HEKA UnicLine Installation, Operator And Technical Manual

Type
Installation, Operator And Technical Manual

HEKA UnicLine is an electrically controlled patient chair designed for dental treatment. It is intended for use with the UnicLine Dental Unit. The UnicLine patient chair has the following capabilities:

  • Electrically controlled movements for backrest, seat, and footrest
  • Manual adjustment of the headrest
  • Integrated thermal cut-out for safety
  • Emergency stop button (accessory)
  • Safety stop for chair lowering (accessory)
  • Programmable chair positions
  • Last position function
  • Compatible with make and break contacts

The UnicLine patient chair is designed to provide optimal functionality and comfort for both patient and dentist. The simple curves and stylish design result from collaboration with dentists, engineers, and designers.

HEKA UnicLine is an electrically controlled patient chair designed for dental treatment. It is intended for use with the UnicLine Dental Unit. The UnicLine patient chair has the following capabilities:

  • Electrically controlled movements for backrest, seat, and footrest
  • Manual adjustment of the headrest
  • Integrated thermal cut-out for safety
  • Emergency stop button (accessory)
  • Safety stop for chair lowering (accessory)
  • Programmable chair positions
  • Last position function
  • Compatible with make and break contacts

The UnicLine patient chair is designed to provide optimal functionality and comfort for both patient and dentist. The simple curves and stylish design result from collaboration with dentists, engineers, and designers.

Heka Dental A/S Catalogue no.: KA-0033-EN Revision 4.2
Installation, operator's and
technical manual
Patient chair
Keep this manual with the patient chair at all times.
Installation, Service and annually maintenance must only be performed by authorized Heka dealer.
RX Only
2
Table of Contents
1. Foreword ............................................................................................................................ 4
1.1 Introduction ................................................................................................................................... 5
1.2 Indications for Use ........................................................................................................................ 5
1.3 Contraindications ......................................................................................................................... 5
1.4 Warnings ......................................................................................................................................... 5
1.5 Cautions .......................................................................................................................................... 6
2. Assembly instructions ................................................................................................. 11
4. Description of individual components .................................................................. 13
5. Safety instructions ........................................................................................................ 14
5.1 Position of the operator ..................................................................................................... 14
5.2 Position of the patient ........................................................................................................ 14
6. Warning signals and emergency stop .................................................................... 15
6.1 Program stop ........................................................................................................................... 15
6.2 Safety stop: Backrest .......................................................................................................... 15
6.3 Limit stop .............................................................................................................................. 15
6.4 Integrated thermal cut-out ............................................................................................... 15
6.5 Emergency stop (accessory) .............................................................................................. 15
6.6 Safety stop, chair lowering (accessory) .......................................................................... 16
7. Starting up the patient chair ..................................................................................... 16
7.1 On/Off .................................................................................................................................... 16
7.2 Adjusting the headrest ....................................................................................................... 16
7.3 Adjusting the position of the headrest ........................................................................... 16
7.4 Operating the armrest ....................................................................................................... 17
7.5 Adjusting the angle of the seat ............................................................................................ 17
7.6 Adjusting the footrest............................................................................................................ 17
8. Description of the joystick ......................................................................................... 18
8.1 Manual adjustment of the chair ...................................................................................... 18
8.2 Seat up .................................................................................................................................... 18
8.3 Seat down .............................................................................................................................. 18
8.4 Backrest down ...................................................................................................................... 18
8.5 Backrest up ........................................................................................................................... 18
9. Automatic adjustment of the chair ......................................................................... 19
10. Programming the chair's positions .................................................................... 19
11. Maintenance................................................................................................................ 20
11.1 Cleaning of upholstery and armrest: ............................................................................. 20
11.2 Cleaning of external surface aluminium components ................................................ 21
11.3 Changing the fuse ................................................................................................................ 21
12. Technical part, only for authorized Technicians. .......................................... 22
12.1 The chair has the following physical dimensions: ...................................................... 22
2.2 Technical specifications: .................................................................................................. 22
12.3 Regulatory Standards Compliance ..................................................................................... 23
3
12.6 Block diagram of the operating system ......................................................................... 25
12.7 Circuit diagram ................................................................................................................... 26
12.5 Position of components ...................................................................................................... 27
13. Coding options ........................................................................................................... 28
13.1 Last position function ......................................................................................................... 28
13.2 Working with make contacts............................................................................................ 28
13.3 Working with break contacts ........................................................................................... 29
13.4 Running the test program ................................................................................................. 29
13.5 Initialization sequence........................................................................................................ 29
14. Warranty terms ......................................................................................................... 30
15. Installation/service card ........................................................................................ 31
16. EMC Information ....................................................................................................... 32
17. Disposal of the unit ............................................................................................................. 36
4
1. Foreword
Dear Customer
Congratulations on the purchase of your new patient chair.
Our aim was to develop an ergonomically correct patient chair, with design, ergonomics,
reliability, simple maintenance, and hygiene being essential considerations during the
development of the product.
The patient chair was specially designed to suit our units.
Together, the patient chair and the unit make the perfect workstation; both convenient and
a pleasure to use.
Heka Dental A/S has developed this patient chair with the aim of providing optimal
functionality and comfort for both patient and dentist. The simple curves and stylish
design result from our collaboration with dentists, the engineer and the designer.
Welcome to Heka Dental's world.
Thomas Steen Nielsen
Managing director
Heka Dental A/S is certificated to ISO 13485.
5
1.1 Introduction
The UnicLine patient chair is electrically controlled equipment for dental treatment.
The UnicLine patient chair is intended for dental treatment performed by dental
professionals. It is intended for use with the UnicLine Dental Unit.
This manual describes how to use and maintain the UnicLine patient chair. Please read
the manual in its entirety before using the unit. This manual is the primary source of
information about the patient chair.
1.2 Indications for Use
The UnicLine Chair is an electrically powered, is a dental chair intended to take part of a
operative unit intended to supplying power to and serve as a base for other dental devices
and instruments. The system is indicated for performing general dental procedures. The
device is intended for operation and use in a clinical setting by legally qualified
professionals.
1.3 Contraindications
There are no known contraindications for the use of this device.
1.4 Warnings
Heka Dental accepts no liability for direct or consequential injury or damage resulting
from improper use, arising in particular through the non-observance of the operating
instructions, or improper preparation and maintenance.
WARNING: Use for intended purposes only. Failure to observe the operating
instructions may result in the patient or user suffering serious injury or the product being
damaged, possibly beyond repair. Before using this product, make sure you have studied
and understood the operating instructions.
WARNING: For use by qualified and trained personnel only.
WARNING: Position the device to allow adequate working space and to prevent
disconnection from the power source. The chair must be positioned on a smooth level
surface.
WARNING: Do not raise the chair using the backrest. Only lift or raise the chair by its
base.
WARNING: Keep fingers away from area between headrest plat and neck clamp
WARNING: The mains tap must be switched off when the unit is not in use.
6
WARNING: All equipment must be cleaned before the patient chair is used.
Please see the section on "Maintenance" for details of approved cleaning agents
and methods.
There is a detailed explanation of methods for cleaning and maintenance
according to the manufacturer's instructions.
Electromagnetic Compatibility (EMC)
Changes or modifications to this product not expressly approved by Heka Dental A/S
may result in increased emissions or decreased immunity performance of the product and
could cause EMC issues with other equipment. This product is designed and tested to
comply with applicable regulation regarding EMC and shall be installed and put into
service according to the EMC information stated below:
WARNING: Use of portable phones or other portable or mobile radio frequency (RF)
emitting equipment near the product may cause unexpected or adverse operation.
WARNING: The use of accessories, transducers and cables other than those supplied
may result in increased emissions or decreased immunity performance of the product.
The product shall not be used adjacent to, or stacked with, other equipment. If adjacent
or stacked use is necessary, the user shall be responsible to test to verify normal operation
in the configuration in which it is being used.
Cautions
1.5 Cautions
CAUTION: Federal law restricts this device to sale by or on the order of a dentist.
CAUTION: Always examine unit components for damage before commencing
treatment. Damaged components must not be used and must be replaced.
CAUTION: Read and understand all accompanying documents for OEM instruments
and equipment purchased for use with the UnicLine Mobile before attempting to connect
for use.
CAUTION: This patient chair maximum load is 268lbs/130Kg.
CAUTION: Ensure this equipment is inspected annually to ensure the device and the
equipment remains fully operational. In the event of faults or defects, this manual is the
primary source of information.
7
The use of additional accessories that do not comply with the safety regulations for this
equipment may impair the safety of the entire unit.
In this context, the following should be taken into consideration:
Use of Accessories:
Documentation proving that the safety of the accessory has been certified in
compliance with the current IEC 60601-1 and/or IEC 60601-1-1 international
standard and the current ISO7494.
The patient chair is a medical device that complies with the standards listed in Section 9.3
of this manual.
The patient chair meets the requirements of the 93/42/EEC Directive.
The manufacturer, the service engineer, and the importer are only responsible for the
safety, reliability, and function of the device if:
Installation, calibration, modification and repair of the device are carried out by
qualified and authorized personnel.
Electrical installations are in compliance with current regulations, e.g. IEC 60364.
Only the OEM instruments specified in the order are used on this patient chair.
The equipment is operated according to the instructions in the user's manual.
It is not permitted:
To modify this equipment without the advance written permission of Heka Dental
A/S.
If this equipment is modified, it must be inspected and tested by qualified and authorized
personnel to ensure that it is suitable for use.
8
Symbols
Type B applied part (Standard IEC 60878)
Alternating current (Standard IEC 60878)
General warning (Standard ISO 7010)
See attached documents and/or catalogues
Protective earth
High voltage
Follow "Instructions for use"
Separate collection of electrical and electronic equipment
pursuant to Directive 2002/96/EEC (WEEE).
9
Transport and stock conditions
Package contents of the UnicLine Chair
FIGURE 1 Chair Carton
FIGURE 2 Chair & Box Containing Upholstery
2
1
10
FIGURE 3 Chair with Box Containing Neck
Support and Instructions for Use
1. Lift off the main carton from the base/pallet (FIGURE 1)
2. Lift off the box containing the upholstery from the chair and set aside until ready
to install (FIGURE 2)
3. Locate the box containing the Neck Support and set aside until ready to install
(FIGURE 3)
4. Proceed to 2. Assembly Instructions on page 11.
3
11
2. Assembly instructions
On receipt of the chair: Remove the transport packaging from the patient chair.
Note: The chair is bolted to the wooden pallet for transport.
Position the patient chair on the spot where it will be installed by lifting it as shown
below.
Lay the chair on its side and mount 1 - 3 joysticks in the optional positions, attach the
cable to the clips and mount the plugs.
Attach the wheels, leg extensions and fittings for raising as indicated in the assembly
sheet in the technical instructions.
Lift the chair into an upright position and adjust the screws to ensure that the chair is
stable.
Mount the armrest. Note: This is an accessory.
Mount the footrest. Note: This is an accessory.
Check that the seat is in the upper position (5° seat angle). See section 5.6.
Insert the headrest rod into the backrest and fit the cover onto the headrest.
Install 4(brass) upholstery screws in the leg support and fit the seat cover, finally
install finger screw under the seat, M6x22 mm without armrest or M6x40 mm with
armrest.
Fit the backrest cover.
Depending on the chair model, connect the patient chair to the portable cabinet's power
point (230VAC/115VAC).
Switch on the mains switch, which is located under the footrest on the front of the
patient chair.
Test all functions.
Go through all of the patient chair's functions and maintenance with the dentist.
The chair must not be raised by the backrest
12
Note: A manufacturer pre-installed telescoping
suction system arm is included on chairs
purchased for use with the UnicLine Mobile
system.
3. Structure of the operator's manual
This operator's manual provides instructions for use of the patient chair.
Therefore, the operator's manual must be read carefully before the patient chair is used.
The manual also includes technical instructions, which may be of use in the event of a
service visit.
In the event of faults or defects, contact an authorised
Heka Dental dealer.
Heka Dental reserves the right to continue to develop the product.
Although we strive to keep the documentation up to date, the chair may
include options that are not described in this operator's manual.
We reserve the right to modify the product's specifications without prior
notice.
13
4. Description of individual components
Nakkebetræk
Armrest
(Accessory)
Backrest cover
Seat cover
Footrest and
Stirrup
(Accessory)
Safety stop, chair lowering
(Accessory)
Backrest plate
Headrest with
handle
Headrest cover
seat angle
Handle,
Programming button
14
5. Safety instructions
The patient chair must only be used under the supervision of dental professionals.
This is to protect the operator, patient, and chair against injury and damage.
To make sure the patient is in the most ergonomic position during treatment, the operator
must maintain a clear view of the chair and patient at all times. In addition, the operator
must make sure the patient is sitting or lying in the chair correctly according to the
instructions below.
This is to protect the operator and patient from being squeezed or crushed.
Before moving the chair into a different position, the operator must check to be sure that
the area under the chair's components is completely clear.
CAUTION: This chair is designed for a maximum load of 268lbs/130Kg.
5.1 Position of the operator
The operator must position him/herself behind the patient chair next to the headrest,
facing the chair and the patient.
From this position, the operator can guide the patient into the correct sitting/lying position
and move the chair to the next position with his/her foot on the joystick.
5.2 Position of the patient
When the patient chair is moved to a new position, the patient must sit/lie with hands on
stomach or on the handle if the chair is equipped with such an accessory.
15
6. Warning signals and emergency stop
The patient chair is equipped with safety systems to protect you against injury and the
chair against damage.
6.1 Program stop
When the chair moves to a programmed position, it can be stopped by moving the chair's
joystick in one of the four directions.
6.2 Safety stop: Backrest
The backrest has a safety
function that stops the chair movement if
the backrest hits a knee or thigh, or if it is obstructed
in any other way.
When the backrest stops, the power to the chair's
motor will cut out and an alarm signal will be emitted.
When the obstruction has been removed, the chair can
be operated normally.
6.3 Limit stop
The chair is equipped with a limit stop to prevent excess strain
on the chair's mechanical components. When the limit stop is reached, the motor cuts out
and the chair can be moved away from the stop by pushing the joystick in the opposite
direction.
6.4 Integrated thermal cut-out
The chair's backrest and seat motors have integrated
thermal fuses that cut power to the motors if they are subjected to
harmful temperatures. This may occur if the chair is in constant operation
or the motors are overloaded.
The motors can be activated again once they have cooled down.
6.5 Emergency stop (accessory)
When the emergency stop button is pressed,
the chair's mains power is cut off.
If the chair moves to a programmed position and
the emergency stop is activated, the chair must be
moved manually to the backrest and seat
outer positions once the emergency stop
has been released.
16
6.6 Safety stop, chair lowering (accessory)
If the chair meets an obstruction when it is lowered,
it will immediately stop and lift slightly.
7. Starting up the patient chair
7.1 On/Off
The On/Off switch is under the footrest.
The chair must be switched off at the end of the
working day.
7.2 Adjusting the headrest
The height of the headrest can be adjusted
manually by pulling or pushing it to the
required height.
Caution! Fingers may get caught between the headrest
and the backrest.
7.3 Adjusting the position of the headrest
The position of the headrest can be adjusted by pulling the handle
away from the backrest.
After the headrest has been adjusted, it must be locked into
position by pushing the handle towards the backrest.
Caution! Fingers may get caught between the headrest plate
and the neck clamp.
17
7.4 Operating the armrest
The armrest on the chair is an accessory (can be
added later).
The armrest can be swivelled away from the chair to
make it easier for the patient to get in and out of the
chair. The armrest must be raised approx. 1 cm and
can then be moved to the position required for the
patient to sit down or leave the chair.
7.5 Adjusting the angle of the seat
The angle of the seat can be adjusted by pulling the release
knob under the seat.
The seat can be levelled out by pressing it down while
pulling on the release knob.
It may be an advantage to level out the seat when treating
children.
The bottom position is intended for use during
operation and must not be used when a patient
is getting into or out of the chair as this will put
an unnecessary strain on the footrest.
7.6 Adjusting the footrest
The footrest on the chair is an accessory (can be added later).
The footrest can be clicked into 3 positions.
Lift the stirrup to change the position.
If the stirrup is raised to the upper position (position 4), it will
be released and will return to the starting position.
Use the stirrup as a footrest only and not as a step.
18
8. Description of the joystick
The chair can be adjusted manually or automatically to the desired position using the foot
joystick.The chair can be equipped with up to 3 joysticks to make adjustment easier.
8.1 Manual adjustment of the chair
If you wish to adjust the chair manually, the joystick must be activated until the desired
position is reached.
8.2 Seat up
When the joystick is in the "Up" position,
the seat will be raised until the foot is
removed from the joystick or the seat has
reached its limit stop.
8.3 Seat down
When the joystick is in the "Down"
position, the seat will be lowered until the
foot is removed from the joystick or the
seat has reached its limit stop.
8.4 Backrest down
When the joystick is in the "Left"
position, the backrest will be lowered
until the foot is removed from the joystick
or the backrest has reached its limit stop.
8.5 Backrest up
When the joystick is in the "Right"
position, the backrest will be raised until
the foot is removed from the joystick or
the backrest has reached its limit stop.
A double-function joystick is available as
an extra accessory, i.e. the backrest and seat can be activated simultaneously.
19
9. Automatic adjustment of the chair
The chair has 4 programs that can be programmed as required.
If one of the 4 programs is used, the joystick will only need to be activated briefly in the
program in question. The chair will automatically move to the programmed position.
The 4 programs in the chair are set at the factory. Once installed, the programs can be
modified to suite the operator’s needs.
10. Programming the chair's positions
The chair must be manually set to the desired position.
Next, hold down the program button (located under the seat of the chair) and,
simultaneously, move the joystick to the desired program setting (P1 - P4).
The chair will emit a high-pitched sound. When the sound stops, the position has been
stored and the programming button can be released.
The same procedure applies to the setting of the other positions.
Once the positions have been programmed, the settings can be checked by briefly moving
the joystick to the program setting in question.
The programs will be saved when the chair is switched off.
The chair can be stopped while moving to a programmed position by
briefly pushing the joystick in one of the four directions.
20
11. Maintenance
The Centers for Disease Control and Prevention (CDC) recommends use of a chemical
germicide registered with the EPA as a Hospital Disinfectant and labeled for
tuberculocidal activity for surfaces that have been soiled with patient material. In
compliance with this guideline, we recommend intermediate-level disinfection between
patients using Sani-Cloth AF3. It is important to follow the manufacturer instructions for
use to ensure effective disinfection.
We recommend cleaning and disinfection at the start and end of each day as well as
between patients.
11.1 Cleaning of upholstery and armrest:
Material: UltraLeather (Imitation Leather upholstery)
Cleaning:
Wipe all surfaces using a Sani-Cloth AF3 cloth at the beginning and end of each workday
and between patients. Inspect and repeat this process if any visible soil or residue exists.
Discard the Sani-Cloth used for cleaning. .For stains, alcohol based cleaners such as
Fantastik® and Formula 409® can be used. To sanitize, use a solution of 1:5 bleach to
water. Rinse with clean water to remove cleaning solution residue. Allow to air dry.
Disinfection:
After each cleaning, disinfect using a fresh Sani-Cloth AF3. Ensure the surfaces remain
wet for the full three minutes. It is crucial that the Sani-Cloth AF manufacturer
instructions are followed EXACTLY to ensure effectiveness.
Agents that must not be used:
The following agents may cause permanent damage to upholstery and the armrest:
Wax polish
Alcohol
Acetone
Perchloroethylene
Trichloroethylene
Powder cleaning.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37

HEKA UnicLine Installation, Operator And Technical Manual

Type
Installation, Operator And Technical Manual

HEKA UnicLine is an electrically controlled patient chair designed for dental treatment. It is intended for use with the UnicLine Dental Unit. The UnicLine patient chair has the following capabilities:

  • Electrically controlled movements for backrest, seat, and footrest
  • Manual adjustment of the headrest
  • Integrated thermal cut-out for safety
  • Emergency stop button (accessory)
  • Safety stop for chair lowering (accessory)
  • Programmable chair positions
  • Last position function
  • Compatible with make and break contacts

The UnicLine patient chair is designed to provide optimal functionality and comfort for both patient and dentist. The simple curves and stylish design result from collaboration with dentists, engineers, and designers.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI