddenoughR-0706SilverNitrateReagenttogiveabrowncolor.Firstchange
fromyellowtoamilkysalmon(brick)redistheendpoint.
5.MultiplydropsofR-0706SilverNitrateReagentby10.Recordaspartspermillion(ppm)
chloride.For14.6mLsample,recorddropsasgrainspergallon(gpg)chloride.
Instr. #5212
DROP TEST
P ALKALINITY (1 drop = 10 ppm)
COMPONENTS:
1 x 5228 Instruction
1 x 9198 Sample Tube, Graduated, 25 mL, plastic w/cap
1 x R-0638 Phenolphthalein Indicator, DB
1 x R-0687 Sulfuric Acid .12N, DB
TO ORDER REPLACEMENT PARTS AND REAGENTS CALL TOLL-FREE 800-TEST KIT
(800-837-8548).
PROCEDURE:
CAREFULLY READ AND FOLLOW PRECAUTIONS ON REAGENT LABELS.
KEEP REAGENTS AWAY FROM CHILDREN.
1.Rinseandll25mLsampletube(#9198)to25mLmarkwithwatertobetested.
NOTE: Forresultsingrainspergallon(gpg),llto14.6mLmark.
2.Add 3 drops R-0638 Phenolphthalein Indicator. Sample will turn pink if P alkalinity is
present.
3.Ifpink,addR-0687SulfuricAcid.12Ndropwise,swirlingandcountingaftereachdrop,until
colorchangesfrompinktocolorless.Alwaysholdbottleinverticalposition.
31LovetonCircle,Sparks,MD21152U.S.A.
800-TESTKIT(837-8548) • 410-472-4340 3/90
4.MultiplydropsofR-0687SulfuricAcidNby10.Recordaspartspermillion(ppm)Palkalinity
as calcium carbonate. For 14.6 mL sample, record drops as grains per gallon (gpg) P
alkalinityascalciumcarbonate.