Invacare IRC 1705 is a portable medical device designed to efficiently deliver physician-prescribed medication to the bronchial lung passages. It consists of a compressor, nebulizer, air tubing, and mouthpiece. With a sound level of 58 to 63 dBA, it operates with a maximum pressure of 36 psi and a flow rate of 8 to 11 L/min. The device is compliant with UL 1431 and CSA 22.2 standards, ensuring safety and reliability. Its lightweight design and intermittent use mode (30 min ON/30 min OFF) make it convenient for personal use.

Invacare IRC 1705 is a portable medical device designed to efficiently deliver physician-prescribed medication to the bronchial lung passages. It consists of a compressor, nebulizer, air tubing, and mouthpiece. With a sound level of 58 to 63 dBA, it operates with a maximum pressure of 36 psi and a flow rate of 8 to 11 L/min. The device is compliant with UL 1431 and CSA 22.2 standards, ensuring safety and reliability. Its lightweight design and intermittent use mode (30 min ON/30 min OFF) make it convenient for personal use.

Dealer: This manual MUST be given to the end user.
User:
BEFORE using this product, read this manual and
save for future reference.
For more information regarding
Invacare products, parts, and
services, please visit
www.invacare.com
Operator’s Manual
Invacare
®
Select
Compressor Nebulizer System
Model IRC 1705
Invacare
®
Select 2 Part No 1148073
WARNING
DONOTusethisproductoranyavailableoptional
equipmentwithoutfirstcompletelyreadingand
understandingtheseinstructionsandanyadditional
instructionalmaterialsuchasownersmanuals,service
manualsorinstructionsheetssuppliedwiththisproductor
optionalequipment.Ifyouareunabletounderstandthe
warnings,cautionsorinstructions,contactahealthcare
professional,dealerortechnicalpersonnelbefore
attemptingtousethisequipment‐otherwise,injuryor
damagemayoccur.
Invacareproductsarespecificallydesignedand
manufacturedforuseinconjunctionwithInvacare
accessories.Accessoriesdesignedbyothermanufacturers
havenotbeentestedbyInvacareandarenotrecommended
forusewithInvacareproducts.
TABLE OF CONTENTS
Part No 1148073 3 Invacare
®
Select
TABLE OF CONTENTS
REGISTER YOUR PRODUCT .............................. 4
SPECIAL NOTES ............................................... 5
IMPORTANT SAFEGUARDS .............................. 6
INTRODUCTION ............................................ 11
TYPICAL PRODUCT PARAMETERS ................. 12
SUPPLEMENTAL INFORMATION ..................... 14
Compressor........................................................................................14
Extension Cords................................................................................14
SECTION 1—HANDLING ................................. 15
Shipping Instructions.........................................................................15
Unpacking............................................................................................15
Inspection............................................................................................15
SECTION 2—OPERATING INSTRUCTIONS ....... 16
Setup.....................................................................................................16
Treatment Instructions....................................................................19
SECTION 3—CLEANING AND DISINFECTING ... 20
Cleaning the Outer Case.................................................................20
Cleaning the Nebulizer ....................................................................21
Cleaning the Compressor ...............................................................23
TABLE OF CONTENTS
TABLE OF CONTENTS
Invacare
®
Select 4 Part No 1148073
Cleaning the Mouthpiece.................................................................23
SECTION 4—MAINTENANCE .......................... 24
Routine Care......................................................................................24
Storage.................................................................................................24
Nebulizer Change..............................................................................25
Filter Change ......................................................................................25
SECTION 5—TROUBLESHOOTING ................... 26
LIMITED WARRANTY .................................... 29
REGISTER YOUR PRODUCT
The benefits of registering include:
1. Safeguarding your investment.
2. Ensuring long-term maintenance and servicing of your product.
3. Receiving updates with product information, maintenance tips and
industry news.
Register ONLINE at warranty.invacare.com
Please have your model number and purchase date available to complete
your registration.
Any registration information you submit will only be used by Invacare
Corporation and protected as required by applicable laws and regulations.
SPECIAL NOTES
Part No 1148073 5 Invacare
®
Select
SPECIAL NOTES
Signalwordsareusedinthismanualandapplytohazardsor
unsafepracticeswhichcouldresultinpersonalinjuryor
propertydamage.Refertothetablebelowfordefinitionsof
thesignalwords.
NOTICE
The information contained in this document is subject to
change without notice.
SIGNAL
WORD
MEANING
DANGER
Danger indicates an imminently hazardous situation which, if
not avoided, will result in death or serious injury.
WARNING
Warning indicates a potentially hazardous situation which, if
not avoided, could result in death or serious injury.
CAUTION
Caution indicates a potentially hazardous situation which, if
not avoided, may result in property damage or minor injury
or both.
IMPORTANT SAFEGUARDS
Invacare
®
Select 6 Part No 1148073
IMPORTANT SAFEGUARDS
WARNING
IMPORTANT SAFEGUARDS contains important
information for the safe operation and use of this product.
CAUTION
Federallawrestrictsthisdevicetosaleorrentalbyoronthe
orderofaphysician,oranyotherpractitionerlicensedbythe
lawoftheStateinwhichhe/shepracticestouseororderthe
useofthisdevice.
“Thisappliancehasapolarizedplug(onebladeiswiderthan
theother).Asasafetyfeature,thisplugwillfitinapolarized
outletonlyoneway.Iftheplugdoesnotfitfullyintheoutlet,
reversetheplug.Ifitstilldoesnotfit,contactaqualified
electrician.Donotattempttodefeatthissafetyfeature.”
WARNING
Whenusingelectricalproducts,especiallywhenchildrenare
present,basicsafetyprecautionsshouldalwaysbefollowed,
includingthefollowing:
IMPORTANT SAFEGUARDS
Part No 1148073 7 Invacare
®
Select
READALLINSTRUCTIONSBEFOREUSING
DANGER
FIRERISK‐DONOTusethisproductinanoxygen
enrichedenvironment(i.eoxygentent).
Riskofelectricshock.DONOTdisassemble.Refer
servicingtoqualifiedservicepersonnel.
To Reduce the Risk of Electrocution:
ALWAYSunplugthisproductIMMEDIATELYafterusing.
DONOTusewhilebathing.
DONOTplaceorstoreproductwhereitcanfallorbe
pulledintoatuborsink.
DONOTplaceinordropintowaterorotherliquid.
DONOTreachforaproductthathasfallenintowater.
UnplugIMMEDIATELY.
IMPORTANT SAFEGUARDS
Invacare
®
Select 8 Part No 1148073
WARNING
To Reduce the Risk of Burns,
Electrocution, Fire and Injury to Persons:
ThisproductshouldNEVERbeleftunattendedwhen
pluggedin.
Closesupervisionisnecessarywhenthisproductisusedby,
on,ornearchildrenandphysicallychallengedindividuals.
Usethisproductonlyforitsintendeduseasdescribedinthis
manualandonlywiththeprescribedmedication.Usethis
productonlyunderphysician’sdirection.
NEVERoperatethisproductwithadamagedcordorplug,if
itisnotworkingproperly,ifitwasdroppedordamaged,orif
itwasdroppedintowater.Callqualifiedservicepersonnelfor
examinationandrepair.
IMPORTANT SAFEGUARDS
Part No 1148073 9 Invacare
®
Select
WARNING
Keepthenebulizertubing,cord,andunitawayfromheated
orhotsurfaces,includingspaceheaters,electricblankets,
stovesandsimilarelectricalappliances.
NEVERdroporinsertobjectsintoanyopeningorhose.
NEVERusewhilesleepingordrowsy.
NEVERblocktheairopeningsoftheproductorplaceitona
softsurface,suchasabedorcouch,wheretheairopenings
maybeblocked.Keeptheairopeningsfreefromlint,hairand
thelike.Blockedopeningsmaycausetheunittoshutdown.
Contactdealerimmediately.
Thecompressornebulizersystemusesthesurroundingairto
effectivelyadministertheprescribedmedicationthroughthe
nebulizer.Theuserandadministratorshouldbeawareof
surroundingssoasnottohavecontaminatedairfiltered
throughthecompressornebulizersystem.(Examplescanbe
aircontaminatedwithpaintorcleaningfumes,weedkiller,
carexhaust,etc.)
DONOTusethecompressornebulizersysteminthese
environments.Otherwise,seriousbodilyinjurymayoccur.
Thisunitisoilless.DONOTlubricate.
IMPORTANT SAFEGUARDS
Invacare
®
Select 10 Part No 1148073
WARNING
Thisequipmentisnotsuitableforuseinthepresenceofa
flammableanestheticmixturewithairorwithnitrousoxide.
Whennotadministeringtreatment,keepoutofthereachof
smallchildren.Someofthepartsofthenebulizeraresmall
enoughforchildrentoswallow.Thetubingmaypresenta
strangulationhazard
NEVERtouchtheunitwithwethands.
DONOTpullonthepowercordtounplugtheunitfroman
electricaloutlet.
DONOToverloadwalloutlets.
MaintenanceMUSTbeperformedbyaqualifiedservice
technician.
SAVE THESE INSTRUCTIONS
INTRODUCTION
Part No 1148073 11 Invacare
®
Select
INTRODUCTION
TheInvacareSelectCompressorNebulizerSystemisa
portablemedicaldevicedesignedtoefficientlydeliver
physicianprescribedmedicationtothebronchiallung
passages.Thenebulizersystemconsistsofthefollowing
components:
Compressor
Nebulizer
•Airtubing
•Mouthpiece
TYPICAL PRODUCT PARAMETERS
Invacare
®
Select 12 Part No 1148073
TYPICAL PRODUCT
PARAMETERS
Electrical Requirements: 115 VAC, 60Hz, Single Phase
Running Current: 1.1 A @ 14 psig @ 115 volts (approx.)
Sound Level: 58 to 63 dBA
Power Wattage: 65 watts @ 14 psig
Electric Motor: Class B in compliance with UL2111.
Foam Intake Filter: Rayon Fiber Filter
Safety Systems: High temperature compressor shutdown.
Width: 6.4 inches
Height: 5.5 inches
Depth: 3.3 inches
Weight: 3.2 Ibs
Shipping Weight: 3.8 lbs
Regulatory Listing: In compliance with UL 1431 Personal
Hygiene and Healthcare Appliances and
CSA 22.2 Electromedical Equipment.
Maximum Pressure: 36 psi
Maximum Flow Rate: 8 to 11 L/min.
Operating Flow Rate: 4.0 L/min. @ 16 psi operating pressure
TYPICAL PRODUCT PARAMETERS
Part No 1148073 13 Invacare
®
Select
NOTE:Productparametersandappearancearesubjecttochange
withoutnoticeforpurposesofimprovement.
NOTE:Nebulizationrateandparticlesizearedependentonthe
characteristicsofthenebulizerchosen.
Storage Conditions: -13°F to 158°F (-25° to 70°C); 10 to 95%
RH
Operating Conditions: 50°F to 104°F (10°C to 40°C); 10 to 95%
RH
Altitude: 0 to 6000 ft (1825 m) above sea level
Mode of Operation: Intermittent Use (30 min. ON / 30 min.
OFF)
SUPPLEMENTAL INFORMATION
Invacare
®
Select 14 Part No 1148073
SUPPLEMENTAL
INFORMATION
Compressor
Iftheunitshutsdownfrequently,youmayhavealowvoltage
situation.
Lowvoltagecanalsobesuspectedwhen:
•Themotordoesnotgetuptofullpowerorspeed.
•Homefusesorcircuitbreakersactivatewhenstarting
compressor.
•Lightsdimorremaindimwhencompressorisstarted.
•Othermotoroperatedappliancesfailtooperateproperly
ortoomanymotoroperatedappliancesareonsame
circuit.
Extension Cords
Ifyoumustuseanextensioncord,useonlyacordhavingthe
sameorhigherelectricalratingasthedevicebeingconnected.
Makecertaintheextensioncordisingoodconditionwithout
anyexposedwiresorhazardouscontacts,andtheextension
cordwireisNOTLESSthan18gaugefor25feet.
SECTION 1—HANDLING
Part No 1148073 15 Invacare
®
Select
SECTION 1—HANDLING
Shipping Instructions
Theshippingcontainerhasbeendesignedtoassure
protectionofthisdevice.
Ifthisdeviceistobereshippedbycommoncarrier,itshould
bepackedinthesamecarton.
Unpacking
1. Checkforanyobviousdamagetothecartonorits
contents.Ifdamageisevident,DONOTuse.Notify
carrier/InvacareoryourlocalInvacaredealer.
2. Removealltheloosepackingfromthecarton.
3. Carefullyremoveallthecomponentsfromthecarton.
NOTE:Retaincartonandpackingmaterialsforuseinshippingor
storingtheunituntiluseisrequired.
Inspection
1. Examineexteriorfordamage.Inspectallcomponents.
SECTION 2—OPERATING INSTRUCTIONS
Invacare
®
Select 16 Part No 1148073
SECTION 2—OPERATING
INSTRUCTIONS
WARNING
DONOToperateunitwithoutfilter.
Avoidoperatingthis productinadustyenvironment.
Otherwise,prematurecompressorwearcouldoccur.
NOTE:Cleananddisinfectthenebulizerandmouthpiecebefore
usingthedeviceforthefirsttime.RefertoCleaningand
Disinfectingonpage 20.
Setup
NOTE:Forthisprocedure,refertoFIGURE2.1onpage 17.
1. Placetheunitonatableorotherflat,stablesurface.Besure
youcaneasilyreachthecontrolswhenyouareseated.
CAUTION
Makesurethepowerswitchisintheoff“O”position.
2. Gentlytwistthenebulizertoseparateitintotwosections.
SECTION 2—OPERATING INSTRUCTIONS
Part No 1148073 17 Invacare
®
Select
3. Besurethepickupfunnelisinplaceinthebottomsection
ofthenebulizer.
CAUTION
DONOTaddover6mlofmedicationtothemedication
cup.Themaximumcapacityofthenebulizermedicationcup
is6ml.
4. Addtheprescribedamountofmedicationtothecupofthe
nebulizerusinganeyedropper,premeasureddoseor
ampoule.
FIGURE 2.1 Adding Medication
5. Gentlytwistthetopandbottomsectionstogetherto
assemblethenebulizer.Besurethetwosectionsfit
securely.
6. Attachoneendoftheairtubingtotheairtubingconnector
locatedonthefrontofthecompressor.
Eyedropper
Nebulizer Cup
SECTION 2—OPERATING INSTRUCTIONS
Invacare
®
Select 18 Part No 1148073
7. Attachtheotherendoftheairtubingtothebaseofthe
nebulizer.
NOTE:Ensurenebulizertubingisfirmlyattachedtocompressor
beforeuse.
WARNING
Inthecaseofhighhumidity,condensationmayforminside
theinternaltubing.Runtheunitfortwominutesbefore
connectingthenebulizertothetubing.
8. Attachthemouthpiecetothetopsectionofthenebulizer.
Refertothenebulizermanualforadditionalinstructions.
NOTE:Beforestartingacoldunit,lettheunitreachroom
temperatureinaheatedarea.
9. Withthepowerswitchoff,plugthepowercordintoan
appropriateelectricaloutlet.
CAUTION
Thisdeviceisforintermittentuse:30min.ON/30min.OFF.
Failuretooperatewithintheselimitscoulddamagethe
devicebeyondrepair.
SECTION 2—OPERATING INSTRUCTIONS
Part No 1148073 19 Invacare
®
Select
Treatment Instructions
NOTE:Forthisprocedure,refertoFIGURE2.2.
1. Turntheuniton,
pressingtheON/OFF
switchtothe“I”position.
2. Placemouthpiece
betweenyourteeth,close
yourmouthandtiltthe
nebulizertowardyour
mouth.
FIGURE 2.2 Using the
Nebulizer
3. Asaerosolbeginstoflow,inhaledeeply,slowly,andhold
breathfortenseconds.Thenexhaleslowlythroughthe
mouthpiece.
NOTE:Ifyouneedtorestduringatreatment,besurethenebulizer
remainsinanuprightposition;otherwise,themedicationmayspill.
4. Whentreatmentiscomplete,turntheunitoffbysetting
theswitchtothe“O”positionandunplugtheunitfrom
theelectricaloutlet.
5. Beforestoringsuppliesbetweentreatments,cleananddry
thenebulizerandaccessories.RefertoCleaningand
Disinfectingonpage 20.
SECTION 3—CLEANING AND DISINFECTING
Invacare
®
Select 20 Part No 1148073
SECTION 3—CLEANING AND
DISINFECTING
Followingthecleaninginstructionsaftereachusewill
preventanyremainingmedicationinthenebulizercupfrom
dryingandcausingthedevicetonebulizeineffectively.
Itisrecommendedthatthenebulizerandthemouthpiecebe
thoroughlyrinsedwithhottapwateraftereachuseand
cleanedwithamilddetergentafterthelasttreatmentofthe
day.ThesamepiecesMUSTalsobedisinfecteddailyby
soakinginavinegar/watersolutionoracommercially
availabledisinfectant.Ifyourphysicianorrespiratory
therapistspecifiesadifferentcleaningprocedure,followtheir
instructions.
Cleaning the Outer Case
WARNING
Unplugunitbeforecleaning.
DONOTsubmergeinwatertoclean.Cleanoutercasewith
dampcloth.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32

Invacare IRC 1705 User manual

Category
Air compressors
Type
User manual

Invacare IRC 1705 is a portable medical device designed to efficiently deliver physician-prescribed medication to the bronchial lung passages. It consists of a compressor, nebulizer, air tubing, and mouthpiece. With a sound level of 58 to 63 dBA, it operates with a maximum pressure of 36 psi and a flow rate of 8 to 11 L/min. The device is compliant with UL 1431 and CSA 22.2 standards, ensuring safety and reliability. Its lightweight design and intermittent use mode (30 min ON/30 min OFF) make it convenient for personal use.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI